K152529 · Focus Medical, LLC · GEX · Feb 4, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152529
Device Name
NaturaLase CO2 Laser System
Applicant
Focus Medical, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 4, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NaturaLase CO2 laser operating in tractionated mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery. NaturaLase CO2 laser operating in fractional mode is indicated for ablative skin resurfacing.
Device Story
RF-excited CO2 laser system; delivers 10.6 μm laser energy via articulated arm to surgical or scanning handpieces. Input: user-defined parameters via touchscreen controller. Transformation: switching-mode power supply converts line voltage to high voltage for laser tube; optical bench aligns beam; scanner/handpiece distributes energy to target tissue. Output: controlled tissue ablation, incision, or coagulation. Used in clinical settings (dermatology, plastic surgery, ENT, etc.) by trained physicians. Benefits: precise, uniform tissue interaction via selective photothermolysis; enables both traditional surgical procedures and fractional skin resurfacing.
Clinical Evidence
Bench testing only. Performance testing demonstrated device functions according to specifications. Conformance to IEC 60601-1 (safety), IEC 62366, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1 (laser safety) confirmed. Biocompatibility consistent with ISO 10993. Software verification and validation testing completed.
Technological Characteristics
RF-excited CO2 laser; 10.6 μm wavelength; 30W max power. 7-joint spring-balanced articulated arm; 3mW @ 650nm aiming beam. Microprocessor-based touchscreen interface. Electrical: 110VAC/60Hz or 220VAC/50Hz. Conforms to IEC 60601-1, IEC 62366, IEC 60825-1. Biocompatibility per ISO 10993.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues (including intraoral) in aesthetic, ENT, gynecology, neurosurgery, dental/oral, and genitourinary surgery; also indicated for ablative skin resurfacing in fractional mode.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 4, 2016
Focus Medical, LLC % Michail Pankratov Principal MMP Medical Associates LLC 16 Appleton Street Waltham, Massachusetts 02453
Re: K152529
Trade/Device Name: Naturalase Co2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX. ONG Dated: January 7, 2016 Received: January 8, 2016
Dear Michail Pankratov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name NaturaLase CO2 Laser System
Indications for Use (Describe)
NaturaLase CO2 laser operating in tractionated mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
NaturaLase CO2 laser operating in fractional mode is indicated for ablative skin resurfacing.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# As required by section 21 CFR 807.92
# I. General Information
| Applicant: | Focus Medical, LLC |
|-------------------|---------------------------------|
| | 2 Francis J Clarke Circle |
| | Bethel, CT 06801 |
| | TEL: 203-730-8885 |
| | FAX: 203-730-8851 |
| Contact Person: | Michail M. Pankratov, MD, PhD |
| | MMP Regulatory Consultants, LLC |
| | 16 Appleton Street |
| | Waltham, MA 02453 |
| | TEL: 617-480-4543 |
| | clregconsult@aol.com |
| Preparation Date: | August 26, 2015 |
# II. Device Information
| Trade Name: | NaturaLase CO2 Laser System |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | CO2 Laser System |
| Classification: | Class II |
| Regulation Number: | 878.4810 |
| Product Code: | GEX; ONG |
| Classification Name: | Powered Laser Surgical Instrument |
| Review Panel: | General & Plastic Surgery |
| Predicate Devices: | K110333 - NaturaLase CO2 Laser System, Focus<br>Medical<br>LLC; K123573 – YOULASER CO2 Laser System, Quanta<br>System S.P.A. |
{4}------------------------------------------------
## III. Device Description:
The NaturaLase CO2 is an RF excited CO2 laser. The main console of the NaturaLase CO2 laser system contains the following components:
- CO2 laser tube
- Power Supply
- Optical bench assembly
- Articulated Arm
- Hand-pieces: scanning and surgical
- Touch screen controller/user interface
The laser source is a sealed CO2 laser tube that is mounted vertically within the main cabinet and the laser beam is passed from the laser tube into the articulated arm. The NaturaLase CO2 laser unit is equipped with a switching-mode power supply, which converts the input line voltage to the high voltage required for laser operation. The optical bench assembly consists of the articulated arm mount, the aiming beam diode laser (3mW @ 650nm) and optics to align the laser beam and the aiming beam to the articulated arm. The laser energy is delivered through an articulated arm to a hand-piece (scanner or surgical) that distributes the laser energy on the target in a controlled and uniformed way. The user-interface microprocessor-based touch screen allows operator to control all operating functions.
