Voluson S6, Voluson S8, Voluson S10

K160184 · GE Healthcare · IYN · Feb 25, 2016 · Radiology

Device Facts

Record IDK160184
Device NameVoluson S6, Voluson S8, Voluson S10
ApplicantGE Healthcare
Product CodeIYN · Radiology
Decision DateFeb 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (Obsterics); Abdominal (including renal and GYN/pelvic); Pediatic; Small Organ (breast, testes, thyroid, salvary gland, lymph nodes, pediatic and neonatal patients); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (TR); Transvaginal (TV).

Device Story

The Voluson S6, S8, and S10 are mobile, full-featured Track 3 ultrasound systems for general radiology and OB/GYN. They utilize various linear, curved linear, and matrix phased array transducers to acquire ultrasound signals, which are processed to produce real-time 2D, 3D, and 4D images. The system includes a console with a keyboard, touch panel, and display for clinician operation. It supports DICOM networking for image storage and reporting. New features include a SIM card for wireless MMS/SMS image/text transmission, the IOTA LR2 GYN measurement model, and SonoAVC (antral/general) software. The device aids clinicians in diagnostic imaging and clinical decision-making, potentially benefiting patients through improved diagnostic accuracy and guidance for procedures like biopsies.

Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with standards such as AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2, and NEMA UD 3.

Technological Characteristics

Mobile console with keyboard, touch panel, and display. Utilizes linear, curved linear, and matrix phased array transducers. Features include 3D/4D imaging, DICOM connectivity, and wireless (SIM card) communication. Complies with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/3, and ISO 10993-1. Beamformer: STB MLA4.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal/obstetrics, abdominal (renal, GYN/pelvic), pediatric, small organ (breast, testes, thyroid, salivary gland, lymph nodes), neonatal/adult cephalic, cardiac (adult/pediatric), peripheral vascular, musculoskeletal (conventional/superficial), transrectal, and transvaginal applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized representation of a human figure with three faces in profile, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 25, 2016 GE Healthcare % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226 Re: K160184 Trade/Device Name: Voluson S6, Voluson S8, Voluson S10 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 20, 2016 Received: January 27, 2016 Dear Ms. Ortiz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Ocks Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are four curved lines emanating from the circle, one at each of the cardinal directions. The logo is simple and recognizable, and it is associated with a well-known and established company. DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. 510(k) Number (if known) #### K160184 Device Name Voluson S6 / Voluson S8 / Voluson S10 Indications for Use (Describe) The device is a general-purpose ultrasound system. Specific clinical applications and exam types include: Fetal (Obsterics); Abdominal (including renal and GYN/pelvic); Pediatic; Small Organ (breast, testes, thyroid, salvary gland, lymph nodes, pediatic and neonatal patients); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (TR); Transvaginal (TV). Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FORM FDA 3881 (1/14) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative flourishes or embellishments around the circle. The logo is presented in a blue color. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a vintage or classic appearance. ### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br><i>Anatomy/Region of Interest</i> | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |-----------------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6,9] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6,9] | | Pediatric | P | P | P | P | P | P | P | P | P | P | [5,6,9] | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | [5,6,9] | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | [5] | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P | [5] | | Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [5] | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [5,6,9] | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6,9] | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [5,6,9] | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6,9] | | Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6,9] | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA [1] Abdominal includes renal, GYN/Pelvic Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (2D/3D/4D) [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with IC9-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | N | N | N | N | N | N | N | N | N | N | [ 6,9] | | | Abdominal[1] | N | N | N | N | N | N | N | N | N | N | [ 6,9] | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | N | N | N | N | N | N | N | N | N | N | [ 6,9] | | | Transvaginal[10] | N | N | N | N | N | N | N | N | N | N | [ 6,9] | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA [1] Abdominal includes renal, GYN/Pelvic Notes: - [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients - [3] Cardiac is Adult and Pediatric - [5] 3D/4D Imaging Mode - [6] Includes imaging of guidance of biopsy (2D/3D/4D) - [7] Includes infertility monitoring of follicle development - [8] Includes urology/prostate - [9] Elastography imaging- Elasticity - [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with 4C-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | | Pediatric | P | P | P | | P | P | P | P | P | P | [6] | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance. The color of the logo is a light blue. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with C1-5-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | | Pediatric | P | P | P | | P | P | P | P | P | P | [6] | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three swirls around the outside of the circle. The logo is simple and recognizable. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with 8C-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | | | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | | | | Pediatric | P | P | P | | P | P | P | P | P | P | | | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | | | | Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | P | P | P | | P | P | P | P | P | P | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with E8C-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6,9] | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6,9] | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | [6,9] | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | [6,9] | | Transvaginal | P | P | P | | P | P | P | P | P | P | [6,9] | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (2D/3D/4D) [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The letters and the border are in a blue color, while the background is white. The logo is simple, yet recognizable, and represents the brand identity of General Electric. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with with 12L-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | P | P | P | [6,9] | | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | [6,9] | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6,9] | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6,9] | | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [6,9] | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue. ### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with 9L-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | | Pediatric | P | P | P | | P | P | P | P | P | P | [6] | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | [6] | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [6] | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance. The color of the logo is a light blue. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with 3Sc-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [6] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [6] | | Pediatric | P | P | P | P | P | P | P | P | P | P | [6] | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P | [6] | | Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [6] | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. #### Diagnostic Ultrasound Indications for Use Form Voluson S6 / Voluson S8 / Voluson S10 with RIC5-9A-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | N | N | N | | N | N | N | N | N | N | [ 5,6,9] | | Abdominal[1] | N | N | N | | N | N | N | N | N | N | [ 5,6,9] | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | N | N | N | | N | N | N | N | N | N | [ 5,6,9] | | Transvaginal | N | N | N | | N | N | N | N | N | N | [ 5,6,9] | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes renal, GYN/Pelvic [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric. [5] 3D/4D Imaging Mode. [6] Includes imaging of guidance of biopsy (2D/3D/4D). [7] Includes infertility monitoring of follicle development. [8] Includes urology/prostate. [9] Elastography imaging- Elasticity [*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue. #### Diagnostic Ultrasound Indications for Use Form #### Voluson S6 / Voluson S8 / Voluson S10 with RAB6-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...