K152167 · World Bio Tech Co., Ltd. · NJM · May 27, 2016 · Dental
Device Facts
Record ID
K152167
Device Name
BIJOU Orthodontic Ceramic Bracket
Applicant
World Bio Tech Co., Ltd.
Product Code
NJM · Dental
Decision Date
May 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BIJOU Orthodontic Ceramic bracket is indicated for orthodontic movement of natural teeth.
Device Story
BIJOU Orthodontic Ceramic Bracket is a single-crystal alumina orthodontic bracket used to facilitate tooth movement. The device consists of a wire slot, a round groove for elastic 'O' rings, and a base for adhesion to the tooth surface. It features colored markings on the wings to assist clinicians with orientation during placement. The device is applied by an orthodontist in a clinical setting. It functions as a mechanical component of an orthodontic system, where the bracket is bonded to the tooth and interacts with orthodontic wires to apply force. The translucent design provides aesthetic benefits to the patient. The device is non-sterile and intended for single use.
Clinical Evidence
No clinical performance testing was performed on the BIJOU ceramic brackets. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1, -5, -10, -12), design conformance (ISO 27020:2010), and adhesion testing (ISO 11405:2015).
Technological Characteristics
Material: Single crystal alumina (Al2O3). Design: Translucent, twin bracket, standard/Roth/MBT configurations. Dimensions: Slot sizes 0.018/0.022 inch; In-out 1.03–1.44 mm; Torque -7 to +17 degrees. Orientation: Colored dot on external surface. Standards: ISO 27020:2010 (design), ISO 11405:2015 (adhesion), ISO 14971:2012 (risk management). Sterilization: Non-sterile. Connectivity: None.
Indications for Use
Indicated for orthodontic movement of natural teeth in patients requiring orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2016
World Bio Tech Co., Ltd % Mr. Peter Chung Plus Global 300, Atwood Pittsburgh, Pennsylvania 15213
Re: K152167
Trade/Device Name: BIJOU Orthodontic Ceramic Bracket Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NJM Dated: April 8, 2016 Received: April 19, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No.0910-0120 Expiration Date:January 31, 2017 See PRA Statement below.
510(k} Number (if known)
K152167
Device Name BIJOU Orthodontic Ceramic Bracket
Indications for Use (Describe)
BIJOU Orthodontic Ceramic bracket is indicated for orthodontic movement of natural teeth.
Type of Use (Select one or both, as applicable)
A Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of infonnation is estimated to average 79 hours per response, including the time to review instructions,search existing data sources, gather and maintain the data needed and complete and review the collection of infonnation. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for World Bio Tech (WBT). The logo consists of the letters "WBT" in a bold, sans-serif font, enclosed within two overlapping ovals. The text "WORLD BIO TECH" is written in a smaller font size below the ovals. The entire logo is in a dark blue color.
#### K152167 510(k) Summary
| Submitter's Name: | WORLD BIO TECH CO., LTD. |
|-------------------------------|---------------------------------------------------------------------------------------------------|
| Submitter's Address: | #302 BizCenter, SknTechnoPark, 124, Sagimakgol-ro,<br>Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-776-2804 |
| Submitter's Fax: | +82-31-776-2805 |
| Submitter's Contact: | Jin-kyu Park |
| Date 510(k) summary prepared: | April 8, 2016 |
| Proprietary Name: | BIJOU Orthodontic Ceramic Bracket |
| Common or Usual Name: | Orthodontic Ceramic Bracket |
| Classification Name: | Orthodontic plastic bracket |
| Regulation Number: | 21 CFR 872.5470 |
| Review Panel: | Dental |
| Product Code: | NJM |
| Classification Code: | II |
| Type of 510(k) submission: | Traditional |
Predicate Device: K073045 / PURE Sapphire / Ortho Technology, Inc.
# Description of the Device
Orthodontic Ceramic bracket, BIJOU is composed of single crystal alumina and consists of three parts: the first part is the slot for the orthodontic wire; the second part is a round groove that is to hold a wire with an elastic "O" ring; the third part is the base that adheres to the tooth surface. A colored marking on wing part of bracket indicates orientation for placement.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for World Bio Tech (WBT). The logo consists of two overlapping ovals, with the letters "WBT" in bold, blue font inside the inner oval. Below the ovals, the words "WORLD BIO TECH" are written in a smaller, blue font. The logo is simple and professional, and the use of blue suggests a sense of trust and reliability.
