K151955 · Biogaming , Ltd. · LXJ · Feb 5, 2016 · Physical Medicine
Device Facts
Record ID
K151955
Device Name
YuGo System
Applicant
Biogaming , Ltd.
Product Code
LXJ · Physical Medicine
Decision Date
Feb 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5360
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
A software system used with the Microsoft Kinect intended to be used to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the lower and upper extremities with audiovisual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.
Device Story
YuGo is a virtual reality software platform utilizing Microsoft Kinect motion sensing technology for physical rehabilitation. It supports upper and lower extremity exercises in clinical or home settings. The system comprises a Clinician Dashboard, Patient Interface, and Reporting System. Clinicians use the dashboard to prescribe personalized exercise protocols, define movement parameters, and monitor patient performance via remote reports. Patients interact with the system through a virtual trainer or games displayed on a screen, which provide real-time audiovisual feedback and graphic movement representations. The system records movement data, analyzes performance, and transmits metrics to the clinician. This enables remote monitoring and adjustment of rehabilitation programs, facilitating home-based therapy and potentially improving patient adherence and recovery outcomes.
Clinical Evidence
Bench testing only. Software validation testing was conducted to verify performance against Software Requirements Specifications (SRS). The Software Test Description (STD) documented test cases, acceptance criteria, and pass/fail results, confirming the system performs according to specifications.
Technological Characteristics
Software system utilizing Microsoft Kinect 3D optical motion sensing. Operates on standard PC hardware. Requires internet connectivity for remote reporting. No energy delivery to patients. No electrical, chemical, mechanical, thermal, or radiation safety concerns. Software-based analysis of limb/body motion.
Indications for Use
Indicated for physical rehabilitation of adults with orthopedic and neurological conditions in clinical or home settings. Requires medical professional assessment, exercise guidance, and approval prior to use.
Regulatory Classification
Identification
Measuring exercise equipment consist of manual devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. These devices also include instrumentation, such as the pulse rate monitor, that provide information used for physical evaluation and physical planning purposes., Examples include a therapeutic exercise bicycle with measuring instrumentation, a manually propelled treadmill with measuring instrumentation, and a rowing machine with measuring instrumentation.
Special Controls
*Classification.* Class II (special controls). The device, when it is a measuring exerciser or an interactive rehabilitation exercise device for prescription use only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2016
BioGaming Ltd. c/o Suzan Onel, Partner and Chair, Global FDA Practice K&L Gates LLP&K 1601 K Street. NW Washington, DC 20006
Re: K151955
Trade/Device Name: YuGo System Regulatory Class: Unclassified Product Code: LXJ Dated: December 31, 2015 Received: January 5, 2016
Dear Suzan Onel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Michael J. Hoffmann -A
- Carlos Peña, PhD, MS for Director, Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151955
Device Name YuGo system
Indications for Use (Describe)
A software system used with the Microsoft Kinect intended to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the lower and upper extremities with audio- visual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
### 510(K) Number K151955
- 5.1 Applicant's Name: BioGaming Itd. 20 Hamagshimim St. Petah-Tikva 4934829, Israel Tel: + 972 72 72 221 8870 Fax: + 972 72 72 221 8876
- 5.2 Contact Person: Suzan Onel K&L Gates LLP 1601K Street, NW Washington, DC 20006 Phone 202/778-9134 Fax: 202/778-9100
- 5.3 Date Prepared: December 31, 2015
- 5.4 Trade Name: YuGo System
- 5.5 Common or Usual Name: Software system utilizing optical position recording for rehabilitation exercises
- 5.6 Classification Name: System, optical position/movement recording
- 5.7 Device Panel: Physical Medicine
- 5.8 Product Code: LXJ
- 5.9 Classification Regulation: None. Reason: YuGo is substantially equivalent to predicates in product code LXJ. The product code is 'unclassified' in FDA's product classification database with the reason being 'pre-amendment'. Refer to URL: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpcd/classifica tion.cfm?ID=4850
{4}------------------------------------------------
## 5.10 Predicate Device:
Jintronix Rehabilitation System ("JRS") (Jintronix Inc.), cleared under K130847: product code: LXJ (System, Optical Position/Movement Recording).
