The sponsor performed a search of the MDR and MAUDE databases to assess the performance and safety of the device materials and indications, using this real-world postmarket surveillance data to support the substantial equivalence determination.
MDR and MAUDE database search; Postmarket surveillance database review
Patients undergoing procedures involving drill bits with similar materials and indications; Number of Sites: Not applicable
Not applicable for this study
Reported incidences of adverse events
Indications for Use
The Renovo Life Hard Carbon Coated Drill Bit is used to cut bone prior to insertion of bone screws or guide pins.
Device Story
Manual surgical drill bit; used to create holes in bone for screw or pin fixation. Device consists of surgical grade stainless steel (ASTM F899) with hard carbon coating on cutting portion. Available in solid or cannulated configurations; diameters 1.1-8.0 mm; lengths 60-300 mm. Operated by surgeons in clinical settings (OR, clinic). Coating intended to reduce bone heating and improve fatigue resistance. Device provided non-sterile; intended for single use.
Clinical Evidence
No clinical testing was performed. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), steam sterilization validation (ANSI/AAMI/ISO 17665-1), accelerated aging (ASTM F1980-07), and mechanical integrity/wear surface analysis (ASTM F1877).
Technological Characteristics
Surgical grade stainless steel (ASTM F899) with hard carbon coating. Manual surgical instrument; standard connection; solid or cannulated form factor. Sterilization via steam (ANSI/AAMI/ISO 17665-1).
Indications for Use
Indicated for cutting bone prior to insertion of bone screws or guide pins in orthopedic, dental, oral, maxillofacial, plastic, ENT, and neurological surgical procedures.
Regulatory Classification
Identification
An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, burr, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.
Predicate Devices
Mahe Surgical Drill Bits (Class I)
Reference Devices
Nobel Biocare Amorphous Diamond Coated Drill (K990846)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Renovo Life LLC % Mr. Donald Gunther Regulatory Consultant Orgenix LLC 111 Hill Road Douglassville, Pennsylvania 19518
Re: K151661
Trade/Device Name: Renovo Life Hard Carbon Coated Drill Bit Regulation Number: 21 CFR 888.4540 Regulation Name: Orthopedic Manual Surgical Instrument Regulatory Class: I Product Codes: HTW Dated: December 5, 2015 Received: December 8, 2015
Dear Mr. Gunther:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151661
#### Device Name
Renovo Life Hard Carbon Coated Drill Bit
Indications for Use (Describe)
The Renovo Life Hard Carbon Coated Drill Bit is used to cut bone prior to insertion of bone screws or guide pins.
Type of Use (Select one or both, as applicable)
| × Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
| Manufacturer: | Renovo Life LLC<br>1104 Spruce Street<br>Belmont, NC 28012 |
|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 14, 2016 |
| Submitted by: | Renovo Life LLC |
| Company Contact | Harold Crowder<br>+1(704) 224-4136<br>hcrowder@renovolife.com |
| Official Correspondent | Orgenix LLC<br>Mr. Donald W. Guthner<br>111 Hill Road<br>Douglassville, PA 19518<br>+1-646-460-2984<br>+1-484-363-5879 (FAX)<br>dg@orgenix.com |
| Proprietary Name: | Renovo Life Hard Carbon Coated Drill Bit |
| Performance standards: | The Renovo Life Hard Carbon Coated Drill Bits are made of<br>surgical grade stainless steel conforming to ASTM F899. |
| Regulation:<br>Common/Usual Name:<br>Classification name:<br>Review Panel:<br>Primary Product Code:<br>Device Class: | 21 CFR 888.4540<br>Drill Bit<br>Orthopedic manual surgical instrument<br>Orthopedic<br>HTW<br>Class I |
| Substantial Equivalence: | Substantial equivalence for the Renovo Life Hard Carbon Coated<br>Drill Bits is based on its similarities in indications for use, design<br>features, operational principles and material composition when compared<br>to the predicate devices. |
| Primary Predicate Device<br>Reference Device: | Mahe Surgical Drill Bits (Class I)<br>K990846 Nobel Biocare Amorphous Diamond Coated Drill (Class II)<br>The subject device is substantially equivalent to similar previously<br>cleared devices. |
{4}------------------------------------------------
| Device Description: | Renovo Life Hard Carbon Coated Drill Bits are available in diameters ranging from 1.1 mm to 8.0 mm, lengths ranging from 60 mm to 300 mm, and either solid or cannulated. The drill bits are designed with a standard connections.<br>Renovo Life Hard Carbon Coated Drill Bits are made of surgical grade stainless steel. The bone-cutting portion of the drill is coated with a hard carbon coating.<br>Renovo Life Hard Carbon Coated Drill Bits are provided non-sterile and are intended for single use only. |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use:: | The Renovo Life Hard Carbon Coated Drill Bit is used to cut bone prior to insertion of bone screws or guide pins. |
| Common use of drill bits: | |
Drill bits are intended for use in the drilling of bone and other bonerelated tissue in a variety of surgical procedures. Procedures for use on bone are often used in orthopedic, dental, oral surgery, maxillofacial, plastics, ENT, and neurological surgeries. Drill bits are also used in the placement of screws, wires, pins, and other fixation devices.
