CELLVIZIO 100 SERIES WITH SYSTEM CONFOCAL MINIPROBES

K141358 · Mauna Kea Technologies · OWN · Sep 25, 2014 · Gastroenterology, Urology

Device Facts

Record IDK141358
Device NameCELLVIZIO 100 SERIES WITH SYSTEM CONFOCAL MINIPROBES
ApplicantMauna Kea Technologies
Product CodeOWN · Gastroenterology, Urology
Decision DateSep 25, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. The CystoFlex UHD R Confocal Miniprobe can be used within anatomical tracts. i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories.

Device Story

Cellvizio 100 Series System is a confocal laser imaging system using fiber optic probes (Miniprobes) to visualize internal tissue microstructure. The CystoFlex UHD R Miniprobe is an accessory designed for insertion into rigid cystoscopes with working channels ≥ 2.8mm. Operated by physicians during endoscopic procedures, the device provides real-time optical imaging of the urinary tract. The system transforms laser light reflected from tissue into visual output, allowing clinicians to examine tissue microstructure in situ. This aids in clinical decision-making during urological examinations. The device benefits patients by enabling minimally invasive, real-time visualization of tissue morphology.

Clinical Evidence

Bench testing only. Verification and validation testing confirmed the device meets design specifications and user needs. Biocompatibility testing confirmed no changes to the device's biocompatibility profile.

Technological Characteristics

Confocal laser imaging system using fiber optic probes. Features a UHD-type distal optical head. Compatible with rigid cystoscopes with working channels ≥ 2.8 mm. System utilizes laser-based optical scanning for tissue microstructure visualization.

Indications for Use

Indicated for imaging internal microstructure of tissues within urinary anatomical tracts (urethra, bladder, ureter) via endoscope or endoscopic accessories.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized human figure with three faces in profile, representing the department's focus on health and human well-being. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 25, 2014 Mauna Kea Technologies % Mr. Michael A. Daniel Daniel and Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460 Re: K141358 Trade/Device Name: Cellvizio 100 Series System with Confocal Miniprobes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN Dated: August 27, 2014 Received: August 29, 2014 Dear Mr. Daniel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K141358 Device Name Cellvizio 100 Series System with Confocal Miniprobes Indications for Use (Describe) The Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. The CystoFlex UHD R Confocal Miniprobe can be used within anatomical tracts. i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Section 7. Premarket Notification 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | 510(k) Number: | K141358 | |----------------|---------| |----------------|---------| Applicant Information: | Date Prepared: | May 6, 2014 | |----------------|------------------------------------------| | Name: | Mauna Kea Technologies | | Address: | 9 rue d'Enghien<br>F-75010 Paris, France | | Phone: | +33 1 48 24 03 45 | | Fax: | +33 1 48 24 12 18 | Contact Person: Michael A Daniel, Consultant | Phone Number: | (415) 407-0223 | |-------------------|----------------| | Office: | (775) 392-2970 | | Facsimile Number: | (775) 392-2972 | Device Information: | Device Trade Name: | Cellvizio® 100 Series System with Confocal<br>Miniprobes™ | |---------------------------|-----------------------------------------------------------| | Common Name: | Endoscope and Accessories | | Classification Name(s): | Confocal Optical Imaging | | Product Code/ Regulation: | OWN / 21 CFR 876.1500 | ### Predicate Device: Classification: - Regarding the Cellvizio Confocal Imaging System, the Cellvizio 100 . Series System has been cleared in K111047. Class II - . Regarding the compatibility of the access during urological procedures in rigid cystoscopes and the visualization and examination of the urinary tract, the CystoFlex UHD R Confocal Miniprobe is substantially equivalent to the previously cleared UroFlex B Confocal Miniprobe (K132389). {4}------------------------------------------------ ### Device Description: The CystoFlex UHD R Confocal Miniprobe, when used as part of the Cellvizio 100 Series system and the Cellvizio system, is intended to allow confocal laser imaging of the internal microstructure of tissues in anatomical tracts, i.e., urinary, including, but not limited to, urethra and bladder, accessed through an endoscope or endoscopic accessories. The CystoFlex UHD R Confocal Miniprobe is similar in design to the UroFlex B Confocal Miniprobe, except for its distal optical head, which is a UHD-type distal head. The CystoFlex UHD R Confocal Miniprobe can be inserted into endoscope lumens with inner diameter equal to or higher than 2.8mm, allowing them to work in currently used rigid cystoscopes. ### Indications for Use: The Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues. The CystoFlex™ UHD R Confocal Miniprobe can be used within anatomical tracts, i.e., urinary, including, but not limited to, urethra, bladder, and ureter, accessed through an endoscope or endoscopic accessories. ### Comparison to Predicate Device: No change is being made to the fundamental technology and operating principle of the previously cleared UroFlex™ B Confocal Miniprobe used in this submission as a predicate device (K132389). The CystoFlex UHD R is identical to this predicate device, only differing by the design of its distal optical head ("UHD-type" for the CystoFlex™ UHD R instead of "Z-type" for the UroFlex™ B). The indications for use is unchanged and remains to image the urinary tract during endoscopic procedures. Due to this UHD-type distal optical head, the CystoFlex™ UHD R is designed to fit into rigid cystoscopes with working channel ≥ 2.8 mm. Verification and validation testing provides proof that the modifications meet the design specifications and user needs, and biocompatibility testing provides evidence that there are no changes to the biocompatibility of the device. ### Summary: Based upon the intended use, indications for use, product technical information, performance testing and biocompatibility information provided in this premarket notification, the CystoFlex™ UHD R Confocal Miniprobe, when used as part of the Cellvizio 100 Series and the Cellvizio Systems, has been shown to be substantially equivalent to the UroFlex™ B Confocal Miniprobe, and can be used as intended to image the internal microstructure {5}------------------------------------------------ #### Mauna Kea Technologies of tissues in the urinary tract during endoscopic procedures. This Confocal Miniprobe has been verified to be compatible with cystoscopes that are designed to be used in these applications.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...