K153210 · Bc Group International, Inc. · DRL · Dec 21, 2015 · Cardiovascular
Device Facts
Record ID
K153210
Device Name
Defibrillator Analyzer Variable Load
Applicant
Bc Group International, Inc.
Product Code
DRL · Cardiovascular
Decision Date
Dec 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5325
Device Class
Class 2
Indications for Use
The DA-2006-VL is used to determine that defibrillators are performing within their performance specifications by providing multiple loads of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms. The DA-2006-VL is used in conjunction with the DA-2006/DA-2006P.
Device Story
DA-2006-VL is a non-powered, selectable resistive load bank used for testing defibrillator performance; operates in conjunction with DA-2006/DA-2006P analyzer. Device provides specific resistive loads (25, 50, 75, 100, 125, 150, 175, 200 Ohms) via series/parallel resistor combinations to verify defibrillator output compliance with IEC 60601-2-4 and AAMI DF80 standards. Used by trained service technicians in clinical or service environments. Output allows technicians to confirm defibrillator energy delivery accuracy; ensures device safety and effectiveness during maintenance.
Clinical Evidence
Bench testing only. Verification and validation testing performed to ensure hardware specifications met. Compliance with IEC 61010-1 (safety requirements for electrical equipment) confirmed. No clinical data.
Technological Characteristics
Non-powered, manually selectable resistive load bank. Materials/construction identical to predicate. Dimensions: 248.9 x 205.7 x 120.7 mm. Weight: 2.27 kg. Operating temperature: 15-40 C. Complies with IEC 60601-2-4 and AAMI DF80. Exempt from EMC testing.
Indications for Use
Indicated for use by trained service technicians to test defibrillator performance against specifications by providing selectable resistive loads (25-200 Ohms).
Regulatory Classification
Identification
A defibrillator tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load. Some testers also provide waveform information.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird-like shape.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2015
BC Group International, Inc. % Mel Roche President 3081 Elm Point Industrial Drive Saint Charles, Missouri 63301
Re: K153210
Trade/Device Name: Defibrillator Analyzer Variable Load, Model DA-2006-VL Regulation Number: 21 CFR 870.5325 Regulation Name: Defibrillator Tester Regulatory Class: Class II Product Code: DRL Dated: July 17, 2015 Received: November 5, 2015
Dear Mel Roche:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153210
Device Name
DA-2006-VL Defibrillator Analyzer Variable Load
#### Indications for Use (Describe)
The DA-2006-VL is used to determine that defibrillators are performing within their performance specifications by providing multiple loads of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms. The DA-2006-VL is used in conjunction with the DA-2006/DA-2006P.
Type of Use (Select one or both, as applicable)
| <span></span> Research Use (Part 21 CFR 361. Subpart B) | <span></span> Over-The-Counter Use (21 CFR 331. Subpart C) |
|---------------------------------------------------------|------------------------------------------------------------|
|---------------------------------------------------------|------------------------------------------------------------|
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### The CDRH Premarket Review Submission Cover Sheet for the DA-2006-VL 5 Variable Load is contained in this section 510(k) Summary (As required by 21 CFR 807.92)
Date Prepared: February 1, 2015 Submitter's Information:
| Submitter's Name/<br>Address | BC Group International, Inc<br>3081 Elm Point Industrial Drive<br>St. Charles, MO 63301 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Owner/Operator | BC Group International, Inc<br>3081 Elm Point Industrial Drive<br>St. Charles, MO 63301 USA |
| Manufacturing Sites | BC Group International, Inc<br>3081 Elm Point Industrial Drive<br>St. Charles, MO 63301 USA<br>FDA Establishment Registration Number: 2939012 |
#### Submission Contact Information:
| Primary Contact | Name | Mel Roche |
|-------------------|----------------|-------------------------|
| | Title | President |
| | Telephone: | (314)-638-3800 |
| | Fax: | (314)-638-3200 |
| | Email Address: | mroche@bcgroupintl.com |
| Alternate Contact | Name | Lucio Simoni |
| | Title | Engineering Manager |
| | Telephone: | (314)-638-3800 |
| | Fax: | (314)-638-3200 |
| | Email Address: | lsimoni@bcgroupintl.com |
#### Proposed Device
Device Proprietary Name: Classification Name Class Product Code
DA-2006-VL Defibrillator Analyzer Variable Load Defibrillator Tester Variable Load Class II/21CFR 870.5325 DRL
## Predicate Devices
Fluke Impulse 7010 Defibrillator Selectable Loads (K083347)
{4}------------------------------------------------
#### Section 514 Compliance
Not applicable. Special Controls have not been established as of this date.
#### Indications for Use
The DA-2006-VL is used to determine that defibrillators are performing within their performance specifications by providing multiple loads of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms. The DA-2006-VL is used in conjunction with the DA-2006/DA-2006P.