### Indications for Use
BIJOU Orthodontic Ceramic bracket are indicated for orthodontic movement of natural teeth.
PURE Sapphire, (K073045), and BIJOU have similar technological characteristics:
| | WORLD BIO TECH CO., LTD. | Ortho Technology, Inc. |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K152167 | K073045 |
| Common/Generic Name | Orthodontic Ceramic Bracket | Sapphire Ceramic Bracket |
| Trade/Proprietary Name | BIJOU | PURE® |
| Intended use | BIJOU Orthodontic Ceramic bracket are indicated for orthodontic movement of natural teeth. | Sapphire Ceramic Brackets are indicated for the orthodontic movement of teeth. The devices are for Orthodontic Use Only, on the order of an Orthodontist. |
| Material composition of Bracket | Al2O3 (single crystal alumina) | Al2O3 (single crystal alumina) |
| Material composition of colorants for bracket placement orientation | Yellow (C.I. Food Yellow 4), Blue (C. I. Acid Blue 9, disodium salt), Red (FD&C Red No.3), and Pink (Carmine B.C.) | Not known |
| Translucent | Yes | Yes |
| Bracket design | Standard, Roth, MBT designs with and without hook, conforming to ISO 27020:2010 Dentistry – Brackets and tube for Use in Orthodontics | Standard, Roth, MBT designs with and without hook |
| Bracket In-out(mm) | 1.03 – 1.44 | 0.53 - 0.89 |
| Bracket Torque(°) | -7 to +17 | - 7 to +17 |
| Available slot sizes | 0.018 / 0.022 inch | 0.018 / 0.022 inch |
| Orientation marking | Colored dot on external surface | Colored dot on external surface |
| Single use | Yes | Yes |
| Non-Sterile | Yes | Yes |
### Substantial Equivalence Table 1: Substantial equivalence comparison
The indication for use on natural teeth stated in the submission indications for use statement, as well as the omission of the statement for orthodontic use only and on order of an orthodontist, do not change the intended use of the device as both the submission device and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for World Bio Tech (WBT). The logo consists of the letters "WBT" in a bold, sans-serif font, colored in a dark blue. The letters are enclosed within a series of overlapping blue circles, creating a visual effect of orbits or rings. Below the circles and letters, the words "WORLD BIO TECH" are written in a smaller, less prominent font, also in blue.
the predicate device are prescription devices for use in orthodontic treatment. The available bracket designs are similar; the slight variation in design of the maxillary in-out ranges is within the acceptable range for ISO 27020. The material used for the preformed translucent block form shapes. While both the submission device and the predicate device are marked with a colorant for orientation during placement and the colorants for the submission device may be different from the predicate device, the submission includes biocompatibility testing. The overall design is identical as a translucent, twin bracket with identical wire slot size;.
## Biocompatibility
Biocompatibility testing including cytotoxicity, sensitization, oral mucosal irritation was completed according to the following standards:
ISO 10993-1 Biological Evaluation of Medical Devices -Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5 Biological Evaluation of Medical Devices - Part 5 Cytotoxicity
ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization
ISO 10993-12 Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials
### Non-clinical Performance Data
Non-clinical performance testing was conducted as follows: design characteristics based on and in accordance with ISO 27020:2010 Dentistry - Brackets and tubes for use in Orthodontics; adhesive strength and analysis of detached teeth surface were conducted in accordance with ISO 11405:2015, Dentistry -Testing of adhesion to tooth structure; in-house comparative performance testing was conducted for the submission device and the predicate device, including wire slot drag test, wire slot torque test, shear test, and bracket removal test. A risk analysis was conducted based on ISO 14971:2012 Medical devices -Application of risk management to medical devices
### Clinical Data
No clinical performance testing was performed on BIJOU, ceramic brackets.
### Conclusion
Based upon the chemical composition, design, indications for use, and results of non-clinical performance testing, we conclude that our device BIJOU is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.