Arcos One-piece Femoral Revision System
K151603 · Biomet, Inc. · LPH · Oct 6, 2015 · Orthopedic
Device Facts
| Record ID | K151603 |
| Device Name | Arcos One-piece Femoral Revision System |
| Applicant | Biomet, Inc. |
| Product Code | LPH · Orthopedic |
| Decision Date | Oct 6, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3358 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. 5. Revision procedures where other treatment or devices have failed. The Arcos One-Piece Femoral Revision System hip components are single-use implants, intended for uncemented use only.
Device Story
Arcos One-piece Femoral Revision System is a monolithic, porous-coated femoral stem implant used in hip replacement surgery. Designed for primary and revision total hip and hemi-hip arthroplasty; intended for uncemented fixation. System includes implant-specific instrumentation. Used by orthopedic surgeons in clinical settings. Device replaces damaged femoral head/neck; restores joint function; provides stability for patients with degenerative disease or fractures. Output is a physical structural replacement for the proximal femur.
Clinical Evidence
No clinical data provided. Substantial equivalence based on non-clinical bench testing, including femoral stem fatigue and range of motion assessments.
Technological Characteristics
Monolithic, porous-coated femoral stem; uncemented fixation. Available in multiple lengths and proximal body sizes. Materials and design consistent with predicate Arcos modular system. Single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring primary or revision total hip or hemi-hip arthroplasty due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or proximal femur fractures (non-union, femoral neck, trochanteric) unmanageable by other techniques.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
- Modular Femoral Revision System (Arcos) (K090757)
- Reach Femoral Component (K000760)
Reference Devices
- Echo Bi-Metric Microplasty Hip Stem (K143009)
- Echo Bi-Metric Microplasty Line Extension (K150503)
- Echo Bi-Metric Press Fit Stems (K070274)
Related Devices
- K133381 — OMNI ARC MONOBLOCK HIP STEM · Omnilife Science · Mar 13, 2014
- K021403 — HA MALLORY/HEAD POROUS FEMORAL STEM & HA MALLORY/HEAD POROUS LATERALIZED FEMORAL STEM · Biomet Orthopedics, Inc. · May 31, 2002
- K221104 — Actera hip system · Conformis, Inc. · Aug 4, 2022
- K963486 — REVISION HIP SYSTEM · Smith & Nephew Richards, Inc. · Nov 27, 1996
- K072417 — SMITH & NEPHEW MIS HIP STEM · Smith & Nephew, Inc. · Jan 10, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure, emphasizing the department's role and national scope. The logo is simple, yet conveys a sense of care and public service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2015
Biomet Incorporated % Mr. Paul Cutlan Regulatory Affairs Manager (Hips) Waterton Industrial Estate Bridgend. Bridgend CF31 3XA UNITED KINDOM
Re: K151603 Trade/Device Name: Acros One-piece Femoral Revision System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, KWL, LWJ, KWZ, JDI, OQG, OQH, OQI, PBI, KWY Dated: September 15, 2015 Received: September 17, 2015
Dear Mr. Cutlan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K151603
Device Name
Arcos One-piece Femoral Revision System
#### Indications for Use (Describe)
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
5. Revision procedures where other treatment or devices have failed.
The Arcos One-Piece Femoral Revision System hip components are single-use implants, intended for uncemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The color of the text is dark gray, and the background is white. The overall impression is clean and modern.
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Arcos One-piece Femoral Revision System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581-0857<br>Phone Number: (574) 267-6639<br>Fax Number: (574) 371-1027<br>Establishment Registration Number: 1825034 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Paul Cutlan<br>Regulatory Affairs Manager |
| Date: | October 2nd 2015 |
| Subject Device: | Trade Name: Arcos One-piece Femoral Revision System<br>Common Name: Hip Prosthesis |
| | Classification Name: |
| | ●<br>LPH - Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3358) |
| | ●<br>LZO - Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (21 CFR 888.3353) |
| | ●<br>KWL - Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis (21 CFR 888.3360) |
| | ●<br>LWJ - Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis (21 CFR 888.3360) |
| | ●<br>KWZ - Hip joint metal/polymer constrained cemented or uncemented<br>prosthesis (21 CFR 888.3310) |
| | ●<br>JDI - Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3350) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "BIOMET" in all caps. The letters are a dark gray color and are in a simple, sans-serif font. The letters are evenly spaced and the word is centered in the image. The background is white.
- . OQG - Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis (21 CFR 888.3358)
- OQH - Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3350)
- OQI - Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR 888.3353)
- . PBI – Hip joint metal/polymer constrained cemented or uncemented prosthesis (21 CFR 888.3310)
- KWY - Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis (21 CFR 888.3390)
#### Legally marketed device(s) to which equivalence is claimed:
- K090757 Modular Femoral Revision System (Arcos) (Biomet)
- K000760 Reach Femoral Component (Biomet)
#### Reference Devices:
Labeling and Biocompatibility:
- · K143009 Echo Bi-Metric Microplasty Hip Stem (Biomet).
Implant Materials/Instrument materials/Type I taper Trunnion:
- K150503 Echo Bi-Metric Microplasty Line Extension (Biomet).
Implant Stem Diameters:
- K070274 Echo Bi-Metric Press Fit Stems (Biomet).
#### Device Description:
The Arcos One-piece Femoral Revision System is to be used as a femoral stem component in hip replacement surgery. The system is designed for both primary and revision total hip and hemi hip arthroplasty. All implants included in this system are intended for single use only. The system also includes implant specific instrumentation for all stem variants.
#### Intended Use and Indications for Use:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
{5}------------------------------------------------
- 2. Rheumatoid arthritis.
- Correction of functional deformity. 3.
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- 5. Revision procedures where other treatment or devices have failed.
The Arcos One-piece Femoral Revision System hip components are single-use implants, intended for uncemented use only.
#### Summary of Technical Characteristics:
The rationale for substantial equivalence is based on consideration of the following:
- The technological characteristics of the Arcos One-piece Femoral Revision System are the same as those of predicate devices (K090757 and K000760) in terms of design, material, and principle of operation with the exception of slight modifications as described within the 510(k).
- . The Arcos One-piece Femoral Revision System is a monolithic variant of the Modular Femoral Revision System (Arcos) (K090757). Like the modular system, the Arcos Onepiece is designed as a porous-coated femoral stem available in multiple lengths and proximal bodies for primary and revision procedures.
#### Summary of Performance Data (Non-clinical and/or Clinical) :
#### Non-clinical Tests:
- Femoral Stem Fatigue
- Range of Motion
#### Clinical Tests:
None provided as a basis for substantial equivalence.
#### Substantial Equivalence Conclusion:
The proposed Arcos One-piece Femoral Revision System has the same intended use and similar indications for use as the predicate devices. The proposed device has similar technological characteristics to the predicates, and the information provided herein demonstrates that:
- . any differences do not raise new questions of safety and effectiveness; and
- the proposed device is at least as safe and effective as the marketed predicate device.