The Hyperion X5 device, 3D version machine, is an image reproducer suitable for expert professionals, which allows to obtain dental images in a simple and automatic way. The image is acquired by means of an X-ray detector and a constant potential X-ray source, powered by a high-voltage and high-frequency generator. The image is then transmitted to a computer in real time. The Hyperion X5 device, 3D version machine, intended to: I. produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity; II. produce tomographic images of the oral cavity and maxillofacial structures and carry out diagnostic examination on teeth, dental arches, structures of the oral cavity and some cranial bones. The system performs tomographic exams with the acquisition of X-ray images through a rotating sequence and the reconstruction of a three-dimensional matrix of the examined volume, thus producing two- and three-dimensional views of the volume itself. This technique is known as CBCT.
Device Story
Dental radiographic imaging system; performs panoramic and cone beam computed tomography (CBCT) scans. Input: X-ray signals acquired via rotating sequence; constant potential X-ray source; high-voltage/high-frequency generator. Processing: X-ray detector captures data; transmitted to computer in real-time; reconstruction of 3D matrix from volumetric data. Output: 2D orthopanoramic and 3D tomographic images. Usage: Clinical setting; operated by dental professionals. Healthcare providers use images for diagnostic examination of teeth, dental arches, maxillofacial structures, and cranial bones. Benefits: Enables visualization of oral anatomy for improved diagnostic accuracy and treatment planning.
Clinical Evidence
Bench testing included safety/EMC (IEC 60601 series), image resolution, geometric performance, and extreme exposure testing using anthropomorphic and geometric phantoms. 3D performance (MTF, NPS) compared against Hyperion X9 using Catphan 500 phantom. Clinical evidence included qualitative evaluation of images for pediatric, edentulous, third molar, and endodontic cases, confirming system functionality.
Technological Characteristics
Dental X-ray system; 2D panoramic and 3D CBCT modes. X-ray source: 60-85 kV (2D), 60-90 kV (3D). Detector: TFT with direct deposition CsI scintillator. Connectivity: Giga Ethernet to PC. Software: NNT/IRYS for reconstruction/viewing. Positioning: 3 laser pointers. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 62366, 60601-2-63, 60825-1, 62304.
Indications for Use
Indicated for expert professionals to produce orthopanoramic and tomographic (CBCT) images of the maxillofacial region, teeth, dental arches, oral cavity structures, and cranial bones in adult and pediatric patients.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2016
CEFLA S.C. % Mr. Maurizio Pantaleoni CEO ISEMED srl Via A. Altobelli Bonetti 3/A Imola, BO 40026 ITALY
Re: K161900
Trade/Device Name: Hyperion X5 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS, MUH Dated: October 21, 2016 Received: October 28, 2016
Dear Mr. Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
#### 510(k) Number (if known)
## K161900
Device Name Hyperion X5
Indications for Use (Describe)
The Hyperion X5 device, 3D version machine, is an image reproducer suitable for expert professionals, which allows to obtain dental images in a simple and automatic way. The image is acquired by means of an X-ray detector and a constant potential X-ray source, powered by a high-yoltage and high-frequency generator. The image is then transmitted to a computer in real time.
The Hyperion X5 device, 3D version machine, intended to:
I. produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;
II. produce tomographic images of the oral cavity and maxillofacial structures and carry out diagnostic examination on teeth, dental arches, structures of the oral cavity and some cranial bones.
The system performs tomographic exams with the acquisition of X-ray images through a rotating sequence and the reconstruction of a three-dimensional matrix of the examined volume, thus producing two- and three-dimensional views of the volume itself. This technique is known as CBCT.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary of the HYPERION X5
This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92.
