PaX-i3D Green Premium (Model: PCT-90LH) is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the maxillofacial areas for a dental and ENT treatment in adult and pediatric dentistry. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by physicians, dentists, ENT specialist and x-ray technicians.
Device Story
Digital radiographic imaging system (CBCT) for dental and ENT applications. Inputs: X-ray projections captured via CMOS digital detector (Xmaru3104CF). Operation: X-ray generator produces radiation; detector captures data; system performs computer reconstruction of 3D volumes or 2D panoramic/cephalometric images. Includes iterative reconstruction algorithm to reduce noise. Output: 2D/3D diagnostic images displayed on PC via EzDent-i and Ez3D-i software. Used in dental/ENT clinics by dentists, physicians, and technicians. Clinical utility: Provides anatomical visualization for treatment planning (implants, orthodontics, oral surgery). Benefits: Enhanced diagnostic detail, noise reduction via iterative processing, and multi-modal imaging (PANO, CEPH, CT, 3D photo) in one system.
Clinical Evidence
Bench testing only. Performance evaluated via non-clinical testing of the Xmaru3104CF detector (MTF, DQE, NNPS) and CT image quality metrics (Contrast, Noise, CNR, MTF). Results demonstrated performance equivalent or superior to the predicate device. Compliance with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63, IEC 60601-1-2, and 21 CFR 1020.30/31/33 confirmed.
Technological Characteristics
Digital CBCT system using CMOS photodiode array detector (Xmaru3104CF) with CsI:Ti scintillator. X-ray generator (1,200W inverter) with 60-120 kV range. Focal spot 0.5 mm. Connectivity: DICOM 3.0 compliant. Software: EzDent-i (2D) and Ez3D-i (3D) viewers. Iterative reconstruction algorithm for noise reduction. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63, IEC 60601-1-2, NEMA PS 3.1-3.18.
Indications for Use
Indicated for producing panoramic, cephalometric, or 3D digital X-ray images to provide diagnostic details of dento-maxillofacial, ENT, sinus, and TMJ anatomy in adult and pediatric patients. Includes carpal imaging for orthodontic treatment. Operated by healthcare professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, resembling a family or group of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2016
VATECH Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K160882
Trade/Device Name: PaX-i3D Green Premium (Model: PCT-90LH) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: September 22, 2016 Received: September 29, 2016
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160882
Device Name PaX-i3D Green Premium (Model: PCT-90LH)
#### Indications for Use (Describe)
PaX-i3D Green Premium (Model: PCT-90LH) is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the maxillofacial areas for a dental and ENT treatment in adult and pediatric dentistry. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by physicians, dentists, ENT specialist and x-ray technicians.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| ( Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 7 – 510(k) Summary
## 1. 510(k) Summary
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
## 2. Date 510K Summary prepared: September 22, 2016
#### 3. Administrative Information
| Official Correspondent: | Dave Kim / Mtech Group |
|-------------------------|-------------------------------------------------------------------------|
| | 8310 Buffalo Speedway, Houston, TX 77025 |
| | Tel: +713-467-2607 |
| | Fax: +713-583-8988 |
| | Contact person: Mr. Dave Kim (davekim@mtech-inc.net) |
| 510(k) Submitter: | VATECH Co., Ltd. |
| | Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, Korea |
| | Tel: +82-31-379-9492 |
| | Fax: +82-31-379-9400 |
| | Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr) |
#### 4. Device Information
| Type of 510(k) Submission: | Traditional |
|----------------------------|----------------------------------------------------|
| Trade or Proprietary Name: | PaX-i3D Green Premium (Model: PCT-90LH) |
| Common or Usual Name: | Dental Computed Tomography X-ray System |
| Regulation Classification: | Computed tomography x-ray system (21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
#### 5. Predicate Device Information
| Manufacturer: | VATECH Co., Ltd. |
|----------------------------|----------------------------------------------------|
| Predicate device: | PaX-i3D Ortho (TON-95LH) / K132983 |
| Common or Usual Name: | X-Ray, Tomography, Computed, Dental |
| Regulation Classification: | Computed tomography x-ray system (21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
※ This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
#### 6. Reference Device Information
| Manufacturer: | VATECH Co., Ltd. |
|----------------------------|----------------------------------------------------|
| Reference device: | PaX-i3D Smart (PHT-30LFO) / K152106 |
| Common or Usual Name: | X-Ray, Tomography, Computed, Dental |
| Regulation Classification: | Computed tomography x-ray system (21 CFR 892.1750) |
| Product Code: | OAS |
| Class of Device: | Class II |
| Panel: | Radiology |
※ This predicate has not been subject to a design-related recall.
