K152337 · AtriCure, Inc. · GEH · Mar 22, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152337
Device Name
cryoFORM cryoICE cryoablation probe
Applicant
AtriCure, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Mar 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
AtriCure's cryoICE cryoFORM cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
Device Story
The cryoFORM cryoablation probe is a sterile, single-use surgical instrument used in conjunction with the AtriCure Cryo Module. It functions by circulating nitrous oxide to a cryotip end effector, which freezes target cardiac tissue to create cryonecrosis and block electrical conduction pathways. The device features a malleable, corrugated stainless steel cryotip and flexible tubing to accommodate varying patient body habitus and surgeon preference. It is operated by surgeons in a clinical or surgical setting. The healthcare provider uses the probe to perform targeted ablation; the resulting tissue lesion interrupts arrhythmia-causing electrical signals, potentially benefiting patients by restoring normal heart rhythm. The device includes a thermocouple for temperature monitoring during the procedure.
Clinical Evidence
No human clinical data. Evidence consists of a GLP animal study using a porcine thigh tissue model (n not specified). Primary endpoints were lesion width and depth, compared against the predicate CRYO2 device. Statistical equivalence was demonstrated using two one-sided tests (TOST). Additional non-clinical bench testing included mechanical, reliability, cryogen performance, sterilization validation, LAL, accelerated aging, and transit testing. Biocompatibility testing followed ISO 10993-1.
Technological Characteristics
Single-use cryosurgical probe; stainless steel corrugated malleable cryotip; flexible tubing; rigid shaft; thermocouple for temperature sensing. Energy source: Nitrous oxide cryogen via AtriCure Cryo Module. Sterilization: Gamma irradiation. Biocompatible materials per ISO 10993-1. External communicating device, tissue/bone contact, <24 hours duration.
Indications for Use
Indicated for patients requiring cryosurgical treatment of cardiac arrhythmias via freezing of target tissues to block electrical conduction pathways.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping heads suggesting a sense of community and interconnectedness. Above the figure are three curved lines, possibly symbolizing growth or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2016
Atricure, Inc % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K152337
Trade/Device Name: cryoFORM cryoICE Cryoablation Probe Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH, OCL Dated: March 7, 2016 Received: March 8, 2016
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152337
Device Name
AtriCure® cryoICE™ cryoFORM cryo-ablation probe (CRYOF)
Indications for Use (Describe)
AtriCure's cryoICE cryoFORM cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the AtriCure logo. The word "Atri" is in blue, and the word "Cure" is in orange. There is a blue dot above the letter "C" in "Cure". The logo is simple and modern.
## 510(k) Summary
#### l. Submitter
| Manufacturer: | AtriCure, Inc.<br>6217 Centre Park Dr.<br>West Chester, OH 45069<br>P: 513-755-4100<br>F: 513-755-4108 | |
|--------------------|--------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Jonathan McElwee, RAC<br>Senior Regulatory Specialist | |
| Alternate Contact: | Jim Taufen<br>Director of Regulatory Affairs | |
| Date Prepared: | 08/11/2015 | |
#### ll. Device
| Name of Device: | AtriCure® cryoICE™ cryoFORM™ cryoablation probe<br>(CRYOF) |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Common Name: | Cryosurgical probe |
| Classification Name: | Unit, Cryosurgical, Accessories<br>Surgical, General and Plastic Surgery, 878.4350 |
| | Surgical Device, For Cutting, Coagulation, And/Or Ablation Of Tissue,<br>Including Cardiac Tissue, 878.4400 |
| Regulatory Class: | Class II |
| Product Code: | GEH and OCL |
#### lll. Predicate Device
The primary predicate device, AtriCure cryolCE cryoablation probe (CRYO2), was cleared via 510(k) K142203 on December 8, 2014 under the Product Code GEH.
A secondary predicate device, Heartport Maze System: Cryoprobe Set, was cleared via 510(k) K970496 on May 9, 1997 under the Product Code OCL.
The predicate devices have not been subject to a design-related recall.
No reference devices were used in this submission.
{4}------------------------------------------------
## IV. Device Description
AtriCure's cryolCE cryoFORM cryoablation probe (CRYOF) is a sterile, single use, cryosurgical device to be used in coniunction with the AtriCure Crvo Module [K111042. K11207. K140058, K142203) to freeze target tissue, blocking the electrical conduction pathways by creating an inflammatory response or cryonecrosis.
#### V. Indications For Use
AtriCure's cryolCE cryoFORM cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
#### VI. Comparison Of Technological Characteristics With The Predicate Device
- Same intended use and; .
