K152032 · Azena Medical, LLC · GEX · Sep 16, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152032
Device Name
ELUMI 810 + 980 Soft Tissue Laser
Applicant
Azena Medical, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ELUMI 810+980 Soft Tissue Laser is intended for the incision, excision, ablation, vaporization, hemostasis, and treatment of oral soft tissue. The following are the oropharyngeal indications for use: Excisional and incisional biopsies; Exposure of unerupted teeth; Fibroma removal; Gingivoplasty; Hemostasis and coagulation; Incision and drainage of abscess; Operculectomy; Pulpotomy; Reduction of gingival hypertrophy; Soft tissue crown lengthening; Vestibuloplasty; Laser Soft Tissue Curettage; Tissue retraction; Frenectomy and Frenotomy; Gingival troughing for crown impressions; Gingivectomy; Gingival incision and excision; Implant recovery; Leukoplakia; Oral papillectomnies; Pulpotomy as an adjunct to root canal therapy; Reduction of bacterial level (decontamination) and inflammation; Treatment of aphthous ulcers; Lesion (tumor) removal; Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa; Removal of Diseased, Inflamed and necrotic soft tissue within the periodontal pocket; Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium; Sulcular debridement (removal of necrotic, diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility); Light activation for bleaching materials for teeth whitening; Laser-assisted whitening/bleaching of teeth.
Device Story
Battery-powered, countertop diode laser system for oral soft tissue surgery and teeth whitening. Input: user-selected settings via interface; Output: infrared laser energy (810nm/980nm) delivered via flexible fiber optic cable and disposable tips. System includes laser unit, anodized aluminum handpiece, wireless footswitch, and auxiliary power supply. Clinician operates device to perform procedures like biopsies, gingivectomy, and sulcular debridement. Red aiming beam (650nm) and white LED illumination assist targeting. Laser energy interacts with tissue for cutting/coagulation; light activation used for bleaching. Output allows precise tissue management, hemostasis, and decontamination, benefiting patients through minimally invasive surgical options and improved clinical indices in periodontal treatment.
Clinical Evidence
No clinical data submitted. Evidence consists of bench testing (power output verification per IEC 60601-2-22, cleaning/sterilization validation) and an animal study (pig liver) comparing histological effects to the predicate device. Cleaning validation achieved < 5.4 µg/cm² protein and < 1.6 µg/cm² hemoglobin. Sterilization validation confirmed 10-6 sterility assurance level.
Technological Characteristics
Class IV GaAlAs diode laser; 810nm/980nm wavelengths; 10W peak power per diode (20W dual); 400 µm quartz glass fiber delivery; wireless footswitch; lithium-ion battery; anodized aluminum handpiece. Standards: 21 CFR 1040.10/11, IEC 60601-2-22, IEC 60825-1, AAMI/ANSI ES60601-1, ISO 10993-5, ISO 17665. Sterilization: gravity steam (135°C, 15 min).
Indications for Use
Indicated for oral soft tissue procedures including incision, excision, ablation, vaporization, hemostasis, and teeth whitening in patients requiring dental soft tissue surgery or treatment of oral mucosal lesions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines above them that resemble a bird in flight. The profiles are stacked, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2015
Azena Medical, LLC % Mr. Dave Yungvirt Third Party Review Group, LLC 45 Rockefeller Plaza Suite 2000 New York, New York 10111
Re: K152032
Trade/Device Name: ELUMI 810 + 980 Soft Tissue Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 27, 2015 Received: September 1, 2015
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152032
### Device Name ELUMI 810 + 980 Soft Tissue Laser
Indications for Use (Describe)
The indications for use are identical to those of the previously cleared Predicate Device. The ELUMI 810+980 Soft Tissue Laser is intended for the incision, excision, ablation, vaporization, hemostasis, and treatment of oral soft tissue.
