CapSure Fixation System - Straight 5mm x 37 cm – 30 Permanent fasteners, CapSure Fixation System - Straight 5mm x 37 cm – 15 Permanent fasteners

K160900 · C.R. Bard, Inc. · GDW · May 31, 2016 · General, Plastic Surgery

Device Facts

Record IDK160900
Device NameCapSure Fixation System - Straight 5mm x 37 cm – 30 Permanent fasteners, CapSure Fixation System - Straight 5mm x 37 cm – 15 Permanent fasteners
ApplicantC.R. Bard, Inc.
Product CodeGDW · General, Plastic Surgery
Decision DateMay 31, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CapSure ™ Permanent Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Device Story

Sterile, single-use, mechanical fixation system; delivers 15 or 30 permanent fasteners via 37 cm straight shaft. Fasteners consist of 316L stainless steel helical coil with PEEK cap. Used in open or laparoscopic surgery (e.g., hernia repair) by surgeons. Device handle-actuated; rotational delivery mechanism drives fasteners into tissue/mesh. Provides permanent fixation (>30 days). Benefits include secure mesh attachment and reduced tissue attachments compared to controls. No software or electronic components.

Clinical Evidence

No clinical studies were performed. Evidence consists of non-clinical bench testing (biocompatibility, mechanical testing including trigger force, mesh compatibility, deployment reliability, MR compatibility, burst testing) and a porcine in vivo implantation study showing significantly fewer tissue attachments (p<0.05) compared to control.

Technological Characteristics

Mechanical fixation system; 37 cm straight shaft; 15 or 30 fastener capacity. Fasteners: 316L Stainless Steel (ASTM F138) helical coil with PEEK (ASTM F2026) cap. Deployment: Handle-actuated, rotational delivery. Sterilization: Ethylene Oxide (EtO). No software, electronics, or energy source.

