EVO platform

K152012 · Quanta System Spa · GEX · Dec 8, 2015 · General, Plastic Surgery

Device Facts

Record IDK152012
Device NameEVO platform
ApplicantQuanta System Spa
Product CodeGEX · General, Plastic Surgery
Decision DateDec 8, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EVO Platform is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows.

Device Story

EVO Platform is a multi-source laser system for dermatology, plastic, and general surgery. It utilizes Q-switched, pulsed, and picosecond laser sources (532, 694, 755, 1064, 1320 nm) and optional IPL handpieces (400-1200 nm range). Input is controlled by a physician via a touch screen interface and footswitch. Optical delivery occurs through an articulated arm (Q-switched) or optical fiber (pulsed). An integrated skin cooling system reduces pain and side effects. The device performs photocoagulation, ablation, and vaporization of soft tissue. Output is visual/thermal effect on tissue, enabling clinicians to treat vascular/pigmented lesions, remove tattoos, reduce hair, and perform skin resurfacing. Benefits include precise tissue management and reduced recovery time for aesthetic/surgical procedures.

Clinical Evidence

Bench testing only. Evidence includes electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2, and compliance with IEC 62366, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1. Biocompatibility is established via predicate devices.

Technological Characteristics

Laser family with Q-switched, pulsed, and picosecond sources (532, 694, 755, 1064, 1320 nm) and IPL (400-1200 nm). Delivery via articulated arm or optical fiber. Integrated contact/air cooling system. Touch screen control interface. Footswitch activation. Compliant with IEC 60601-1, 60601-1-2, 60601-2-22, and 60825-1.