The following are the oropharyngeal indications for use:
- · Excisional and incisional biopsies
- · Exposure of unerupted teeth
- Fibroma removal
- · Gingivoplasty
- · Hemostasis and coagulation
- · Incision and drainage of abscess
- · Operculectomy
- · Pulpotomy
- · Reduction of gingival hypertrophy
- · Soft tissue crown lengthening
- · Vestibuloplasty
- · Laser Soft Tissue Curettage
- · Tissue retraction
- Frenectomy and Frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- · Gingival incision and excision
- · Implant recovery
- · Leukoplakia
- · Oral papillectomnies
- Pulpotomy as an adjunct to root canal therapy
- · Reduction of bacterial level (decontamination) and inflammation
- Treatment of aphthous ulcers
- · Lesion (tumor) removal
- · Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- · Removal of Diseased, Inflamed and necrotic soft tissue within the periodontal pocket
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
· Sulcular debridement (removal of necrotic, diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- · Light activation for bleaching materials for teeth whitening
- · Laser-assisted whitening/bleaching of teeth
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Traditional 510(k)
# Section 5 510(k) Summary
# 510(k) Summary of Safety and Effectiveness
# Traditional 510(k) Premarket Notification
## Submitter:
Azena Medical, LLC 21 Massolo Dr, 2nd Floor – Unit C Pleasant Hill, CA 94523
# Regulatory Authority:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990, 21 CFR 807.92.
- 1. Submitter's name, address, telephone number, contact person, and date summary prepared:
| Submitter: | Azena Medical, LLC<br>21 Massolo Drive<br>FI2 Unit C<br>Pleasant Hill, CA 94523 |
|-----------------|-----------------------------------------------------------------------------------------------|
| Contact Person: | Alexandre B. Di Sessa<br>Founder & CEO<br>Phone: 800-466-5273<br>Email: 510k@azenamedical.com |
- Date of Preparation: April 6, 2015
- 2. Name of device, including the trade name and classification name:
| Trade Name: | <i>ELUMI 810 + 980 Soft Tissue Laser</i> |
|-------------------------|-------------------------------------------------------------------------------------|
| Common Name(s): | Diode laser, powered laser surgical instrument |
| Classification Name(s): | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Regulation Number: | §878.4810 |
Device Class: Class II for all requested indications
{5}------------------------------------------------
Product Code: GEX
Classification Panel: General and Plastic Surgery & Others
- 3. ldentification of the predicate device or leqally marketed device or devices to which substantial equivalence is being claimed:
## Predicate Device:
| Company: | QuickLase Limited |
|---------------|-------------------|
| Device: | QuickLase DUAL+ |
| 510(k): | K100474 |
| Date Cleared: | March 16, 2010 |
- 4. A description of the device that is the subject of the 510(k), including an explanation of how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):
The ELUMI 810 + 980 Soft Tissue Laser, or ELUMI, is a battery-powered counter top surgical and therapeutic device designed for a wide variety of oral soft tissue procedures and dental teeth whitening.
The ELUMI comprises of five main assemblies; of a Laser Unit, a flexible fiber delivery system tethered to an anodized aluminum hand piece, disposable single-use fiber tips, a wireless footswitch, and an auxiliary power supply. Infrared laser energy is emitted from the fiber tips when the wireless footswitch is depressed. The ELUMI Laser Unit contains two single-emitter solid-state laser diodes of 10.0-watt peak output power each (Class IV laser), one lasing at 810nm and the other at 980nm wavelength. The laser diodes can simultaneously emit laser energy when ELUMI is set to Dual Wavelength mode. The laser diodes are directly coupled to the flexible fiber optic cable that connects the Laser Unit to the surgical hand piece and disposable fiber tip that emits the energy to the target area. The system also contains a 5mW power, 650nm laser diode coupled to the same fiber optic cable to produce the red aiming light. The laser's visible light is designed to aid the clinician in aiming the tip of the delivery fiber onto the target tissue. Additionally, a bright white light from two LEDs in the hand piece illuminates tarqet tissue during procedures through the translucent disposable laser tips. The auxiliary power supply connects to the Laser Unit and is used to charge the system battery and as a power source for the system in the event of low battery capacity.
#### Statement of intended use: 5.
The indications for use are identical to those of the previously cleared Predicate Device.
The ELUMI 810+980 Soft Tissue Laser is intended for the incision, excision, ablation,
{6}------------------------------------------------
vaporization, hemostasis, and treatment of oral soft tissue.
