3D PROSTATE SUITE 3DTRUS PROSTATE MAPPING, MRI/3DTRUS FUSION OPTION, SECOND LOOK 3DTRUS FUSION OPTION
K131448 · Koelis · LLZ · Jul 26, 2013 · Radiology
Device Facts
Record ID
K131448
Device Name
3D PROSTATE SUITE 3DTRUS PROSTATE MAPPING, MRI/3DTRUS FUSION OPTION, SECOND LOOK 3DTRUS FUSION OPTION
Applicant
Koelis
Product Code
LLZ · Radiology
Decision Date
Jul 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K131448 · Jul 26, 2013
3D PROSTATE SUITE 3DTRUS PROSTATE MAPPING, MRI/3DTRUS FUSION OPTION, SECOND LOOK 3DTRUS FUSION OPTION
Koelis
Retrospective clinical patient volumes; Clinical biopsy session data
The sponsor used retrospective clinical data from multiple sites to validate the accuracy and repeatability of the elastic registration and organ tracking algorithms for the Urostation system.
Elastic registration validation; Organ tracking accuracy; Retrospective clinical cohort
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Brolis et al, Internal Technical Review, 2012; Retrospective analysis of clinical patient volumes; Study Period: 2012
49 patients; Sample Size: 49 patients; Number of Sites: 5
Not applicable for this study
Registration quality (distance between contours and landmarks)
Baumann et al, Medical Image Analysis 16 (2012) 562-576; Retrospective analysis of clinical biopsy sessions; Study Period: 2012
47 patients; Sample Size: 786 biopsy volumes from 47 patients
Not applicable for this study
Registration failure rate, accuracy using fiducials
Indications for Use
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS fusion option and with Second Look 3DTRUS fusion option is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate.
Device Story
Urostation 3D Prostate Suite is a PC-based software application for transrectal prostate biopsy guidance. It receives 3D digital ultrasound images from external scanners via Ethernet and DICOM MRI data. The device uses 'Organ-Based Tracking' technology to perform elastic registration and fusion of 3DTRUS images with MRI or prior 3DTRUS reference images. Physicians use the system intraoperatively to visualize 3D biopsy mapping relative to the prostate anatomy. The software provides multiplanar reconstruction, surface/volume rendering, and organ delineation to assist in targeting biopsy sites. Output includes 3D biopsy maps and clinical reports for storage or printing. The system is operated by physicians in clinical or hospital environments to improve targeting accuracy during prostate procedures.
Clinical Evidence
Evidence includes bench testing and clinical validation. Elastic MR/3DTRUS fusion validated on 49 patients (112 landmark pairs) showed 1.7±0.7mm mean landmark distance and 0.7±0.3mm surface accuracy. Elastic 3DTRUS/3DTRUS fusion validated on 47 patients (786 volumes) showed 0.76±0.52mm accuracy. Targeting accuracy evaluated in 6 prostate phantoms showed 100% hit rate for hypoechoic lesions (2.35mm overall error) and 84% for isoechoic lesions (2.92mm overall error). Total system RMS error is 2.05mm.
Technological Characteristics
PC-based software application running on Windows 7 or higher. Interfaces with 3D ultrasound scanners via Ethernet. Features include elastic registration, multiplanar reconstruction, surface/volume rendering, and 3D image fusion. Connectivity via Ethernet/DICOM. No measurement features. Software-based image processing.
Indications for Use
Indicated for use by physicians in clinic or hospital settings for 2D/3D visualization of the prostate gland and 3D transrectal ultrasound (TRUS) based fusion of multiple imaging modalities (ultrasound, MRI) to map the prostate gland. Includes patient data management, multimodal communication, multiplanar reconstruction, surface/volume rendering, organ/ROI delineation, 3D registration, and reporting.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a handwritten string of characters. The string appears to be a combination of letters and numbers. The string is "K13/448". The characters are written in a cursive style with thick, dark lines.
