K152748 · Kaz USA, Inc (A Subsidiary of Helen of Troy, Inc) · FLL · Mar 1, 2016 · General Hospital
Device Facts
Record ID
K152748
Device Name
Braun Thermoscan PRO 6000 Ear Thermometer
Applicant
Kaz USA, Inc (A Subsidiary of Helen of Troy, Inc)
Product Code
FLL · General Hospital
Decision Date
Mar 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.
Device Story
Handheld infrared ear thermometer for professional use in hospitals/clinics; measures thermal infrared radiation from tympanic membrane and ear canal. Device features heated tip to minimize ambient temperature influence; includes technique compensation feature using capacitive rings in speculum to detect insertion depth and outer ear temperature to correct for misuse. Microcontroller processes sensor inputs; displays temperature on LCD. Healthcare providers use device to obtain rapid temperature readings; aids in clinical assessment of patient febrile status. Benefits include improved accuracy and reduced measurement errors due to improper technique.
Clinical Evidence
Clinical comparison study performed per ASTM E1965-03 across four age groups (0-12 months, 12 months-<5 years, <18 years, 18+ years). Compared test device to predicate (K103800) using SureTemp Plus (K030580) as gold standard. Results demonstrated non-inferiority in bias and standard deviation. Clinical repeatability was <0.3°C (0.58°F), meeting clinical acceptability criteria.
Technological Characteristics
Handheld infrared thermometer; impact-resistant casing; biocompatible resins/metals (ISO 10993-5, ISO 10993-10). Sensing: Infrared sensor + capacitive rings for insertion depth detection. Power: 2x AA alkaline or NiMH battery pack. Connectivity: Standalone. Software: Embedded firmware for predictive algorithms and technique compensation. Standards: ASTM E1965-09, IEC 60601-1, IEC 60601-1-2, EN 12470-5, ISO 80601-2-56.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature for patients of all ages in professional use environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Braun Thermoscan® PRO 4000 Ear Thermometer (K103800)
Reference Devices
SureTemp Plus Oral/Rectal and Axillary Contact Thermometer (K030580)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are simple and abstract, with flowing lines suggesting hair or head coverings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2016
KAZ USA, Inc (a Subsidiary of Helen of Troy, Inc) Mr. Raj Kasbekar Vice President, Regulatory Affairs 250 Turnpike Road Southborough, Massachusetts 01772
Re: K152748
Trade/Device Name: Braun Thermoscan® PRO 6000 Ear Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: December 4, 2016 Received: December 8, 2016
Dear Mr. Kasbekar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use Statement Section 4:
## Indications for Use
510(k) Number (if known): K152748
Braun Thermoscan® PRO 6000 Ear Thermometer Device Name:
Indications for Use:
The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Kaz USA, Incorporated. The logo is a stylized "kaz" in white letters against a blue background. Below the logo is the text "Kaz USA, Incorporated" followed by the address "250 Turnpike Road, Southborough, MA 01772".
## K152748 510(k) Summary
# Braun Thermoscan® PRO 6000 Ear Thermometer
#### 1.0 Preparation Date:
February 19, 2016
#### 2.0 Submitted By:
KAZ USA, Inc., a Helen of Troy Company 400 Donald Lynch Boulevard, Suite 300 Marlborough, MA 01752
# Primary Contact Person/Prepared by:
Raj S Kasbekar, Global VP Regulatory & Clinical Affairs
KAZ USA, Inc., a Helen of Troy Company 400 Donald Lynch Boulevard, Suite 300 Marlborough, MA 01752
#### 3.0 Device Identification:
#### 3.1 Trade Name
Braun Thermoscan® PRO 6000 Ear Thermometer
## Models:
US: PRO6000SCMN (Thermometer with Small Cradle), PRO6000LCMN (Thermometer with Large Cradle), PRO6000BSNA (Charge Dock) [Braun Thermoscan® PRO 6000 Ear Thermometer]
#### 3.2 Common Name
Infrared Ear Thermometer
#### 3.3 Classification Name
Thermometer, Clinical, Electronic
{4}------------------------------------------------
(21CFR 880.2910: Product code - FLL)
### 4.0 Predicate Device:
| Predicate | Manufacturer | Docket<br>Number |
|-----------------------------------------------|--------------------------------------|------------------|
| Braun Thermoscan® PRO 4000 Ear<br>Thermometer | Kaz USA, Inc<br>Contract: Keytronics | K103800 |
### 5.0 Device Description:
The Braun Thermoscan® PRO 6000 Ear Thermometer is a hand held instrument that measures human body temperature through the opening of the auditory canal. It is a single mode ear thermometer that measures the natural thermal infrared radiation emitted from the tympanic membrane and adjacent surfaces with a built-in correction algorithm to compensate the influence of ambient temperature using a heated tip. The Braun Thermoscan® PRO 6000 Ear Thermometer also includes a technique compensation feature to detect depth of insertion to compensate for common methods of misuse, this feature uses capacitive rings mounted in the speculum of the thermometer to detect insertion and depth and measurements of the outer ear temperature during insertion to detect misuse and correct accordingly. The Braun Thermoscan® PRO 6000 Ear Thermometer is meant for professional use in hospitals and healthcare or professional office settings.