The following are the oropharyngeal indications for use:
- . Excisional and incisional biopsies
- . Exposure of unerupted teeth
- . Fibroma removal
- Gingivoplasty
- . Hemostasis and coaqulation
- . Incision and drainage of abscess
- . Operculectomy
- . Pulpotomy
- . Reduction of gingival hypertrophy
- . Soft tissue crown lengthening
- Vestibuloplasty
- . Laser Soft Tissue Curettage
- . Tissue retraction
- . Frenectomy and Frenotomy
- . Gingival troughing for crown impressions
- Gingivectomy
- . Gingival incision and excision
- Implant recovery
- . Leukoplakia
- Oral papillectomnies
- . Pulpotomy as an adjunct to root canal therapy
- . Reduction of bacterial level (decontamination) and inflammation
- Treatment of aphthous ulcers
- Lesion (tumor) removal
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- . Removal of Diseased, Inflamed and necrotic soft tissue within the periodontal pocket
- . Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
- . Sulcular debridement (removal of necrotic, diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Light activation for bleaching materials for teeth whitening .
- Laser-assisted whitening/bleaching of teeth ●
### 6. Statement of how the technological characteristics of the device compare to those of the predicate or legally marketed device:
The proposed ELUMI 810 + 980 Soft Tissue Laser has the following technological similarities
{7}------------------------------------------------
to the Predicate Device:
- the same laser type: solid state diode,
- the same laser medium: GaAlAs,
- . the same working beam wavelengths: 810nm or 980nm, dual wavelength (810nm + 980nm),
- the same average power output up to 2.0 Watts,
- . the same increments of average power available: 0.1 Watts,
- . the same maximum peak output power: 10 Watts @ 810nm, 10 Watts @ 980nm, 20 Watts @ Dual Wavelength,
- . the same pulse type: gated,
- . the same delivery system and fiber diameter: 400 µm quartz glass fiber,
- the same cutting method: tissue contact,
- the same patient contacting components: quartz glass fiber,
- . the equivalent user interface, and
- the same activation means: foot switch.
#### 7. Statement of how the functional characteristics of the device compare to those of the predicate or legally marketed device:
The proposed ELUMI 810 + 980 Soft Tissue Laser has the following functional similarities to the Predicate Device:
- the same indications for use,
- . the same operating principle,
- the same basic construction,
- . the same shelf life, and
- . the same packaging materials and processes.
#### 8. Performance Data:
The ELUMI 810 + 980 Soft Tissue Laser was tested in accordance, and found to be in compliance, with the following national and international standards:
- . 21 CFR 1040.10 & 1040.11 except for deviations pursuant to laser notice 50 dated June 24, 2007
{8}------------------------------------------------
- IEC 60601-2-22
- IEC 60825-1
- . AAMI/ANSI ES60601-1
- . IEC 60601-1-2
- AAMI/ANSI ST81:2004/(R)2010
- AAMI/ANSI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013
- . AAMI/ANSI/ISO 17665-1:2006
- . AAMI/ANSI/ISO 17665-2:2009
- AAMI/ANSI/ISO 10993-5:2009
Cleaning validation was conducted according to FDA Reprocessing Guidance. The recovery method for the Anodized Aluminum Surgical Hand Piece with Fiber Connector achieved 78.5% and 77.6% recoveries for protein and hemoglobin respectively. All cleaned devices for the Anodized Aluminum Surgical Hand Piece with Fiber Connector were found to contain < 5.4 μg/cm² and < 1.6 µg/cm² for protein and hemoglobin respectively. The cleaning procedure is validated for the reprocessing of the Anodized Aluminum Surgical Hand Piece with Fiber Connector.
Sterilization validation was conducted to validate a fifteen minute gravity steam sterilization cycle at 135°C for the Anodized Aluminum Hand Piece Shell. There was no growth of the biological indicators that had been exposed to steam with the test article. The verified half cycle indicates that a full gravity cycle of not less than 15 minutes at 135°C is capable of a 12 log reduction and will provide a 10-6 sterility assurance level of a worst case population. The gravity cycle of 135°C at 15 minutes is validated for the Anodized Aluminum Hand Piece Shell.