# 510(k) Summary for UROSTATION – 3D Prostate Suite MRI/3DTRUS FUSION OPTION SECOND LOOK 3DTRUS FUSION OPTION
The 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Owner: | KOELIS<br>5, avenue du Grand Sablon 38700 La Tronche<br>FRANCE | |
|----------------|--------------------------------------------------------------------|--|
| | Phone: +33 476 637 588 Fax: +33 476 549.561 | |
| Contact Name: | Ms Laetitia GERVAIS<br>Quality Manager<br>Mail: gervais@koelis.com | |
| Date Prepared: | 04/30/2013 | |
# Proposed Device:
| Trade Name: | UROSTATION - 3D PROSTATE SUITE |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Common Name: | Medical Image Processing Software System |
| Classification Name: | System, Image processing, Radiological<br>Picture archiving and communication system,<br>21 CFR PART 892.2050 |
| Device Class | II |
| Product Code | LLZ |
# Cleared Device:
The Urostation system is substantially equivalent to:
| 510(k) Number | Device Name: |
|---------------|------------------------------|
| K100793 | UROSTATION-3D Prostate Suite |
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "KOELIS" in a sans-serif font. To the left of the word is a black arrow pointing to the left. The letters are all capitalized and evenly spaced. The image is black and white.
## Intended Use:
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS fusion option and with Second Look 3DTRUS fusion option is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate.
## Indications for Use:
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS fusion option and with Second Look 3DTRUS fusion option is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of the prostate gland and for the 3D transrectal ultrasound based fusion of multiple imaging modalities (ultrasound, MRI) in order to map such prostate gland. Additional software features include patient data management, multimodal data communication, multiplanar reconstruction, surface and volume rendering, organ delineation, region of interest delineation, 3D image registration and data reporting.
## Device Description:
UROSTATION - 3D PROSTATE SUITE is a computer-based software application designed to process, visualize and record 3D digital ultrasound images of the prostate, and to manage patient and clinical data in the context of transrectal prostate biopsy. Software options provide the fusion of 3DTRUS with MRI or with other 3DTRUS modalities.
Hardware Platform and Operating System
The application runs on a standard Personal Computer under Microsoft Windows® operating system (version 7 or higher).
#### Peripheral and accessories
The application is controlled by a footswitch and manual input devices (mousc, kevboard),
It is designed to work in connection with commercially available 3D ultrasound scanners with Ethernet connection, 3D transrectal ultrasound probe and needle guide.
## Software Features
UROSTATION - 3D PROSTATE SUITE implements image fusion and display algorithms to provide 3D representation of prostate biopsies.
A typical workflow enables the physician to intraoperatively visualize the 3D mapping of biopsies with respect to a 3DTRUS reference image of the patient's prostate.
For that purpose. 3D digital images may be transferred at any time from the 3D ultrasound scanner to the Urostation for registration and display, while the physician keeps track of the organ on the ultrasound scanner using the usual 2D live B mode.
Optionally, MRI/3DTRUS fusion allows the elastic registration of the 3DTRUS reference image with other imaging modalities (MRI here) in order to display the 3D biopsy mapping on multiple imaging modalities.
Optionally, Second Look 3DTRUS Fusion allows the elastic registration of the 3DTRUS reference image with a previously acquired 3DTRUS reference image of the same patient in order to superimpose two 3D biopsy mappings on a unique 3DTRUS reference image.
Alternatively, UROSTATION - 3D PROSTATE SUITE also provides a review mode that allows the mapping of histologic results on the said 3D reference image of the patient's prostate. Patient
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "KOELIS" in a sans-serif font. To the left of the word is a black arrow pointing to the left. The letters are all capitalized and evenly spaced. The image is simple and clear, with a focus on the text and arrow.
information, images and 3D biopsy mapping may be stored or printed for future retrieval and examination.
