Smith and Nephew Suture Anchors - 5.0mm Absorbable Polymer Anchor, Smith and Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317, 7209318, Smith and Nephew Suture Anchors - Bioraptor 2.3 PK Suture Anchor, Smith and Nephew Suture Anchors - Osteoraptor Suture Anchor, Smith and Nephew Suture Anchors - Twinfix Ultra HA Suture Anchor, Smith and Nephew Suture Anchors - OSTEORAPTOR Curved Suture Anchors, Smith and Nephew Suture Anchors - Ultra Fast-Fix Meniscal Repair System, Ultra Fa

K151105 · Smith & Nephew, Inc. · MAI · Jul 28, 2015 · Orthopedic

Device Facts

Record IDK151105
Device NameSmith and Nephew Suture Anchors - 5.0mm Absorbable Polymer Anchor, Smith and Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317, 7209318, Smith and Nephew Suture Anchors - Bioraptor 2.3 PK Suture Anchor, Smith and Nephew Suture Anchors - Osteoraptor Suture Anchor, Smith and Nephew Suture Anchors - Twinfix Ultra HA Suture Anchor, Smith and Nephew Suture Anchors - OSTEORAPTOR Curved Suture Anchors, Smith and Nephew Suture Anchors - Ultra Fast-Fix Meniscal Repair System, Ultra Fa
ApplicantSmith & Nephew, Inc.
Product CodeMAI · Orthopedic
Decision DateJul 28, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Smith & Nephew 5.0 Absorbable Polymer Anchor is intended for use only for the reattachment of soft tissue to bone for the following indications: Shoulder: 1. Bankart lesion repairs 2. SLAP lesion repairs 3. Acromioclavicular separation repairs 4. Rotator cuff tear repairs 5. Capsular shift or capsulolabral reconstructions 6. Biceps tenodesis 7. Deltoid repairs Foot and Ankle: 1. Hallux valgus repairs 2. Medial or lateral instability repairs/reconstructions 3. Achilles tendon repairs/reconstructions 4. Midfoot reconstructions 5. Metatarsal ligament/tendon repairs/reconstructions Elbow, Wrist, and Hand: 1. Scapholunate ligament reconstructions 2. Ulnar or radial collateral ligament reconstructions 3. Lateral epicondylitis repair 4. Biceps tendon reattachment Knee: 1. Extra-capsular repairs: a. Medial collateral ligament b. Lateral collateral ligament c. Posterior oblique ligament 2. Iliotibial band tenodesis 3. Patellar realignment and tendon repairs, including vastus medialis obliquous advancement

Device Story

Suture anchors for reattachment of soft tissue to bone; used in orthopedic procedures (shoulder, knee, hip, foot/ankle, elbow/wrist/hand). Devices consist of absorbable polymer anchors pre-assembled on inserters; some configurations include non-absorbable sutures, needles, guides, or dilators. Operated by surgeons in OR settings (percutaneous or endoscopic). Anchors provide mechanical fixation of soft tissue to bone; facilitate meniscal repair and allograft transplant. Benefit: secure tissue fixation during healing.

Clinical Evidence

Bench testing only. Performance data includes ship testing and post-shipment mechanical functional testing for insertion and pullout (fixation) force, demonstrating equivalence to predicate devices.

Technological Characteristics

Composite absorbable materials; pre-assembled on inserters; sterile, single-use. Includes configurations with non-absorbable sutures, stainless steel needles, guides, and threaded dilators. Mechanical fixation principle.