## 5.11 Intended Use / Indication for Use:
A software system used with the Microsoft Kinect intended to be used to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the lower and upper extremities with audiovisual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.
## 5.12 Device Description:
The YuGO device, developed by Biogaming Ltd. is a virtual reality platform used with the Microsoft Kinect motion sensing technology, which provides exercise programs for the body's lower and upper extremities. It allows physical therapists to define professional, supervised, and personalized exercise programs via video games or a virtual trainer.
The system was designed to serve a number of purposes:
Allowing the treating clinician a direct access to the training program a) performed at the patients' home. by determining the allowed movements. extent and difficulty levels of the training protocol.
Encourage physical therapy home-training through the use of camerab) sensors, visualizing and promoting body movements using a game displayed on screen.
Collecting movement and performance related data from the homec) user; analyzing it and instantly preparing a report on a designated website, to allow treating clinicians (the patients' Physical Therapist or a Physician) access to a "performance report", in order to monitor, adjust and otherwise modify the training protocol to suit the needs of the patient in his rehabilitation process.
The system includes three (3) main functions: Clinician Dashboard, Patient Interface and Reporting System. Using the Microsoft Kinect for Windows to track motion, the YuGo Patient Interface records performance metrics providing them to qualified medical professionals via the Clinician Dashboard, in the Reporting system. Medical professionals can monitor patients performance, assign or modify rehabilitation exercises in Patient Interface for their patients through the Clinician Dashboard allowing for patients to perform their prescribed rehabilitation program even from the comfort of their home.
{5}------------------------------------------------
## 5.13 Substantial Equivalence:
### Intended Use and Indications for Use
The subject YuGo System and predicate Jintronix Rehabilitation System (JRS) have the same intended use and identical indications for use, except for the fact that the JRS is intended for the upper extremity and trunk only and the YuGo is also intended for the lower extremities, but not for the trunk. The additional lower extremity exercises available with the YuGo system do not present additional safety concern to the device, and do not change the substantial equivalence determination.
### Comparison of Technological Characteristics
The YuGo system, like its predicate, is software used with the Microsoft Kinect intended to be used to support the physical rehabilitation of adults in the clinic/ at home, and require an internet connection to operate. The YuGo and JRS are based on a 3D sensor that analyses the patient's movements in real time and provides feedback on performance quality, guidance for improving performance and exercise results summary. The exercise is recorded and analyzed by the system for further monitoring by the therapist.
The following technological differences exist between the subject and predicate devices:
- YuGo clinician selects exercises and the software incorporates them . into the patient interface (virtual trainer and games), while in the JRS system clinician is required to select the specific games (activities) based on his familiarity with them as suitable for the patient. This difference is a technological software progress and does not have any clinical affect, and therefore no new questions of safety and effectiveness are raised.
- The two products differ in the games currently included in them. These . games serve only as a virtual environment for performing the prescribed rehabilitation exercises, and therefore this difference does not raise new questions of safety and effectiveness.
- Slightly lower minimal PC system requirements (such as Windows ● version, processor, memory, Kinect version, resolution) of the YuGo system. These minor differences have no clinical effects and therefore do not raise any new questions of safety and effectiveness.
### Performance Testing
Software validation testing was conducted to verify that the device performs according to its specifications as described in the Software Requirements Specifications (SRS). The Software Test Description (STD) for the YuGo system presents the methodology for the validation and describes the test cases along with their acceptance criteria and the detailed test procedure. The
{6}------------------------------------------------
STD also includes the test log (including individual grade of Pass/Fail). The SRS and STD demonstrate that the YuGo system performs according to its specifications.
#### Summary
The YuGo and predicate device are software systems that share in common intended use/indications for use, target populations, anatomical sites, location of use and technological characteristics like principle of operation, design and technology used. The devices are intended for the physical rehabilitation of patients with orthopedic and neurological conditions. The two systems track limb and body motion while providing visual feedback. They are prescription use devices that enable home based rehabilitation exercises with remote medical professional control. The YuGo and equivalent predicate device make use of optical motion sensing technology and a computer operating system. YuGo and JRS also neither deliver energy to patients nor pose any issues in terms of electrical, chemical, mechanical, thermal, radiation safety or compatibility. YuGo is validated for system compatibility and performance. Therefore, BioGaming Ltd believes that the YuGo is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.