# Technological Characteristics compared with the cleared device:
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS Fusion option and Second Look 3DTRUS Fusion option utilizes the same technological characteristics as the cleared device:
- Both are PC-based software applications that provide 2D/3D medical image acquisition ﮯ and visualization for enhanced observation and analysis of the prostate gland.
- Both software architectures follow a workflow adapted to the physician's practice. -
- Both systems implement a dedicated algorithm to visualize biopsies with respect to a single 3DTRUS reference image of the prostate.
- Both systems have no measurement features. -
- Both systems provide in addition patient and clinical data management features. -
- Both systems deal with 3D ultrasound images received from commercially available imaging devices.
- Both systems provide 3D image registration and data reporting features -
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS Fusion option and Second Look 3DTRUS Fusion option provide also additional features are based on the same fundamental scientific technology and do not raise any safety or effectiveness concern in comparison with the cleared device:
- The MR/3DTRUS Fusion option implements the same principles of 3D model-based । iterative matching of prostate data as the cleared device. More specifically, the implemented algorithm to fuse multimodal images (Ultrasound. MRI) is based on userdelineated 3D prostate contours. The accuracy and repeatability are equivalent.
- -The Second Look 3DTRUS Fusion option implements the same principles and imagebased matching of prostate data as the cleared device. The accuracy and repeatability are equivalent.
- -The system provides patient data management. multimodal data communication, multiplanar reconstruction, surface and volume rendering, organ delineation, region of interest delineation. 3D image registration and data reporting
{4}------------------------------------------------
| Company: | KOELIS | KOELIS | KOELIS |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------|
| System: | 3D PROSTATE SUITE<br>3DTRUS prostate<br>mapping | 3D PROSTATE SUITE<br>MRI/3DTRUS FUSION<br>OPTION | 3D PROSTATE SUITE<br>SECOND LOOK<br>3DTRUS FUSION<br>OPTION |
| 510(k) number: | K100793 | | |
| Function | 2D/3D image:<br>acquisition viewing/reviewing processing registration storage Multi-Planar Reformatting (MPR)<br>Volume rendering<br>Patient and clinical data management<br>Patient and clinical data reporting<br>Printing | - Surface rendering<br>- Organ delineation<br>- Region of interest<br>delineation | |
| Intended Use: | To process, visualize and record 3D digital ultrasound images of the<br>prostate | | |
| Data Source: | 3D TRUS scanners | | |
| | • 3D TRUS scanners | - 3D TRUS scanners<br>- Dicom MR images | • 3D TRUS scanners |
| Physical Characterization: | Software package<br>- operates on standard PC-based hardware<br>- Windows operating system | | |
# Substantial Equivalence Comparison Chart
# Conclusion:
The results of comparing the intended uses, functions and technological characteristics of the UROSTATION - 3D PROSTATE SUITE including MRI/3DTRUS Fusion option and Second Look 3DTRUS Fusion option with its cleared device shows that the system is as safe and as effective as its cleared device.
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS fusion option and Second Look 3DTRUS fusion option is substantially equivalent to existing product currently on the market.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "KOELIS" in a sans-serif font. To the left of the word is a black arrow pointing to the left. The letters are spaced apart, and the font is simple and clean. The image is in black and white.
## Performance summary
UROSTATION - 3D PROSTATE SUITE's 3D display and image fusion capabilities are based on a patented technology called Organ-Based Tracking. The ability of Urostation to register biopsy cores, and optionally MR lesions, on a reference 3DTRUS image of the prostate would consequently allow to assist the physician in targeting different regions of the prostate.
The performance and accuracy of KOELIS fusion technology has been validated on phantom and patient data, as shown in the following studies.
