K153395 · In2bones Sas · JDR · Feb 18, 2016 · Orthopedic
Device Facts
Record ID
K153395
Device Name
OS2-C Compression Staple
Applicant
In2bones Sas
Product Code
JDR · Orthopedic
Decision Date
Feb 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OS2®-C Compression Staples are indicated for fixation of arthrodesis, osteotomies and fractures in hand or foot surgery. The number and size of the OS2®-C Compression Staples must be adapted to the indication.
Device Story
OS2®-C Compression Staple is an osteosynthesis implant designed to provide constant, reproducible bone compression. It is manufactured from PEEK-OPTIMA® polymer (ASTM F2026). The device is supplied sterile via gamma irradiation, pre-assembled on an inserter for single-use application. It is intended for use in hospitals or outpatient surgery centers by surgeons. The staple is available in various interaxis lengths (11mm to 15mm) to accommodate specific anatomical requirements. During surgery, the clinician implants the device to stabilize bone segments, facilitating healing through compression. The device is designed for long-term implantation (>30 days).
Clinical Evidence
Bench testing only. Mechanical performance was assessed via comparison with predicate devices according to ASTM F564-06. Animal and clinical testing were deemed not applicable.
Technological Characteristics
Material: PEEK-OPTIMA® polymer (ASTM F2026). Design: Compression staple with various interaxis (11-15mm) and leg lengths. Energy source: None (mechanical). Sterilization: Gamma irradiation. Single-use. No coating.
Indications for Use
Indicated for fixation of arthrodesis, osteotomies, and fractures in hand or foot surgery for patients requiring bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Biotech International / EOS Eleos Memory Staple (K112794)
Smith and Nephew Bioraptor 2.3 PK suture anchor (K071586)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of an eagle or bird with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2016
In2Bones SAS % Dr. Norman Estrin, Ph.D. Managing Partner Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K153395
Trade/Device Name: OS2® -C Compression Staple Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: November 22, 2015 Received: November 24, 2015
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K153395
Device Name
OS2®-C Compression Staple
Indications for Use (Describe)
The OS2®-C Compression Staples are indicated for fixation of arthrodesis, osteotomies and fractures in hand or foot surgery.
The number and size of the OS2®-C Compression Staples must be adapted to the indication.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span>☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IN2BONES. The logo is in a gray oval shape with a white line that curves around the left side of the oval. The text "IN2BONES" is in white, with the "2" in orange.
# For In2Bones OS2®-C ComMARY
For In2Bones OS2®-C Compression Staple
| Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment registration<br>number | 3010470577 |
| Date of preparation | November 20, 2015 |
| Contact person | Norman F. Estrin. Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent in the<br>United States | Norman F. Estrin, Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | OS2®-C Compression Staple |
| Common name | Bone Compression staple |
| Device classification<br>regulation | 21 CFR 888.3030: Single/multiple component metallic bone fixation<br>appliances and accessories<br>Class II |
| Device Product<br>Code and Panel | JDR: staple, fixation, bone<br>87 orthopedics |
{4}------------------------------------------------
| Device Description | The OS2®-C Compression Staple is an osteosynthesis staple, enabling a<br>constant and reproducible compression.<br>The OS2®-C Compression Staple is available in multiple lengths and<br>interaxis. It is delivered sterile, ready to use, pre-assembled on its<br>inserter.<br>The implant is manufactured from PEEK-OPTIMA®, polymer from<br>Invibio®, and is designed for single use only.<br>Sizes:<br>The OS2®-C Compression Staple is available in various interaxis from<br>11mm to 15mm.<br>Material:<br>The OS2®-C Compression Staple is manufactured from PEEK<br>according to standard ASTM F2026.<br>It does not have any coating.<br>Single use:<br>The OS2®-C Compression Staple is designed for single use only.<br>Sterilization:<br>The OS2®-C Compression Staple is supplied sterile, using gamma<br>irradiation.<br>Place of use:<br>The OS2®-C Compression Staple is indicated for use in a hospital, or<br>outpatient surgery center where sterile field may be created and<br>maintained. |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Devices | Biotech International / EOS Eleos Memory Staple (K112794)<br>Integra / Newdeal UNI-CLIP (K011716)<br>Memometal Easyclip (K070031)<br>Smith and Nephew Bioraptor 2.3 PK suture anchor (K071586) - PEEK |
| Indications for use: | The OS2®-C Compression Staples are indicated for fixation of<br>arthrodesis, osteotomies and fractures in hand or foot surgery.<br>The number and size of the OS2®-C Compression Staples must be<br>adapted to the indication. |
| Comparison of the<br>indications for use with<br>the predicate devices: | As with the predicate devices, the OS2®-C Compression Staple is<br>indicated for surgical implantation longer than 30 days in the fixation<br>of bone fractures or for bone reconstruction in hand and foot. |
| Comparison of<br>Technological<br>characteristics | The technological characteristics of the OS2®-C Compression Staple<br>are the same as the characteristics of predicate devices in terms of<br>intended use and design. All these implants have the following features:<br>- Insertion into bone: The OS2®-C Compression Staple and all<br>predicate devices are intended for surgical implantation into<br>bone for longer than 30 days.<br>- Design: The OS2®-C Compression Staple has partially similar<br>design, when compared to the Integra / Newdeal UNI-CLIP |
| | (K011716), Memometal Easyclip (K070031), Biotech<br>International / EOS Eleos Memory Staple (K112794) and<br>Memometal Easyclip (K070031); all are compression staples<br>with various interaxis and leg lengths in the range. Any<br>differences of design from staple to staple do not affect the<br>Safety and Effectiveness of the OS2®-C Compression Staple.<br>Material: The OS2®-C Compression Staple are manufactured<br>from the same raw material when compared to the Smith and<br>Nephew Bioraptor 2.3 PK suture anchor (K071586) and meet<br>appropriate ASTM standard.<br>Equivalent size range: The OS2®-C Compression Staple has<br>similar range size when compared to the Biotech International<br>Eleos Memory Staple (K112794), Integra / Newdeal UNI-CLIP<br>(K011716) and Memometal Easyclip (K070031). |
| Substantial Equivalence<br>Summary | The OS2®-C Compression Staple intended use, design, material,<br>technological characteristics and principles of operation are<br>substantially equivalent to those of predicate devices, if applicable.<br>The OS2®-C Compression Staple has similar mechanical properties<br>when compared to the Biotech International Eleos Memory Staple<br>(K112794). |
| Summary Performance<br>Data | Performance assessment for the OS2®-C Compression Staple was made<br>through mechanical comparison with predicate devices, animal and<br>clinical testing being considered not applicable.<br>Mechanical testing for both OS2®-C Compression Staple and predicate<br>device was performed according to ASTM F564-06.<br>The OS2®-C Compression Staple is substantially equivalent to the<br>predicate devices identified in the 510(k) submission. |
| CONCLUSION | Based on the evaluations performed, the design and indications of<br>the OS2®-C Compression Staple are substantially equivalent to the<br>predicate devices identified in the 510(k) submission. No new<br>materials or processes are used in the development of this implant.<br><br>In addition, the results of the testing performed by the test<br>laboratory indicated that the implants performed as expected for<br>each test.<br><br>The OS2®-C Compression Staples are acceptable for the<br>application. |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.