The Everest 20cc Inflation Device/ Survival Kit is to be used to facilitate the use of catheters and guide wires during interventional procedures. The Everest 20cc Inflation Device is designed to be used to inflate/ deflate balloon catheters as well as to monitor pressure within the balloon. The Y/Tri-Adaptor with Hemostasis Valve is designed to be used on a guiding catheter or dilatation catheter to control backbleeding and to provide a port for introduction of fluids into the interventional system. The Guide Wire Insertion Tool is designed to facilitate placement of a guide wire tip through the Y/ Tri-Adapter and into the wire lumen of an interventional catheter. The Guide Wire Steering Handle is designed to hold a small diameter guide wire and provide a handle for manipulating the wire.
Device Story
Sterile 20cc disposable inflation device; used in interventional procedures to inflate/deflate balloon catheters and monitor pressure. Features trigger-operated locking mechanism; allows manual piston manipulation when released. Includes manometer (20 or 30 bar/atm), high-pressure connecting tube, 3-way stopcock. Survival Kit versions include Y/Tri-Adapter with hemostasis valve, guide wire insertion tool, and steering handle. Operated by clinicians in clinical settings. Provides pressure monitoring and fluid introduction ports; assists in guide wire placement and manipulation. Benefits patient by enabling controlled balloon catheterization and reducing backbleeding during procedures.
Clinical Evidence
No clinical investigations performed. Substantial equivalence supported by bench performance testing (lubrication break away, pressurization, tensile tests, media compatibility) and biocompatibility testing per ISO 10993-1.
Indicated for use in interventional procedures to facilitate catheter and guide wire placement, inflate/deflate balloon catheters, monitor balloon pressure, control backbleeding, and manipulate guide wires.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures in profile, arranged in a row. The figures are connected by flowing lines, creating a sense of unity and continuity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2016
Medtronic, Inc. Nisarg Shah Regulatory Affairs Specialist 37a Cherry Hill Drive Danvers, Massachusetts 01923
Re: K153038
Trade/Device Name: Everest 20 Disposable Inflation Device. Everest 20 Survival Kit, Everest 30 Disposable Inflation Device, Everest 30 Survival Kit Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: MAV Dated: March 11, 2016 Received: March 14, 2016
Dear Nisarg Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Kenneth J. Cavanaugh -S
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Food and Drug Administration | |
| <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120 |
| | Expiration Date: January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number (if known) | |
| K153038 | |
| Device Name | |
| Medtronic's Everest <sup>TM</sup> Inflation Devices | |
| Indications for Use (Describe) | |
| The Everest 20cc Inflation Device/ Survival Kit is to be used to facilitate the use of catheters and guide wires during interventional procedures. The Everest 20cc Inflation Device is designed to be used to inflate/ deflate balloon catheters as well as to monitor pressure within the balloon. The Y/Tri-Adaptor with Hemostasis Valve is designed to be used on a guiding catheter or dilatation catheter to control backbleeding and to provide a port for introduction of fluids into the interventional system. The Guide Wire Insertion Tool is designed to facilitate placement of a guide wire tip through the Y/ Tri-Adapter and into the wire lumen of an interventional catheter. The Guide Wire Steering Handle is designed to hold a small diameter guide wire and provide a handle for manipulating the wire. | |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
PSC Publishing Services (301) 443-6740 EF
ချောင်းများ မြန်မာနိုင်ငံ ရွာများ ဧရာဝတီ
{3}------------------------------------------------
## Special 510(k) Summary
| Submitter: | Medtronic Vascular<br>37A Cherry Hill Drive,<br>Danvers, MA 01923, USA. | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Nisarg Shah<br>Senior Regulatory Affairs Specialist<br>37A Cherry Hill Drive,<br>Danvers, MA 01923, USA.<br>Phone: (978) 739-6632<br>Fax: (978) 750-8204<br>E-mail: nisarg.g.shah@medtronic.com | |
| Date Prepared: | March 11, 2016 | |
| Device Trade Name<br>& Model Numbers: | Trade Name | Model Number |
| | Everest™ 20 Inflation Device | AC2200 |
| | Everest™ 20 Survival Kit | AC2205P |
| | Everest™ 30 Inflation Device | AC3200 |
| | Everest™ 30 Survival Kit | AC3205P |
| Common Name: | Syringe, Balloon Inflation | |
| Classification Name: | Angiographic injector and syringe<br>Class II per 21 CFR §870.1650<br>Product Code: MAV | |
