New Deantronics Smoke Evacuation Adapter with 10mm connector. New Deantronics Smoke Evacuation Adapter with 22mm connector. New Deantronics Button Switch Electrosurgical Pencil with SA01 Adapter. New Deantronics Button Switch Electrosurgical Pencil with SA02 Adapter.
K152031 · New Deantronics Taiwan , Ltd. · GEI · Sep 17, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152031
Device Name
New Deantronics Smoke Evacuation Adapter with 10mm connector. New Deantronics Smoke Evacuation Adapter with 22mm connector. New Deantronics Button Switch Electrosurgical Pencil with SA01 Adapter. New Deantronics Button Switch Electrosurgical Pencil with SA02 Adapter.
Applicant
New Deantronics Taiwan , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Indications for Use
The New Deantronics Smoke Evacuation Adapter can be attached directly to an electrosurgical pencil, improving visibility and capture smoke and odors directly at the surgical site when connected to a smoke evacuator.
Device Story
Device is a single-use accessory for electrosurgical pencils; attaches to pencil handle via clamp; includes suction tubing and connectors. Operates by connecting to a user-provided external smoke evacuator; suction pump maintains low-pressure condition to remove surgical smoke and odors from the site. Used in hospitals and clinics by healthcare professionals. Improves surgical site visibility; benefits patient and staff by reducing exposure to surgical smoke.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993), sterilization validation (ISO 11137), packaging validation (ISO 11607), and electrical/safety testing (IEC 60601-1, IEC 60601-2-2).
Technological Characteristics
Single-use electrosurgical accessory. Materials: EVA (tubing), ABS/PC (adapter). Dimensions: 10mm inlet; 10mm/22mm connectors. Sterilization: Gamma radiation. Passive suction principle; requires external smoke evacuator. Complies with ISO 10993, ISO 11137, ISO 11607, IEC 60601-1, IEC 60601-2-2, and IEC 62366.
Indications for Use
Indicated for use by healthcare professionals in hospital or clinic settings to improve visibility and capture smoke and odors at the surgical site during electrosurgical procedures when attached to an electrosurgical pencil and connected to a user-provided smoke evacuator.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Valleylab Inc., Smoke Evacuation Attachment for Electrosurgical Pencils (K945260)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting hair or clothing. The profiles are stacked slightly offset from each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the top half of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2015
New Deantronics Taiwan, LTD. % Mr. Lewis Ward L. W. Ward and Associated Incorporated 4655 Kirkwood Court Boulder, Colorado 80301
Re: K152031
Trade/Device Name: Smoke Evacuation Adapter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 20, 2015 Received: July 22, 2015
Dear Mr. Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152031
Device Name Smoke Evacuation Adapter
The New Deantronics Smoke Evacuation Adapter can be attached directly to an electrosurgical pencil, improving visibility and capture smoke and odors directly at the surgical site when connected to a smoke evacuator.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for New Deantronics Group. The logo features the word "NEW" in a light gray outlined font, followed by a blue circle with a red dot in the center. To the right of the circle is the word "DEANTRONICS" in a bold, blue font. Below the logo, the words "New Deantronics Group" are written in a smaller, blue font.
# Section 5 510(k) Summary
### Submitter:
New Deantronics Taiwan Ltd. 12F., No.51, Sec. 4, Chong Yang Rd., Tu Cheng Dist. New Taipei City 23675, Taiwan R.O.C. (886) 2-2268-1726 Phone (886) 2-2268-3800 Fax
大稳企業股份有限公司 New Deantronics Taiwan, Ltd 土城市中央路4段51號12 12F .. No.51. St a Yang Rd., Tu Ch Tel: 886 2 2268-1726 Fax: 886 2 2268-3800
Taiwan
Contact Person: Lewis Ward L.W. Ward and Associates, Inc. 4655 Kirkwood Court Boulder, CO 80301 USA 303-530-3279 lwward@gwest.net
Date Prepared: July 20, 2015
#### Trade Name: Smoke Evacuation Adapter I.
New Deantronics Ltd 1990 North California Blvg
U.S.
II. Common Name: Electrosurgical accessory, Smoke Evacuation Adapter
III. Classification: 21 CFR 878.4400, electrosurgical cutting and coagulation device and accessories. Class 2, General and Plastic Surgery Panel 79
# IV. Product Code: GEI
# V. Indications for Use:
The Smoke Evacuation Adapter can be attached directly to an electrosurgical pencil, improving visibility and capture smoke and odors directly at the surgical site when connected to a smoke evacuator.
VI. Predicate Device 510(k) 945260, Valleylab Inc., Smoke Evacuation Attachment for Electrosurgical Pencils
VII. Device Description and Technological Characteristics
The Smoke Evacuation Adapter is a single-use device provided to healthcare professionals.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for New Deantronics. The logo features the word "NEW" in a light blue, outlined font, followed by a blue circle with a red dot in the center. To the right of the circle, the word "DEANTRONICS" is written in a bold, blue font. Below the word "DEANTRONICS", the words "New Deantronics Group" are written in a smaller, light blue font.
