ECGxxx(z)(+) is a high-performance, multi-channel, interpretative resting electrocardiograph. The ECG signal is acquired with a 10-wires patient cable and is displayed in real time on a LCD screen integrated in the device. The electrocardiograph can analyse and store the ECG traces, send them to an external peripheral via network or via USB, print the 12 lead ECG in automatic or manual mode by means of a thermal printer. ECGxxx(z)(+) is intended for assessment and diagnosis of cardiac functions. In any case the results of analysis performed by the electrocardiograph must be validated by a Physician. ECGxxx(z)(+) is intended for use in hospitals, in medical clinics and doctor's offices of any size. - The device is indicated for use to acquire, analyse, display and print electrocardiograms. - The device is intended to provide the physician with an automatic interpretation of the ECG to be reviewed by a physician. - The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. - The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data. - The device is indicated for use on adult and pediatric populations. - The device is not intended to be used as a vital signs physiological monitor.
Device Story
ECG100+ and ECG200+ are 12-lead resting electrocardiographs. Input: 12-lead ECG signals via 10-wire patient cable. Processing: Real-time display on integrated LCD; optional Glasgow resting ECG interpretation algorithm (age/sex/race-specific criteria) provides diagnostic messages. Output: Printed ECG reports (thermal printer), stored traces (up to 1000), or digital export (USB/LAN/DICOM/HL7). Used in hospitals/clinics by physicians or trained personnel. Output supports physician diagnosis; requires mandatory physician overread. Benefits: Automated interpretation and standardized reporting for cardiac function assessment.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 60601-2-25 (performance), and EN 60601-1-2 (EMC) verified. Performance testing included defibrillation protection, input impedance, CMRR, filter response, and pacemaker detection.
Technological Characteristics
12-lead resting ECG; 24-bit A/D conversion; 32 kHz sampling; 0.05-150 Hz bandwidth; >100 dB CMRR. Features: 4.3" or 7" color LCD, thermal printer (8 dot/mm), mechanical keypad. Connectivity: USB, LAN (optional). Power: AC/battery (NiMH). Standards: IEC 60601-1, IEC 60601-2-25, AAMI/ANSI/AHA/ACC.
Indications for Use
Indicated for adult and pediatric patients for the acquisition, analysis, display, and printing of 12-lead resting electrocardiograms in clinical settings (hospitals, clinics, doctor's offices). Not intended as a vital signs monitor or sole diagnostic tool; requires physician overread.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
DEPARTMENT
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 12, 2016
Cardioline S.p.A Mr. Alessandro Peluso Official Correspondent Via Prati 1/2 Zola Predosa-localita Ponte Ronca-bologna, 40069 IT
Re: K160840 Trade/Device Name: ECG100+, ECG200+ Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: March 11, 2016 Received: April 6, 2016
Dear Mr. Alessandro Peluso.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160840
Device Name ECG100+, ECG200+
### Indications for Use (Describe)
ECGxxx(z)(+) is a high-performance, multi-channel, interpretative resting electrocardiograph.
The ECG signal is acquired with a 10-wires patient cable and is displayed in real time on a LCD screen integrated in the device. The electrocardiograph can analyse and store the ECG traces, send them to an external peripheral via network or via USB, print the 12 lead ECG in automatic or manual mode by means of a thermal printer.
ECGxxx(z)(+) is intended for assessment and diagnosis of cardiac functions. In any case the results of analysis performed by the electrocardiograph must be validated by a Physician.
ECGxxx(z)(+) is intended for use in hospitals, in medical clinics and doctor's offices of any size.
- The device is indicated for use to acquire, analyse, display and print electrocardiograms.
- The device is intended to provide the physician with an automatic interpretation of the ECG to be reviewed by a physician.
- The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
- The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data.
- The device is indicated for use on adult and pediatric populations.