Indications for Use

Indicated for soft tissue approximation and surgical mesh fixation during open or laparoscopic procedures, such as hernia repair, in patients requiring such surgical intervention.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a traditional symbol of medicine. The symbol is composed of three figures in profile, representing health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 31, 2016 C.R. Bard Inc. Mr. Andrew Harrell Senior Regulatory Affairs Specialist 100 Crossings Blvd Warwick, Rhode Island 02886 Re: K160900 Trade/Device Name: Capsure Fixation System - Straight 5mm X 37 Cm - 30 Permanent Fasteners, Capsure Fixation System - Straight 5mm X 37 Cm - 15 Permanent Fasteners Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: March 31, 2016 Received: April 1, 2016 Dear Mr. Harrell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number: K160900 Device Name: CapSure™ Permanent Fixation System The CapSure ™ Permanent Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. Prescription Use _ X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 PREMARKET NOTIFICATION FOR CAPSURE™ PERMANENT FIXATION SYSTEM {3}------------------------------------------------ #### Section 5 510(k) Summary ### SUBMITTER Davol Inc. 100 Crossings Boulevard Warwick, RI 02886 Contact Person: Andrew Harrell Senior Regulatory Affairs Specialist Phone: (401) 825-8472 (401) 825-8447 Fax: e-Mail: Andrew.Harrell(@crbard.com Date Prepared: March 31, 2016 ## II. DEVICE | Name of Device: | CapSure™ Permanent Fixation System (product code GDW) | |----------------------|-------------------------------------------------------| | Common Name: | Implantable Staple | | Classification Name: | Implantable staple (21 CFR §878.4750) | | Regulatory Class: | II | | Product Code: | GDW | ## III. PREDICATE DEVICE K142808 CapSure™ Permanent Fixation System (Davol Inc.) - Cleared March 9, 2015 # IV. DEVICE DESCRIPTION The CapSure™ Permanent Fixation System is a sterile single use device that delivers either 15 or 30 permanent fasteners via a straight shaft. The shaft of the CapSure™ Permanent Fixation System is 37 cm in length. The fasteners are designed with a 316L Stainless steel helical coil and polyetheretherketone (PEEK) cap on the proximal end to support mesh or tissue. ### V. INDICATIONS FOR USE The CapSure™ Permanent Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. The Indications for Use statement for the subject device is unchanged as a result of this 510(k) and remains identical to the device cleared in K142808. Both the subject and predicate devices retain the identical intended use for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. {4}------------------------------------------------ # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | Device Comparison | | | |----------------------------------------------|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Features | CapSureTM<br>(Subject Device) | CapSureTM<br>(Predicate - K142808) | | Intended Use | Identical to predicate | Permanent soft tissue fixation | | Indication For use | Identical to predicate | Indicated for the approximation of<br>soft tissue and fixation of surgical<br>mesh to tissues during open or<br>laparoscopic surgical procedures,<br>such as hernia repair. | | Mesh/Tissue<br>Retention Feature<br>Material | Identical to predicate | Polyetheretherketone (PEEK) cap<br>(injection molded) | | Fastener Material | Identical to predicate | 316L Stainless Steel (metal wire<br>fabrication) | | Fastener Body<br>Contact | Identical to predicate | Long term implant (>30 days)<br>contacting tissue and/or bone | | Fastener<br>Shape/Design | Identical to predicate | Helical Coil/Screw with retention<br>feature (proximal cap) | | Fastener<br>Dimensions | Identical to predicate | 4.2 mm overall fastener length<br>4.0 mm overall cap diameter | | Fastener Quantity<br>per Device | Identical to predicate | 15 & 30 fasteners | | Deployment<br>component -Shaft<br>Length | Identical to predicate | 37 cm length | | Deployment<br>component Handle<br>design | Identical to predicate | Handle<br>actuated-level<br>delivery<br>device | | Fastener Delivery<br>System | Identical to predicate | Rotational - driven by inner shaft<br>assembly | | Device<br>Sterilization | Identical to predicate | EtO | # VII. PERFORMANCE DATA # Performance Standards No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act. The following performance data are provided in support of the substantial equivalence determination. ### Biocompatibility testing The biocompatibility evaluation for the CapSure™ Permanent Fixation System was reported in K142808, in accordance with the FDA Blue Book Memorandum #G95-1 "Use of {5}------------------------------------------------ International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: Device Testing . Cytotoxicity - Sensitization . - Irritation - . Pyrogen Testing Implant Materials Testing - . Cytotoxicity - . Maximization Sensitization Study - Intracutaneous Study - . Acute Systemic Toxicity Study - . Pyrogenicity - . Hemolysis - Complement Activation Assay - . Intramuscular Implant - 12 Weeks - . Rabbit Femoral Bone Implant - 12 Weeks - Rabbit Femoral Bone Implant - 26 Weeks - o Bacterial Reverse Mutation (Ames) Assay - In Vitro Mouse Lymphoma Mutation Assay - o Mouse Peripheral Blood Micronucleus Test - . Subacute (14-Day) Intraperitoneal Toxicity Study Mice - . Subchronic (14-Day) Intravenous Toxicity Study Mice - . Subchronic (13-Week) Toxicity Study in Rats - . In Vivo Neurotoxicitv - o Non-Volatile Residue - . Residue on Ignition - Turbidity - . UV Absorption The deployment device of the CapSure™ Permanent Fixation System is determined to be tissue contacting for duration of less than 24 hours. while the fasteners are determined to be permanent implants. The implantable fastener material conforms to ASTM F138 (316L Stainless Steel) and ASTM F2026 (PEEK). All samples tested met the acceptance criteria. # Mechanical testing The following non-clinical tests were completed for the CapSure™ Permanent Fixation System in K142808 and passed all the test requirements and showed substantial equivalence to the results of the predicate device identified in K142808 – ProTack™ - . Trigger Force - Mesh Compatibility ● - . Deployment Reliability {6}------------------------------------------------ - . MR Compatibility - . Burst Testing All samples tested met the acceptance criteria. ## Animal Studies The following in vivo study was completed for this premarket submission - . Porcine implantation study As compared to the control. CapSure™ had significantly (p<0.05) fewer tissue attachments to fasteners, indicative of a role of the CapSure™ Device in minimizing tissue attachments to fasteners. # Clinical Studies Clinical studies were not performed for this device nor were clinical studies performed for the predicate device, as they are not necessary to adequately assess the safety and effectiveness of Class II products. ## VIII. CONCLUSIONS The CapSure™ Permanent Fixation System is substantially equivalent to the legally marketed predicate device for the following reasons: - A) The intended use and indications for use are not changed from the predicate device. - B) No technological characteristics have changed related to this premarket notification. - C) The update to the instructions for use has been supported by adequate performance data. As demonstrated in the completed tests that were conducted by the company in K142808 and this submission, the CapSure™ Permanent Fixation System is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...