The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The MectaLIF Anterior Stand-Alone system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
Device Story
Anterior interbody fusion device; used for spinal stabilization in DDD patients. System comprises PEEK disc spacer (optionally titanium-coated), titanium alloy plate, and titanium bone screws. Spacer interior packed with autograft/autologous bone graft. Plate secured to spacer via interlocking mechanism; assembly fixed to vertebral body with screws. Provides stand-alone fixation; no supplementary fixation required. Used by surgeons in clinical settings to facilitate fusion; benefits patient by stabilizing spinal segment and reducing discogenic pain.
Clinical Evidence
Bench testing only. Performance testing conducted per ASTM F2077, ASTM F2267, ASTM F1877, and ISO 17853, including static/dynamic axial compression, static/dynamic compression-shear, static/dynamic torsion, subsidence, and expulsion testing.
Technological Characteristics
Materials: PEEK (ASTM F2026), CPTi coating (ASTM F1580), Ti6Al4V ELI (ISO 5832-3/ASTM F136). Components: Disc spacer, plate (Flush, Long, L5-S1, Hybrid), bone screws. Tantalum markers (ISO 13782/ASTM F560). Stand-alone fixation via interlocking plate-spacer mechanism. Sterilization method not specified.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade I spondylolisthesis or retrolisthesis. Patients must have had six months of non-operative treatment. May include patients with previous non-fusion spinal surgery at the involved level(s).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2016
Medacta International % Ms. Roshana Ahmed. RAC Mapi USA, Incorporated 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
Re: K160605
Trade/Device Name: MectaLIF Anterior Stand-Alone Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: June 15, 2016 Received: June 15, 2016
Dear Ms. Ahmed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160605
Device Name MectaLIF Anterior Stand-Alone
#### Indications for Use (Describe)
The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
The MectaLIF Anterior Stand-Alone system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in blue, with three blue triangles to the left of the word. There is a yellow line underneath the word "Medacta", and the word "International" is written in blue below the yellow line. To the right of the word "International" is a red square with a white cross in the center.
## 510(k) Summary
| | Applicant/Sponsor: Medacta International SA |
|-----------------|------------------------------------------------|
| | Strada Regina |
| | 6874 Castel San Pietro (CH) |
| | Switzerland |
| | Phone (+41) 91 696 60 60 |
| | Fax (+41) 91 696 60 66 |
| Contact Person: | Roshana Ahmed, M.A., RAC |
| | Mapi USA, Inc. |
| | 2343 Alexandria Drive, Suite 100 |
| | Lexington, KY 40504 |
| | Sr. Manager, Regulatory Affairs |
| | Phone: 657-248-7361 |
Date Prepared: June 19, 2016
### DEVICE INFORMATION
| Trade/Proprietary Name: | MectaLIF Anterior Stand-Alone |
|--------------------------|---------------------------------------------------------------|
| Common or Usual<br>Name: | Intervertebral body fusion device |
| Classification Name: | Intervertebral Fusion Device with Integrated Fixation, Lumbar |
| Product Code: | OVD |
| Regulation Number: | 888.3080 |
| Device Class: | 2 |
### PREDICATE DEVICE INFORMATION
Primary Predicate:
MectaLIF Anterior, K124034, Medacta International SA
Additional Predicates:
MectaLIF TiPEEK, K133192, Medacta International SA Mecta-C TiPEEK, K142744, Medacta International SA Vu aPOD, K101310, Theken Spine
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Medacta International. The logo consists of two blue triangles on the left, followed by the name "Medacta" in blue. Underneath the name is a yellow line, and below that is the word "International" in gray. To the right of the word "International" is a red square with a white cross in the center.
### DEVICE DESCRIPTION
The MectaLIF Anterior Stand-Alone is a stand-alone anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. MectaLIF Anterior Stand-Alone is intended to be used with bone screws provided and requires no additional supplementary fixation.
MectaLIF Anterior Stand-Alone consists of disc spacers made of PEEK Implant Grade Polyetheretherketone (ASTM F 2026) body. The spacer is available without any coatings and with a commercially pure titanium (CPTi, ASTM F 1580) coating. Both coated and uncoated spacers contain tantalum markers (ISO 13782 / ASTM F 560), and include bone screws made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136), and a plate made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136).
The interior of the disc spacer can be packed with autograft or autologous bone graft. The plate comes in four different designs (Flush, Long, L5-S1, and Hybrid) and is secured to the disc spacer via an interlocking mechanism. The disc spacer and attached plate are secured to the vertebral body with the bone screws. The Flush, Long, and Hybrid plates are used with four bone screws while the L5-S1 plate is used with three bone screws.
The purpose of this submission is to alter the indications for use, and introduce new plate and screw designs.
### INDICATIONS FOR USE
The MectaLIF Anterior Stand Alone system is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior Stand Alone can be packed with autograft or autologous bone graft.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
The MectaLIF Anterior Stand-Alone system is a system intended to be used with bone screws provided and requires no additional supplementary fixation. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Medacta International. The logo consists of three blue triangles arranged in a row, followed by the word "Medacta" in blue. Below the triangles and the word is a yellow line, and below that is the word "International" in gray, followed by a red and white Swiss flag.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The MectaLIF Anterior Stand Alone is identical to the identified predicate devices in terms of cage size footprint, plate sizes, lordosis, screw plate interface, screw sizes, materials of construction, and sterilization method.
#### PERFORMANCE TESTING
The following tests were conducted on the MectaLIF Anterior Stand-Alone in accordance with ASTM F2077, ASTM F2267, ASTM F1877, and ISO 17853:
- 1. Static Axial Compression
- 2. Dynamic Axial Compression
- 3. Static Compression-Shear
- Dynamic Compression-Shear 4.
- న. Static Torsion
- 6. Dynamic Torsion
- 7. Subsidence
- 8. Expulsion
#### CONCLUSION:
Based on the information presented above and within the submission, the MectaLIF Anterior Stand Alone is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.