This device is an accessory to an electrosurgical pencil. The Smoke Evacuation Adapter can be attached directly to an electrosurgical pencil improving visibility and capture smoke and odors directly at the surgical site when connected to a smoke evacuator. The device components include a pencil holding clamp, an electrosurgical pencil wire holder, smoke evacuator tubing pathway to the smoke evacuator, and a connector to the smoke evacuator. The clamp secures the smoke evacuation pencil adapter to the bottom of the pencil handle. Suction to remove smoke and odors is created by a smoke evacuator device furnished by the user organization. The smoke evacuator maintains a low pressure condition by a suction pump.
#### Taiwan
球企業股份有限公司 Fax: 886 2 2268-38
U.S
New Deantronics Ltd.
1990 North Califomia Blvd.
Suite 1040
Walnut Creek, CA. 94596
Tel: (925) 280 8388
Fax: (925) 280 1788
## Table 1: New Deantronics Smoke Evacuation Adapters
Four models are available, branded as "E Surgical".
| Item | Catalog No. | ND P/N | Description |
|------|-------------|----------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | SA01 | AE710SE1 | Smoke Evacuation Adapter, with 10mm (3/8")<br>x 20.35cm (0.70 ft) tubing with a connector to<br>10mm (3/8") hose, Single Use, Sterile |
| 2 | SA02 | AE711SE1 | Smoke Evacuation Adapter, with 10mm (3/8")<br>x 163cm (5.5ft) tubing with a connector to 22<br>mm (7/8") x 145cm (5 ft) hose, Single Use,<br>Sterile |
Table 2: New Deantronics Electrosurgical Pencil with Smoke Evacuation Adapter
| Item | Catalog No. | ND P/N | Description |
|------|-------------|----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | PEN11SA01 | PB314SE1 | Button Switch Electrosurgical Pencil, Smoke<br>Evacuation Adapter, with 10mm (3/8") x<br>20.35cm (0.70 ft) tubing with a connector to<br>10 mm (3/8") hose, Single Use, Sterile |
| 2 | PEN11SA02 | PB314SE2 | Button Switch Electrosurgical Pencil, Smoke<br>Evacuation Adapter, with 10mm (3/8") x<br>163cm (5.5ft) tubing with a connector to 22<br>mm (7/8") x 145cm (5 ft) hose, Single Use,<br>Sterile |
VIII. Non-Clinical Testing
IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
IEC 60601-2-2: 2009 (Fifth Edition) + C1:2014, Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment.
ISO 14971:2007, Medical Device- Application of Risk Management to Medical Devices
ISO 10993-1: 2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process.
ISO 10993-5: 2009, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for New Deantronics. The logo features the word "NEW" in a light blue, outlined font, followed by a blue circle with a red dot in the center. To the right of the circle, the word "DEANTRONICS" is written in a bold, blue font. Below the word "DEANTRONICS", the words "New Deantronics Group" are written in a smaller, blue font.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization ISO 10993-12: 2012, Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials.
ISO 11137-2:2013, Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose.
# Taiwan
U.S.
New Deantronics Ltd
1990 North California Blvd
Valnut Creek, CA 9459
大蕴企業股份有限公司 New Deantronics Taiwan, Ltd 縣土城市中央路4段51號12: 12F No.51.9 hong Yang Rd., Tu Ch Fax: 886 2 2268-38
ISO 11607:2006. Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including: Amendment 1 (2014)]; Part 2: Validation requirements for forming, sealing and assembly processes [Including: Amendment 1 (2014)].
ISTA 2A Packaged-Products 150 lb (68 kg) or Less
IEC 62366:2007 Medical Devices - Application of Usability Engineering to Medical Devices
### Non-clinical testing conclusion:
The non-clinical tests were successfully performed on the Smoke Evacuation Adapter to demonstrate that the device is as safe, as effective, and performs as well or better than the predicate device.
## IX. Substantial Equivalence
The Smoke Evacuation Adapter is substantially equivalent to the Valleylab (Covidien) Smoke Evacuation Attachment for Electrosurgical Pencils (K945260) based on design, intended use, technology, materials, comparison testing, and performance. Both devices require a user-furnished smoke evacuator to remove the smoke and odor.
| X. Key Features Comparison to Predicate | | |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| FEATURE | New Deantronics Smoke Evacuation Adapter | Valleylab (Covidien) Smoke Evacuation Attachment for Electrosurgical Pencils (K945260) |
| Use Environment | Single Use, hospital, clinic | Single Use, hospital, clinic |
| Sterilization Method | Gamma radiation | Gamma radiation |
| Construction Design | Pencil holder, suction tube, wire holders, swivel connector, and connector to smoke evacuator | Pencil holder, suction tube, wire holders, swivel connector, and connector to smoke evacuator |
| Inlet Diameter | Approx. 10 mm | Approx. 10 mm |
| Material of suction tube | EVA | EVA |
| Connector Tube | 10 mm and 22 mm | ~10 mm |
| Materials of pencil adapter | ABS + PC | Polystyrene (PS) |
| Requires smoke evacuator suction device with a 10 mm or 22 mm port | 10 mm and 22 mm | 10 mm |
#### X Kev Features Comparison to Predicate
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.