#### Elastic MR/3DTRUS image fusion:
#### [Brolis et al, Internal Technical Review, 2012]
From 17 contoured TRUS patient volumes, 850 simulated deformations were registered using Koelis clastic registration method. The overall time for registration was 3.9s. After clastic registration the mean distance between the reference and deformed prostates was 0.09mm The RMS error stemming from deformations in the probe area is reduced from 4.1mm down to 1.9mm. Clinical validation has been performed on 49 patients in 5 clinical sites. A total of 112 landmark pairs were identified and approved by physicians. The registration quality is measured using the distance between contours and between landmarks. Surfaces were superimposed after elastic registration with 0.7±0.3mm mean accuracy. The mean distance between corresponding landmarks was 1.7+0.7 mm after elastic registration, and the RMS error was 1.9mm.
#### Elastic 3DTRUS/3DTRUS image fusion and organ tracking:
## [Baumann et al, Medical Image Analysis 16 (2012) 562-576]
A volume-swept 3DTRUS based tracking system for fast and accurate estimation of prostate tissue motion is proposed. Prostate deformations are estimated with elastic registration to maximize accuracy. The system is robust with only 17 registration failures out of 786 (2%) biopsy volumes acquired from 47 patients during biopsy sessions. Accuracy was evaluated to 0.76mm±0.52mm using fiducials on 687 registered volumes stemming from 40 patients.
## Targeting Accuracy of Urostation
## [Ukimura et al, J. Urol. Vol. 187, 1080-1086, 2012]
The accuracy of the Urostation 3DTRUS system for image-based mapping biopsies was evaluated in 6 prostate phantoms. 3 containing 3 hypoechoic lesions and 3 containing 3 isoechoic but MRJ-visible lesions, to perform MR fusion guided biopsy. Three targeted biopsies were done per lesion. A total of 27 ultrasound guided biopsies were targeted into 9 hypocchoic lesions. All 27 biopsies (100%) successfully hit the target lesion with a procedural targeting error of 1.52±0.78mm and a system registration error of 0.83mm, resulting in an overall error of 2.35mm. Of the 27 MR fusion biopsies 24 (84%) hit the lesion. For isoechoic lesions mean procedural targeting error was 2.09±1.28mm, resulting in an overall error of 2.92mm.
## Conclusion on performance and accuracy:
The overall fusion RMS error of the UROSTATION - 3D PROSTATE SUTE. combines preoperative MRI/3DTRUS fusion (1.9mm error) and intraoperative 3DTRUS/3DTRUS fusion used for organ tracking (0.76mm error). The total RMS error of the visualization process is hence 2.05mm at any time during the procedure, giving the physician a probability greater than 95% to hit the lesion according to /Karnick et al. Med. Phys. 37 (2). 802-8131.
KOELIS therefore concluded that the accuracy provided by the UROSTATION fusion technology is clinically acceptable.
{6}------------------------------------------------
Public Health Service
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes that resemble birds in flight or flowing lines. The logo is printed in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
DEPARTMENT OF HEALTH & HUMAN SERVICES
LAETITIA GERVAIS QUALITY MANAGER KOELIS 5, AVENUE DU GRAND SABLON LA TRONCHE 38700
Re: K131448
FRANCE
Trade/Device Name: Urostation - 3D Prostate Suite, MRI/3DTRUS Fusion Option, Second Look 3DTRUS Fusion Option Regulation Number: 21 CFR 892.2050
July 26, 2013
Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 20, 2013 Received: June 03, 2013
Dear Ms. Gervais:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Ms. Gervais
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131448
UROSTATION 3D Prostate Suite Device Name: MRV3DTRUS FUSION OPTION SECOND LOOK 3DTRUS FUSION OPTION
#### Indications for Use:
UROSTATION - 3D PROSTATE SUITE with MRI/3DTRUS fusion option and with Second Look 3DTRUS fusion option is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of the prostate gland and for the 3D transrectal ultrasound based fusion of multiple imaging modalities (ultrasound, MRI) in order to map such prostate gland. Additional software features include patient data management, multimodal data communication, multi-planar reconstruction, surface and volume rendering, organ delineation, region of interest delineation, 3D image registration and data reporting.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7?)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> K131448 510(k)
> > Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.