FUJINON EPX-4440HD DIGITAL VIDEO PROCESSOR

K102466 · Fujinon, Inc. · FET · May 25, 2011 · Gastroenterology, Urology

Device Facts

Record IDK102466
Device NameFUJINON EPX-4440HD DIGITAL VIDEO PROCESSOR
ApplicantFujinon, Inc.
Product CodeFET · Gastroenterology, Urology
Decision DateMay 25, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VP-4440HD unit is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope (a video camera in an endoscope). This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope. light source, monitor, recorder and various peripherals. The XL-4450 light source is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to provide illumination to an endoscope. The light source also functions as a pump to supply air through the endoscope while inside the body to assist in obtaining clear visualization to facilitate diagnostic examination. This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, digital video processor, monitor, recorder and various peripherals.

Device Story

System comprises VP-4440HD Video Processor, XL-4450 Light Source, and DK-440E Keyboard. Processor receives electronic signals from video endoscope; transforms signals into visual output for monitor; provides optional spectral/structural image enhancement via proprietary software; supports digital/analog storage. Light source uses 300W Xenon lamp for illumination and air pump for visualization. Keyboard allows user entry of procedural data and control of processor features. Used in clinical settings by physicians/technicians for endoscopic procedures. Output enables real-time visualization, diagnosis, and treatment; digital storage facilitates record-keeping. Benefits include improved visualization and procedural documentation.

Clinical Evidence

No clinical data. Non-clinical testing included software validation, electrical safety testing per IEC 60601, and physician-led visual comparison/performance evaluations of spectral enhancement capabilities.

Technological Characteristics

Steel-polycarbonate housing. 300W Xenon lamp with 75W Halogen emergency backup. Connectivity: Analog/digital outputs, Ethernet for off-site storage, internal/CF card storage. Connectors for 400 and 500 series endoscopes. Software-based spectral and structural image enhancement. AC powered (120V/60Hz).

Indications for Use

Indicated for all patients requiring endoscopic examination, diagnosis, or treatment using Fujinon endoscopes, light sources, and peripheral equipment.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary - # Fuiinon EPX-4440 HD Digital Video Processor MAY 25 2011 Date Prepared: May 25, 2011 # Sponsor: Fujinon, Inc. 10 High Point Drive Wavne, NJ 07474 # Contact: Gina Walljasper Director, Quality and Regulatory Compliance Tel: (973) 633-5600 Fax: (973) 633-8818 E-mail: gina walljasper@fujinon.com Trade name: Fujinon EPX-4440HD Video Processor and Light Source Common name: Endoscopic Video Processor and Light Source # Classification name: - Endoscopic Video Imaging System/Component, Gastroenterology-Urology (Product Code FET) and Light Source, Endoscope, Xenon Arc (Product Code GCT) - Both product codes are covered under 21 CFR 876.1500 - Predicate device(s): Fujinon 400 Series (K944620) # Device description: This system is intended to be used in conjunction with Fujinon endoscopes to provide illumination, visual display and data storage during endoscopic procedures. The EPX-4440HD Digital Video Processor consists of three components used in conjunction with one another: The VP-440HD Video Processor: The Processor relays the image from the endoscope to a video monitor. Projection can be either analog or digital at the user's {1}------------------------------------------------ preference. The Processor also incorporates internal or external digital storage capacity. The Processor also controls the light projected to the body cavity. The Processor also provides for optional structural enhancement at the user's option. Spectral and structural enhancements are achieved through proprietary software. The device is AC operated at a power setting of 120V/60MzJO.8A. The Processor is housed in a steel-polycarbonate case measuring 390X105X460mm. The XL-4450 Light Source: The Fujinon endoscope employs fiber bundles to transmit light from the light source and subsequently to the body cavity. The Light Source employs a 300W Xenon lamp with a 75W emergency back-up Halogen lamp. Brightness control is performed by the user. The device is AC operated at a power setting of 120V/60MzJ3.3A. The Processor is housed in a steel-polycarbonate case measuring 390X155X485mm. The DK-440E Keyboard: The Keyboard is used to enter pertinent procedural information (patient, physician, date, etc) for display on the video monitor and digital/analog storage systems. The Keyboard is also used to control operational features of the VP-4440HD Processor. The Keyboard resembles a standard computer keyboard in size and shape. ### Intended Use(s): The VP-4440HD unit is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope (a video camera in an endoscope). This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope. light source, monitor, recorder and various peripherals. The XL-4450 light source is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to provide illumination to an endoscope. The light source also functions as a pump to supply air through the endoscope while inside the body to assist in obtaining clear visualization to facilitate diagnostic examination. This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, digital video processor, monitor, recorder and various peripherals. Technological characteristics: The subject device is a modification of the current Fujinon Video Processor. Basic technological characteristics have not changed. The Video Processor, Light Source and Keyboard remain essentially unchanged from the predicate device with the exception of additional cabling connections. The processor has been updated to allow for both digital and analog outputs. Previously these outputs were analog only with no digital storage capacity. Digital storage capability can be achieved by on-site storage (internal or CF card) or off-site (Ethernet). {2}------------------------------------------------ The Processor now incorporates a new connector for the 500 series of Fujinon endoscopes as well as the original series 400 scopes. The above changes were made for the purposes of user preference and general technological advancement: they were not the result of product recalls or · adverse events. # Non-clinical testing: The EPX-4440HD Digital Video Processor is non-sterile and has no potential for patient contact. Testing of the VP-4440HD consisted of software validation electrical safety in accordance with IEC 60601 requirements and performance testing using physician evaluators. All testing criteria were met. Non-clinical testing of the subject device consisted of electrical testing in accordance with IEC 60601 requirements, software validation in accordance with IEC 62304 requirements and physician evaluation of the spectral enhancement capability. Functional testing consisting of visual comparison tests were conducted using physician evaluators. In all cases they passed the required testing regimens. The sponsor believes that the testing regimen demonstrates the continued safety of the subject device. # Clinical testing: # NA # Conclusions: Modifications to the predicate device consisted of: Software additions: Allowing for either digital or analog outputs Addition of optional digital memory storage Adding an additional port for acceptance of newer generation endoscopes Changes to the endoscopes were made for purposes of ergonomics and user, preferences. The changes were not made for reasons of product safety or adverse events. Testing was performed to demonstrate the continued electrical safety of the device and the reliability of the software. The sponsor believes that the modified device maintains the same safety and performance levels as the previous generation processor while adding additional features for user convenience. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Fujinon, Inc. c/o Robert Schiff, Ph.D., RAC, CQA, FRAPS Regulatory Affairs Consultant Schiff & Company, Inc. 1129 Bloomfield Avenue WEST CALDWELL NJ 07006 MAY 25 2011 Re: K102466 Trade/Device Name: Fujinon EPX-4440HD Digital Processor and Light Source Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET and GCT Dated: May 19, 2011 Received: May 20, 2011 Dear Dr. Schiff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {4}------------------------------------------------ #### Page 2 adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael Lemon MD Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications For Use Statement 510(k) Number (If known): K102466 EPX-4440HD Digital Video Processor and Light Source Device Name: #### Indications for Use (VP-4440HD): The VP-4440HD unit is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope (a video camera in an endoscope). This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, light source, monitor, recorder and various peripheral devices. #### Indications for Use (XL-4450): The XL-4450 light source is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to provide illumination to an endoscope. The light source also functions as a pump to supply air through the endoscope while inside the body to assist in obtaining clear visualization to facilitate diagnostic examination. This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, video processor, monitor, recorder and various peripheral devices. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K102466
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...