Stingray LP Catheter

K152401 · Boston Scientific Corporation · DQY · Dec 4, 2015 · Cardiovascular

Device Facts

Record IDK152401
Device NameStingray LP Catheter
ApplicantBoston Scientific Corporation
Product CodeDQY · Cardiovascular
Decision DateDec 4, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The Stingray LP Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the coronary and peripheral vasculature. When used as part of the system consisting of the CrossBoss™ Catheter, Stingray LP Catheter, and Stingray Guidewire, the Stingray LP Catheter is indicated for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention.

Device Story

Stingray LP Catheter is a percutaneous catheter used in coronary and peripheral vasculature; facilitates guidewire placement, support, and steering. Device features a distal balloon for fluoroscopic orientation and two side-ports identified by radiopaque markers; side-ports allow guidewire exit at an angle to the central lumen. Used by physicians in clinical settings; compatible with 6F guide catheters and ≤0.014 in guidewires. Operates as part of a system with CrossBoss Catheter and Stingray Guidewire. Physician uses fluoroscopic imaging to visualize radiopaque markers for orientation; enables precise guidewire steering to navigate complex stenotic lesions or CTOs. Facilitates successful guidewire crossing of lesions, potentially improving procedural success rates for PTCA or stent intervention.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, sensitization, systemic toxicity, hemocompatibility, etc.) and in-vitro performance (tensile strength, inflation/deflation time, crossing profile, torque, burst pressure, and coating integrity) per FDA guidance for PTCA catheters.

Technological Characteristics

Percutaneous catheter; polymer construction with hydrophilic coating; distal balloon for orientation; radiopaque markers for side-port identification. Compatible with 6F guide catheters (min ID 0.070 in) and ≤0.014 in guidewires. Sterilized via Ethylene Oxide. Mechanical design includes manifold and strain relief.