- The device is not intended to be used as a vital signs physiological monitor.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K160840
#### 510(k) SUMMARY ECGxxx(z)(+)
### 1. SUBMITTER
CARDIOLINE S.p.A Via Linz nr. 19-20-21 38121 Spini di Gardolo (TN), Italy
T +39 0463 850125 F +39 0463 850088
Contact Person: Mr Emanuele Ercoli Date preparated: September 15, 2015
# 2. DEVICE
Name of Device: Cardioline ECGxxx(z)(+) Common or Usual Name: ECGxxx(z)(+) Electrocardiograph Model name: ECG100+, ECG200+ Classification Name: Electrocardiograph Regulatory Class: II Product Code: CFR 870.2340 Electrocardiograph, DPS
#### 3. PREDICATE DEVICE
| Manufacturer name | Applicant Name | Predicate Device | 510(k) Number |
|-------------------|-------------------|--------------------------|---------------|
| Cardioline S.p.A. | Cardioline S.p.A. | ET MEDICAL DEVICES SPA | K051534 |
| Cardioline S.p.A. | Cardioline S.p.A. | MORTARA INSTRUMENTS INC. | K101403 |
#### 4. DEVICE DESCRIPTION
ECGxxx(z)(+) is a family of high-performance, multi-channel, interpretative resting electrocardiograph. ECG100+ and ECG200+ are two models of that electrocardiographs family.
The device is a 12-lead diagnostic electrocardiograph which displays, acquires, prints and stores EGG tracings for adults and children. It also calculates the main overall ECG parameters.
The device can be supplied with the optional 12-lead Glasgow resting ECG interpretation algorithm, with specific criteria by age, sex and race. If this option is enabled, the algorithm can provide an over-reading physician with a second opinion generating diagnostic messages in the ECG report.
For further information on the resting ECG interpretation algorithm, see the Guidance for the physician on the application on adults and children (see accessories list)
The device can be configured with a larger memory, with bidirectional connectivity (LAN) and with DICOM® functionality.
The device can be powered by battery or the electrical mains.
{4}------------------------------------------------
The printing formats supported include: standard or Cabrera 3, 3+1, 3+3, 6 or 12 channels in automatic mode and 3, 6 or 12 (only for ECG200+) channels rhythm strip printing.
The ECG signal is acquired with a 10-wires patient cable and is displayed in real time on a LCD screen integrated in the device. The electrocardiograph can analyze and store the ECG traces, send them to an external peripheral via network or via USB, print the 12 lead ECG in automatic or manual mode by means of a thermal printer.
ECGxxx(z)(+) is intended for control and diagnosis of cardiac functions. In any case the results of analysis performed by the electrocardiograph must be validated by a Cardiologist.
ECGxxx(z)(+) is intended for use in hospitals, in medical clinics and offices of any size.
ECGxxx(z)(+) is a family of electrocardiographs which are a 12-lead diagnostic electrocardiographs and which are intended to displays, acquires, prints and stores ECG tracings for adults and children. It also calculates the main overall ECG parameters.
The devices have the following characteristics:
- mains and internal battery operation
- manual and automatic acquisition of the 12 Standards Leads
- simultaneous acquisition of the 12 Standards Leads
- internal storage up to 100 ECGs. Can storage up to 1000 ECGs optionally
- multichannel ECG printout on thermal paper:
- a) (for ECG100+) 3, 6 channels, 5/10/25/50 mm/s
- b) (for ECG200+) 3, 6 or 12 channels, 5/10/25/50 mm/s
- high resolution thermal printer:
- a) (for ECG100+) 8 dot/mm 108mm; Z-fold 100x150mm
- b) (for ECG200+) 8 dot/mm 216mm; A4 Z-fold
- for autoprint mode:
- a) (for ECG100+) Standard or Cabrera; 3, 3+1, 6 channels Patient Demographic, Global Measurements, Optional Interpretation (Glasgow University - Prof. MacFarlane) Adult, Paediatric, STEMI
- b) (for ECG200+) Standard or Cabrera; 3, 3+1, 6, 12 channels Patient Demographic, Global Measurements, Optional Interpretation (Glasgow University - Prof. MacFarlane) Adult, Pediatric, STEMI
- display:
- a) (for ECG100+) Backlit, colour LCD display, 4.3" ECG waveform real-time tracing
- b) (for ECG200+) 7" backlit LCD colour display, displays the ECG waves in real time
- Keyboard (for both devices): Mechanical keypad with alphanumeric keys and special function keys -
- filters (for both devices): diagnostic fully digital high pass filter; adaptive digital AC interference filter (50/60 Hz); digital low pass filter muscular filter 25 and 40 Hz (only for display and printing)
- connectivity (for both devices): USB device and LAN (optional)
- patient cable (for both devices): standard 15D, 10-wires
- -Data export (for both devices): SCP (standard format), XML-PDF-GDT (included in standard connectivity option), DICOM (included in DICOM connectivity option), HL7 (optional).