The Renovo Life Hard Carbon Coated Drill Bits are designed to drill holes in bones
The Renovo Life Hard Carbon Coated Drill Bits have undergone mechanical testing to show Non-Clinical substantial equivalence to the predicate device.
Testing A comprehensive series of tests were conducted on the Renovo Life drill bits coated with a hard carbon coating (manufactured by BioMedFlex LLC.). Testing was performed on uncoated, coated, and aged coated Renovo Life drill bits. The test results were analyzed and are summarized in the body of this report to determine if the coating provided an adequate interface for drilling bone and demonstrated reduced heating and improved resistance to fatigue (longevity). Table 1 outlines the series of tests conducted on the coated drill bit to assess biocompatibility and mechanical integrity of the coating.
{5}------------------------------------------------
| TEST METHOD | TEST STANDARDS |
|-------------------------------------|------------------------------------------------------|
| Biocompatibility ISO 10993 | |
| Cytotoxicity | ISO 10993 |
| Maximization Sensitization | ISO 10993 |
| Intracutaneous Study | ISO 10993 |
| Systemic Toxicity | ISO 10993 |
| Steam Sterilization - Part I | ANSI/AAMI/ISO 17665-1:2006<br>(EN ISO 17665-1:2006) |
| Steam Sterilization - Part II | ANSI/AAMI/ISO 17665-1:2006<br>(EN ISO 17665-1:2006) |
| Accelerated Aging - 1 year | ASTM F1980-07 |
| Mechanical Integrity of Coating - | No Standards for this test |
| Wear surface Analysis of Drill Bits | ASTM F1877 - guideline for<br>methods and comparison |
#### TABLE 1: - SUMMARY OF TESTS AND ANALYSES CONDUCTED
No clinical testing was performed.
Clinical Testing Comparison Table
| Information | Subject Device | Primary Predicate | Reference Device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Renovo Life LLC | Mahe Medical | Nobel Biocare |
| Trade Name | Renovo Life Hard<br>Carbon Coated Drill<br>Bits | Mahe Surgical Drill<br>Bits | Amorphous Diamond<br>Coated Drill Bits |
| 510(k) Number | K151661 | N/A (Class I) | K990846 |
| Product Code | HTW | HTW | DZI |
| Classification | 888.4540 | 888.4540 | 872.4120 |
| Materials | Stainless Steel<br>Hard Carbon Coating | Stainless Steel | Stainless Steel<br>Amorphous Diamond |
| Indications for Use | The Renovo Life Hard<br>Carbon Coated Drill<br>Bit is used to cut bone<br>prior to insertion of<br>bone screws or guide<br>pins | Drill Bits are used to<br>cut bone prior to<br>insertion of bone<br>screws or guide pins | The drill is used to<br>cut into the maxima or<br>mandible to create an<br>osteotomy for endosseous<br>dental implant placement. |
| Design | Standard square head | Standard square head | Latch end to use with hand<br>piece |
| Diameters | 1.1 to 8.0 mm | 1.1 to 8.0 mm | 2.0 mm |
| Lengths | 60 to 300 mm | 60 to 300 mm | 18 mm |
| Styles | Solid, Cannulated | Solid, Cannulated | |
| Cutting Flutes | Two | Two | Three |
| Sterility | Non-sterile | Non-sterile | Non-sterile |
| Packaging | Plastic bag | Plastic Bag | Plastic vial with foam<br>inserts |
Substantial equivalence for the Renovo Life Hard Carbon Coated Drill Bits is based on its similarities in indications for use, design features, operational
{6}------------------------------------------------
principles and material composition when compared to the primary and reference predicate devices:
- 1. Mahe Surgical Drill Bits (Class I)
Substantial equivalence for the Renovo Life Hard Carbon Coated Drill Bits is based on its similarities in material composition when compared to the reference device:
- 2. Nobel Biocare Amorphous Diamond Coated Drill (K990846)
A search of the MDR and MAUDE databases showed no reported incidences with these materials for this indication, indicating a certain level of performance of the product.
Thus, the Renovo Life Hard Carbon Coated Drill Bits are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.