Voluson E6 / Voluson E8 / Voluson E10 Diagnostic Ultrasound Systems

K142472 · GE Healthcare · IYN · Feb 10, 2015 · Radiology

Device Facts

Record IDK142472
Device NameVoluson E6 / Voluson E8 / Voluson E10 Diagnostic Ultrasound Systems
ApplicantGE Healthcare
Product CodeIYN · Radiology
Decision DateFeb 10, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

Device Story

Voluson E6/E8/E10 are mobile diagnostic ultrasound systems for general radiology and OB/GYN. Systems use various linear, curved, and matrix phased array transducers (mechanical/electronic scanning) to acquire ultrasound signals. Data is processed to produce real-time 2D, 3D, and 4D images and fluid flow analysis. Operated by clinicians in clinical settings. Features include color LCD/TFT touch panels, DICOM networking, and image storage. Output is displayed for clinical review to support diagnostic decision-making. New features include wireless MMS/SMS image/text transmission via SIM card, expanded Fetal Cardio presets, and IOTA LR2 GYN measurement tool. Benefits include high-accuracy real-time imaging and comprehensive analysis for patient assessment.

Clinical Evidence

Bench testing only. Device evaluated for acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, electromagnetic, and mechanical safety. Conforms to AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 3, ISO 10993-1, NEMA UD 2, ISO 14971, and DICOM standards.

Technological Characteristics

Mobile console with keyboard and color LCD/TFT touch panel. Uses linear, curved linear, and matrix phased array transducers (mechanical/electronic scanning). Supports 2D, 3D, 4D, B, M, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. Connectivity via DICOM and wireless SIM card (MMS/SMS). Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, musculoskeletal, and peripheral vascular applications. Used for transvaginal and transrectal exams. Patient population includes adults, pediatric, and neonatal patients.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black and white, and the text is in uppercase letters. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 10, 2015 GE Healthcare % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226 Re: K142472 Trade/Device Name: Voluson E6 / Voluson E8 / Voluson E10 Diagnostic Ultrasound Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 13, 2015 Received: January 14, 2015 Dear Ms. Ortiz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. This determination of substantial equivalence applies to the following transducers intended for use with the Voluson E6 / Voluson E8 / Voluson E10. as described in your premarket notification: #### Transducer Model Number | RAB2-5-D | RIC5-9-D | RNA5-9-D | 4C-D | |----------|-----------|-----------|--------| | IC5-9-D | RSP6-16-D | RIC6-12-D | 9L-D | | P2D | P6D | 11L-D | C1-5-D | | ML6-15-D | RM6C | 3Sp-D | C4-8-D | | RAB6-D | eM6C | S4-10-D | | {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A Ochs Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. 510(k) Number (if known) K142472 Device Name Voluson E6 / Voluson E8 / Voluson E10 Diagnostic Ultrasound Systems Indications for Use (Describe) The device is a general purpose ultrasound system. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) - Over-The-Counter Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular shape. The color of the logo is a light blue. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular shape. The circle is made up of swirling lines, giving it a dynamic and flowing appearance. The logo is colored in a light blue hue. # Indications for Use Forms The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson E Series system and for all of its probe/mode combinations. Combinations identified as "P" represents those previously cleared with another GE Ultrasound system. This modification does not add to indications to the previously cleared system level or transducer indications or clinical applications. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The blue circle has a white border and a few white accents that give the impression of movement or energy around the letters. # Diagnostic Ultrasound Indications for Use Form GE Voluson E Series (Voluson E6 / Voluson E8 / Voluson E10) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color#<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |----------------------------------------------------|---|---|---------------|---------------|-------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------| | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Pediatric | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | [ 5 ] | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P | [ 5 ] | | Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [ 5 ] | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Transvaginal[10] | P | P | P | P | P | P | P | P | P | P | [ 5,6,9] | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [9] Elastography imaging- Elasticity [10] Transvaginal includes GYN [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD [*] 4D color Doppler (only with eM6C probe) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The blue circle has a white border and a few white accents that give the impression of movement or energy around the letters. ### Diagnostic Ultrasound Indications for Use Form ### GE Voluson E Series with RAB2-5-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [ 5,6] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [ 5,6] | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [ 5,6] | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (3D/4D) [7] Includes infertility monitoring of follicle development [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue. # Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with RIC5-9-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [ 5,6,9] | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | [ 5,6,9] | | | Transvaginal[10] | P | P | P | | P | P | P | P | P | P | [ 5,6,9] | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (3D/4D) [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [9] Elastography Imaging- Elasticity [10] Transvaginal includes GYN [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known company. ### Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with RNA5-9-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6] | | Pediatric | P | P | P | P | P | P | P | P | P | P | [5,6] | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | [5,6] | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | [5] | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [5] | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [5,6] | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [5,6] | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E [1] Abdominal is Neonatal and pediatric Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Neonatal and Pediatric [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (3D/4D) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance. ### Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with 4C-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [6] | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [6] | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [6] | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology [6] Includes imaging of guidance of biopsy (2D) [7] Includes infertility monitoring of follicle development [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The blue circle has a white border and a few white accents that give the impression of movement or energy around the letters. # Diagnostic Ultrasound Indications for Use Form ### GE Voluson E Series with IC5-9-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|--------------------|------------------|-------------|---------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined<br>Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [ 6,9] | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | [ 6,9] | | | Transvaginal[10] | P | P | P | | P | P | P | P | P | P | [ 6,9] | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E [6] Includes imaging of guidance of biopsy (2D) Notes: [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [9] Elastography Imaging- Elasticity [10] Transvaginal includes GYN [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. ### Diagnostic Ultrasound Indications for Use Form ### GE Voluson E Series with RSP6-16-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | P | P | P | [5,6] | | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | [5,6] | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [5,6] | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6] | | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [5,6] | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (3D/4D) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The blue circle has a white border and a few white accents that give the impression of motion or energy around the letters. # Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with RIC6-12-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |-------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | Anatomy/Region of Interest | | | | | | | | | | | | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] | | Abdominal[1] | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6] | | Transvaginal[10] | P | P | P | | P | P | P | P | P | P | [5,6] | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [5] 3D/4D Imaging Mode [6] Includes imaging of guidance of biopsy (3D/4D) [7] Includes infertility monitoring of follicle development [8] Includes urology/prostate [10] Transvaginal includes GYN [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known company. ### Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with 9L-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [6] | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | P | P | P | [6] | | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | [6] | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | | | | | | | | | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [6] | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [6] | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients Notes: [6] Includes imaging of guidance of biopsy (2D) [7] Includes infertility monitoring of follicle development [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known company. ### Diagnostic Ultrasound Indications for Use Form ### GE Voluson E Series with P2D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | P | | | | | | | | | Cardiac[3] | | | | P | | | | | | | | | Peripheral Vascular | | | | P | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [3] Cardiac is adult and Pediatric (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) {15}------------------------------------------------ Image /page/15/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known company. # Diagnostic Ultrasound Indications for Use Form #### GE Voluson E Series with P6D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac[3] | | | | P | | | | | | | | | | Peripheral Vascular | | | | P | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal | | | | | | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...