{5}------------------------------------------------
More specifically, the equipment family is based on two model variants characterized by different print and display capabilities.
Both devices offers full ECG acquisition meeting the standards used in clinical and diagnostic applications (AAMI, ANSI, AHA, ACC). When battery powered both devices have a duration of more than 500 ECGs and the recharging time is 4 hours to 85% of full charge.
The package includes (for both devices):
- 1. Patient cable
- 2. AC Power supply (100-240 VAC 50/60 Hz)
- 3. Paper
- 4. Pack of electrodes
- 5. Banana/clip adapter set
- 6. Guidance for the physician on the application on adults and children (with interpretative key)
- 7. User manual
The common family name is ECGxxx(z)(+). Where:
xxx = printer size
- + = model with network connectivity
- z = models with different interfaces
(refer to Technical File for more details)
The results of the analysis must always be validated by qualified, trained medical personnel and the devices are intended for use in a medical environment. ECG100+ and ECG200+ are intended to be used on adult and all pediatric patients. The devices must be handled with care by taking all the necessary precautions in order to prevent and avoid shocks, vibrations, heat sources, liquids and anything else that may damage it.
{6}------------------------------------------------
# 5. INDICATION FOR USE
ECGxxx(z)(+) is a high-performance, multi-channel, interpretative resting electrocardiograph.
The ECG signal is acquired with a 10-wires patient cable and is displayed in real time on a LCD screen integrated in the device. The electrocardiograph can analyse and store the ECG traces, send them to an external peripheral via network or via USB, print the 12 lead ECG in automatic or manual mode by means of a thermal printer.
ECGxxx(z)(+) is intended for assessment and diagnosis of cardiac functions. In any case the results of analysis performed by the electrocardiograph must be validated by a Physician.
ECGxxx(z)(+) is intended for use in hospitals, in medical clinics and doctor's offices of any size.
- -The device is indicated for use to acquire, analyse, display and print electrocardiograms.
- । The device is intended to provide the physician with an automatic interpretation of the ECG to be reviewed by a physician.
- -The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
- -The interpretations of ECG offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.
- -The device is indicated for use on adult and pediatric populations.
- -The device is not intended to be used as a vital signs physiological monitor.
{7}------------------------------------------------
# 6. TABULAR COMPARISON WITH PREDICATE DEVICES
For the comparison below, are present two tables:
TABLE 1: the ECG100+ is compared with AR 1200 of ET MEDICAL and ELI 250 of MORTARA INSTRUMENTS INC.
TABLE 2: the ECG200+ is compared with AR 2100 of ET MEDICA and ELI 250 of MORTARA INSTRUMENTS INC.
| FEATURES | CARDIOLINE ECG100+ | AR 1200 | ELI 250 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>use | ECGxxx(z)(+) is a high-<br>performance, multi-channel,<br>interpretative resting<br>electrocardiograph.<br>The ECG signal is acquired with a<br>10-wires patient cable and is<br>displayed in real time on a LCD<br>screen integrated in the device.<br>The electrocardiograph can<br>analyse and store the ECG traces,<br>send them to an external<br>peripheral via network or via USB,<br>print the 12 lead ECG in automatic<br>or manual mode by means of a<br>thermal printer.<br>ECGxxx(z)(+) is intended for<br>assessment and diagnosis of<br>cardiac functions. In any case the<br>results of analysis performed by<br>the electrocardiograph must be<br>validated by a Physician.<br>ECGxxx(z)(+) is intended for use in<br>hospitals, in medical clinics and<br>doctor's offices of any size.<br>- The device is indicated<br>for use to acquire,<br>analyse, display and<br>print<br>electrocardiograms.<br>- The device is intended to<br>provide the physician<br>with an automatic<br>interpretation of the<br>ECG to be reviewed by a<br>physician.<br>- The device is indicated<br>for use in a clinical<br>setting, by a physician or<br>by trained personnel<br>who are acting on the<br>orders of a licensed<br>physician. It is not<br>intended as a sole<br>means of diagnosis.