## 7. Device Description
PaX-i3D Green Premium (Model: PCT-90LH), a digital radiographic imaging system, acquires and processes multi FOV diagnostic images for a dentist and an ENT specifically designed for dental and ENT radiography, PaX-i3D Green Premium (Model: PCT-90LH) is a complete digital X-ray system equipped with imaging viewers, Xray generator and a dedicated SSXI detector.
The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of head, neck and craniofacial anatomy for ENT, oral surgery, implant and orthodontic treatment. PaX-i3D Green Premium (Model: PCT-90LH) can also acquire 2D diagnostic image data in panoramic and cephalometric mode.
#### Key components of the device
- 1) PaX-i3D Green Premium (Model: PCT-90LH) digital x-ray equipment
- 2) SSXI detector: Xmaru3104CF
| Item | Description |
|-----------------------------|---------------------------------------------------|
| Model | Xmaru3104CF |
| Detector Type | CMOS photodiode array |
| Pixel Size | 99 µm @ 2X2 binning : PANO |
| | 198 µm @ 4X4 binning : Dental CT, ENT CT, i-CEPH |
| Active Area | 310.4 x 36.4 mm |
| Frame Rate | 107 fps (2x2 Binning) : PANO |
| | 308 fps (4x4 Binning) : Dental CT, ENT CT, i-CEPH |
| Analogue-Digital Conversion | 14 bits |
| Converter | Csl:Ti |
| Energy Range | 60 ~ 120 kV |
| Readout Type | Charge amplifier array |
| Video Output | Optic |
#### 3) X-ray generator
#### - Option 1
| Item | | Description |
|---------------------------|--------------------|-------------|
| High Voltage<br>Generator | Model | DG-08C22S1 |
| | Rated output power | 1,200 W |
| | Type | Inverter |
| | Normal/Pulse kV | 60 ~ 120 kV |
{5}------------------------------------------------
| Item | | Description |
|------------|---------------------|----------------------------------------------|
| | mA | 4 ~ 10 mA |
| | Cooling | Air (Optional fan cooling, ≥ 60 ℃ ) |
| | Total filtration | Min. 2.5 mm Al |
| | Added filtration | 2.0 mm Al |
| X-ray Tube | Manufacturer | Superior |
| | Model | SXR-130-15-0.5 (Stationary Anode type) |
| | Focal spot size | 0.5 mm |
| | Target Angle | 15 degree |
| | Inherent Filtration | At least 1.1 mm Al equivalent at 80 kV |
| | Anode Heat Content | 21 kJ |
| | Duty Cycle | 1:20 or more (Exposure time : Interval time) |
### - Option 2
| Item | Description | |
|---------------------------|--------------------------|----------------------------------------------|
| High Voltage<br>Generator | Model | DG-08C22S2 |
| | Rated output power | 1,200 W |
| | Type | Inverter |
| | Normal/Pulse<br>kV<br>mA | 60 ~ 120 kV<br>4 ~ 10 mA |
| | Cooling | Air (Optional fan cooling, ≥ 60 °C) |
| | Total filtration | Min. 2.5 mm Al |
| | Added filtration | 2.0 mm Al |
| X-ray Tube | Manufacturer | C.E.I |
| | Model | OX/115-05 (Stationary Anode type) |
| | Focal spot size | 0.5 mm |
| | Target Angle | 15 degree |
| | Inherent Filtration | 0.5mm Al |
| | Anode Heat Content | 30 kJ |
| | Duty Cycle | 1:60 or more (Exposure time : Interval time) |
## 4) PC system
| Item | Description |
|---------------------|---------------------------------------------------------------------------------------------------------------------------|
| Operating System | Windows 8.1 Professional 64-Bit OS |
| CPU | Intel Xeon E5-1620v3 3.5GHz 2133 10MB cache CPU |
| RAM | 32GB DDR4-2133 ECC RAM |
| HDD | 1TB SATA 1st HDD |
| Graphics board | NVIDIA GEFORCE GTX980 Ti OC D5 6GB or greater |
| Ethernet interface | Broadcom 5761 Gigabit PCIe NIC |
| Serial Port (RS232) | HP Serial Port Adapter Kit |
| Power Supply | ≥ 700 Watts (90% Efficiency) |
| Slots | 1 PCI Express Gen3 x 8 Slot<br>2 PCI Express Gen3 x 16 slot<br>1 PCI Express Gen2 x 8 Slot<br>1 PCI Express Gen2 x 4 Slot |
# 5) Imaging software
- ① EzDent-i: 2D viewer and patient management software
|
- ・ Manufacturer: EWOO SOFT Co., Ltd.
