Fujifilm Duodenoscope Model ED-530XT

K152257 · Fujifilm Medical Systems U.S.A, Inc. · FDT · Jul 21, 2017 · Gastroenterology, Urology

Device Facts

Record IDK152257
Device NameFujifilm Duodenoscope Model ED-530XT
ApplicantFujifilm Medical Systems U.S.A, Inc.
Product CodeFDT · Gastroenterology, Urology
Decision DateJul 21, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Fujifilm Duodenoscope Model ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Story

Flexible video duodenoscope; used for visualization, diagnosis, and treatment of upper digestive tract (esophagus, stomach, duodenum). Input: optical images captured via distal CCD sensor; illumination provided by internal glass fiber bundles. Operation: physician-controlled distal angulation (up/down/left/right) via operation section; channels provide air/water, suction, and access for endoscopic accessories (e.g., biopsy forceps). Output: real-time video displayed on external monitor via connected video processor/light source. Used in clinical settings (e.g., endoscopy suites) by trained physicians. Benefits: enables minimally invasive visualization and therapeutic intervention in the upper GI tract.

Clinical Evidence

Bench testing only. Evaluated per IEC 60601-1-2:2007, ANSI/AAMI ES60601-1:2005, IEC 60601-2-18:2009, ISO 10993 (biocompatibility), and ISO 8600-1:2013. Reprocessing validated per AAMI TIR12:2010, AAMI TIR30:2011, and FDA 2015 guidance. Additional testing confirmed specifications for field of view, bending, air/water/suction rates, resolution, light output, and seal integrity.

Technological Characteristics

Flexible endoscope with operation section, insertion portion, and umbilicus. Features CCD image sensor, glass fiber bundles for illumination, and multiple channels (air/water, suction, forceps). Dimensions and materials modified from predicate. Connectivity: connects to external video processors/light sources. Sterilization: validated reprocessing procedures. Standards: IEC 60601-1-2, ANSI/AAMI ES60601-1, IEC 60601-2-18, ISO 10993, ISO 8600-1.

Indications for Use

Indicated for visualization, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum in patients requiring upper gastrointestinal endoscopy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with a stylized wing-like shape extending from the top profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 21, 2017 FUJIFILM Medical Systems U.S.A., Inc. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, DC 20004 Re: K152257 > Trade/Device Name: Fujifilm Duodenoscope Model ED-530XT Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDT Dated: May 30, 2017 Received: May 30, 2017 Dear John J. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ Page 2 - John Smith, M.D., J.D. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and appears to be centered. The background is a light color, possibly white, with a faint watermark or design element visible behind the text. Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K152257 Device Name Fujifilm Duodenoscope Model ED-530XT Indications for Use (Describe) The Fujifilm Duodenoscope Model ED-530XT is intended for the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. | Type of Use (Select <i>one</i> or <i>both</i> , as applicable) | |----------------------------------------------------------------| |----------------------------------------------------------------| 区 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY ## FUJIFILM Medical Systems U.S.A., Inc.'s Fujifilm Duodenoscope Model ED-530XT #### Submitter's Information FUJIFILM Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293 ## Contact Person: Shraddha More Specialist, Regulatory Affairs and Quality Assurance Telephone: (973) 686-2627 Facsimile: (973) 633-8818 E-Mail: smore@fujifilm.com Larry Picciano Senior Director, Quality and Regulatory Affairs Telephone: (973) 686-2479 Facsimile: (973) 633-8818 E-Mail: larry.picciano@fujifilm.com #### Date Prepared: May 30, 2017 #### ldentification of the Subject Device: | Proprietary/Trade Name: | Fujifilm Duodenoscope Model ED-530XT | |-------------------------|--------------------------------------| | Common Name: | Video Endoscope | | Device Class: | Class II | | Review Panel: | Gastroenterology/Urology | #### Classification Information: Endoscope and Accessories (Flexible/Rigid), 21 C.F.R. § 876.1500 Product Code: FDT ### Primary Predicate Fujinon G5 Duodenoscope ED-450XT5 / ED-250XT5 (K042076) {4}------------------------------------------------ # Intended Use / Indications for Use The Fujifilm Duodenoscope Model ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. # Technological Characteristics The Fujifilm Duodenoscope Model ED-530XT is comprised of three main sections: an operation section, an insertion portion, and an umbilicus. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion portion contains glass fiber bundles, several channels and a CCD image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD sensor to capture an image and display it on the monitor. The endoscope also contains several channels to deliver air/water and provide suction, as well as a forceps channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps during the procedure. The umbilicus section consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The subject device is used in combination with Fujifilm's EPX-4400 (K944620), EPX-4400HD (K944620), and EPX-4440HD (K102466, K140149) video processors and light sources, as well as peripheral devices such as water tank, endoscope accessories, monitor, printer, electrosurgical instruments, foot switch, and cart. All of the peripheral accessories are substantially similar to those cleared in the predicate submission, with the exception of a new single use Elevator Cleaning Brush WB1318DE. The differences between the subject and predicate device are as follows: - Distal end design differences (e.g., changes in materials and dimensions); - . Reduction in the diameter of the flexible portion and insertion portion; - Optical differences (e.g., modified viewing direction and CCD sensor); and ● - Modifications to the device's labeled reprocessing procedure. ## Performance Data EMC and electrical safety of the subject devices were evaluated using the following consensus standards: IEC 60601-1-2:2007; ANSI/AAMI ES60601-1:2005; and IEC 60601-2-18:2009. Biocompatibility of the subject device was evaluated in accordance with ISO 10993. Cleaning, disinfection, and sterilization were evaluated according to the following consensus standards: AAMI TIR12:2010, AAMI TIR30:2011. Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's quidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015. Endoscope specific testing was conducted using the consensus standard ISO8600-1:2013. {5}------------------------------------------------ The subject device met performance specifications in the following additional testing: - Field of View ● - Bending capability - Rate of air supply ● - Rate of water supply ● - Rate of suction - Resolution ● - Light Guide (LG) Output ● Performance testing and design analysis was conducted to demonstrate the safety and effectiveness of the elevator wire channel seal, as well as distal end cap seal strength and integrity. # Substantial Equivalence The ED-530XT is substantially equivalent the ED-450XT5. The ED-530XT has the same intended use, indications, and principles of operation, and similar technological characteristics, as its predicate device. The minor technological differences between the ED-530XT and its predicate devices were made for the purpose of overall product enhancement and general technological advancement, and do not raise different questions of safety or effectiveness. Performance data demonstrate that the ED-530XT is substantially equivalent to the ED-450XT5. Thus, the ED-530XT is substantially equivalent to the predicate device. ## Conclusions Based on device description, comparison to predicate device, and performance bench data described above, Fujifilm believes that the ED-530XT is substantially equivalent to the predicate device ED-450XT5 based on intended use/indications for use and technological characteristics. The minor technological differences between the subject endoscope and its predicate device raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject endoscope is as safe and effective, and performs as well or better, than the predicate device. Thus, the device is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...