Specular Microscope CEM- 530

K151706 · Nidek Co., Ltd. · NQE · Mar 17, 2016 · Ophthalmic

Device Facts

Record IDK151706
Device NameSpecular Microscope CEM- 530
ApplicantNidek Co., Ltd.
Product CodeNQE · Ophthalmic
Decision DateMar 17, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2
AttributesPediatric

Indications for Use

The Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.

Device Story

Non-contact ophthalmic device; captures high-magnification images of corneal endothelium; performs optical pachymetry. Features auto-tracking, auto-shooting, and manual analysis modes. Input: optical images of corneal endothelium. Processing: Center Point Method allows operator to mark cell centers via touch screen; system identifies cell walls and calculates cell density, area, and hexagonality. Output: endothelial images, cell analysis metrics, corneal thickness measurements. Used in clinics by eye care professionals for intraocular/corneal surgery, postoperative follow-up, and monitoring corneal health. Results displayed on integrated LCD, printed via thermal printer, or exported via LAN/USB. Benefits: provides objective, repeatable assessment of corneal endothelial health and thickness to guide clinical management.

Clinical Evidence

Prospective clinical study (n=74) compared CEM-530 Center Point Method to Konan Cellchek Plus. Populations: young healthy (n=28), adult healthy (n=27), pathologic adult (n=19). Primary endpoints: agreement and precision (repeatability/reproducibility) of endothelial cell density, coefficient of variation (CV), and % hexagonality. Results showed high correlation (R2=0.97 for cell density) and comparable precision metrics, with CEM-530 demonstrating lower variability than the predicate in several parameters. Bench testing confirmed pachymetry accuracy of ±10 microns.

Technological Characteristics

Non-contact specular microscope and optical pachymeter. Materials: standard medical-grade components. Sensing: CCD camera, infrared LED illumination, cyan LED flash. Pachymetry: optical, ±10 microns accuracy. Connectivity: LAN, USB-A, BNC video output. Interface: 8.4" SVGA color LCD with touch panel. Software: includes automated and manual analysis modes (Center Point Method). Safety: IEC 60601-1, IEC 60601-1-2, ISO 15004-1, ISO 15004-2.

