The Konan Specular Microscope XIV, Cellchek Plus, is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.
Device Story
Non-contact ophthalmic specular microscope, optical pachymeter, and camera; captures images of corneal endothelium using CCD camera and infrared focusing. Device features 5-axis mechanical alignment unit; touch screen interface; computer-assisted cell counting and analysis software. Used in clinical settings by eye care professionals. Inputs: reflected light from patient eye. Outputs: visual images of corneal endothelium, corneal thickness measurements, and cell distribution analysis (density, hexagonality, coefficient of variation). Healthcare providers use output to assess corneal health and make clinical decisions regarding ocular condition. Benefits include non-invasive examination and automated analysis of endothelial cell morphology.
Clinical Evidence
Clinical study compared four analysis methods (Manual, PC-Assist with/without redrawing, Center) using 40 eye images. Evaluated agreement and variability for cell density, hexagonality, and coefficient of variation. Results showed good correlation for cell density between methods; other parameters showed weaker correlation. Variability (standard deviation as % of mean) for cell density ranged from 0.78% to 1.23%. Study confirms analysis methods are consistent with predicate.
Indicated for examination of the corneal endothelium and measurement of corneal thickness in patients requiring ophthalmic assessment. Prescription use only.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
K120264
APR 1 1 2012
### 510(K) SUMMARY
### 11.1 SUBMITTER INFORMATION
| A. | Company Name: | Konan Medical, Inc. |
|----|------------------------|-----------------------------------------------------------|
| B. | Company Address: | 10-29, Miyanishicho, Nishinomiya<br>Hyogo 662-0976, Japan |
| C. | Company Phone: | +81-798-36-3456 |
| D. | Company Facsimile: | +81-798-26-1028 |
| E. | Contact Person: | Tatsuya Kasahara |
| F. | Date Summary Prepared: | January 24, 2012 |
### 11.2 DEVICE IDENTIFICATION
| A. | Device Trade Name: | Konan Specular Microscope XIV, Cellchek Plus |
|----|-------------------------------|----------------------------------------------|
| B. | Common Name: | Specular Microscope |
| C. | Classification Name(s): | AC-powered Slit Lamp Biomicroscope |
| D. | Classification Regulation(s): | 886.1850 |
| E. | Device Class: | Class 2 |
| F. | Product Codes: | NQE |
| G. | Advisory Panel: | Ophthalmic |
## 11.3 IDENTIFICATION OF PREDICATE DEVICE
The predicate device is the Konan Medical, Inc. Noncon Robo Pachy F&A non-contact specular microscope, which was cleared by FDA under 510(k) number K062763 on February 22, 2008.
### 11.4 DEVICE DESCRIPTION
The Konan Specular Microscope XIV, CellChek Plus, is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea. Cell counting and analysis programs are included, and allow for analysis of the cell distribution of the eye (the same analysis software present in the predicate device is also present in the modified device).
Konan Medical, Inc.
Confidential
{2}------------------------------------------------
When photographing the corneal endothelium, the equipment performs the alignment and automatically focuses by capturing the reflected light from the patient's eye with the CCD camera. The device permits visual inspection and photography of the corneal endothelium and measurement of the corneal thickness without any object contacting the eye. It features focusing by means of infrared techniques, and computer-assisted cell counting and cell analysis capabilities. The computer functions are also used to aid in setting up the various features of the machine and to aid in photography. Photographic images are temporarily stored in the system's memory, and are preserved by using a printer.
Both the image of the corneal endothelium and the various computerized control functions are displayed on the touch screen.
The parts of the device that come into contact with a patient are the forehead rest and the chin rest. The head rest is comprised of Polytetrafluoroethylene (PTFE), which is known as Teflon®, and the material that comprises the chin rest is Acrylonitrile butadiene styrene (ABS).
### 11.5 INDICATIONS FOR USE
The Konan Specular Microscope XIV, Cellchek Plus, is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.
### 11.6 TECHNOLOGICAL CHARACTERISTICS
The Konan Specular Microscope XIV, Cellchek Plus is technically equivalent to the Konan Noncon Robo Pachy F&A non-contact specular microscope. The principal design modification for the device is that the mechanical unit used for alignment has been changed. The modified mechanical unit now consists of five axes (the predicate device had 3 axes) and the optical unit can be placed in any position over the cornea to obtain an image of endothelial cells. A smaller computer mother board was implemented, and a new computer program was installed for the modified mechanical system. The system now operates by means of a touch screen.
