K152420 · Ak Beauty Enterprises, LLC · NGX · Dec 1, 2016 · Physical Medicine
Device Facts
Record ID
K152420
Device Name
AK Body Toning Device
Applicant
Ak Beauty Enterprises, LLC
Product Code
NGX · Physical Medicine
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AK Body Toning Device is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin-contact electrodes in order to improve or facilitate muscle performance.
Device Story
Battery-operated, hand-held muscle stimulation system; intended for home use by healthy individuals. Device features rounded unit with handle and three rows of axels; two outer rows contain stainless steel powered electrodes; middle row contains plastic wheels for mobility. User applies electroconductive gel to skin; rolls device across treatment area (abdomen, thighs, buttocks). Device delivers monophasic, rectangular electrical pulses via electrodes spaced ~4 inches apart. Intensity adjusted via manual knob. Rolling electrode design allows stimulation of broader areas while potentially reducing skin burn risk compared to fixed-electrode systems. Output affects muscle performance through induced contraction. No software or microprocessor control.
Clinical Evidence
Bench testing only. Study used ultrasound imaging to observe and compare anterior thigh muscle movement during stimulation between subject and predicate devices; results showed similar muscle stimulation performance. Biocompatibility testing performed per ISO 10993. Electrical safety and EMC testing conducted per IEC 60601-1, IEC 60601-1-2, UL 60601-1, CSA C22.2 No. 601.1, and IEC 60601-2-10.
Technological Characteristics
Battery-operated (3 x 1.5V AAA); monophasic rectangular waveform; manual intensity control. Housing material: ABS. Electrodes: stainless steel. Dimensions: 3.12 x 3.51 x 7.0 inches. Weight: 800g. Standards: IEC 60601-1:2005, IEC 60601-1-2:2007, UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-10:2012. No software, firmware, or microprocessor control.
Indications for Use
Indicated for healthy persons to apply transcutaneous electrical muscle stimulation (EMS) to improve or facilitate muscle performance in the abdomen, thighs, and buttocks. Contraindicated for use on unhealthy muscles.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three faces layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
AK Beauty Enterprises, LLC % Thomas C. Knott Senior Regulatory Advisor Benjamin L. England and Associates, LLC 810 Landmark Dr. Suite 126 Glen Burnie, Maryland 21061
Re: K152420
Trade/Device Name: AK Body Toning Device Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: October 29, 2016 Received: November 1, 2016
Dear Thomas C. Knott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michae|J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152420
Device Name AK Body Toning Device
Indications for Use (Describe)
The AK Body Toning Device is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin-contact electrodes in order to improve or facilitate muscle performance.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (21 CFR 807.92)
#### l. SUBMITTER
AK Beauty Enterprises, LLC 436 North Bedford Dr. Suite 203 Beverly Hills, CA 90210
Phone: (310) 550-7747 Fax: (310) 550-8822
Contact Person: Robert Applebaum, MD Date Prepared: August 20, 2015
#### II. DEVICE
Name of Device: AK Body Toning Device Common or Usual Name: Muscle Stimulation Device Classification Name: Powered Muscle Stimulator for Muscle Conditioning Regulatory Class: 21 CFR 890.5850, Class II Product Code: NGX
#### lll. PREDICATE DEVICE
- . K083140; Compex Sport Plus manufactured by DJO, LLC
These predicates have not been subject to a design-related recall. 2
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The AK Body Toning Device is a battery-operated, hand-held, single-program, muscle stimulation system to tone healthy muscles of the abdomen, thighs, and buttock. The device is self- contained with three rows of axels under a rounded unit with a handle that is shaped to allow a comfortable hand-grip for the user. Each axel has three knobbed wheels. The two sets of stainless on the ends are the powered electrodes and the middle set of plastic wheels allows for easier movement across the skin.
The device produces transcutaneous electrical muscle stimulation through the powered electrodes that are approximately four inches apart. An on-off switch applies the power and an adjustable knob sets the intensity of the treatment. An electroconductive gel is applied to the skin in the area to be treated and the device is rolled across the treatment area. The knob is adjusted until the treatment is effective and may be increased after initial use.
#### V. INDICATIONS FOR USE
The AK Body Toning Device is intended to for use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin-contact electrodes in order to improve or facilitate muscle performance.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
<sup>1</sup> On July 9, 2012, section 605 of FDASIA (Pub. L. 112-144) added section 518A to the FD&C Act, which directs FDA to establish a program to routinely and systematically assess information regarding device recalls, and to use that information to proactively identify strategies for mitigating health risks presented by defective or unsafe devices. FDA believes that one way to carry out this directive is to provide greater transparency on recalled devices. Identifying whether a predicate was recalled is optional, but doing so would help the Agency achieve this FDASIA directive.