#### 6.0 Indications for Use:
The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.
#### 7.0 Validation Results:
## Clinical Study to show substantial equivalence:
A comparison study and clinical repeatability testing was performed on the following four age groups: 0-12 months, 12 months- <5 years, <18 years, and 18 years and older in accordance with ASTM E1965-03 to compare the Braun Thermoscan® PRO 6000 Ear Thermometer (test thermometer) with the predicate Braun Thermoscan® PRO 4000 Ear Thermometer (K103800). The reference or the gold standard used was the SureTemp Plus Oral/Rectal and Axillary Contact Thermometer in the monitoring mode (K030580). This clinical comparison study demonstrated that the Braun Thermoscan® PRO 6000 Ear Thermometer is as good as (non-inferior or substantially equivalent to) the previously approved Braun Thermoscan® PRO 4000 Ear Thermometer in all age groups with respect to the bias and standard deviation in comparison to the Reference SureTemp Plus Oral/Rectal and Axillary Contact Thermometer in the monitoring mode (K030580). The temperatures obtained with the test Infrared Ear Thermometer were highly related when compared to the predicate device, where temperatures were
{5}------------------------------------------------
measured in the oral mode for children above 5 years) and axillary mode (for children under 5 years) . The clinical bias with stated uncertainty and clinical repeatability as defined in the ASTM E1965-03 standard were within clinical acceptability (bias less than predicate device when compared to reference). The clinical repeatability of the Infrared Ear Thermometer was statistically and clinically acceptable (less than 0.3 deg C or 0.58 deg F).
#### 8.0 Device Comparison with Predicate
### 8.1 Intended Use
The predicate device, the Braun Thermoscan® PRO 4000 Ear Thermometer is intended for the intermittent determination of the human's body temperature for people of all ages. The intended use and indications for use of the predicate and the Braun Thermoscan® PRO 6000 Ear Thermometer are similar. The predicate device as well as the KAZ Braun Thermoscan® PRO 6000 Ear Thermometer use the measurement site.
### 8.2 Materials
Materials used in the manufacture of the Braun Thermoscan® PRO 6000 Ear Thermometer are similar (derived from similar resins) to the predicate device. All skin contacting materials used in the new thermometer have been tested for biocompatibility (cyto-toxicity in accordance with ISO 10993-5 as well as irritation and sensitization in accordance with ISO 10993-10) and FDA Blue book memo G 95-1 for both thermometers and the corresponding test reports are included in this submission.
### 8.3 Design
The industrial design of the KAZ Braun Thermoscan® PRO 6000 Ear Thermometer is identical to the predicate device except for a new outer shell.
### 8.4 Operational Principles
The Braun Thermoscan® PRO 6000 Ear Thermometer is a handheld device, containing an On/Off switch, sensor area, microcontroller and LCD screen to control the device and take measurements. The operating principle is based on detection of infrared energy and use of predictive algorithms to estimate the body temperature.
#### 8.5 Technology
The technology of the Braun Thermoscan® PRO 6000 Ear Thermometer is identical to the predicate device. The technique compensation feature is additional technology that is integrated into the Braun Thermoscan® PRO 6000 Ear Thermometer using additional hardware (three capacitive rings) and software.