Indications for Use

Indicated for reattachment of soft tissue to bone in shoulder, elbow, wrist, hand, hip, knee, foot, and ankle procedures, including ligament/tendon repairs, reconstructions, and meniscal repair/allograft transplant.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of a human face in profile, with three overlapping faces suggesting a sense of community and connection. The logo is black and white. July 28, 2015 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Smith & Nephew, Incorporated Ms. Katherine Marcaccio Regulatory Affairs Specialist II 150 Minuteman Road Andover, Massachusetts 01810 Re: K151105 Trade/Device Name: Smith and Nephew Suture Anchors - 5.0mm Absorbable Polymer Anchor; Smith and Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317; 7209318, Smith and Nephew Suture Anchors -Bioraptor 2.3 PK Suture Anchor; Smith and Nephew Suture Anchors - Osteoraptor Suture Anchor; Smith and Nephew Suture Anchors -Twinfix Ultra HA Suture Anchor; Smith and Nephew Suture Anchors - OSTEORAPTOR Curved Suture Anchors; Smith and Nephew Suture Anchors - Ultra Fast-Fix Meniscal Repair System; Ultra Fast-Fix AB Meniscal Repair System; Fast-Fix 360 Meniscal Repair System; Smith and Nephew Suture Anchors - Healicoil Absorbable Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, HWC, JDR, MBI, GAT Dear Ms. Marcaccio: Dated: April 22, 2015 Received: April 24, 2015 We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1}------------------------------------------------ #### Page 2 - Ms. Katherine Marcaccio If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ 510(k) Number (if known): ________________ Smith & Nephew Suture Anchors – 5.0mm Absorbable Polymer Anchor Device Name: Indications for Use: The Smith & Nephew 5.0 Absorbable Polymer Anchor is intended for use only for the reattachment of soft tissue to bone for the following indications: | Shoulder: | Elbow, Wrist, and Hand: | |----------------------------------------------------------|---------------------------------------------------------------------------------------------| | 1. Bankart lesion repairs | 1. Scapholunate ligament reconstructions | | 2. SLAP lesion repairs | 2. Ulnar or radial collateral ligament reconstructions | | 3. Acromioclavicular separation repairs | 3. Lateral epicondylitis repair | | 4. Rotator cuff tear repairs | 4. Biceps tendon reattachment | | 5. Capsular shift or capsulolabral reconstructions | | | 6. Biceps tenodesis | | | 7. Deltoid repairs | | | Foot and Ankle: | Knee: | | 1. Hallux valgus repairs | 1. Extra-capsular repairs: | | 2. Medial or lateral instability repairs/reconstructions | a. Medial collateral ligament | | 3. Achilles tendon repairs/reconstructions | b. Lateral collateral ligament | | 4. Midfoot reconstructions | c. Posterior oblique ligament | | 5. Metatarsal ligament/tendon repairs/reconstructions | 2. Iliotibial band tenodesis | | | 3. Patellar realignment and tendon repairs, including vastus medialis obliquous advancement | Prescription Use ___X____ AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use _ (Part 21 CFR 801 subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {3}------------------------------------------------ 510(k) Number (if known): ___________ Device Name: Smith & Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317, 7209318 Indications for Use: - 1. Bankart Lesion Repair - 2. Rotator Cuff Repair - 3. Capsular Stabilization - 4. Anterior Shoulder Instability Repair - 5. Repair of ligaments and tendons of the elbow, foot, and ankle including the treatment of: - Bunionectomy - Lateral ankle instability । - Biceps tendon reattachment - Prescription Use X AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use _ (Part 21 CFR 801 subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {4}------------------------------------------------ Smith & Nephew Suture Anchors - Bioraptor 2.3 PK Suture Anchor Device Name: Indications for Use: These suture anchors are intended for the fixation of soft tissue to bone in the Hip, Shoulder, Foot, Ankle, Elbow, Wrist, Hand and Knee as follows: #### Hip: Hip capsule repair । Acetabular labrum reattachment #### Shoulder: Capsular stabilization - Bankart repair - - -Anterior shoulder instability - SLAP lesion repairs - - Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis # Elbow, Wrist, and Hand: Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair #### Knee: Extra-capsular repairs: - Medial collateral ligament - - -Lateral collateral ligament - -Posterior oblique ligament Patellar realignment and tendon repairs - Vastus medialis obliquous advancement Iliotibial band tenodesis #### Foot and Ankle: Hallux valgus repairs Medial or lateral instability repairs/ reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use (Part 21 CFR 801 subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {5}------------------------------------------------ Smith & Nephew Suture Anchors – Osteoraptor Suture Anchor Device Name: Indications for Use: The Smith & Nephew OSTEORAPTOR Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: #### Elbow, Wrist, and Hand Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair ### Foot and Ankle Hallux valgus