| Predicate Device: | The following Medtronic Everest™ Inflation Devices legally<br>marketed currently were used as predicate devices in this<br>510(k) premarket notification: | |
| | 1. K942269 (Medtronic Everest™ 20 Inflation Device)<br>2. K960983 (Medtronic Everest™ 30 Inflation Device,<br>Survival Kit) | |
| Device<br>Description: | Medtronic's Everest™ Disposable Inflation Device is a sterile<br>20cc inflation device with a locking mechanism that is<br>operated via a trigger. Normally, the locking mechanism is<br>engaged. Once the trigger is pulled back, the locking<br>mechanism is released and the piston can be manually<br>manipulated. | |
| | The Everest™ 20 Device is outfitted with a manometer with<br>measuring pressures ranging from vacuum to 20 bars in 0.5bar<br>increments. The Everest™ 30 Device is outfitted with a | |
| | manometer with measuring pressure reading from vacuum to<br>30bars in 1 bar increments. A high pressure connecting tube<br>with a male rotating adapter and a disposable 3-way stopcock<br>are also included to aid in preparation of the device. When<br>purchased as a “Survival Kit", the package includes a Y-/ Tri-<br>Adapter with hemostasis valve, a Guide Wire Insertion Tool<br>and a Steering Handle. | |
| Statement of<br>Intended Use: | The Everest 20cc Inflation Device/ Survival Kit is to be used<br>to facilitate the use of catheters and guide wires during<br>interventional procedures. The Everest 20cc Inflation Device<br>is designed to be used to inflate/ deflate balloon catheters as<br>well as to monitor pressure within the balloon. The Y/Tri-<br>Adaptor with Hemostasis Valve is designed to be used on a<br>guiding catheter or dilatation catheter to control backbleeding<br>and to provide a port for introduction of fluids into the<br>interventional system. The Guide Wire Insertion Tool is<br>designed to facilitate placement of a guide wire tip through<br>the Y/Tri-Adapter and into the wire lumen of an<br>interventional catheter. The Guide Wire Steering Handle is<br>designed to hold a small diameter guide wire and provide a<br>handle for manipulating the wire. | |
| Summary of<br>Technological<br>Characteristics: | Medtronic's EverestTM Disposable Inflation Device is a sterile<br>20cc inflation device designed to be used during<br>interventional procedures to inflate/ deflate balloon catheters<br>as well as monitor pressure within the balloon. Medtronic<br>offers the EverestTM Inflation Device with a 20 atm or 30 atm<br>pressure gauge. The EverestTM Inflation Device is constructed<br>of the following key design components: | |
| | 1. Syringe body with 20cc capacity<br>2. Body cap<br>3. Compression spring<br>4. Piston or lead screw<br>5. Half nut assembly<br>6. Rubber Plunger Tip<br>7. Plunger Insert<br>8. Gauge or Manometer (20 atm or 30atm)<br>9. High pressure tube with a male rotating adaptor<br>10. Trigger | |
| | The difference between the subject and predicate devices is<br>the change in material of the rubber plunger tip component. | |
| Summary of Non- | The device performance testing and biocompatibility testing | |
{4}------------------------------------------------
{5}------------------------------------------------
| clinical Data: | were performed in accordance to the relevant FDA guidance in order to demonstrate substantial equivalence to the legally marketed predicate devices. |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>Performance/ Bench Testing:</b> The following tests performance or functional tests were performed to demonstrate substantial equivalence to the predicate devices: |
| | 1. Lubrication break away test |
| | 2. Pressurization test |
| | 3. Rubber Tip and Insert Tensile test |
| | 4. Compatibility with Contrast Media, Saline or any combination. |
| | <b>Biocompatibility testing:</b> The Biocompatibility testing was performed pursuant to the requirements of ISO 10993-1: Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process. |
| | Based on the results of the Performance testing and Biocompatibility testing, no new concerns of safety and effectiveness were raised for Medtronic's Everest <sup>TM</sup> Inflation Devices. The test data demonstrate that the modified Everest <sup>TM</sup> Inflation Device is safe, effective, and performs as well or better than the predicate devices. |
| Summary of<br>Clinical Data: | No clinical investigations have been performed on the modified device. |
| Conclusion from<br>Data: | Medtronic Vascular has demonstrated that the modified Everest <sup>TM</sup> Inflation Devices are substantially equivalent to the legally marketed predicate devices based on the intended use and technological characteristics. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.