{6}------------------------------------------------
- · 510(k) Number: K150747
- · Classification Number: 21CFR892.2050
- Product code: LLZ
- ② Ez3D-i: 3D image viewer
- Manufacturer: EWOO SOFT Co., Ltd.
- · 510(k) Number: K150761
- · Classification Number: 21CFR892.2050
- · Product code: LLZ
#### 8. Indication for use
PaX-i3D Green Premium (Model: PCT-90LH) is intended to produce panoramic, cephalometric or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, ENT, sinus and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
{7}------------------------------------------------
| | Subject Device | Predicate Device | Reference Device | |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Device Name | PaX-i3D Green Premium<br>(Model: PCT-90LH) | PaX-i3D Ortho (TON-95LH) | PaX-i3D Smart (PHT-30LFO) | |
| Applicant Name | VATECH Co., Ltd. | VATECH Co., Ltd. | VATECH Co., Ltd. | |
| 510(k) Number | N/A | K132983 | K152106 | |
| Device Classification Name | X-Ray, Tomography, Computed,<br>Dental | X-Ray, Tomography, Computed,<br>Dental | X-Ray, Tomography, Computed,<br>Dental | |
| Classification Product Code | OAS | OAS | OAS | |
| Regulation Number | 21 CFR 892.1750 | 21 CFR 892.1750 | 21 CFR 892.1750 | |
| Indications for Use | PaX-i3D Green Premium (Model: PCT-<br>90LH) is intended to produce<br>panoramic, cephalometric or 3D<br>digital x-ray images.<br>It provides diagnostic details of the<br>dento-maxillofacial, ENT, sinus and<br>TMJ for adult and pediatric patients.<br>The system also utilizes carpal images<br>for orthodontic treatment. The device<br>is to be operated by healthcare<br>professionals. | PaX-i3D Ortho (TON-95LH) is a<br>computed tomography x-ray system<br>intended to produce cross-sectional<br>images of the oral anatomy by<br>reconstructing a three dimensional<br>radiographic images from the same<br>axial plane taken at different angles.<br>It provides diagnostic details of the<br>maxillofacial areas for a dental<br>treatment in adult and pediatric<br>dentistry. The system also acquires<br>carpal images for orthodontic<br>treatment. The device is operated<br>and used by physicians, dentists,<br>and x-ray technicians. | PHT-30LFO is a computed tomography<br>x-ray system intended to produce<br>panoramic, cephalometric or cross-<br>sectional images of the oral<br>anatomy on a real time basis by<br>computer reconstruction of x-ray<br>image data from the same axial<br>plane taken at different angles. It<br>provides diagnostic details of the<br>anatomic structures by acquiring<br>360° rotational image sequences of<br>oral and maxillofacial area for a<br>precise treatment planning in adult<br>and pediatric dentistry. The device is<br>operated and used by physicians,<br>dentists, and x-ray technicians. | |
| Performance Specification | Panoramic, Cephalometric and<br>computed tomography | Computed tomography | Panoramic, cephalometric and<br>computed tomography | |
| Input Voltage | AC 100 - 240 V | AC 100 - 240 V | AC 100 - 240 V | |
| X-Ray source | (Option 1) SXR-130-15-0.5<br>(Option 2) OX/115-05 | SXR-130-15-0.5 | D-052SB | |
| Tube Voltage | 60 - 120 kV | 50 - 120 kV | 50 - 99 kV | |
| Tube Current | 4 - 10 mA | 4 - 10 mA | 4 - 16 mA | |
| Focal Spot Size | 0.5 mm | | 0.5 mm | |
| Exposure Time | Max. 18 s | | Max. 18 s | |
| Slice Width | Min. 0.1 mm | Min. 0.1 mm | Min. 0.1 mm | |