#### 1. General Information
| Submitter: | CEFLA S.C.<br>Via Selice Prov.le 23/a<br>Imola, BO40026ITALY<br>Tel. +390542 653441<br>Fax +39 0542 653607 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Consultant/ Contact: | Maurizio Pantaleoni<br>ISEMED srl<br>Via Altobelli Bonetti 3/a<br>Imola, BO 40026 ITALY<br>Mob. +393484435155<br>Tel. +390542683803<br>Fax+390542698456<br>Email: regulatory@isemed.eu |
Date 510K summary prepared: July 1, 2016
## 2. Names
| Device Name: | HYPERION X5 |
|------------------|-----------------------------------------|
| Common Name: | Dental Computed Tomography X-ray System |
| Regulation Name: | Computed Tomography X-ray System |
| Product Code: | OAS/MUH |
| Classification: | 21CFR 892.1750, Class II |
## 3. Predicate Devices
The legally marketed device to which substantial equivalence is being claimed is as follows:
| Applicant | Device name | Premarket Notification<br>Number | Product code |
|------------|-------------|----------------------------------|--------------|
| CEFLA S.C. | HYPERION X9 | K123381 | OAS/ MUH |
| CEFLA S.C. | HYPERION X5 | K152162 | MUH |
{4}------------------------------------------------
## 4. Device Description
The subject device Hyperion X5 is a dental radiographic imaging system, that consists of two image acquisition modes: panoramic and cone beam computed tomography (CBCT). Specifically designed for dental radiography of the teeth, jaws and oral structures, the subject device:
produces orthopanoramic images of the maxillofacial region and carry out diagnostic l. examination on teeth, dental arches and other structures in the oral cavity;
produces tomographic images of the oral cavity and maxillofacial structures and carry out II. diagnostic examination on teeth, dental arches, structures of the oral cavity and some cranial bones. Hyperion X5 has been developed according to FDA-Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices".
The subject device can be sold under different brands and commercial names for commercial needs, without changing any of the safety, electrical and functional features. The variants are:
| # | Product Name | Brand | Manufacturer | REF – TYPE |
|---|------------------|------------|--------------|------------|
| 1 | Hyperion X5 | Myray | Cefla s.c. | 708H |
| 2 | NewTom GO | NewTom | Cefla s.c. | 708K |
| 3 | X-Radius Compact | Castellini | Cefla s.c. | 708N |
## 5. Indications for Use
The Hyperion X5 device, 3D version machine, is an image reproducer suitable for expert professionals, which allows to obtain dental images in a simple and automatic way. The image is acquired by means of an X-ray detector and a constant potential X-ray source, powered by a highvoltage and high-frequency generator. The image is then transmitted to a computer in real time.
The Hyperion X5 device, 3D version machine, intended to:
- I. produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;
- II. produce tomographic images of the oral cavity and maxillofacial structures and carry out diagnostic examination on teeth, dental arches, structures of the oral cavity and some cranial bones.
The system performs tomographic exams with the acquisition of X-ray images through a rotating sequence and the reconstruction of a three-dimensional matrix of the examined volume, thus producing two- and three-dimensional views of the volume itself. This technique is known as CBCT.
## 6. Comparison of technological characteristics with the predicate device
The subject device Hyperion X5 allows 2D as well as 3D images representing a further development of the cleared Hyperion X5 (K152162) which allows 2D images only.
For panoramicx-ray exposures (also referred to as "2D mode"), the subject device provides similar technological characteristic, operating principles, components and programs of the previous version Hyperion X5 (K152162). For CBCT image acquisition and 3D processing (also referred to as "3D
{5}------------------------------------------------
| mode"), the subject device provides the same operating principles and similar technology of |
|---------------------------------------------------------------------------------------------|
| Hyperion X9 Computed Tomography X-ray System (K123381). In particular: |