## Description of Device
The BC Group International DA-2006-VL Variable Load is a precision instrument for testing defibrillators, and for ensuring that defibrillators comply with international standards IEC 60601-2-4 and AAMI DF80.
The DA-2006-VL is a non-powered, selectable resistive load bank. Through various series and parallel resistor combinations. the DA-2006-VL allows a user to select 25, 50, 75, 100, 125, 150, 175, and 200 Ohm loads.
The DA-2006-VL is used in conjunction with the DA-2006/DA-2006P Defibrillator Analyzer.
NOTE: The instrument is intended for use by trained service technicians.
## Comparison to Predicate Device
The DA-2006-VL Variable Load uses the same exact approach and DUT evaluation procedures as the predicate device.
#### Summary of Technical Characteristics
The DA-2006-VL Variable Load shares the same technical characteristics; design, materials, and composition as the predicate device.
{5}------------------------------------------------
| Table 2: Predicate Feature Comparison | | | |
|---------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device Feature or<br>Characteristic | Fluke Impulse 7010 | BC Biomedical<br>DA-2006-VL<br>Specifications |
| General Specifications | Maximum voltage | 5000 V | 5000 V |
| | Maximum Continuous<br>Power | 12 W, 10 defib pulses of<br>360 J every 5 minutes | 12 W, 10 defib pulses of 360<br>J every 5 minutes |
| | Inductance | < 2 μH, @25 Ω<br>< 3 μH, @50 Ω<br>< 4 μH, @75 Ω and 100 Ω<br>< 5 μH, @125 Ω<br>< 6 μH, @150 Ω<br>< 7 μH, @175 Ω<br>< 8 μH, @200 Ω | < 10 μH |
| | Electrical Specifications | Load Settings | 25, 50, 75, 100, 125, 150,<br>175, 200 Ohms, ±1% |
| Accuracy (50 Ω) | | ± 1% of reading + 0.1 J | High Range<br>± 2 % of reading for >100<br>Joules<br>± 2 Joules for ≤100 Joules<br>Low Range<br>± 2 % of reading for >20<br>Joules<br>± 0.4 Joules for ≤20 Joules |
# Summary of Technological Characteristics in Comparison with the Predicate Device
{6}------------------------------------------------
| Table 2: Predicate Feature Comparison | | | |
|---------------------------------------|-------------------------------------|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device Feature or<br>Characteristic | Fluke Impulse 7010 | BC Biomedical<br>DA-2006-VL<br>Specifications |
| | Accuracy (25 Ω, 75-<br>200 Ω) | ± 2% of reading + 0.1 J | High Range<br>± 3 % of reading for >100<br>Joules<br>± 3 Joules for ≤100 Joules<br>Low Range<br>± 3 % of reading for >20<br>Joules<br>± 0.6Joules for ≤20 Joules |
| Pysical | Dimensions | 138.7 mm x 154 mm x 272 mm | 248.9 mm x 205.7 mm x<br>120.7 mm |
| | Weight | 1.54 kg | 2.27 kg |
| | Operating<br>Temperature | 10 to 40 C | 15 to 40 C |
| | Storage Temperature | -20 to +60 C | -20 to +65 C |
- there are 7 differences, .
- there are 0 features where a comparison cannot be made due to lack of predicate ● specifications,
- there are 3 features where the DA-2006-VL is identical to the predicate. .
Overall, the devices are very similar, with the 6 differences not having any effect on the indications for use or the device's safety. There are a few minor effects on the performance characteristics of the device; some of these are discrepancies in size, weight, and accuracy. The DA-2006-VL is heavier than the predicate device and is slightly less accurate.
None of the differences raise new questions of safety or effectiveness.
{7}------------------------------------------------
#### Summary of Non-Clinical Testing/Statement of Equivalence
The DA-2006-VL Defibrillator Tester was tested to meet international standards for electrical safety. Since the DA-2006-VL is a non-powered, manually selectable resistive load, it is exempt from electromagnetic compatibility testing. Verification and Validation tests were performed to ensure that all of the product specifications were met.
The testing performed indicates that the DA-2006-VL is as safe and as effective as the predicate device. Table 2 below shows the extent of the testing that was performed.
| Test | Description | Result |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------|--------|
| IEC61010-1 | Safety requirements for<br>electrical equipment for<br>measurement, control, and<br>laboratory use Part 1:<br>General Requirements | Pass |
| Device Validation | Validate device function of<br>each hardware specification | Pass |
| Device Verification | Ensure device function<br>during assembly. | Pass |
## Table 3: Device Test Summary
#### Conclusion
In conclusion, the results from the nonclinical testing demonstrate that the DA-2006-VL Defibrillator Tester is as safe, as effective, and performs as well as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.