### 11.7 SUMMARY OF TESTING
#### Non Clinical Tests A.
The following testing was performed on the Konan Specular Microscope XIV, Cellchek Plus:
- The modified CELLCHEK PLUS device was subjected to electrical safety testing . in accordance with IEC 60601-1.
{3}------------------------------------------------
- The modified CELLCHEK PLUS device was subjected to electromagnetic compatibility (EMC) testing in accordance with IEC 60601-1-2.
- . The modified CELLCHEK PLUS device was subjected to performance testing, including optical radiation safety testing in accordance with ISO 15004-1 and 15004-2.
- The modified CELLCHEK PLUS device was subjected to software validation . testing in accordance with IEC 60601-1-4.
#### B. Clinical Tests
A Clinical test was performed with four "classifiers", each analyzing the same 40 images of eyes by using the Konan predicate device Noncon Robo Pachy F&A noncontact specular microscope. Each classifier analyzed each image three times. The analyses of the classifiers were analyzed for cell density, hexagonality, and coefficient of variation.
Agreement and variability of the analysis methods was obtained using a sample that included virtually no eyes with Percent Hexagonality <45, Coefficient of Variation >0.41, or Cell Density <2100. Agreement and variability of the analysis methods is not known for eyes with parameters beyond these values.
Agreement Between Methods of Analysis
For a single image from each eye, cell density, coefficient of variation and percent hexagonality were determined by each analysis method. For a given parameter and pair of methods of image analysis, agreement between outputs was assessed. This was done by taking the difference between the two outputs for each image and then calculating the mean difference, and the 95% limits of agreement. These measures of agreement between analysis methods were calculated for each parameter. Note that each measure estimates the degree of agreement between the different methods of analysis applied to a single image. It does not take into account variation due to repeated image capture. Variations between different images of the same eye will significantly reduce agreement. The listed values should not be taken as estimates of the agreement of the measurements associated with repeated image capture.
Confidential
{4}------------------------------------------------
| Agreement Between Methods of Analysis for Cell Density | | | |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------|
| | Mean Difference1<br>(%)<br>Note : a negative<br>number indicates<br>that first method<br>gives lower results<br>than the second | Limits of Agreement2 (%) : ~95% of<br>Differences should fall between these<br>figures | |
| Analysis Methods | | Lower 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) | Upper 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) |
| Manual vs. PC-<br>Assist (with<br>redrawing) | 0.06 | -0.78 | 0.9 |
| Manual vs. PC-<br>Assist (without<br>redrawing) | 0.00 | -2.3 | 2.3 |
| Center vs. PC-<br>Assist (with<br>redrawing) | -0.11 | -1.47 | 1.25 |
| Center vs. PC-<br>Assist (without<br>redrawing) | -0.17 | 2.71 | 2.37 |
| Manual vs. Center | 0.16 | -1.1 | 1.42 |
| Agreement Between Methods of Analysis for Coefficient of Variation | | | |
|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------|
| | Mean Difference1<br>(%)<br>Note : a negative<br>number indicates<br>that first method<br>gives lower results<br>than the second | Limits of Agreement2 (%) : ~95% of<br>Differences should fall between these<br>figures | |
| Analysis Methods | | Lower 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) | Upper 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) |
| Manual vs. PC-<br>Assist (with<br>redrawing) | 0.43 | -2.35 | 3.21 |
| Manual vs. PC-<br>Assist (without<br>redrawing) | 0.24 | -5.62 | 6.10 |
| Center vs. PC-<br>Assist (with<br>redrawing) | 2.38 | -4.18 | 8.94 |
| Center vs. PC-<br>Assist (without<br>redrawing) | 2.18 | -4.18 | 8.94 |
| Manual vs. Center | -1.96 | -8.58 | 4.66 |
Konan Medical, Inc.
