NuVasive CoRoent Thoracolumbar System

K153419 · Nu Vasive, Incorporated · MAX · Apr 13, 2016 · Orthopedic

Device Facts

Record IDK153419
Device NameNuVasive CoRoent Thoracolumbar System
ApplicantNu Vasive, Incorporated
Product CodeMAX · Orthopedic
Decision DateApr 13, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K153419 · Apr 13, 2016NuVasive CoRoent Thoracolumbar SystemNu Vasive, IncorporatedClinical literature reviewA clinical literature review was conducted to support the safety and effectiveness of the device when used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.Literature review; Multilevel degenerative scoliosis; Adjunct to fusion

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical literature review for multilevel degenerative scoliosis; Literature reviewPatients diagnosed with multilevel degenerative scoliosisNot applicable for this studySupport for use as an adjunct to fusion

Indications for Use

The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The CoRoent Thoracolumbar System (XL platform) implants are in interbody fusions at one or two contiguous levels in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain). The CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use at one or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent Thoracolumbar System (XL and L platforms) implants can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.

Device Story

Intervertebral body fusion device; used as adjunct to supplemental internal spinal fixation. Implants feature hollow core for autogenous bone graft packing; surface spikes/teeth provide fixation to adjacent vertebrae to resist migration. Available in various sizes and lordotic angles. Used by surgeons in clinical settings to treat degenerative disc disease and scoliosis. Benefits include stabilization of spinal segments to facilitate fusion.

Clinical Evidence

No new performance testing performed. Clinical literature review conducted to support use as an adjunct to fusion in patients with multilevel degenerative scoliosis.

Technological Characteristics

Materials: PEEK-Optima LT-1 (ASTM F2026) or titanium alloy Ti-6Al-4V (ASTM F136). Radiographic markers: titanium alloy (ASTM F136/F1472) or tantalum (ASTM 560/ISO 13782). Mechanical fixation via surface spikes/teeth. No software or electrical components.

Indications for Use

Indicated for skeletally mature patients with symptomatic disc degeneration (DDD), thoracic disc herniation, or DDD with up to Grade I spondylolisthesis, or multilevel degenerative scoliosis. Requires at least six months of failed non-operative treatment. Used for interbody fusion at one or two contiguous levels in thoracic (T1-T12), thoracolumbar (T12-L1), or lumbar (L1-S1) spine.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure, emphasizing the department's role and national scope. The logo is simple, yet conveys a sense of care and public service. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 13, 2016 NuVasive, Incorporated Ms. Michelle Cheung Associate, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121 Re: K153419 Trade/Device Name: NuVasive® CoRoent® Thoracolumbar System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, PHM Dated: March 18, 2016 Received: March 21, 2016 Dear Ms. Cheung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Vincent J. Devlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153419 Device Name NuVasive® CoRoent® Thoracolumbar System ### Indications for Use (Describe) The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The CoRoent Thoracolumbar System (XL platform) implants are in interbody fusions at one or two contiguous levels in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain). The CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use at one or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent Thoracolumbar System (XL and L platforms) implants can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. Type of Use (*Select one or both, as applicable*) | <div> <div> <b> X</b> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> </div> | |--------------------------------------------------------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------------------------------------------------------------------------| ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font and purple color. # 510(k) Summary A. In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided: Submitted by: Michelle Cheung, Esq. Associate, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-3360 Date Prepared: April 12, 2016 #### B. Device Name | Trade or Proprietary Name: | NuVasive® CoRoent® Thoracolumbar System | |----------------------------|----------------------------------------------| | Common or Usual Name: | Intervertebral Body Fusion Device | | Classification Name: | Spinal Intervertebral Body Fixation orthosis | | Device Class: | Class II | |-----------------|-------------------| | Classification: | 21 CFR § 888.3080 | | Product Code: | MAX, PHM | #### C. Predicate Devices The subject NuVasive CoRoent Thoracolumbar System is substantially equivalent to primary predicate CoRoent Lumbar System (K151472), and additional predicate CoRoent Thoracolumbar Implants (K150994). #### D. Device Description The subject NuVasive CoRoent Thoracolumbar System are interbodies manufactured from PEEK-Optima LT-1 conforming to ASTM F2026 or titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The PEEK devices contain titanium alloy radiographic markers conforming to ASTM F136 or ASTM F1472 or tantalum markers conforming to ASTM 560 or ISO 13782. The device's hollow core or graft aperture allows for packing of autograft to help promote a solid fusion. Small spikes or teeth on each end of the device serve to grip the adjacent vertebrae to resist migration and expulsion of the device. The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. #### E. Intended Use The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous {4}------------------------------------------------ Image /page/4/Picture/1 description: The image is a logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in purple. bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions at one or two contiguous levels in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain). The CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use at one or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The CoRoent Thoracolumbar System (XL and L platforms) implants can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. #### F. Technological Characteristics As was established in this submission, the subject CoRoent Thoracolumbar System is substantially equivalent to predicate device systems already cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment. #### G. Performance Data No new device designs and no new worst case sizes are being introduced to the subject CoRoent Thoracolumbar System. Therefore, no performance testing was performed for this 510(k) submission. A clinical literature review was performed to support the use of the subject device as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. #### H. Conclusions Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject CoRoent Thoracolumbar System has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...