Indications for Use

Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissue; tattoo removal; hair removal/reduction; treatment of vascular lesions (e.g., port wine stains, telangiectasias, hemangiomas, leg veins); benign pigmented lesions (e.g., lentigines, nevi, cafe-au-lait spots); acne/acne scars; wrinkles; and psoriasis. Applicable to various Fitzpatrick skin types (I-VI) depending on wavelength and mode. Contraindications include conditions where laser treatment is not appropriate as determined by a physician.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine and healthcare. The symbol is composed of three intertwined figures, possibly representing people, and a staff-like element. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Ouanta System Spa Francesco Dell'antonio Compliance Manager Via Iv Novembre, 116 Solbiate Olona (va), 21058 Italy December 8, 2015 Re: K152012 Trade/Device Name: Evo Platform Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 11, 2015 Received: November 12, 2015 Dear Francesco Dell'antonio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Joshua C. Nipper -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152012 Device Name EVO Platform Indications for Use (Describe) The EVO Platform is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows. Indications for use 1064 & 532 nm (Q-Switched, nanosecond mode) The EVO Platform is intended for treatment of vascular lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of: 532 nm (Q-Switched, nanosecond mode) Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos Treatment of vascular lesions including, but not limited to: - port wine birthmarks - telangiectasias - spider angiomaa - Cherry angioma - Spider nevi Treatment of benign pigmented lesions including, but not limited to: - cafe-au-Iait birthmarks - Ephalides, solar lentigines - senile lentigines - Becker's nevi - freckles - common nevi - nevus spilus - Ota Nevus Treatment of seborrheic keratosis Treatment of post inflammatory hyperpigmentation Skin resurfacing procedures for the treatment of acne scars and wrinkles. 1064 nm (O-Switched, nanosecond mode) Removal of dark ink (black, blue and brown) tattoos Removal of benign pigmented lesions including; - nevus of Ota - Café au lait spot - Ephalides, solar lentigo (lentigines) - Becker Nevus - Nevus spilus Treatment of common nevi Removal or lightening of unwanted hair Skin resurfacing procedures for the treatment of acne scars and wrinkles {3}------------------------------------------------ 1064 nm (non Q-Switched - free running mode) Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types. Fitzpatrick I-VI, including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Treatment of wrinkles. Treatment of mild to moderate inflammatory acne vulgaris. 532 nm (picosecond mode) Indicated for the removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. 1064 nm (picosecond mode) Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. 1064 nm (pulsed) Dermatology/Plastic Surgery: Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lessons such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. The laser is also indicated for the treatment of facial wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The laser is also indicated for pigmented lesions to reduce lesion size, for patients with lessions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. It is indicated for use on all skin types (Fitzpatrick I-VI ) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types. 532 nm (pulsed) Intended for the coagulation and hemostasis of vascular lesions. ## Dermatology/Plastic Surgery: For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or ex-tremities telangiectasias, venous anomalies, leg veins]; benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, {4}------------------------------------------------ #### blanching, flattening, reduction of lesion size). 694 nm (Q-Switched) ## Indicated for: Tattoo removal: Suggested for blue, sky blue, black, green and violet ink - Pigmented lesion removal (benign): - Cafe au lait spot - Ephalides, solar lentigo lentigines) - Becker Nevus - Ota and Ito Nevus - Nevus spilus - Mongolian spot 694 nm (non q-switch - free running mode) Intended to remove benign dermal and epidermal pigmented lesions, and, to effect hair removal of patients with skin types 1-4 through selective targeting of melanin in hair follicles in dermatology and plastic surgery. # 755 nm (pulsed) Indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles. 1320 nm (pulsed) Indicated for use in general surgery and dermatology for the incision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars. ## IPL 590-1200nm; 625-1200nm; 650-1200nm Indicated for permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime ## IPL 550-1200nm; 570-1200nm Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions. ## IPL 400-1200nm Indicated for inflammatory acne (acne vulgaris). ## Integrated Skin Cooler The intended use of the integrated cooling system in the EVO Platform is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments. {5}------------------------------------------------ Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {6}------------------------------------------------ K152012 Page 1 of 6 ## 5. 510(K) SUMMARY #### Introduction: This document contains the 510(k) Summary for the EVO Platform. The content of this summary is based on the requirements of 21 CFR 807.92(c). | Applicant / Manufacturer Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058 | |--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact Person: | Francesco Dell'Antonio<br>Compliance Manager<br>Quanta System SPA<br>Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 | | Date Prepared: | July 15th 2015 | | Device Name: | EVO Platform | | Classification: | Class II | | Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology. | | Regulation Number: | 21 CFR 878.4810 | | Product Code: | GEX | #### Predicate Devices: EVO Platform is substantially equivalent to the following legally marketed predicate devices: - Main predicate device: MDK Multi-Applications Platform (K130356) Quanta System SpA - . Q-Plus T+IPL (K123168) Quanta System SpA - PicoWay Laser System (K150326) Candela Corporation - Fotona Qx Nd:Yag/Ktp Laser System Family (K083889) FOTONA D.D. ● - . Elite Mpx Laser System With Xpl Handpiece (K090235) CYNOSURE, INC - Polaris Q-Switch Ruby System (K121162) Sandstone Medical Technologies, Llc - ELITE + LASER SYSTEM (K141425) CYNOSURE {7}------------------------------------------------ ## Description of the device: The EVO Platform is a laser family that includes Q-Switched and/or Pulsed laser sources, emitting at one or more of the following wavelengths: 532 