| | Characteristics | Subject device | HYPERION X5<br>K152162 | HYPERION X9<br>K123381 |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Indication for use statement | The Hyperion X5 device, 3D version machine, intended to:<br>I. <b>produce orthopanoramic images</b> of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;<br>II. <b>produce tomographic images</b> of the oral cavity and maxillofacial structures and carry out diagnostic examination on teeth, dental arches, structures of the oral cavity and some cranial bones.<br>The system performs tomographic exams with the acquisition of X-ray images through a rotating sequence and the reconstruction of a three-dimensional matrix of the examined volume, thus producing two- and three-dimensional views of the volume itself. This technique is known as CBCT | HYPERION X5 is an extraoral X-ray system for digital panoramic X-rays suitable for:<br>(i) production of <b>orthopanoramic images</b> of the maxillofacial region, diagnostic examination of the dentition (teeth), arches and other structures of the oral cavity. | HYPERION X9 is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, indicated for use in:<br>(i) producing panoramic X-ray images of the maxillofacial area, for diagnostic examination of dentition (teeth), jaws and oral structures; and<br>(ii) producing tomographic images of the oral and maxillofacial structure, for diagnostic examination of dentition (teeth), jaws ,oral structures and some cranial bones if equipped with CBCT option.<br>(iii) producing radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, when equipped with tele-radiographic arm (CEPH); | |
| Performance features | | | | |
| Performance<br>Specification | | • Panoramic<br>• Computed tomography | • Panoramic<br>• Computed tomography<br>• Cephalometric | |
| Patient population | | Adult<br>Paediatric | Adult<br>Paediatric | |
| Exposition selectable | | 2D : PAN, DENT, SIN, TMJ, PAN BITEWING<br><br>3D: Computed tomography (CBCT) | 2D: PAN DENT, , SIN, TMJ,PAN BITEWING, CEPHALOMETRIC<br><br>3D: Computed tomography (CBCT) | |
| Patient type selectable | | Male<br>Female<br>Child | Male<br>Female<br>Child | |
| Patient size selectable | | Male: Large, medium, small<br>Female: Large, medium, small<br>Child: Fixed | Male: Large, medium, small<br>Female: Large, medium, small<br>Child: Fixed | |
| Dose reduction for<br>child/adult male-<br>standard exposition | 0.7 | 0.7 | 0.6 | |
| Technical & Functional features comparison: (A) X-ray emission | | | | |
| TUBE VOLTAGE | • Panoramic: 60 - 85 kV, -<br>continuous emission<br>• CBCT: 60 - 90kV pulsed | Panoramic: 60 – 85 kV,<br>- continuous emission . | • Panoramic: 60 – 85 kV, -<br>continuous emission .<br>• CBCT: 60 - 90 kV pulsed | |
| TUBE CURRENT | 4 – 15 mA | 4 - 15 mA | 1 – 10 mA | |
| EXPOSURE TIME | • Panoramic: 1s - 15s<br>continuous radiation<br>• CBCT: 1,6- 5,6 effective<br>radiation time, | Panoramic: 1s - 15s<br>continuous radiation | • Panoramic. 1s - 14s<br>continuous exposure<br>CBCT: 3.6 s- 9 s effective<br>radiation time | |
| SHAPE OF X-RAY<br>BEAM | • Panoramic: Fan-shaped beam<br>• CBCT: cone beam | Panoramic: Fan-shaped<br>beam | • Panoramic: Fan-shaped<br>beam<br>• CBCT: cone beam | |
| FOCAL SPOT | • Panoramic: 0.5 mm<br>• CBCT: 0.6 mm | Panoramic: 0.5 mm | • Panoramic: 0.5 mm<br>• CBCT: 0.5 mm | |
| COLLIMATOR | One collimator, adjustable in<br>function of selected projection. | One collimator, fixed. | One primary collimator,<br>adjustable in function of<br>selected projection for CBCT<br>(One secondary collimator for<br>CEPH) | |
| FOV | Max: 10x10<br>Child: 8x7 | NA | Max 11x13<br>Child: 8x8: full dentition / 8x5:<br>whole dental arch | |
| Technical & Functional features comparison: (B) SSD Detector & IMAGE acquisition | | | | |
| IMAGE DETECTOR<br>TECHNOLOGY | • Panoramic: TFT detector<br>• CBCT: TFT detector | Panoramic: CMOS<br>detector | • Panoramic: CCD detector<br>• CBCT: TFT detector | |
| SCINTILLATOR<br>MATERIAL | Direct deposition Csl | Direct deposition Csl | Direct deposition Csl | |
| IMAGE DETECTORS<br>DIMENSION | • Panoramic: 6 x 146 mm<br>• CBCT: 146 x 146 mm | Panoramic: 6 x 151 mm | • Panoramic: 6 x 146 mm<br>• CBCT: 130x130 mm | |
| PIXEL SIZE | • 127 μm² | Panoramic: 100 μm² | • Panoramic: 96 μm² (2x2<br>binning mode)<br>• CBCT: 127 μm² | |
| MTF | 58% @ 1lp/mm (1x1) | 58% @ 1lp/mm | 60% @ 1lp/mm (1x1) | |
| DQE | 70% @ 0lp/mm (1x1) | 70% @0lp/mm | 80% @ 0lp/mm (1x1) | |
