Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon): The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Choice Spine VEO™ Lateral Access & Interbody Fusion System: The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine.
Device Story
Spinal interbody fusion and vertebral body replacement implants; families include Sabre, Shark, Hornet, Harpoon, VEO, and Interbody Fusion System. Implants feature hollow centers for bone graft placement; rectangular shapes; various heights/lengths/angulations. Delivered via posterior, transforaminal, lateral, anterior, or anterolateral surgical approaches. Used by orthopedic or neurosurgeons in clinical settings to stabilize spine, facilitate fusion, or replace collapsed/unstable vertebral bodies. Implants provide structural support; used with supplemental fixation and autograft. Benefit: stabilization of spinal segments, pain reduction, and restoration of vertebral height.
Clinical Evidence
No clinical data. Bench testing only; engineering analysis confirmed that the worst-case design for each implant family remains unchanged by the use of titanium-alloy versus PEEK.
Technological Characteristics
Materials: PEEK-Optima (ASTM F2026), Tantalum markers (ASTM F560), Ti-6Al-4V ELI (ASTM F136), stainless steel, aluminum, Radel R. Form factor: Rectangular spacers with hollow centers. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) or vertebral body replacement (T1-L5) for tumor/trauma. DDD defined as discogenic back pain with degeneration confirmed by history/radiographs; may include Grade 1 spondylolisthesis or retrolisthesis. Requires 6 months non-operative treatment. Contraindications: none stated.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Alphatec Battalion Universal Spacer System (K143740)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a way that suggests unity and connection. The profiles are black and are set against a white background. Encircling the profiles are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
Choice Spine, LP Ms. Kim Finch Manager of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919
Re: K153107
Trade/Device Name: Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™), Choice Spine VEO™ Lateral Access & Interbody Fusion System, Choice Spine Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: April 22, 2016 Received: April 25, 2016
Dear Ms. Finch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153107
#### Device Name
Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™) Choice Spine VEOTM Lateral Access & Interbody Fusion System Choice Spine Interbody Fusion System
#### Indications for Use (Describe)
Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon):
The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
### Choice Spine VEOTM Lateral Access & Interbody Fusion System:
The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine.
#### Choice Spine Interbody Fusion System:
The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| Date | May 17, 2016 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Choice Spine, LP<br>400 Erin Drive<br>Knoxville, TN 37919 |
| Phone: | 865-246-3333 |
| Fax: | 865-246-3334 |
| Contact Person: | Kim Finch, Manager of Regulatory Affairs |
| Proposed<br>Proprietary Trade<br>Name: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™)<br>Choice Spine VEO™ Lateral Access & Interbody Fusion System<br>Choice Spine Interbody Fusion System |
| Product Class: | Class II |
| Classification<br>Name: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™):<br>888.3080- Spinal Intervertebral Body Fusion Device 888.3060- Spinal Intervertebral Body Fixation Orthosis Choice Spine VEO™ Lateral Access & Interbody Fusion System:<br>888.3080- Spinal Intervertebral Body Fusion Device Choice Spine Interbody Fusion System:<br>888.3080- Spinal Intervertebral Body Fusion Device |
| Device Product<br>Code: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™):<br>MAX, MQP Choice Spine VEO™ Lateral Access & Interbody Fusion System:<br>MAX Choice Spine Interbody Fusion System:<br>MAX |
| Purpose of<br>Submission: | The purpose of this submission is to gain clearance for additional devices<br>manufactured from titanium-alloy. The previously cleared system are Choice<br>Spine Lumbar Spacer System, K140142 and K073669, Choice Spine VEO™<br>Lateral Access & Interbody Fusion System (Formerly TranS1 Lateral Interbody<br>Fusion System) K100210 and K123997, and Choice Spine Interbody Fusion |
TranS1 systems were acquired by Choice Spine in 2015.
{5}------------------------------------------------
Device Description: This submission includes three lumbar implant families. The device description for each family is listed below.
Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon):
The Choice Spine Lumbar Spacer System consists of interbody fusion devices (specifically, SABRE, SHARK™, HORNET™ & HARPOON™ Lumbar Spacers) comprised of polyetheretherketone (PEEK - Optima®) with tantalum markers (ASTM F2026 and ASTM F560) or Ti-6AI-4V ELI (ASTM F136). The spacers have a basic rectangular shape, a hollow center for placement of bone graft and a smooth bullet shaped anterior surface. They are available in an assortment of height, length and anteroposterior angulation combinations to accommodate many different anatomic requirements. These implants are delivered via a posterior or transforaminal approach.
Choice Spine VEO™ Lateral Access & Interbody Fusion System:
The Choice Spine VEO™ Lateral Access & Interbody Fusion System is a multicomponent system including instrumentation made of stainless steel. Aluminum. and Radel R and implants made of tantalum (ASTM F560) and PEEK Optima® (ASTM F2026) or Ti-6Al-4V ELI (ASTM F136). The implants are delivered via lateral approach.
#### Choice Spine Interbody Fusion System:
The Choice Spine Interbody Fusion System is a family of implants intended to aid in spinal fixation of the lumbar spine. This system includes implants made of Zeniva ZA-500 PEEK (ASTM F2026) with Tantalum markers (ASTM F560) or Ti-6Al-4V ELI (ASTM F136), which are delivered via anterior or anterolateral approach.
#### Intended Use: Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon):
The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiquous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
#### Choice Spine VEO™ Lateral Access & Interbody Fusion System:
The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with
{6}------------------------------------------------
degenerative disc disease (DDD) at one or two contiquous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondvlolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine. Comparison of The implants included in this submission are equivalent to the primary predicate Technological Alphatec Battalion Universal Spacer System (K143740), and additional predicates Characteristics: are Choice Spine Lumbar Spacer System (K140142 and K073669), Choice Spine Interbody Fusion System (K120991), and the Choice Spine VEO™ Lateral Access & Interbody Fusion Device System (K100210 and K123997). Non-Clinical Testing: The worst case design for each family does not change if the implant was manufactured from Titanium instead of PEEK. Therefore, no new non-clinical testing is needed. Conclusion: The subject implants are substantially equivalent in intended use, anatomic location, indications, method of stabilization, and material of manufacture.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.