nm, 1064 nm, 1320 nm (Nd:YAG laser), 694 nm (Ruby laser), or 755 nm (Alexandrite laser). The EVO Platform systems, through the special universal Twain connector, can be equipped with intense pulsed light handpieces (Twain IPL) emitting at the following wavelengths: 650-1200nm, 625-1200nm, 590-1200nm, 570-1200nm, 550-1200nm, 400-1200nm. The EVO Platform systems, when operating with Pulsed laser sources and IPL, can be used in combination with optional contact, or air, cooling systems. The optical delivery system is an articulated arm with fixed handpieces for Q-switched sources and an optical fiber with focusing handpieces for pulsed sources. The optical delivery system for the IPL system is a handpiece (Twain IPL) with fixed or interchangeable light filters at different wavelengths. All the models belonging to the EVO platform have the same components and the same control software. The only difference between different models is the optical bench that depends on the sources installed. The EVO Platform is controlled via a touch screen display housed in the front of the device. Emission is triggered by means of a footswitch. ## Intended use ## General intended use The EVO Platform is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows. ## Indications for use # 1064 & 532 nm (Q-Switched, nanosecond mode) The EVO Platform is intended for treatment of vascular lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of: ## 532 nm (Q-Switched, nanosecond mode) Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos Treatment of vascular lesions including, but not limited to: - l port wine birthmarks - telangiectasias - spider angiomaa l - -Cherry angioma - -Spider nevi Treatment of benign pigmented lesions including, but not limited to: - cafe-au-lait birthmarks {8}------------------------------------------------ K152012 Page 3 of 6 - -Ephalides, solar lentigines - senile lentigines - -Becker's nevi - freckles - - common nevi - nevus spilus - -Ota Nevus Treatment of seborrheic keratosis Treatment of post inflammatory hyperpigmentation Skin resurfacing procedures for the treatment of acne scars and wrinkles. #### 1064 nm (Q-Switched, nanosecond mode) Removal of dark ink (black, blue and brown) tattoos Removal of benign pigmented lesions including; - י nevus of Ota - Café au lait spot י - -Ephalides, solar lentigo (lentigines) - -Becker Nevus - -Nevus spilus Treatment of common nevi Removal or lightening of unwanted hair Skin resurfacing procedures for the treatment of acne scars and wrinkles ## 1064 nm (non Q-Switched - free running mode) Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Treatment of wrinkles. Treatment of mild to moderate inflammatory acne vulgaris. ## 532 nm (picosecond mode) Indicated for the removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, vellow and orange. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. ## 1064 nm (picosecond mode) Indicated for the removal of tattoos for all skin types (Fitzpatrick skin types I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. ## 1064 nm (pulsed) {9}------------------------------------------------ K152012 Page 4 of 6 ## Dermatology/Plastic Surgery: Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. The laser is also indicated for the treatment of facial wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. lt is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The laser is also indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. lt is indicated for use on all skin types (Fitzpatrick I-VI ) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types. ## 532 nm (pulsed) Intended for the coagulation and hemostasis of vascular lesions. #### Dermatology/Plastic Surgery: For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or ex-tremities telangiectasias, venous anomalies, leg veins]; benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, flattening, reduction of lesion size). #### 694 nm (Q-Switched) Indicated for: Tattoo removal: Suggested for blue, sky blue, black, green and violet ink Pigmented lesion removal (benign): - । Cafe au lait spot - -Ephalides, solar lentigo lentigines) - । Becker Nevus - -Ota and Ito Nevus - -Nevus spilus - -Mongolian spot {10}------------------------------------------------ ## 694 nm (non q-switch – free running mode) Intended to remove benign dermal and epidermal pigmented lesions, and, to effect hair removal of patients with skin types 1-4 through selective targeting of melanin in hair follicles in dermatology and plastic surgery. # 755 nm (pulsed) Indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles. # 1320 nm (pulsed) Indicated for use in general surgery and dermatology for the incision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of periorbital and perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars. # IPL 590-1200nm; 625-1200nm; 650-1200nm Indicated for permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime ## IPL 550-1200nm; 570-1200nm Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions. ## IPL 400-1200nm Indicated for inflammatory acne (acne vulgaris). ## Integrated Skin Cooler The intended use of the integrated cooling system in the EVO Platform is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments. ## Performance data: The following performance data are provided in support of the substantial equivalence determination: {11}------------------------------------------------ #### Safety and electromagnetic compatibility (EMC) Electrical safety and electromagnetic compatibility (EMC) testing for the EVO Platform was conducted by an independent test laboratory in accordance with IEC 60601-1, Medical electrical equipment, Part 1: General requirements for basic safety and essential performance and with IEC 60601-1-2, Collateral Standard: Electromagnetic Compatibility - Requirements and Tests, 3rd ed. The EVO Platform was determined to be in conformance with applicable IEC standards (IEC 62366, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1). #### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". #### Biocompatibility: The biocompatibility of the EVO Platform is established based on the predicate devices. #### Substantial Equivalence: The EVO Platform has the same intended use and the same indications for use as its predicate devices. The EVO Platform presents the same or similar technological characteristics as its predicate devices, including the laser type, wavelengths, pulse width, frequency and spot sizes. Any minor differences do not present any new types of safety or effectiveness questions since the parameters are the same as the predicate devices. Therefore, the EVO Platform is substantially equivalent to its predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...