{4}------------------------------------------------
The AK Body Toning Device is substantially equivalent to other legally marketed muscle stimulation devices cleared under product code NGX. The intended use of the AK Body Toning Device is the same as the predicate devices. All are intended for use on healthy muscles to improve muscle tone and are not intended to treat any diseases and are contraindicted for use on unhealthy muscles.
The AK Body Toning Device uses the same electrical stimulation principals as the predicate device. The AK Body Toning Device delivers the stimulation through rolling electrodes versus fixed electrodes in the predicate devices. However, the basic unit characteristics and output specifications of the AK Body Toning Device are substantially equivalent to the predicate devices. The rolling electrodes do not raise any new questions of safety or effectiveness. On the other hand, because the electrodes are moved across the treatment area and do not stay in one place, the AK Body Toning Device provides and extra measure of safety from consequences like skin burns, while effectively delivering electrical stimulation to a broader area during a treatment session. Thus, the AK Body Toning Device is substantially equivalent to existing predicate devices with the NGX product code.
| Parameter | AK Beauty Enterprises,<br>LLC | Compex Sport Plus | |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| <b>BASIC UNIT CHARACTERISTICS</b> | | | |
| 510(k) Number | K152420 | K083140 | |
| Device Name and Model | AK Body Toning Device | Compex Sport Plus | |
| Manufacturer | AK Beauty Enterprises,<br>LLC | DJO LLC | |
| FDA Product Code | NGX | NGX | |
| Power Source(s)† | Battery<br>3 x 1.5 V<br>AAA batteries | Rechargeable Ni-Mh<br>Battery 4.6V (4 cells<br>AA=R6) | |
| - Method of Line Current Isolation | N/A - Internal Power<br>Type BF | N/A (battery operated<br>device) | |
| - Patient Leakage Current†† | N/A (battery operated<br>device) | N/A (battery operated<br>device) | |
| - Normal Condition (μΑ) | N/A (battery operated<br>device) | N/A (battery operated<br>device) | |
| - Single Fault Condition (μΑ) | N/A (battery operated<br>device) | N/A (battery operated<br>device) | |
| Number of Output Modes††† | 1 | One (NMES) | |
| Number of Output Channels:†††† | Synchronous or<br>Alternating? | N/A Single channel. | |
| | Method of Channel<br>Isolation | N/A Single Channel | |
| | | 4<br>Synchronous, but never<br>2 channels activated at<br>the same time | |
| | | Each channel is the<br>middle of a H-Bridge.<br>Except when it is<br>activated, each channel<br>is always in high<br>impedance state | |
| Regulated Current or Regulated Voltage? | Voltage | Regulated current (all<br>channels) | |
| Parameter | AK Beauty Enterprises,<br>LLC | Compex Sport Plus | |
| Software/Firmware/Microprocessor<br>Control? | No | Yes | |
| Automatic Overload Trip? | No | Yes | |
| Automatic No-Load Trip? | No | Yes | |
| Automatic Shut Off? | Yes | "On/Off" switch | |
| User Override Control? | Yes | Yes | |
| Indicator Display: | On/Off Status?<br>Yes<br>Low Battery?<br>No<br>Voltage/Current<br>Level?<br>Proportional Voltage level indicator | Yes<br>Yes<br>Yes, unit = [Energy] | |
| Timer Range (minutes) | 5 min | Maximum = 55 minute<br>Screen shows remaining<br>time in minutes and<br>seconds and displays<br>countdown timer | |
| Compliance with Voluntary Standards? | -IEC 60601 1: 2005 +<br>CORR. 1 (2006) + CORR.<br>2 (2007) + AM1 (2012)<br>-IEC 60601-1-2:2007<br>(Third Edition)<br>-UL 60601-1 (2003)<br>CSA C22.2 No. 601.1<br>(2003)<br>-IEC 60601-2-10: 2012<br>(Second Edition) | IEC 60601-1<br>IEC 60601-2-10 | |
| Compliance with 21 CFR 898? | N/A (electrodes are<br>integral with the device,<br>there are no separate<br>leads) | | |
| Weight (Ibs., oz.) | 800 g | 300 [g] | |
| Dimensions (in.) [W x H x D] | 3.12 x 3.51x 7.0 (IN) | 99 x 142 x 36 [mm]<br>3.9 x 5.6 x 1.4 [in] | |
| Housing Materials and Construction | ABS | | |
| OUTPUT SPECIFICATIONS | | | |
| Mode or Program Name | | | |
| Waveform (e.g., pulsed monophasic,<br>biphasic) | Monophasic | Symmetrical Biphasic | |