:
·
:
{5}------------------------------------------------
| Agreement Between Methods of Analysis for Percent Hexagonality | | | |
|----------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------|
| | Mean Difference1<br>(%)<br>Note : a negative | Limits of Agreement2 (%) : ~95% of<br>Differences should fall between these<br>figures | |
| Analysis Methods | number indicates<br>that first method<br>gives lower results<br>than the second | Lower 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) | Upper 95%Limit of<br>Agreement (Mean<br>Difference - 2*sd) |
| Manual vs. PC-<br>Assist (with<br>redrawing) | -0.08 | -3.32 | 3.16 |
| Manual vs. PC-<br>Assist (without<br>redrawing) | 1.47 | -3.63 | 6.57 |
| Center vs. PC-<br>Assist (with<br>redrawing) | 2.13 | -3.03 | 7.29 |
| Center vs. PC-<br>Assist (without<br>redrawing) | 3.67 | -3.13 | 10.47 |
| Manual vs. Center | -2.23 | -7.03 | 2.57 |
Footnotes:
1. Mean Difference is the average across images of:
Cell Density Method1 – Cell Density Method2 100 ×
Cell Density Method1 + Cell Density Method2
- 2. Approximately 2.5% of differences would be expected to fall below the Lower limit of Agreement and about 2.5% would be expected to fall above the Upper Limit of Agreement. The Limits of Agreement are defined as:
The Mean Difference (footnote 1, above) ±2(standard deviation). The standard deviation is calculated across all of the 40 Differences as defined in footmote 1, above.
{6}------------------------------------------------
| Agreement Between Methods of Analysis | | | |
|-------------------------------------------------|--------------------------|-----------|-----------|
| | Correlation Coefficients | | |
| Manual vs. PC-<br>Assist (with<br>redrawing) | R2=0.8719 | R2=0.2709 | R2=0.0006 |
| Manual vs. PC-<br>Assist (without<br>redrawing) | R2=0.9813 | R2=0.8185 | R2=0.6659 |
| Manual vs. Center | R2=0.9636 | R2=0.2041 | R2=0.4307 |
| Center vs. PC-<br>Assist (with<br>redrawing) | R2=0.8544 | R2=0.201 | R2=0.1846 |
| Center vs. PC-<br>Assist (without<br>redrawing) | R2=0.955 | R2=0.2318 | R2=0.3811 |
Special 510(K) Premarket Notification-CellChek Plus
. .
Chart gives the coefficients of correlation of the parameters indicated at the top of the column when the parameters are computed according to the method indicated in the first column. These tables indicate that cell density has a good correlation with the Manual Method (an exact method); the other parameters are only weakly-correlated, especially without redrawing.
{7}------------------------------------------------
Variability Associated with the Analysis Methods The variability of the results associated with each analysis method was assessed. For each of the forty images, cell density, coefficient of variation and percent hexagonality were determined by each method. Each analysis was repeated on the same image three times (presentation was randomized) by each of four "classifiers." For a given analysis method, the standard deviation of the within-image results was calculated as a measure of variability.
Note that this measure estimates the variability associated only with repeated application of the analysis method to a single image. Differences between two or more images of the same eye will significantly increase variability. The listed values should not be taken as estimates of the variability of the measurements associated with repeated image capture.
| Variability Associated with Analysis Method | | | |
|---------------------------------------------|-----------------------------------------------------------------------------------|--------------------------|----------------------|
| | Standard Deviation (within-image)<br>Expressed as a percentage of the mean value* | | |
| Analysis Methods | Cell Density | Coefficient of Variation | Percent Hexagonality |
| PC-Assist (with redrawing) | 1.14 | 4.23 | 3.87 |
| PC-Assist (without redrawing) | 0.78 | 8.02 | 12.67 |
| Center | 1.23 | 6.60 | 6.42 |
* Value in table is the mean across 40 images of the variability standard deviation (square root of the sum of the "within-observer" variance plus the "between-observer" variance) divided by the mean value for the image (expressed as a percentage).
### 11.8 CONCLUSIONS DRAWN FROM STUDIES
The conclusions drawn from the non-clinical and clinical testing demonstrate that the modified device, Konan Specular Microscope XIV, Cellchek Plus, is substantially equivalent to the Konan predicate device Noncon Robo Pachy F&A non-contact specular microscope.
Confidential
{8}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration. 10903 New Hamnshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Konan Medical, Inc. c/o Mr. David S. Fernquist 2976 Calle Gaucho San Clemente, CA 92673
APR 1 1 2012
Re: K120264
Trade/Device Name: Konan Specular Microscope XIV, Cellchek Plus Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered Slit Lamp Biomicroscope Regulatory Class: II Product Code: NOE Dated: March 8, 2012 Received: March 12, 2012
Dear Mr. Fernquist:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{9}------------------------------------------------
### Page 2 - Mr. David S. Fernquist
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# INDICATIONS FOR USE
### 510(k) Number (if known):
Konan Specular Microscope XIV. Cellchek Plus Device Name:
#### Indications for Use:
The Konan Specular Microscope XIV, Cellchek Plus, is a non-contact ophthalmic. microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ALCA
K120264
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number ***_***
Konan Medical, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.