| Total Filtration | (Option 1) Min. 2.5 mm Al | Min. 2.8 mm Al | 2.8 mm Al | |
| | | (Option 2) Min. 2.5 mm Al | | |
| Mechanical | | Compact design | Compact design | Compact design |
| Electrical | | LDCP logic circuit | LDCP logic circuit | LDCP logic circuit |
| | Software | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
| | 2D Image Viewing Program | EzDent-i | EzDent-i | EasyDent |
| 3D Image Viewing Program | | Ez3D-i | Ez3D-i Ortho | Ez3D Plus |
| Anatomical Sites | | Maxillofacial | Maxillofacial | Maxillofacial |
| Image<br>Receptor | Computed<br>Tomography | | Xmaru2430CF Master Plus | Xmaru1404CF |
| | Panoramic | Xmaru3104CF | Reconstructed from 3D image data | |
| | Cephalometric | | Reconstructed from 3D image data | Xmaru2301CF<br>1210SGA<br>910SGA<br>Xmaru2301CF-O |
| Size of Imaging Volume | | Max. 210 x 190 mm | Max. 240 x 210 mm | Xmaru1404CF: Max. 10 x 8.5 mm |
| Pixel<br>Resolution | Computed<br>Tomography | 2.5 lp/mm -4x4 binning | 5 lp/mm -2x2 binning<br>2.5 lp/mm -4x4 binning | Xmaru1404CF:<br>- 5.0 lp/mm - 2x2 binning<br>- 2.5 lp/mm - 4x4 binning |
| | Panoramic | 5 lp/mm -2x2 binning | N/A | Xmaru1404CF:<br>- 5.0 lp/mm - 2x2 binning<br>- 2.5 lp/mm - 4x4 binning |
| | Cephalometric | 2.5 lp/mm -4x4 binning | N/A | Xmaru2301CF: 5 lp/mm<br>1210SGA: 3.9 lp/mm<br>910SGA: 3.9 lp/mm<br>Xmaru2301CF-O: 5 lp/mm |
| | | | | |
| Pixel Size | Computed<br>Tomography | 198 $\mu$ m -4x4 binning | 99 $\mu$ m-2x2 binning<br>198 $\mu$ m-4x4 binning | Xmaru1404CF:<br>- 99 $\mu$ m- 2x2 binning<br>- 198 $\mu$ m- 4x4 binning |
| | Panoramic | 99 $\mu$ m -2X2 binning | N/A | Xmaru1404CF:<br>- 99 $\mu$ m- 2x2 binning<br>- 198 $\mu$ m- 4x4 binning |
| | Cephalometric | 198 $\mu$ m -4x4 binning | N/A | Xmaru2301CF: 100 x 100 $\mu$ m<br>1210SGA: 127 x 127 $\mu$ m<br>910SGA: 127 x 127 $\mu$ m<br>Xmaru2301CF-O: 100 x 100 $\mu$ m |
# 9. Comparison of Technological characteristics with the predicate device
______________________________________________________________________________________________________________________________________________________________________________
______________________________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
## VATECH Co., Ltd. Premarket Notification 510(k)
{9}------------------------------------------------
## 10. Performance Data
## Summary of Performance Testing
The PaX-i3D Green Premium (Model: PCT-90LH) digital X-ray system described in this 510(k) is similar to the predicate device, PaX-i3D Ortho (TON-95LH) in terms of indications for use, materials, safety characteristics, and X-ray source.
Furthermore, the following information further substantial equivalence between the subject device and predicate device:
The fundamental technological characteristics of the subject and predicate device are similar.
The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the predicated device for the purpose of comparing the performance between the predicate device. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.
The differences are as follows.
- The subject device is equipped with Xmaru3104CF, a SSXI detector, which is used to capture PANO, i-। CEPH, CT (Dental / ENT) images.
- । The subject device utilizes iterative reconstruction algorithm, a CBCT imaging processing algorithm.
- -In addition to th…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.