| Shape (e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Rectangular | |
| Maximum Output Voltage (volts) (+/-<br>10%) | 10.4 V @500Ω<br>35.6 V @ 2 kΩ<br>88.0 v @10 kΩ | 60 V @500Ω<br>165 V @ 2 kΩ<br>165 V @ 10 kΩ | |
| Maximum Output Current (specify units)<br>(+/- 10%) | 22 mA @500Ω<br>18 mA @ 2 kΩ | 120 mA<br>82 mA @ 2 kΩ | |
| Parameter | | AK Beauty Enterprises, LLC | Compex Sport Plus |
| Duration of primary (depolarizing) phase† (µsec) | | 9 mA @10 kΩ | 16 mA @ 10 kΩ |
| Max Pulse Duration† (µsec) | | 1000 µsec | 800 µsec |
| Frequency† (Hz) [or Rate† (pps)] | | 75 Hz | 1 - 120 Hz |
| For interferential modes only: Beat Frequency† (Hz) | | N/A | |
| For multiphasic waveforms only: | Symmetrical phases? | N/A (not multiphasic) | Symmetrical, 400 µsec ea phase |
| | Phase Duration† (include units), | | |
| | (state range, if applicable), | | |
| | (both phases, if asymmetrical) | | |
| Net Charge (microcoulombs (µC) per pulse) (If zero, state method of achieving zero net charge.) | | 22.5 µC @500Ω | 0 µC @ 500Ω<br>Excitation pulse fully compensated |
| Maximum Phase Charge, (µC) | | 22.5 µC @500Ω | 48 [µC] @ 500Ω |
| Maximum Current Density,†† (mA/cm², r.m.s.) | | 4.17 mA/cm² @500Ω ? | 1.49 mA/cm² @ 500Ω |
| Maximum Average Power Density,†† (W/cm²), (using smallest electrode conductive surface area) | | 0.0089W/cm² @ 500Ω | 27.6 mW/cm² @500Ω<br>0.0276 W/cm² @ 500Ω |
| Burst Mode††† | (a) Pulses per burst | N/A, no burst mode | |
| (i.e., pulse trains): | (b) Bursts per second | | |
| | (c) Burst duration (seconds) | | |
| | (d) Duty Cycle: Line (b) x Line (c) | | |
| ON Time (seconds) | | | |
| OFF Time (seconds) | | | |
| Additional Features (specify, if applicable) | | Electrodes are integrated within the unit | |
| Parameter | AK Beauty Enterprises,<br>LLC | Compex Sport Plus | |
| Intended Use | The AK Body Toning<br>Device is intended to for<br>use by healthy persons<br>to apply transcutaneous<br>electrical muscle<br>stimulation (EMS)<br>through skin-contact<br>electrodes in order to<br>improve or facilitate<br>muscle performance. | The Compex® Sport Plus is<br>intended to stimulate<br>healthy muscles in order<br>to improve or facilitate<br>muscle performance. | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Performance testing
A study observed muscle movement by ultrasound imaging. Real-time stimulation of the anterior thigh muscle was recorded using the AK device (while being held stationary and while being rolled) and the predicate device. The muscle stimulation was similar for the AK Body Toning Device and the predicate device.
### Biocompatibility testing
The biocompatibility evaluation for the skin contact materials used in the AK Body Toning Device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
# Electrical safety and electromagnetic compatibility (EMC)
Performance and electrical safety testing were conducted on the AK Body Toning Device. The test results demonstrate that the device safely performs as intended and raised no new questions of safety or performance compared to the predicate devices. Internationally recognized standards include:
- . IEC 60601 1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012)
- . IEC 60601-1-2:2007 (Third Edition)
- UL 60601-1 (2003)
- . CSA C22.2 No. 601.1 (2003)
- IEC 60601-2-10: 2012 (Second Edition) ●
#### VIII. CONCLUSIONS
The AK Body Toning Device is substantially equivalent the predicate device. They both are intended to stimulate healthy muscles to improve or facilitate muscle performance. They stimulate the muscles by delivering electrical current through electrodes to the skin to cause muscle contraction. The predicate is indicated for several muscle groups but the AK Body Toning Device is limited to the abdomen, buttocks and thighs. Their technological characteristics differ in that the AK Body Toning Device uses rolling electrodes but the predicate device uses fixed electrodes. The data indicates that the AK Body toning Device works as safely and effectively as the predicate device.
{8}------------------------------------------------
--- END OF 510(K) SUMMARY ---
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.