- . Same operating principle
- Same fundamental scientific technology
- Used with the same cryo module unit
- . The basic design of the proposed device and the previously cleared devices are the same. The devices are disposable, single-use instruments including a cryotip end effector, shaft. handle, thermocouple, inlet tube, and exhaust tube.
- . The modifications to the proposed cryolCE probe are designed to provide increased options for surgeons based on patient body habitus and surgeon preference. Major modifications compared to the predicate include a more malleable corrugated stainless steel cryotip. more flexible tubing, and a different colored rigid shaft.
{5}------------------------------------------------
# AtriCure®
The AtriCure cryoFORM cryolCE cryoablation probe has the following similarities to the technological characteristics of the previously cleared predicate AtriCure cryolCE cryoablation probe K142203:
| Feature | AtriCure cryolCE™ cryoablation probe<br>(CRYO2)<br>(K142203) | AtriCure cryoFORM cryolCE cryoablation probe<br>(CRYOF)<br>(Proposed Device) |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | AtriCure, Inc. | Same |
| Proprietary Name | AtriCure cryoICE cryoablation probe | AtriCure cryoFORM cryoICE cryoablation probe |
| Product Code(s) | CRYO2 | CRY OF |
| 510k Reference | K142203 | Subject of this submission |
| Intended Use | AtriCure's cryoICE cryoablation probes are<br>sterile, single use devices intended for use in<br>the cryosurgical treatment of cardiac<br>arrhythmias by freezing target tissues,<br>creating an inflammatory response<br>(cryonecrosis) that blocks the electrical<br>conduction pathway. The probe is also<br>intended for use in blocking pain by<br>temporarily ablating peripheral nerves. | AtriCure's cryoFORM cryolCE cryoablation probes<br>are sterile, single use devices intended for use in<br>the cryosurgical treatment of cardiac arrhythmias<br>by freezing target tissues, creating an inflammatory<br>response (cryonecrosis) that blocks the electrical<br>conduction pathway. |
| Device Description | The probe is comprised of a cryotip end<br>effector, shaft, handle, thermocouple, inlet<br>tube, and exhaust tube. | The probe is comprised of a cryotip end effector,<br>shaft, handle, thermocouple, inlet tube, and<br>exhaust tube. |
| Cryotip | Material: Aluminum alloy<br>Construction: Smooth malleable probe | Material: Stainless Steel<br>Construction: Corrugated malleable probe |
| Method of Cryogen<br>Activation | Switch on AtriCure Cryo Module | Switch on AtriCure Cryo Module |
| Cryogen | Nitrous Oxide | Nitrous Oxide |
| Packaging | Sterile - single use, disposable device | Sterile - single use, disposable device |
| Biocompatibility | Biocompatible patient contacting materials | Biocompatible patient contacting materials |
| Sterilization | Gamma Irradiation | Gamma Irradiation |
{6}------------------------------------------------
### VII. Performance Data
The modified cryolCE cryoablation probe was successfully tested on an animal model to confirm the modifications do not affect the ability to successfully ablate cardiac tissue.
## Thigh Prep Study Testing
AtriCure conducted a GLP animal study to evaluate the performance of the CRYOF device in a live porcine thigh tissue model and to compare performance to the predicate CRYO2 device. Lesions were created by each of the cryoablation probes on porcine thigh muscle. The lesions were assessed using histopathology and morphometric measurements and the results demonstrated statistical equivalence via two one-sided tests (TOST) for lesion width and depth.
Additional testing per 21 CFR 820.30 and AtriCure's Quality System was performed to verify the modified cryoablation probe's conformance to design controls and specification. Testing determined that the modified cryoablation probe conformed to design controls and product specifications.
Non-clinical Bench Testing
- Mechanical Testing
- Reliability Testing ●
- Cryogen Performance Testing
- . Bench Testing on an Animal Model
- . Sterilization Validation
- . LAL Testina
- Accelerated Aging Testing .
- . Transit Testing
## Biocompatibility Testing
The biocompatibility evaluation for the cryolCE crypablation probe was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA. The battery of testing included the following tests:
- . Cytotoxicity
- Sensitization .
- . Irritation
- Systemic Toxicity ●
- . Material Mediated Pyrogenicity
The cryolCE cryoFORM cryoablation probe is considered an "External Communicating Device," contact for "Tissue/Bone" and contact duration for "under 24 hours."
#### VIII. Conclusions
The proposed cryolCE cryoFORM cryoablation probe (CRYOF) is substantially equivalent to the previously cleared cryolCE cryoablation probe (CRYO2) as there is no change to intended use, the operating principle, or the basic design of the cryoablation probe. The totality of evidence supports a decision that the modifications to the cryoablation probe have been successfully evaluated for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.