Indications for Use

Indicated for examination of corneal endothelium and measurement of corneal thickness in patients, including those with endothelial disorders or contact lens wearers.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 17, 2016 Nidek Co., Ltd. % Mr. Ryan Bouchard Ora, Inc. 300 Brickstone Square Andover, Massachusetts 01810 Re: K151706 Trade/Device Name: Specular Microscope Cem- 530 Regulation Number: 21 CFR 886.1850 Regulation Name: Ac-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: NQE Dated: February 4, 2016 Received: February 5, 2016 Dear Mr. Bouchard: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151706 Device Name Specular Microscope CEM-530 Indications for Use (Describe) The Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------| | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510(k) Summary This summary of the 510(k) premarket notification for the NIDEK Specular Microscope CEM-530 is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR S 807.92. Date Prepared: March 4, 2016 ### SPONSER/ 510(k) OWNER/ MANUFACTURER NIDEK Co., Ltd. 34-14 Maehama, Hiroishi-cho, Gamagori, Aichi, 443-0038 Japan Telephone: +81-533-67-8901 Facsimile: +81-533-67-6628 E mail: yoneji mizuno@nidek.co.jp Establishment Registration Number: 8030392 ### CONTACT PERSON Ryan Bouchard Ora, Inc. 300 Brickstone Square Andover, MA 01810 Telephone: (978) 332-9574 Facsimile: (978) 689-0020 E-mail: rbouchard@oraclinical.com ### NAME OF DEVICE Trade Name: Specular Microscope CEM-530 Common Name: Specular Microscope #### DEVICE CLASSIFICATION/FDA REVIEWING BRANCH The Ophthalmic Branch has classified AC Powered Slit Lamp Biomicroscopes as Class II devices pursuant to 21 C.F.R. §886.1850. #### PRODUCT CODE: CLASSIFICATION / CFR TITLE NQE, 21 CFR 886.1850 ### PREDICATE DEVICES Nidek Specular Microscope CEM-530 (K130565) Konan Medical, Inc. Cellchek Plus (K120264) {5}------------------------------------------------ #### INDICATIONS FOR USE The Nidek Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea. #### PRODUCT DESCRIPTION The Nidek Specular Microscope CEM-530 provides non-contact, high magnification image capture of the endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of the endothelium are used in intraocular or corneal surgery, postoperative follow-up, and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses. Observation is possible in the central area (visual angle: 5°) and peripheral area (visual angle: 27°) using a periphery capture function as well as in the center of the captured images and analysis results can be printed on the built-in printer or optional video printer, or output to an external device over LAN connection. In addition to the specular microscopy, the corneal thickness can be optically measured in a non-contact method. The CEM-530 has auto-tracking and auto-shooting functions. Results can be printed using the built-in thermal printer or captured images can be transferred to a filing system via LAN connection. The Specular Microscope CEM-530 cleared in this 510(k) is identical to the Specular Microscope CEM-530 cleared in K130565 with the addition of a new analysis mode: Center Point Method. All other aspects of the cleared device remain unchanged. #### SUBSTANTIAL EQUIVALENCE The Specular Microscope CEM-530 and the predicate devices are all non-contact ophthalmic microscopes, optical pachymeters, and cameras intended for examination of the corneal endothelium and for measurement of the thickness of the cornea. Both the Specular Microscope CEM-530 and the predicate device offer automatic capture features and manual capture modes. The Specular Microscope CEM-530 which is the subject of this 510(k) has identical technological characteristics to the Specular Microscope CEM-530 which was cleared in K130565. The only difference is the addition of Center Point Method. With Center Point method, the operator touches the center of a cell on the endothelial image with the