Cellvizio 100 Series System with Confocal Miniprobes

K151593 · Mauna Kea Technologies · OWN · Oct 23, 2015 · Gastroenterology, Urology

Device Facts

Record IDK151593
Device NameCellvizio 100 Series System with Confocal Miniprobes
ApplicantMauna Kea Technologies
Product CodeOWN · Gastroenterology, Urology
Decision DateOct 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures.

Device Story

Confocal laser system using fiber optic probes; inputs optical signals to image internal tissue microstructure. Device consists of CelioFlex™ UHD 5 Confocal Miniprobe, a modification of CystoFlex™ UHD R featuring a distal 'Grabule' component (stainless steel 316L) and 3-meter fiber length. Grabule acts as a grip for laparoscopic forceps, allowing surgeons to position the probe tip precisely against target tissue during laparoscopic procedures. Used in OR/clinic settings by physicians; inserted via trocar (≥4.95mm). Output displayed on laparoscope monitor for real-time visualization. Facilitates tissue imaging during surgery; benefits include enhanced control and protection of probe tip from forceps-related wear.

Clinical Evidence

Bench testing only. Verification and validation testing included sterilization efficacy, reprocessing sterilization validation, and biocompatibility (cytotoxicity) validation. All tests passed requirements.

Technological Characteristics

Confocal laser imaging system; fiber optic probe. Materials: Stainless steel 316L (Grabule), epoxy glue (Epotek 301). Dimensions: 3-meter fiber length, compatible with trocars ≥4.95mm. Connectivity: Part of Cellvizio 100 Series System. Sterilization: Validated reprocessing.

Indications for Use

Indicated for qualified physicians to visualize body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures, and to image internal tissue microstructure.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 23, 2015 Mauna Kea Technologies Mr. Michael A. Daniel Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460 Re: K151593 Trade/Device Name: CelioFlex " UHD 5 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN, GCJ Dated: September 24, 2015 Received: September 25, 2015 Dear Mr. Daniel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # 6. Indications for Use Statement DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. #### 510(K) Number (if known) K151593 Device Name CelioFlex™ UHD 5 · Indications for Use (Describe) Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FORM FDA 3881 (1/14) {3}------------------------------------------------ # 7. Premarket Notification 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. # 510(k) Number: #### Applicant Information: | Date Prepared: | June 9, 2015 | |-------------------|-------------------------------------------------------------| | Name: | Mauna Kea Technologies | | Address: | 9 rue d'Enghien<br>F-75010 Paris, France | | Phone: | +33 1 48 24 03 45 | | Fax: | +33 1 48 24 12 18 | | Contact Person: | Michael A Daniel, Consultant<br>madaniel@clinregconsult.com | | Phone Number: | (415) 407-0223 | | Office: | (775) 392-2970 | | Facsimile Number: | (610) 545-0799 | #### Device Information: | Device Trade Name: | CelioFlex™ UHD 5 Confocal Miniprobe to be used with<br>Cellvizio® 100 Series System | |---------------------------|-------------------------------------------------------------------------------------| | Common Name: | Endoscope and Accessories | | Classification Name(s): | Confocal Optical Imaging | | Product Code/ Regulation: | OWN / GCJ 21 CFR 876.1500 | | Classification: | Class II | # Predicate Device: - Circon's USA Series™ Laparoscope has been cleared with K013165 #### Reference Devices: - The Cellvizio 100 Series System with Confocal Miniprobes cleared via K111047. - . The CystoFlex™ UHD R has been cleared with K141358 # Device Description: Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures. The CelioFlex™ UHD 5 confocal Miniprobe is a CystoFlex™ UHD R Confocal Miniprobe with an additional "Grabule" component and with an optical fiber of 3 meters vs. 2 meters. The Grabule is a metallic part glued to the distal tip and the adjacent part of the sheath of the CelioFlex™ UHD 5 Confocal Miniprobe. The {4}------------------------------------------------ Grabule has been designed as a grip for laparoscopic forceps, facilitating user control of the Miniprobe distal tip during laparoscopic procedures. Both the stainless steel 316L metal comprising the Grabule and the epoxy glue (Epotek 301) used to cement the Grabule to the Miniprobe are currently used in the CystoFlex™ UHD R Miniprobe reference device. The Grabule shields the Miniprobe tip from potential wear that repeated handling by forceps might cause. During standard laparoscopic procedures, CelioFlex™ UHD 5 Confocal Miniprobe is inserted through a pre-installed trocar. When the CelioFlex™ UHD 5 appears on the laparoscope monitor, the surgeon grasps the distal tipped Grabule with the laparoscopic fenestrated forceps to position the Confocal Miniprobe facing the tissue to be imaged. # Indications for Use: Cellvizio® 100 Series System with Confocal Miniprobes is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures. # Comparison to Previous Devices: No change is being made to the fundamental technology and operating principle of the previously cleared CystoFlexTM UHD R Confocal Miniprobe (K141358). The CelioFlex™ UHD 5 is identical to this reference device, only differing by the addition of the Grabule on its distal tip and by the fiber length (2m for CystoFlex™ UHD R and 3m for CelioFlex™ UHD 5). The Indications for Use and Intended Use of CelioFlex™ UHD R are the same as Circon's USA Series™ Laparoscopes (K013165) predicate and CystoFlex™ UHD R (K141358) reference devices, respectively. Due to the addition of the Grabule on its distal tip, the CelioFlex™ UHD 5 is designed to fit into trocars of 4.95mm or greater. Attached verification and validation testing demonstrates that the modifications meet the design specifications and user needs. Sterilization and biocompatibility testing provides evidence that there are no changes to the biocompatibility of the device and no changes to the efficacy and compatibility of the subject device with sterilization processes. #### Testing Completed: | Sterilization Efficacy Verification: | All requirements met, Test Passed | |----------------------------------------|-----------------------------------| | Reprocessing Sterilization Validation: | All requirements met, Test Passed | | Biocompatibility (Cytotox)Validation: | All requirements met, Test Passed | {5}------------------------------------------------ # Summary: Based upon the Intended Use, Indications for Use, product technical information, performance, sterilization testing, and biocompatibility validation provided in this premarket notification, the CelioFlex™ UHD 5 Confocal Miniprobe, when used as part of the Cellvizio 100 Series, has been shown to be substantially equivalent to the Circon's USA Series™ Laparoscopes, and is basically identical in technological characteristics to the CystoFlexTM UHD R Confocal Miniprobe reference device. The subject device can safely and effectively be used to visualize and image the internal microstructure of tissues during endoscopic and laparoscopic surgical procedures.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...