Indications for Use

Indicated for patients requiring guidewire access to coronary and peripheral vasculature, including those with stenotic coronary lesions or chronic total occlusions (CTOs) requiring intraluminal guidewire placement prior to PTCA or stent intervention.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a way that resembles a bird in flight. The profiles are depicted in black. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 4, 2015 Boston Scientific Corporation Kurtis Hunsberger Principal Regulatory Affairs Specialist, Interventional Cardiology One Scimed Place Maple Grove, MN 55311 Re: K152401 Trade/Device Name: Stingray LP Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 5, 2015 Received: November 6, 2015 Dear Kurtis Hunsberger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K152401 Device Name Stingray™ LP Catheter Indications for Use (Describe) The Stingray LP Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the coronary and peripheral vasculature. When used as part of the system consisting of the CrossBoss™ Catheter, Stingray LP Catheter, and Stingray Guidewire, the Stingray LP Catheter is indicated for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><span style="font-size:20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary per 21 CFR §807.92 | Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311<br>Phone: 763-494-1700<br>Fax: 763-494-2222 | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Name<br>and<br>Information | Kurtis Hunsberger<br>Principal Regulatory Affairs Specialist<br>Phone: 763-494-1204<br>Fax: 763-494-2222<br>e-mail: kurtis.hunsberger@bsci.com | | Date Prepared | August 24, 2015 | | Proprietary<br>Name | StingrayTM LP Catheter | | Common Name | Percutaneous Catheter | | Product Code | DQY | | Classification | Class II, 21 CFR Part 870.1250 | | Predicate<br>Device | BridgePoint Medical<br>Stingray TM Orienting Balloon Catheter<br>K102725 May 10, 2011 | | Device<br>Description | The Stingray LP Catheter facilitates the placement, support and steering<br>of a guidewire into discrete regions of the coronary and peripheral<br>vasculature through the central guidewire lumen or through one of two<br>side-ports. These side-ports are on opposite sides of the balloon and are<br>identified by radiopaque markers. The side-ports communicate with the<br>central guidewire lumen and facilitate guidewire steering (at an angle to<br>the central lumen) by allowing the guidewire to exit the Stingray LP<br>Catheter. The Stingray LP Catheter contains a small balloon used for<br>fluoroscopic orientation on the distal tip of a flexible shaft. The distal end<br>of the catheter is hydrophilic coated. The Stingray LP Catheter is<br>compatible with 6F guide catheters with minimum inner diameter of 0.070<br>in (1.7 mm), and may be used with guidewires ≤0.014 in (0.36 mm). | | Intended Use /<br>Indications for<br>Use | The Stingray LP Catheter is indicated for directing, steering, controlling,<br>and supporting a guidewire in order to access discrete regions of the<br>coronary and peripheral vasculature.<br>When used as part of the system consisting of the CrossBossTM Catheter,<br>Stingray LP Catheter, and Stingray Guidewire, the Stingray LP Catheter is<br>indicated for use to facilitate the intraluminal placement of conventional<br>guidewires beyond stenotic coronary lesions (including chronic total<br>occlusions [CTOs]) prior to PTCA or stent intervention. | {4}------------------------------------------------ | Comparison of<br>Technological<br>Characteristics | The Stingray™ LP Catheter incorporates substantially equivalent device<br>materials and design, packaging materials and design, fundamental<br>technology, manufacturing processes, sterilization process and intended<br>use as those featured in the Stingray™ Orienting Balloon Catheter<br>K102725 cleared May 10, 2011. Similarities and differences in<br>technological characteristics between the predicate and subject device<br>are listed below.<br>Similarities:<br>Polymer material construction Hydrophilic coating Radiopaque proximal markers Manifold and strain relief Identical balloon segment Ethylene Oxide sterilization Packaging design with same function Differences: | | |---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------| | | Shaft Outer Diameter: Stingray LP has a lower profile than the<br>predicate Stingray. | | | Performance<br>Data | The Stingray™ LP Catheter was subjected to testing according to the<br>requirements of <i>Guidance for Industry and FDA Staff – Class II Special<br/>Controls for Certain Percutaneous Transluminal Coronary Angioplasty<br/>(PTCA) Catheters</i> , September 8, 2010. Bench testing and<br>biocompatibility testing were performed to support a determination of<br>substantial equivalence. The results of these tests provide reasonable<br>assurance that the proposed device has been designed and tested to<br>assure conformance to the requirements for its intended use. No new<br>safety or performance issues were raised during the testing; therefore,<br>this device may be considered substantially equivalent to the predicate<br>device. | | | | The following biocompatibility and chemical characterization tests were<br>completed on the Stingray™ LP Catheter: | | | | Cytotoxicity | Hemolysis (Extract Method) | | | Sensitization | Partial Thromboplastin Time | | | Intracutaneous Reactivity | <i>In Vitro</i> Hemocompatibility | | | Acute Systemic Toxicity | Complement Activation | | | Materials Mediated Pyrogenicity | <i>In Vivo</i> Thromboresistance | | | Hemolysis (Direct Contact) | USP <661> Physicochemical | {5}------------------------------------------------ | | The following in-vitro performance tests were completed on the Stingray™<br>LP Catheter: | | |------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------| | | Effective Length | Inflation and Deflation Time | | | Shaft Outer Diameter | Full Unit Tensile | | | Crossing Profile | Distal Reflow Bond Yield Strength | | | Delivery | Kink – Proximal End | | | Catheter Torque / Guidewire<br>Movement | Kink – Distal End of Gradient Outer | | | Guidewire Movement<br>at Rated Burst Pressure | Kink – Distal End | | | Withdrawal into a Guide | Catheter Torque | | | Two Catheters in a 7F Guide | Marker Visibility | | | Rated Burst Pressure (RPB) | System Radiopacity | | | Balloon Burst Mode | Coating Integrity (Balloon and Shaft) | | | Repeat Inflation | Particulates (Sim Use) | | | Balloon Dimensional | Corrosion Resistance | | Conclusion | Based on the indications for use, technological characteristics, and safety<br>and performance testing, the Stingray™ LP Catheter has been shown to<br>be appropriate for its intended use and is considered to be substantially<br>equivalent to the Stingray™ Orienting Balloon Catheter. | |
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%