touch screen pen. A dot is displayed on the endothelial image. Shapes are surrounded by the vertical lines to the lines made by connecting each dot and is recognized as a cell. Connected dots are top of cells and vertical lines are cell walls. Therefore, this discussion will focus primarily on the substantial equivalence to the Konan Cellchek Plus cleared in K120264. Both the CEM-530 and the Konan Cellchek Plus have a built-in CCD camera. Both the CEM-530 and the Konan Cellchek Plus include an optical pachymeter with an accuracy of ± 10 microns. There were no modifications made to the optical pachymeter in the Specular Microscope CEM-530 compared with the device cleared in K130565. {6}------------------------------------------------ Regarding image analysis, both the CEM-530 and the predicate device offer automatic image analysis. The CEM-530, which is the subject of this 510(k), also offers manual analysis of images, as does the Konan predicate device. Clinical performance data is provided which evaluates the precision and agreement of the manual measurements performed by the CEM-530 compared to manual measurements performed with the Konan predicate device. The clinical performance data demonstrates the substantial equivalence of the CEM-530 manual mode to the Konan predicate device's manual mode. Both the CEM-530 and the predicate device comply with applicable electrical safety and light safety standards. Therefore, in regards to technological characteristics, the Specular Microscope CEM-530 is similar to the Konan but there are some minor differences between the devices which have been evaluated with non-clinical and clinical performance data. The differences are shown in Table 1. | SUBSTANTIAL EQUIVALENCE CHART | | | | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Nidek | Nidek | Konan | | Device Name | CEM-530 | CEM-530 | Cellchek Plus | | 510(k) Number | NA | K130565 | K120264 | | Product<br>Classification | Class II, NQE | Class II, NQE | Class II, NQE | | Indications for Use | Non-contact<br>ophthalmic microscope,<br>optical pachymeter, and<br>camera intended for<br>examination of the<br>corneal endothelium<br>and for measurement of<br>the thickness of the<br>cornea. | Non-contact ophthalmic<br>microscope, optical<br>pachymeter, and camera<br>intended for examination of<br>the corneal endothelium and<br>for measurement of the<br>thickness of the cornea. | Non-contact ophthalmic<br>microscope, optical<br>pachymeter, and camera<br>intended for examination of<br>corneal endothelium and for<br>the measurement of the<br>thickness of the cornea. | | Device Type | Non-contact specular<br>microscope and<br>pachymeter | Non-contact specular<br>microscope and pachymeter | Non-contact specular<br>microscope and pachymeter | | Capturing Method | Auto capture, auto<br>alignment, auto focus<br>(3D, 2D)/Manual | Auto capture, auto<br>alignment, auto focus (3D,<br>2D)/Manual | Auto capture, auto alignment,<br>auto focus(3D)/Manual | | Capture Field | 0.25x 0.55 mm | 0.25x 0.55 mm | 0.24 x 0.4 mm | | Fixation Lamp<br>(Central) | 1 point | 1 point | 1 point | | Fixation Lamp<br>(Paracentral) | 8 points (5 degrees of<br>visual angle for each<br>point) | 8 points (5 degrees of visual<br>angle for each point) | 4 points (direction of 12, 2,<br>10 and 6 o'clock) | | Fixation Lamp<br>(Periphery) | 6 points (27 degrees of<br>visual angle for<br>direction of 12, 2, 10,<br>6, 4 and 8 o'clock) | 6 points (27 degrees of visual<br>angle for direction of 12, 2,<br>10, 6, 4 and 8 o'clock) | | | Camera | Built-in CCD camera | Built-in CCD camera | Built-in CCD image sensing<br>element camera | | Flash | Cyan LED | Cyan LED | Konan Xe tube | | Illumination for | InfraRed LED | InfraRed LED | Konan halogen lamp | #### TABLE 1 SPECULAR MICROSCOPE CEM-530 SUBSTANTIAL EQUIVALENCE CHART {7}------------------------------------------------ | focusing | | | | |-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------| | Pachymetry<br>Technology | Optical | Optical | Optical | | Measurement Range<br>Pachymetry | 300 to 1000 microns | 300 to 1000 microns | Unknown | | Accuracy of<br>pachymetry | $\pm$ 10 microns | $\pm$ 10 microns | $\pm$ 10 microns | | Auto Analysis | Yes | Yes | Yes | | Manual Analysis | Yes | No | Yes | | Analysis Speed | 2 seconds | 2 seconds | Unknown | | Record (1 patient) | Single mode: 1 endothelial image for<br>both right and left eyes<br>Paracentral and<br>Peripheral modes: 15<br>endothelial images for<br>both right and left eyes | Single mode: 1 endothelial<br>image for both right and left<br>eyes<br>Paracentral and Peripheral<br>modes: 15 endothelial<br>images for both right and left<br>eyes | 1 endothelial image for both<br>right and left eyes | | Built-in Printer | Thermal Line Printer<br>with auto cutter | Thermal Line Printer with<br>auto cutter | NA | | External Interface | USB-A: barcode, card<br>reader, LAN for data<br>output, Video output:<br>BNC terminal | USB-A: barcode, card<br>reader, LAN for data output,<br>Video output: BNC terminal | Video output: BNC terminal<br>Input port for mouse and<br>remote control | | Connection to Filing<br>System | Connectable | Connectable | Connectable | | Monitor | 8.4" SVGA color LCD<br>with touch panel | 8.4" SVGA color LCD with<br>touch panel | 19" color LCD monitor | | Input Power | 100 VA | 100 VA | 70 VA | | Dimensions | 291 (W) x 495 (D) x<br>457 (H) mm | 291 (W) x 495 (D) x 457 (H)<br>mm | 334 (W) x 486 (D) x 420 (H)<br>mm | | Weight | 20 Kg | 20 Kg | 20.5 Kg | | Compliance with<br>Safety Standards | IEC 60601-1<br>IEC 60601-1-2<br>ISO 15004-1<br>ISO 15004-2 | IEC 60601-1<br>IEC 60601-1-2<br>ISO 15004-1<br>ISO 15004-2 | IEC 60601-1<br>IEC 60601-1-2<br>ISO 15004-1<br>ISO 15004-2 | # NON-CLINICAL PERFORMANCE SUMMARY The performance testing conducted using the NIDEK Specular Microscope CEM-530 verified that the device operates as intended. The specifications to which the CEM-530 was verified to are substantially equivalent to the predicate devices and therefore, support a determination of substantial equivalence. The pachymetry functionality was evaluated in model eyes and the measurement accuracy of + 10 microns was confirmed. Additionally, the CEM-530 was subjected to electrical safety testing in accordance with IEC 60601-1, electromagnetic compatibility (EMC) testing in accordance with IEC 60601-1-2, and optical radiation safety testing in accordance with ISO 15004-1 and ISO 15004-2. {8}------------------------------------------------ ### CLINICAL PERFORMANCE SUMMARY A prospective clinical study was conducted to assess the agreement, accuracy and precision of the Specular Microscope CEM-530 Center Point Method by comparing results across three machines/operators to those obtained with the predicate device, the Cellchek Plus. Three populations were studied: young (18-28 years of age) and adult (29-80 years of age) healthy subjects and pathologic adult eyes (29-80 years of age). There were 79 subjects enrolled in the study: 28 in the non-pathologic young eye population, 28 in the non-pathologic adult eye population, and 23 in the pathologic adult eye population. Of those, 74 are included in the effectiveness population (28, 27, and 19, respectively). The agreement population consisted of all 74 subjects (28, 27 and 19, respectively). The precision population included a subset of the effectiveness population, 47 total with 15, 16 and 16 respectively. The pathologic adult subjects consisted of 19 subjects with the following pathologies which qualified them for the subgroup: Guttata (16, 84.2%) and Long Term Fuch's Dystrophy (3, 15.8%). The mean (SD) age of the non-pathologic young population was 22.2 (3.28) years; that of the non-pathologic adult population was 53.0 (13.42) years, and that of the pathologic adult population was 63.9 (12.39) years. All subjects combined had a mean age of 44.1 (20.54) years. The gender distribution of the total population was 