Bench testing was performed to measure the ELUMI 810 + 980 Soft Tissue Laser's average output power side-by-side with the Predicate Device. The intended performance of these devices, based on IEC 60601-2-22, requires that the laser average power output only vary from the device's settings by less than ±20% of the setting. The results of the bench testing demonstrate that both the ELUMI 810 + 980 Soft Tissue Laser and the Predicate Device satisfy this IEC 60601-2-22 requirement.
Bench testing was also conducted to compare features and functions of the ELUMI 810+980 Soft Tissue Laser to the Predicate Device, and found that ELUMI is substantially equivalent to the Predicate Device by having identical features and functions as identified in 21 CFR 1040.10.
Additionally, an animal study was conducted to evaluate the histological effects on pig liver soft tissue of the Elumi 810+980 Soft Tissue Laser system, and compare the results to those of the Predicate Device. The results of this new animal study demonstrate that the Elumi 810+980
{9}------------------------------------------------
Soft Tissue Laser and the Predicate Device have similar effects on soft tissue, despite slight differences in average output power and operation modes.
This performance data, along with conformity to the recognized national and international standards cited above, demonstrates that the ELUMI 810 + 980 Soft Tissue Laser is as safe and as effective as its Predicate Device.
No clinical data was submitted for this Traditional 510(k).
#### Technological Characteristics: 9.
The ELUMI 810 + 980 Soft Tissue Laser has the same fundamental technological characteristics as the Predicate Device.
| | ELUMI 810 + 980 Soft<br>Tissue Laser | QuickLase DUAL+ |
|----------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Laser Classification | IV (4) | IV (4) |
| Type of Laser | Diode Laser | Diode Laser |
| Laser Medium | GaAIAs | GaAIAs |
| Wavelength | 810 ± 10nm; or<br>980 ± 10nm; or<br>810nm and 980nm ±<br>10nm | 810 ± 10nm; or<br>980 ± 10nm; or<br>810nm and 980nm ±<br>10nm |
| Average Output<br>Power | Adjustable 0.1 - 2 Watts | Adjustable 0.1 - 10 Watts |
| Max Peak Output<br>Power | 10 Watts @ 810nm;<br>10 Watts @ 980nm;<br>20 Watts @ 810nm +<br>980nm | 10 Watts @ 810nm;<br>10 Watts @ 980nm;<br>20 Watts @ 810nm +<br>980nm |
| Increments of<br>Power Available | 0.1 Watts | 0.1 Watts |
| Operating Voltage | 100-240 VAC | 100-240 VAC |
| Current Frequency | 50-60 HZ | 50-60 HZ |
| Operation Mode | Pulsed | Continuous, Pulsed |
| Pulse Type | Gated | Gated |
| Battery | Lithium Ion Rechargeable | Lithium Ion Rechargeable |
| Delivery System | Quartz glass fiber & tip* | Quartz glass fiber |
| Fiber/Tip Diameter | 400 μm core | 400 μm core |
| Fiber Aiming Beam | 5mW laser diode, 650nm,<br>Class 1 | 5mW laser diode, 645nm,<br>Class 1 |
| Activation Means | Wireless Foot Switch,<br>with electronic access key | Wired Foot Switch, with<br>electronic access key |
{10}------------------------------------------------
### Traditional 510(k)
*The quartz glass fiber within ELUMI's single-use fiber tip is the only system component that comes in direct or indirect contact with patients. This fiber is exactly the same fiber, sourced from the same supplier, and used in the same manner as the fiber used with another FDA cleared dental soft tissue laser, the SL3 (K102639).
#### Conclusions: 10.
The ELUMI 810 + 980 Soft Tissue Laser has the equivalent indications for use and technological characteristics as that of the Predicate Device. The minor technological differences that exist between the ELUMI 810 + 980 Soft Tissue Laser and its predicate devices do not alter the fundamental scientific technology of the device and raise no new questions of safety or effectiveness. Performance data demonstrates that the ELUMI 810 + 980 Soft Tissue Laser is as safe and as effective as its Predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.