K123573 · Quanta System Spa · GEX · Apr 15, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123573
Device Name
YOULASER CO2 LASER
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for: incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.
Device Story
YOULASER CO2 is a 30-watt, 10.6 µm CO2 laser system. It consists of a power supply, laser source with air cooling, optical bench, articulated arm with CO2 scanner, and control electronics. Operated by clinicians in medical settings, the device delivers laser energy to soft tissues for surgical procedures or skin resurfacing. The system features a touch screen interface for parameter control. The microcontroller-based electronics manage power supply and laser emission. The device provides therapeutic benefits through precise tissue ablation and coagulation, facilitating various surgical and aesthetic interventions.
Clinical Evidence
Bench testing only. Testing evaluated ablation, coagulation depth, and skin healing evolution at typical settings to support substantial equivalence.
Technological Characteristics
30W, 10.6 µm CO2 laser; air-cooled; articulated arm with CO2 scanner; metallic shell with polyurethane panel; touch screen interface; microcontroller-based control electronics; footswitch and remote interlock connectivity.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, excision, ablation, vaporization, coagulation) in dermatology, plastic surgery, ENT, gynecology, neurosurgery, dental/oral surgery, and genitourinary surgery; also indicated for ablative skin resurfacing in fractionated mode.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## K 123573
## 510(K) SUMMARY
## APR 1 5 2013
## Introduction:
This document contains the 510(k) Summary for the YOULASER CO2 laser system.
The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Maurizio Bianchi<br>Regulatory Affairs Manager<br>Quanta System SPA |
| | Email: Maurizio.bianchi@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | November 16th, 2012 |
| Device Name: | YOULASER CO2 |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
#### Predicate Devices:
The YOULASER CO2 Laser System is claimed to be substantially equivalent to the following legally marketed predicate devices:
| Lutronic Corp - MOSAIC e CO2 (K080496) |
|---------------------------------------------------|
| Lutronic Corp - MOSAIC e CO2 (K091115) |
| Lasering Srl - SLIM Evolution Family (K063001) |
| Lasering Srl - SLIM Evolution II Family (K110984) |
| El.En S.p.A - Smart CO2 Laser System (K072159) |
{1}------------------------------------------------
#### Performance Standards:
There are no mandatory performance standards for this device.
A specific investigation to evaluate the safety and the effectiveness has been planned and performed :
A specific non clinical testing was conducted on the proposed YOULASER CO2 device to support a determination of substantial equivalence
The testing has evaluated the ablation, coagulation depth and the healing evolution of the skin at the typical settings of use of Youlaser CO2.
In conclusion the results of these tests provided reasonable assurance that the YOULASER CO2 is as safe as effective as the predicated devices and support a determination of substantial equivalence.
#### General Device Description:
The YOULASER CO2 Laser System includes a single model named YOULASER CO2 emitting a maximum power of 30 Watt at 10.6 µm.
The YOULASER CO2 Laser is composed externally of a metallic shell with a frontal polyurethane panel containing the touch screen display. On this panel the key switch, emergency red push button and the operation led are inserted too. On the rear panel the footswitch connector, the remote interlock, the power switch are located.
The laser system is composed of power supply, CO2 laser source with air cooling system, optical bench, articulated arm with CO2 scanner, the control electronics.
The electronic, based on a microcontroller, manages the voltage power supply and the CO2 laser source.
#### Indications for Use:
YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for:
incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.
YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.
#### Comparison of Technological Characteristics:
The laser specifications for the YOULASER CO2 laser system are substantially equivalent to the laser specifications for its identified predicate device with respect to the laser source, wavelengths, maximum energy, spot size, fluence,
{2}------------------------------------------------
pulse width, repetition rate, beam delivery, power monitor, actuator, and aiming beam.
## Comparison of Intended Use:
The intended use of the YOULASER CO2 laser system is the same as the intended use of its previously cleared devices.
#### Substantial Equivalence:
The Quanta System YOULASER CO2 laser system is as safe and effective as the predicate devices.
The YOULASER CO2 laser system has the same intended use and similar technological characteristics and principles of operation as its predicate devices. The minor technological differences between the YOULASER CO2 laser system and its predicate devices raise no new issues of substantial equivalence or safety and effectiveness.
Thus, the YOULASER CO2 laser system is substantially equivalent to its identified predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is centered horizontally. The words are all capitalized and evenly spaced.
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2013
Quanta System, S.P.A. % Mr. Maurizio Bianchi Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058
Re: K123573 Trade/Device Name: YOULASER CO2 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 8, 2013 Received: March 11, 2013
Dear Mr. Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 – Mr. Maurizio Bianchi
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR
# Peter D. Ruimm -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K123573
Device Name: YOULASER CO2
### Indications for Use:
YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for: incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.
YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.
X ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use AND/OR (Part 21 C.F.R. 801 Subpart D)
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.04.105 09:32:07 -04'00'
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K123573
{6}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K123573
YOULASER CO2 Device Name:
#### Indications for Use:
YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for:
incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.
YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.
Prescription Use X Use (Part 21 C.F.R. 801 Subpart D)
(21 C.F.R. 807 Subpart C)
Over-The-Counter
(Please Do Not WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
FOR
Peter D. Rumm, 2013.04.15 -04'00'
- (Division Sign-Off) for MXM Neil R. Ogden, M.S. Chief, General Surgery Devices Branch I Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.