27 males (36.5%) and 47 females (63.5%). The gender distribution of the non-pathologic young population was 15 males (53.6%) and 13 females (46.4%); that of the non-pathologic adult population was 8 males (29.6%) and 19 females (70.4%), and that of the pathologic adult population was 4 males (21.1%) and 15 females (78.9%). The majority of subjects were white (66/74 subjects, 89.2%). All of the pathologic adult subjects were white (19/19). The non-pathologic adult were white (21/27), American Indian (1/27), Asian (1/27), African American (1/27) and other (3/27). The nonpathologic young subjects were white (26/28) and other (2/28). With regard to iris color, the highest percentage of subjects had study eyes with brown irides (43.2%), followed by blue irides (28.4%), green irides (6.8%), black irides (1.4%) and gray irides (1.4%). # The result of Agreement study Table 2 provides a summary of the agreement data for all subjects with the Center Point Method while Tables 3, 4 and 5 provide the same data by subgroup (non-pathologic young eyes, nonpathologic adult eyes, pathologic adult eyes, respectively). {9}------------------------------------------------ | | CD | CV | % HEX | |--------------------------------------------------------|----------------------|------------------|--------------------| | Nidek Specular Microscope CEM-530 | | | | | N | 74 | 74 | 74 | | Mean (SD) | 2784.3 (428.52) | 22.1 (3.36) | 59.8 (7.30) | | Median | 2795.0 | 21.5 | 60.0 | | Min-Max | 1754 - 3637 | 15 - 35 | 41 - 81 | | Konan CellChek Plus | | | | | N | 74 | 74 | 74 | | Mean (SD) | 2738.6 (412.67) | 29.6 (4.16) | 62.6 (7.42) | | Median | 2754.5 | 29.0 | 63.0 | | Min- Max | 1757 - 3484 | 22 - 42 | 41 - 76 | | Device Comparisons | | | | | Mean Difference (SD) | 45.7 (104.11) | -7.6 (3.00) | -2.9 (5.90) | | Mean Difference (SD) as a % of the<br>CellChek reading | 1.71% (3.976%) | -25.20% (8.391%) | -4.07% (9.662%) | | 95% LOA | (-162.5, 253.9) | (-13.6, -1.6) | (-14.6, 8.9) | | Correlation (R2) | 0.9701 | 0.7019 | 0.6791 | | Deming Regression Intercept (95% CI) | -62.9 (-224.4, 98.6) | 0.1 (-6.4, 6.7) | -1.3 (-20.2, 17.5) | | Deming Regression Slope<br>(95% CI) | 1.0 (1.0, 1.1) | 0.7 (0.5, 1.0) | 1.0 (0.7, 1.3) | Table 2 Center Point Method: Four Corneal Specular Microscopic Variables Assessed with the Two Devices - All Configurations - All Subjects - Effectiveness Population ### Table 3 Center Point Method: Corneal Specular Microscopic Variables – All Configurations – Non-Pathologic Young Eyes - Effectiveness Population | | CD | CV | % HEX | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|-------------------|---------------------| | Nidek Specular Microscope CEM-530 | | | | | N | 28 | 28 | 28 | | Mean (SD) | 3069.5 (346.41) | 20.3 (1.90) | 63.0 (6.33) | | Median | 3104.0 | 21.0 | 63.0 | | Min-Max | 2437 - 3637 | 15 - 24 | 52 - 81 | | Konan CellChek Plus XL | | | | | N | 28 | 28 | 28 | | Mean (SD) | 3030.3 (296.93) | 27.4 (2.99) | 67.0 (5.03) | | Median | 3053.5 | 27.0 | 67.5 | | Min- Max | 2398 - 3484 | 22 - 34 | 57 - 76 | | Device Comparisons | | | | | Mean Difference (SD) | 39.2 (93.37) | -7.2 (2.94) | -4.0 (5.31) | | Mean Difference (SD)<br>as a % of the<br>CellChek reading | 1.19% (3.069%) | -25.59%, (8.660%) | -5.79% (7.933%) | | 95% LOA | (-147.5, 226.0) | (-13.1, -1.3) | (-14.6, 6.7) | | Correlation (R2) | 0.9695 | 0.3398 | 0.5841 | | Deming Regression Intercept (95% CI) | -482.7 (-892.1, -73.4) | 11.4 (-0.8, 23.5) | -35.9 (-82.1, 10.3) | | Deming Regression Slope<br>(95% CI) | 1.2 (1.0, 1.3) | 0.3 (-0.1, 0.8) | 1.5 (0.8, 2.2) | | For subjects in the Precision and Agreement cohort, the measurements from the first acceptable images from each machine within the same<br>configuration are used for the agreement analyses. | | | | The mean differences are calculated as (Nidek Specular Microscope CEM-530) - (Konan CellChek Plus XL). The mean differences as a % of the CELLCHEK reading are calculated for each subject first and then