K151766 · Becton, Dickinson and Company · FMF · Sep 2, 2015 · General Hospital
Device Facts
Record ID
K151766
Device Name
BD Single Use, Hypodermic Syringe
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
Sep 2, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
Device Story
Three-piece, sterile, single-use hypodermic syringe; consists of lubricated plastic barrel with graduated scale, lubricated synthetic rubber stopper, and plastic plunger rod. Features 6% (Luer) male connector (Luer Slip or Luer Lok). Operated by healthcare professionals to aspirate or inject fluids by manual manipulation of plunger rod within barrel. Used in clinical settings for general fluid management. Benefits include sterile, single-use delivery of fluids.
Clinical Evidence
No clinical data. Bench testing only. Verification testing included break out force, sustaining force, flange bend force, dimensional stability, scale permanency, and impact testing. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, hemolysis, acute systemic toxicity, intracutaneous reactivity, sensitization, pyrogenicity) and chemical extractable analysis per ISO 10993-18.
Technological Characteristics
Materials: Polypropylene barrel/plunger, silicone lubricant, rubber stopper. 6% Luer connector. Sterilization: Ethylene Oxide (EtO) to SAL 10^-6. Shelf life: 5 years. Sizes: 1ml, 3ml, 10ml, 20ml.
Indications for Use
Indicated for use by health care professionals for general purpose fluid aspiration and injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping to create a sense of depth and connection. The logo is simple, yet recognizable, and represents the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2015
Becton, Dickinson and Company Murtaza Rana Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, NJ 07417
Re: K151766
Trade/Device Name: BD Single Use, Hypodermic Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: June 25, 2015 Received: June 30, 2015
#### Dear Murtaza Rana:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151766
Device Name BD Single Use, Hypodermic Syringe
Indications for Use (Describe)
The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| Submitted By: | Murtaza Rana |
|-------------------|----------------------------------------------------|
| | Regulatory Affairs Specialist |
| | Becton, Dickinson and Company |
| | 1 Becton Drive |
| | Franklin Lakes, NJ 07417 |
| | Phone: (201)847-6980 |
| | Fax: (201)847-5307 |
| Date Prepared: | August 3, 2015 |
| Device: | |
| | Trade Name: BD Single Use, Hypodermic Syringe |
| | Common Name: Piston Syringe |
| | Classification: Class II device, 21 CFR §880.5860, |
| | Piston Syringe |
| | Product Code: FMF (Syringe, Piston) |
| Predicate Device: | K980987 |
| | BD Single Use, Hypodermic Syringe |
#### Device Description
The BD Single Use, Hypodermic Syringe is a three-piece, sterile, single use hypodermic syringe with a 6% (Luer) male connector in 1ml Luer Slip. 3ml. 10ml and 20ml Luer Lok syringe sizes. The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. The syringe assembly consists of a lubricated plastic barrel with a graduated scale, a lubricated synthetic rubber stopper and a plastic plunger rod. The plunger rod is pulled along the inside of the barrel to aspirate fluids and pushed along the inside of the barrel to inject or expel fluids. The syringe barrel incorporates a male 6% (Luer) connector which is connectable to a compatible female 6% (Luer) connector. The BD Single Use, Hypodermic Syringe are provided sterile (EtO) in a syringe only configuration or with a hypodermic needle. The modified BD Single Use, Hypodermic Syringe includes a new resin material in the barrel of the syringe. The syringe performance characteristics are equivalent to the predicate device.
#### Intended Use
The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
#### Technological Characteristics
The subject BD Single Use, Hypodermic Syringe is equivalent to that of the predicate BD Single Use, Hypodermic Syringe in intended use, materials and performance characteristics.
{4}------------------------------------------------
| Element of<br>Comparison | Subject Device | Predicate Device |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use/Intended Use | The BD Single Use,<br>Hypodermic Syringe<br>is intended for use by<br>health care<br>professionals for<br>general purpose fluid<br>aspiration/injection. | The BD Single Use,<br>Hypodermic Syringe<br>is intended for use by<br>health care<br>professionals for<br>general purpose fluid<br>aspiration/injection. |
| Syringe materials | Barrel<br>Polypropylene | Polypropylene |
| | Barrel<br>Silicone | Silicone |
| | Lubricant | |
| | Plunger Rod<br>Polypropylene | Polypropylene |
| | Stopper<br>Rubber | Rubber |
| | Stopper<br>Silicone | Silicone |
| | Lubricant | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide |
| SAL | $10^{-6}$ | $10^{-6}$ |
| Shelf Life | 5 Years | 5 Years |
# Non-Clinical Testing
BD has performed the following non-clinical/design verification testing based on the risk analysis conducted and the results of these tests demonstrate that the BD Single Use, Hypodermic Syringe performed in an equivalent manner to the predicate device.
| Performance Characteristic | Acceptance Criteria | |
|----------------------------------|-----------------------------------------------------------|----------------------------------------------------|
| Functional Testing | Break Out Force | Equivalence to Predicate |
| | Sustaining Force | Equivalence to Predicate |
| | Flange Bend Force | Equivalence to Predicate |
| | Dimensional Stability of<br>Barrel ID (inner<br>diameter) | Equivalence to Predicate |
| | Barrel Scale Permanency | Equivalence to Predicate |
| | Barrel Impact Test | Equivalence to Predicate |
| | Biocompatibility<br>Testing | Cytotoxicity |
| Hemolysis | | Per ISO10993-4, Non-toxic |
| Acute Systemic Toxicity | | Per ISO10993-11, Non-toxic |
| Intracutaneous Reactivity | | Per ISO10993-10, Non-Irritant |
| Sensitization | | Per ISO10993-10, Non-Sensitizer |
| Pyrogenicity | | Per ISO 10993-11 and USP 151, Non-Pyrogenic |
| Chemical Extractable<br>Analysis | | Per ISO 10993-18, acceptable level of<br>extracts. |
{5}------------------------------------------------
## Clinical Testing
Clinical testing was not required for this submission.
#### Substantial Equivalence
The BD Single Use, Hypodermic Syringe is substantially equivalent to the predicate device in intended use, principles of operation, technology, design, materials and performance.
### Conclusion
The BD Single Use, Hypodermic Syringe has been verified to meet the established performance criteria above. The results of the non-clinical/design verification testing demonstrate that the BD Single Use, Hypodermic Syringe performs as intended and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.