A retrospective clinical analysis was conducted to support the use of the NuVasive Interfixated Interbody System as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.
Retrospective clinical analysis; Multilevel degenerative scoliosis; Adjunct to fusion
Patients diagnosed with multilevel degenerative scoliosis
Not applicable for this study
Support for use as an adjunct to fusion in patients with multilevel degenerative scoliosis
Indications for Use
The NuVasive® Interfixated Interbody System is indicated for the following: CoRoent XL-F System: The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle or facet screws) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment. The CoRoent XL-F System is intended for use in interbody fusions in the lumbar spine from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent XL-F System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. Brigade System: The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis, and when used in these patients at multiple levels, the Brigade Standalone System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
Device Story
Intervertebral body fusion system; implants manufactured from PEEK-Optima LT-1; includes titanium alloy (Ti-6Al-4V) radiographic markers and integrated screws; features hollow core for autograft packing; includes surface spikes/teeth to resist migration. Used by surgeons in spinal fusion procedures; implanted in lumbar spine. Provides structural support and stabilization during fusion process; used with supplemental internal fixation (e.g., pedicle screws) for specific configurations. Benefits patients by facilitating fusion in degenerative disc disease and scoliosis cases.
Clinical Evidence
Retrospective clinical analysis performed to support use as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.
Technological Characteristics
Materials: PEEK-Optima LT-1 (ASTM F2026) and titanium alloy (ASTM F136). Form factor: Interbody implants with integrated screws, hollow core for graft, and surface spikes/teeth. No software or electrical equipment.
Indications for Use
Indicated for skeletally mature patients with symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels (L2-L5 for CoRoent XL-F; L2-S1 for Brigade) or multilevel degenerative scoliosis, following at least six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected lines or shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2016
NuVasive, Incorporated Ms. Michelle Cheung Associate, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K160051
Trade/Device Name: NuVasive® Interfixated Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: March 23, 2016 Received: March 23, 2016
Dear Ms. Cheung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
K160051
Page 1 of 1
510(k) Number (if known)
K160051
Device Name NuVasive® Interfixated Interbody System
### Indications for Use (Describe)
The NuVasive® Interfixated Interbody System is indicated for the following:
CoRoent XL-F System
The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle or facet screws) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.
The CoRoent XL-F System is intended for use in interbody fusions in the lumbar spine from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent XL-F System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.
### Brigade System
The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angels of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis, and when used in these patients at multiple levels, the Brigade Standalone System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape resembling a stylized leaf or eye, followed by the company name "NUVASIVE" in gray capital letters. Below the name, there is a tagline that says "Speed of Innovation" in a smaller font.
### 510(k) Summary
A.
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
Submitted by: Michelle Cheung, Esq. Associate, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-3360
Date Prepared: January 8, 2016
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® Interfixated Interbody System |
|----------------------------|-----------------------------------------|
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification Name: | Intervertebral Body Fusion Device |
| Device Class: | Class II |
|-----------------|-------------------|
| Classification: | 21 CFR § 888.3080 |
| Product Code: | OVD |
#### C. Predicate Devices
The subject NuVasive Interfixated Interbody System is substantially equivalent to primary predicate NuVasive CoRoent XL-F System (K140479), and additional predicates NuVasive CoRoent Single Tab System (K131723), CoRoent XLR Standalone System (K100043), Brigade Hyperlordotic System (K123045), and CoRoent Thoracolumbar System (K153419).
#### D. Device Description
The subject NuVasive Interfixated Interbody System are interbodies manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026 PEEK-Optima® LT-1, with radiographic markers manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The devices all contain integrated screws made of titanium alloy, and include a hollow core or graft aperture which allows for packing of autograft to help promote a solid fusion. The subject implants contain small spikes or teeth on each end of the device which serve to grip the adjacent vertebrae to resist migration and expulsion.
The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. In addition to the integrated screws, the CoRoent XL-F System and Brigade Hyperlordotic System devices are intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font, with the word "Speed" in purple and the rest of the phrase in gray. The logo is clean and modern, with a focus on innovation and technology.
#### E. Intended Use
The NuVasive® Interfixated Interbody System is indicated for the following:
### CoRoent XL-F System:
The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle or facet screws) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.
The CoRoent XL-F System is intended for use in interbody fusions in the lumbar spine from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent XL-F System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.
### Brigade System:
The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) with up to Grade I spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis, and when used in these patients at multiple levels, the Brigade Standalone System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
#### F. Technological Characteristics
As was established in this submission, the subject Interfixated Interbody System is substantially equivalent to predicate device systems already cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Nuvasive. The logo features a stylized purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray, block letters. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller, italicized font.
#### G. Performance Data
No new device designs for the CoRoent XL-F System and no new worst case sizes for the Brigade System are being introduced to the subject Interfixated Interbody System. Therefore, no performance testing was performed for this 510(k) submission.
A retrospective clinical analysis was performed to support the use of the subject device as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.
#### H. Conclusions
Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject Interfixated Interbody System has been shown to be substantially equivalent to legally marketed predicate devices, and as effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.