summarized. Table 4 Center Point Method: Corneal Specular Microscopic Variables - All Configurations - Non-Pathologic Adult Eyes – Effectiveness Population {10}------------------------------------------------ | | CD | CV | % HEX | |-----------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------|-------------------| | Nidek Specular Microscope CEM-530 | | | | | N | 27 | 27 | 27 | | Mean (SD) | 2678.6 (391.33) | 23.3 (3.27) | 57.1 (6.66) | | Median | 2772.0 | 23.0 | 56.0 | | Min-Max | 1956 - 3399 | 19 - 35 | 47 - 73 | | Konan CellChek Plus XL | | | | | N | 27 | 27 | 27 | | Mean (SD) | 2627.3 (348.93) | 31.1 (4.29) | 59.3 (8.11) | | Median | 2674.0 | 31.0 | 60.0 | | Min- Max | 2033 - 3344 | 22 - 42 | 45 – 74 | | Device Comparisons | | | | | Mean Difference (SD) | 51.3 (100.35) | -7.8 (3.04) | -2.2 (6.10) | | Mean Difference (SD) as a % of<br>the CellChek reading | 1.82% (3.753%) | -24.70% (7.780%) | -2.94% (10.324%) | | 95% LOA | (-149.4, 252.0) | (-13.9, -1.7) | (-14.4, 10.0) | | Correlation (R2) | 0.9697 | 0.7075 | 0.6750 | | Deming<br>Regression<br>Intercept<br>(95% CI) | -278.5 (-566.9, 9.9) | 2.0 (-13.1, 17.1) | 12.7 (-7.5, 32.9) | | Deming Regression Slope<br>(95% CI) | 1.1 (1.0, 1.2) | 0.7 (0.2, 1.2) | 0.7 (0.4, 1.1) | | For subjects in the Precision and Agreement cohort, the measurements from the first acceptable images from each machine within the same | | | | configuration are used for the agreement analyses. The mean differences are calculated as (Nidek Specular Microscope CEM-530) - (Konan CellChek Plus XL). The mean differences as a % of the CELLCHEK reading are calculated for each subject first and then summarized. | Table 5 Center Point Method: Corneal Specular Microscopic Variables – All Configurations – | | |---------------------------------------------------------------------------------------------|--| | Pathologic Adult Eves - Effectiveness Population | | | | | | | CD | CV | % HEX | |--------------------------------------------------------|-----------------------|--------------------|-----------------------| | Nidek Specular Microscope CEM-530 | | | | | N | 19 | 19 | 19 | | Mean (SD) | 2514.2 (351.02) | 23.0 (4.11) | 58.7 (7.96) | | Median | 2522.0 | 22.0 | 60.0 | | Min-Max | 1754 - 3078 | 17 - 33 | 41 - 71 | | Konan CellChek Plus XL | | | | | N | 19 | 19 | 19 | | Mean (SD) | 2466.9 (392.01) | 30.8 (4.26) | 60.8 (6.41) | | Median | 2564.0 | 30.0 | 61.0 | | Min- Max | 1757 - 3058 | 24 - 40 | 41 - 68 | | Device Comparisons | | | | | Mean Difference (SD) | 47.3 (127.36) | -7.8 (3.11) | -2.1 (6.47) | | Mean Difference (SD) as a % of the<br>CellChek reading | 2.32% (5.367%) | -25.34% (9.218%) | -3.15% (11.085%) | | 95% LOA | (-207.5, 302.0) | (-14.1, -1.6) | (-15.0, 10.8) | | Correlation (R2) | 0.9472 | 0.7237 | 0.6134 | | Deming Regression Intercept (95% CI) | 318.7 (-127.8, 765.3) | -6.4 (-23.1, 10.4) | -27.7 (-176.4, 121.0) | | Deming Regression Slope<br>(95% CI) | 0.9 (0.7, 1.1) | 1.0 (0.4, 1.5) | 1.4 (-1.0, 3.8) | For subjects in the Precision and Agreements from the first acceptable images from each machine within the same configuration are used for the agreement analyses. The mean differences are calculated as (Nidek Specular Microscope CEM-530) - (Konan CellChek Plus XL). The mean differences as a % of the CELLCHEK reading are calculated for each subject first and then summarized. The result of Precision study {11}------------------------------------------------ Table 6 provides a summary of the precision data for all subjects with the Center Point Method while Tables 7, 8 and 9 provide the same data by subgroup (non-pathologic young eyes, nonpathologic adult eyes, pathologic adult eyes, respectively). Data from a previous Nidek study, CEM-530-US-001, was used as the comparative data with respect to the Konan Cell Chek Plus results for the precision analysis detailed below. In the evaluation of the suitability of using the historical data it was found that the sample