{6}------------------------------------------------
| 8.6 | Comparison table between test | and Predicate Devices | |
|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| SE<br>Comparisons | Subject (Braun Thermoscan® PRO<br>6000 Ear Thermometer) | Predicate (Braun<br>Thermoscan® PRO 4000)<br>K103800 | Comment |
| Classification | 21CFR 880.2910 | 21CFR 880.2910 | Same |
| Product Code | FLL | FLL | Same |
| FDA Class | II | II | Same |
| Intended Use | A non-sterile, re-useable clinical<br>thermometer intended for the intermittent<br>determination of the human's body<br>temperature for people of all ages | A non-sterile, re-useable<br>clinical thermometer<br>intended for the<br>determination of the<br>human's body temperature<br>for people of all ages. | Same |
| Operation | Hand held-Manually operated | Hand held-Manually<br>operated | Same |
| Sensor | Infrared | Infrared | Same |
| Materials | Common Materials- including an impact<br>resistant casing. Biocompatible metals<br>and resins. | Common Materials-<br>including an impact<br>resistant casing.<br>Biocompatible metals and<br>resins. | Similar - validated<br>for cytotoxicity per<br>ISO10993-5 and<br>irritation as well as<br>sensitization per ISO<br>10993-10 |
| Dimensions | 150mm x 60 mm x 35mm | 150mm x 45 mm x 30 mm | Similar |
| Accuracy | 0.3 deg C for <35 deg C<br>0.2 deg C for 35 deg C to 42 deg C<br>0.3 deg C for >42deg C | 0.3 deg C for <35.5 deg<br>C<br>0.2 deg C for 35.5 deg C<br>to 42 deg C<br>0.3 deg C for >42 deg C | Meets ASTM E1965,<br>EN12470-5 and<br>ISO 80601-2-56 |
| Precision (SD) | <0.6 deg C | <0.6 deg C | Same |
| Repeatability | <0.3 deg C | <0.3 deg C | Meets EN 12470-5 |
| Operating<br>temperature | 10 to 40 deg C ambient temperature<br>(amb) and up to 95% RH | 10 to 40 deg C ambient<br>and up to 95% RH | Same |
| Measurement<br>Range | 20 deg C to 42.2 deg C | 20 deg C to 42.2 deg C | Same |
| Display | LCD | LCD | Same |
| Response<br>Time | 2-3 sec | 2-3 sec | Same |
| Measurement<br>Site | Ear | Ear | Same |
| Scale | Deg F/Deg C | Deg F/Deg C | Same |
| Power Supply | Two AA Alkaline Batteries or Custom<br>Nickel Metal Hydride Battery Pack | Two AA Alkaline Batteries<br>or custom rechargeable<br>Nickel Metal Hydride<br>Battery Pack | Same |
| Standards<br>Met for Bench<br>and Clinical<br>Performance | 1) ASTM E1965-1998: 2009 Infrared<br>Thermometers for Intermittent<br>Determination of Patient temperature<br>2) Clinical accuracy test requirements<br>established in the standard ASTM<br>E1965-09 (Clinical part only)- Standard<br>Specification for Infrared Thermometer | Same | All standards verified<br>and validated for test<br>device and met<br>acceptance criteria. |
{7}------------------------------------------------
| For Intermittent Determination of Patient<br>Temperature; |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3) IEC 60601-1 3rd edition:2005 Medical<br>Electrical Equipment: General<br>requirements for Safety, Requirements<br>and Tests. |
| 4) IEC 60601-1-2: 2007 Medical Electrical<br>Equipment- Part 1: General<br>Requirements for Safety, Electromagnetic<br>Compatibility- Requirements and Tests. |
| 5) AAMI / ANSI / ISO 10993-1:2003(E),<br>Biological evaluation of medical devices -<br>- Part 1: Evaluation and testing- and<br>appropriate Parts |
| 6) ISO 10993-5: Biological Evaluation of<br>Medical Devices: Part 5: Tests for In-vitro<br>cytotoxicity. |
| 7) ISO 10993-10: Biological Evaluation of<br>Medical Devices: Part 10: Tests for<br>Irritation and Sensitization. |
| 8) ISO 14971:2012: Application of Risk<br>Management to Medical Devices |
| 9) EN IEC 62304:2006 Medical Device<br>Software – Software life-cycle processes |
| 10) IEC EN62366:2007 Applicability of<br>Usability Engineering to Medical Devices<br>– Collateral Standard: Usability |
| 11) ISO 14155:2011 Clinical investigation<br>of medical devices for human subjects –<br>Good Clinical Practice (also includes<br>21CFR812, parts 50 and 56) |
# 9. Conclusion:
Based on the performance testing and compliance with acceptable voluntary standards, we believe that the Braun Thermoscan® PRO 6000 Ear Thermometer is substantially equivalent to its predicate device cited above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.