| SOURCE TO IMAGE<br>DETECTOR DISTANCE<br>(SID) | • Panoramic: 500 mm<br>• CBCT: 500 mm | Panoramic: 500 mm | • Panoramic: 550mm:<br>• CBCT: 660 mm | |
| ACQUISITION PATH | 2 axis - 210° | 2 axis - 210° | 3 axis - 360° | |
| IMAGE TRANSFER TO<br>PC | Direct Giga ethernet | Direct Giga Ethernet | Direct-Giga Ethernet | |
| Technical & Functional features comparison: (C) Laser & positioning | | | | |
| NUMBER OF LASER<br>POINTER | 3 | 3 | 4 | |
| LASER OPTICAL<br>CLASS | Class 1 for IEC 60825-1: 2003 | Class 1 for IEC 60825-1: 2003 | Class 1 for IEC 60825-1: 2003 | |
| NUMBER OF POINT<br>OF CEPHALOSTAT | 3 (adjustable along X-axis) | 2 (fixed) | 3 (fixed) | |
| Technical & Functional features comparison: (D) Control & Viewing Software | | | | |
| CONTROL SW | Firmware + VKB (on PC or Tablet) | Firmware + VKB (on PC) | Firmware | |
| VIEWING &<br>RECONSTRUCTION<br>SOFTWARE | NNT / IRYS | NNT / iRYS (optional) | NNT / IRYS | |
{6}------------------------------------------------
{7}------------------------------------------------
According to table above, the subject device has similar technology and features than the predicate devices K123381 and K152162, pointing out only few differences that have been addressed by dedicated performance tests demonstrating that the subject device is able to produce images with comparable performances.
## 7. Performance Data
The following tests were performed for determination of substantial equivalence:
- 1) Non clinical tests performed on the subject device:
- A. Safety and EMC test in compliance with:
- IEC 60601-1
- · IEC 60601-1-2,
- · IEC 60601-1-3,
- ●IEC 60601-1-6
- IEC 62366,
- · IEC 60601-2-63,
- IEC 60825-1
- · IEC62304 (software)
Obtained results demonstrate compliance to the standards listed.
- B. Images resolution (panoramic X-rays): The subject device (in 2D mode) was tested in comparison with the predicate K152162 using a specific QUART phantom. Obtained results demonstrated substantial equivalence in images resolution.
- C. Geometrical performance: The subject device (in 2D mode) was tested in comparison with the predicate K152162 using a specific geometric phantom. Obtained results demonstrated substantial equivalence in geometrical performance.
- D. Performance in extreme expositions (under and over exposition): The subject device (in 2D mode) was tested in comparison with the predicate K152162 using an anthropomorphic phantom. Obtained results demonstrated substantial equivalence in performance in extreme conditions (under/over exposition).
- E. 3D Performance evaluation of the subject device in comparison with Hyperion X9 (K123381) has been developed, in order to asses 3D MTF and 3D NPS parameters,
{8}------------------------------------------------
comparing them with the same parameters of the predicate device Hyperion X9 (K123381). A number of axial images extracted by the volumetric reconstructions of phantom Catphan® 500, for both devices was acquired with different scan modalities in order to have broader information. The two devices give substantially equivalent results in terms of spatial resolution and noise power spectrum. This assures that the diagnostic images extracted from volumes acquired with subject device hyperion X5 are of good quality and effective in comparison to the predicate device.
## 2) Clinical tests performed on the subject device:
Images resolution and quality during in CBCT acquisition (3D configuration) - The subject device was tested in comparison with the predicate K123381 performing :
- a quantitative evaluation by using a specific phantom as clinical case and measuring ● the resolution and noise.
- . a qualitative evaluation of images obtained with the two devices focusing on relevant clinical conditions: pediatric patient, Edentulous, third molar, Upper arch endodontics.
Clinical images were provided for the 2d mode of operation; these images were not necessary to establish substantial equivalence based on the modifications to the device (note digital image detector that is equivalent to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended."
## 8. Conclusions
In light of evidence summarized above and based on classification, intended use, technological characteristics, and performance data, the subject device may be found substantial equivalent to the predicate devices K152162 and K123381.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.