size, age, gender, ethnicity/race, iris color and clinical diagnosis of the pathologic sub group were all similar to allow the usage of the data. Table 6 Center Point Method: Precision Analyses – All Subjects – Effectiveness Population {12}------------------------------------------------ | Variable | NIDEK CEM-530<br>N=47 | Konan CellChek Plus<br>N=62 | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------------| | Endothelial Cell Density | | | | Repeatability SD | 48.9 | 62.4 | | Repeatability SD as a % of the Mean | 1.8% | 2.4% | | Repeatability Limit | 136.8 | 174.8 | | Repeatability Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.7829 | --- | | Reproducibility SD | 60.9 | 95.2 | | Reproducibility SD as a % of the Mean | 2.2% | 3.7% | | Reproducibility Limit | 170.4 | 266.7 | | Reproducibility Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.6391 | --- | | Coefficient of Variation of Endothelial Cell<br>Area (CV) | | | | Repeatability SD | 1.5 | 2.7 | | Repeatability SD as a % of the Mean | 6.7% | 8.5% | | Repeatability Limit | 4.1 | 7.5 | | Repeatability Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.5521 | --- | | Reproducibility SD | 1.8 | 2.7 | | Reproducibility SD as a % of the Mean | 8.4% | 8.6% | | Reproducibility Limit | 5.1 | 7.6 | | Reproducibility Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.6778 | --- | | % Hexagonality | | | | Repeatability SD | 3.5 | 5.4 | | Repeatability SD as a % of the Mean | 5.8% | 8.8% | | Repeatability Limit | 9.9 | 15.0 | | Repeatability Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.6586 | --- | | Reproducibility SD | 4.0 | 5.4 | | Reproducibility SD as a % of the Mean | 6.5% | 8.9% | | Reproducibility Limit | 11.1 | 15.2 | | Reproducibility Ratio (NIDEK CEM-530/<br>Konan CellChek Plus XL) | 0.7267 | --- | | N represents the total number of subjects in each eye population in the precision and agreement cohort. If any variance<br>component was negative, it was reported as 0. The repeatability limit is 2.8 times the repeatability standard deviation.<br>which is the square root of the residual within subject variance component. The reproducibility limit is 2.8 times the<br>reproducibility standard deviation, which is the square root of the sum of the variance components of operator+device, | | | #### Center Point Method: Precision Analyses – Non-Pathologic Young Eye Table 7 — Effectiveness Population operator+device x subject interaction, and residual within subject. | Variable | NIDEK CEM-530 | Konan CellChek Plus | |----------|---------------|---------------------| |----------|---------------|---------------------| {13}------------------------------------------------ | | N=15 | XL<br>N= 20 | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-------------| | Endothelial Cell Density | | | | Repeatability SD | 46.0 | 59.3 | | Repeatability SD as a % of the Mean | 1.5% | 2.1% | | Repeatability Limit | 128.8 | 166.0 | | Repeatability Ratio (NIDEK CEM-530/ Konan<br>CellChek Plus XL) | 0.7758 | --- | | Reproducibility SD | 51.9 | 70.2 | | Reproducibility SD as a % of the Mean | 1.7% | 2.5% | | Reproducibility Limit | 145.4 | 196.6 | | Reproducibility Ratio (NIDEK CEM-530/ Konan<br>CellChek Plus XL) | 0.7395 | --- | | Coefficient of Variation of Endothelial Cell Area (CV) | | | | Repeatability SD | 1.3 | 2.6 | | Repeatability SD as a % of the Mean | 6.4% | 8.9% | | Repeatability Limit | 3.5 | 7.4 | | Repeatability Ratio (NIDEK CEM-530/ Konan<br>CellChek Plus XL) | 0.4806 | --- | | Reproducibility SD | 1.5 | 2.7 | | Reproducibility SD as a % of the Mean | 7.5% | 9.1% | | Reproducibility Limit | 4.2 | 7.6 | | Reproducibility Ratio (NIDEK CEM-530/ Konan<br>CellChek Plus XL) | 0.5515 | --- | | % Hexagonality | | | | Repeatability SD | 3.1 | 4.2 | | Repeatability SD as a % of the Mean | 4.8% | 6.4% | | Repeatability Limit | 8.6 | 11.7 | | Repeatability Ratio (NIDEK CEM-530/ Konan<br>CellChek Plus XL)…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...