# IV. Indication for Use:
NaturaLase CO2 laser operating in traditional non-fractionated mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynecology, urology, neurosurgery, dental and oral surgery and genitourinary surgery. NaturaLase CO2 laser operating in fractional mode is indicated for ablative skin resurfacing.
#### V. Technological Characteristics:
The NaturaLase CO2 laser technical specifications are very similar to the predicate devices. All have the same intended use, technological characteristics and operating principles as the previously cleared NaturaLase CO2 Laser system (K110333) and Quanta System YOULASER Laser System (K123573). The NaturaLase CO2 laser design and components are very similar to those of the predicate devices. Their treatment handpieces connected to the articulated arm that is connected to the main console where the user interface is located. Differences between the NaturaLase CO2 laser and predicate devices in specific maximum energy levels and fluence delivered do not raise any new safety or effectiveness issues because the NaturaLase CO2 laser parameters fall within the range of parameters delivered by the predicate devices.
#### VI. Testing Performance
All performance testing demonstrated that the NaturaLase CO2 laser performs according to specifications and functions as intended. The NaturaLase CO2 laser system was determined to be in conformance with all applicable standards. Electrical safety and electromagnetic compatibility (EMC) testing for the NaturaLase CO2 laser system was conducted by an independent laboratory
{5}------------------------------------------------
in accordance with IEC 60601-1. Medical electrical equipment, part 1: general requirements for basic safety and essential performance. The NaturaLase CO2 laser was determined to be in conformance with applicable IEC standards (IEC 62366, 60601-1-6, 60601-1-6, 60601-2-22, and IEC 60825-1). The biocompatibility of the NaruraLase CO2 laser is consistent with the ISO 10993.
Software verification and validation testing results are acceptable for software release.
### VII. Summary of Substantial Equivalence
Review of the technological characteristics for the primary predicate device and the NaturaLase CO2 Laser shows close similarity in parameters principally for generating tissue effects (see Table below).
The NaturaLase CO2 laser has the same intended use as the predicate devices with similar indications for use. It presents similar technological characteristics as the predicate devices including the laser type, wavelengths, device design, pulse width, frequency, spot sizes and system components. Few differences between the NaturaLase CO2 laser and its predicate devices in terms of maximum pulse energy, fluence, pulse duration, etc, do not present any new concerns of safety and effectiveness. The NaturaLase CO2 laser and its predicates all operate with the same mechanism of action based on selective photothermolysis. Therefore, the NaturaLase CO2 laser has the same intended use and similar indications for use, technological characteristics, and principles of operation as predicate devices. The NaturaLase CO2 laser is substantially equivalent to the predicate devices.
{6}------------------------------------------------
| 510(k) | | K123573 | K110333 | |
|----------------------------------|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | | Quanta System S.P.A | Focus Medical LLC | Focus Medical LLC |
| Device Name | | YOULASER CO2 Laser | NaturaLase CO2 | NaturaLase CO2 |
| Device<br>Classification<br>Name | | Powered Laser Surgical<br>Instrument | Powered Laser Surgical<br>Instrument<br>Microbeam/Fractional Output | Powered Laser Surgical<br>Instrument<br>Microbeam/Fractional Output |
| Regulation Number | | 878.4810 | 878.4810 | 878.4810 |
| Product Code | | GEX | ONG | GEX, ONG |
| Laser Source | | CO2 | CO2 | CO2 |
| Wavelength | | 10.6 μm | 10.6 μm | 10.6 μm |
| Power | | Up to 30 W | Up to 30 W | Up to 30 W |
| Pulse<br>width | Fraction | 0.2 - 20 ms | 0.2-10 ms | 0.2-10 ms |
| | Surgical | 0.5 – 80 ms | - | .2 to 10ms |
| Spot Size | Fraction | 300 μm | 120 and 300 μm | 120 and 300 μm |
| | Surgical | 200, 400, 1000 μm | | 300 μm |