<br>- The interpretations of<br>ECG offered by the<br>device are only<br>significant when used in<br>conjunction with a<br>physician over-read as | CARDIOLINE AR 600 and AR 1200 and<br>AR 2100 are a family of<br>electrocardiograph recorders<br>provided with a program for<br>automated ecg analysis and with a<br>graphic LCD display.<br>The equipments are intended for use<br>in routine ecg recording in physician<br>practice and/or hospital. The<br>electrical heart activity is detected by<br>means of two or more<br>electrocardiograph electrodes and is<br>recorded on thermal paper.<br>Intended use for non interpretive<br>applications covers the full range of<br>patient population with no<br>limitations with respect to age, sex<br>and race of the patient.<br>The interpretation program is<br>intended to provide a diagnostic<br>support to the physician for the ecg<br>evaluation on rhythm and<br>morphology.<br>Interpretation Statements must be<br>overviewed and approved by trained<br>Physician's. Interpretation<br>statements just represent a partial<br>qualitative and quantitative<br>information of the general patient<br>cardiovascular condition: no therapy<br>or drugs can be subministrated<br>based solely on Interpretation<br>statements.<br>The equipments are intended to be<br>used by trained medical personnel or<br>physician's.<br>Indication for use of the modified<br>device has not been changed with<br>respect to the predicate device AB<br>CARDIETTE DAEDALUS VIEW base<br>and Hes K002074. | - The device is indicated for<br>use to acquire, analyze,<br>display and print<br>electrocardiograms.<br>- The device is indicated for<br>use to provide interpretation<br>of the data for consideration<br>by a physician.<br>- The device is indicated for<br>use in a clinical setting, by a<br>physician or by trained<br>personnel who are acting on<br>the orders of a licensed<br>physician. It is not intended<br>as a sole means of diagnosis.<br>- The interpretations of ECG<br>offered by the device are<br>only significant when used in<br>conjunction with a physician<br>over-read as well as<br>consideration of all other<br>relevant patient data.<br>- The device is indicated for<br>use on adult and pediatric<br>populations.<br>- The device is not intended to<br>be used as a vital signs<br>physiological monitor |
| | well as consideration of<br>all other relevant patient<br>data.<br>- The device is indicated<br>for use on adult and<br>pediatric populations.<br>- The device is not<br>intended to be used as a<br>vital signs physiological<br>monitor. | | |
| Target<br>population | Adults and pediatric patients<br>Adults and pediatric patients for<br>analisys | Adults and pediatric patients<br>Adults for analisys | Adults and pediatric patients<br>Adults and pediatric patients for<br>analisys |
| Safety<br>standards | IEC 60601-1<br>IEC 60601-2-25<br>CB scheme | IEC 60601-1<br>IEC 60601-2-25 | IEC 60601-1<br>IEC 60601-2-25 |
| EMC<br>standards | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| ECG Leads | 12 Leads Standard / Cabrera | 12 Leads Standard / Cabrera | 12 Leads Standard / Cabrera |
| Sampling<br>Rate | 500 samples/second/channel | 1000 samples/s/channel printing and<br>filters<br>500 samples/s/channel in calculation<br>and filters | 1000 s/sec/channel used for<br>recording and analysis |
| Leads<br>Connector | Single block | Single block | Single block |
| Standard<br>Leads<br>Acquired | I, II, III, aVR, aVL, aVF, V1, V2, V3,<br>V4, V5, V6 | I, II, III, aVR, aVL, aVF, V1, V2, V3, V4,<br>V5, V6 | I, II, III, aVR, aVL, aVF, V1, V2, V3,<br>V4, V5, V6 |
| A/D<br>Conversion | 24 bit, 32 KHz | 12 bit | 20 bit |
| Bandwidth | 0.05 - 150 Hz | 0,05 - 150 Hz | 0.05Hz - 300Hz |
| CMRR | >100 dB | >100 dB | Not specified |
| Defibrillator<br>Protection | AAMI/IEC standards | AAMI/IEC standards with proprietary<br>applied part | Defibrillation proof applied part |
| Pacemaker<br>detection | Hardware detection coupled with<br>convolution digital filtering | Recognize pacemaker impulse<br>according to IEC applicable standards | Not specified |
| Patient Cable | 10 wire single connector | 10 wire single connector | 10 wire cable single connector |
| Mains Power<br>Supply | Medical grade AC power supply<br>(100-240 VAC, 50/60 Hz); | Internal power supply 90-264 VAC,<br>47 – 63 Hz | Universal AC power supply (100-<br>240 VAC at 50/60 Hz) 50 VA…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.