| Energy Emission | | Pulse/single pulse/cw | Pulse/single pulse/cw | Pulse/single pulse/cw |
| Energy to tissue | | | 10-45 mJ per micro spot | 10-45 mJ per micro spot |
| Peak power | | | 60 W | 60 W |
| Aiming beam | | <4 mW @ 650 nm | 3 mW @ 650 nm | 3 mW @ 650 nm |
| Articulated arm | | 7-mirrors | 7-joint spring-balanced | 7-joint spring-balanced |
| Electric Power | | 230 VAC?50-60 Hz | 110VAC/60 Hz or 220 VAC/50 Hz | 110VAC/60 Hz or 220 VAC/50 Hz |
| Dimensions | | 48x55x120 cm | 34x46x126 cm | 34x46x126 cm |
| Weight | | 30 Kg | 35 Kg | 35 Kg |
| Indications for use | | YOULASER CO2 laser when used in traditional non-fractionated mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues including intraoral tissues, in medical | The NaturaLase CO2 Laser System is indicated for use in dermatological procedures requiring ablation (removal), resurfacing or coagulation of soft tissue, additionally, the | NATURALASE CO2 laser operating in traditional non-fractionated mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues including intraoral tissues, in medical |
# TABLE - Comparison of the NaturaLase CO2 with the Predicate Devices
{7}------------------------------------------------
| specialties including aesthetic<br>(dermatology and plastic surgery),<br>otolaryngology<br>(ENT),<br>gynecology,<br>neurosurgery, dental and oral<br>surgery and genitourinary surgery.<br><b>YOULASER CO2</b> laser when<br>used in fractional mode (dot<br>scanner) is indicated for ablative<br>skin resurfacing | fractional spot sizes are used in the<br>treatment of wrinkles; rhytides,<br>furrows, fine lines, textural<br>irregularities, pigmented lesions<br>and vasculardyschromia. | specialties including aesthetic<br>(dermatology and plastic surgery),<br>otolaryngology<br>(ENT),<br>urology,<br>neurosurgery, dental and oral surgery<br>and genitourinary surgery.<br><b>NATURALASE CO2</b> laser<br>operating in fractional mode is<br>indicated for ablative skin<br>resurfacing. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description<br>The YOULASER CO2 Laser<br>System includes a single model<br>named YOULASER CO2 emitting<br>a maximum power of<br>30 Watt at 10.6 µm. The<br>YOULASER CO2 Laser<br>composed externally of metallic<br>shell with a frontal polyurethane<br>panel containing the touch screen<br>display. On this panel the key<br>switch, emergency red push button<br>and the operation led are inserted<br>too. On the rear panel the<br>footswitch connector, the remote<br>interlock, the power switch are<br>located. The system is composed<br>of power supply, CO2 laser source<br>with air cooling system, optical<br>bench, articulated arm with CO2<br>scanner, the control electronics.<br>The electronic, based on a<br>microcontroller, manages the<br>voltage power supply and the CO2<br>laser source. | The NaturaLase CO2 is an RF<br>excited CO2 laser. The laser energy<br>is delivered through an articulated<br>arm and with a scanner, that<br>distributes the laser energy on the<br>skin in a controlled way with<br>uniform distribution. The scanner<br>design allows the system to ablate<br>small micro spots of tissue. The<br>system has the ability to deliver<br>micro spots with a size of 120<br>microns and up to 45 mJ of optical<br>energy per micro spot.<br>The NaturaLase CO2 laser system<br>consists of<br>-CO2 laser tube,<br>-Power supply,<br>-Closed loop cooling system,<br>- user interface control<br>- an articulated arm<br>- a scanning hand piece<br>- a traditional hand piece | The NaturaLase CO2 is an RF<br>excited CO2 laser. The laser energy<br>is delivered through an articulated<br>arm and with a scanner, that<br>distributes the laser energy on the<br>skin in a controlled way with<br>uniform distribution. The scanner<br>design allows the system to ablate<br>small micro spots of tissue. The<br>system has the ability to deliver<br>micro spots with a size of 120<br>microns and up to 45 mJ of optical<br>energy per micro spot.<br>The NaturaLase CO2 laser system<br>consists of<br>-CO2 laser tube,<br>-Power supply,<br>-Closed loop cooling system,<br>- user interface control<br>- an articulated arm<br>- a scanning hand piece<br>- a traditional hand piece |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.