repairs Medial or lateral instability repairs/ reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy #### Hip Hip capsule repair - Acetabular labrum reattachment ### Knee Extra-capsular repairs: - Medial collateral ligament - - -Lateral collateral ligament - Posterior oblique ligament - Patellar realignment and tendon repairs - Vastus medialis obliquous advancement Iliotibial band tenodesis ### Shoulder Capsular stabilization - -Bankart repair - Anterior shoulder instability - - SLAP lesion repairs - - -Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use _ (Part 21 CFR 801 subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {6}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Smith & Nephew Suture Anchors – Twinfix Ultra HA Suture Anchor Device Name: Indications for Use: The Smith & Nephew TWINFIX Ultra HA Suture Anchor is intended for use for the reattachment of soft tissue to bone for the following indications: #### Shoulder: - Bankart repairs SLAP lesion repairs Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis #### Foot and Ankle: Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions ### Knee: Extra-capsular repairs: - । Medial collateral ligament - Lateral collateral ligament - - Posterior oblique ligament - Patellar realignment and tendon repairs: Vastus medialis obliquous advancement । Iliotibial band tenodesis #### Elbow: Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Biceps tendon reattachment AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 subpart D) Over-the-Counter Use (Part 21 CFR 801 subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {7}------------------------------------------------ 510(k) Number (if known): ________________ Device Name: Smith & Nephew Suture Anchors – OSTEORAPTOR Curved Suture Anchors Indications for Use: The Smith & Nephew OSTEORAPTOR Curved Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: #### Shoulder: Capsular stabilization - -Bankart repair - Anterior shoulder instability - - -SLAP lesion repairs - Capsular shift or capsulolabral . reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use _ (Part 21 CFR 801 subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {8}------------------------------------------------ Device Name: Smith & Nephew Suture Anchors – Ultra Fast-Fix Meniscal Repair System, Ultra Fast-Fix AB Meniscal Repair System, Fast-Fix 360 Meniscal Repair System Indications for Use: The ULTRA FAST-FIX Meniscal Repair System is intended for use as suture retention devices to facilitate percutaneous or endoscopic soft tissue procedures. The ULTRA FAST-FIX System is indicated for use in meniscal repairs and allograft transplant procedures. The ULTRA FAST-FIX System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. The ULTRA FAST-FIX AB Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The ULTRA FAST-FIX AB System is indicated for use in meniscal repairs and allograft transplant procedures. The ULTRA FAST-FIX AB System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. The FAST-FIX 360 Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The FAST-FIX 360 System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. Prescription Use X AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use (Part 21 CFR 801 subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {9}------------------------------------------------ Smith & Nephew Suture Anchors – Healicoil Absorbable Suture Anchor Device Name: Indications for Use: Bankart lesion repairs Rotator cuff tear repairs Slap lesion repairs The Smith & Nephew HEALICOIL Absorbable Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: #### Shoulder: Knee: Extra-capsular repairs: - Medial collateral ligament । - -Lateral collateral ligament - Posterior oblique ligament - Patellar realignment and tendon repairs: Vastus medialis obliquous advancement -Iliotibial band tenodesis ### Foot and Ankle: Biceps tenodesis Deltoid repairs Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstruction Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs ### Elbow: Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Biceps tendon reattachment Prescription Use AND/OR (Part 21 CFR 801 subpart D) Over-the-Counter Use (Part 21 CFR 801 subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) {10}------------------------------------------------ # 510(k) Summary # Date Prepared: May 15, 2015 | Submitter Information | Contact Information | |-----------------------|----------------------------------| | Smith & Nephew, Inc. | Katherine Marcaccio | | 150 Minuteman Road | Regulatory Affairs Specialist II | | Andover, MA 01810 | Phone: (508) 261-3602 | | | Fax: (978) 749-1443 | | Device Name (Unmodified) | | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade or proprietary name | Smith & Nephew Absorbable Suture Anchors | | Common or usual name | HWC - Screw, Fixation, Bone<br>JDR - Staple, Fixation, Bone<br>MAI - Fastener, Fixation, Biodegradable, Soft Tissue<br>MBI - Fastener, Fixation, Nondegradable, Soft Tissue<br>GAT - Suture, Nonabsorbable, Synthetic, Polyethylene | | Classification name | 21 CFR §888.3030 – Single/multiple component metallic bone<br>fixation appliances and accessories<br>21 CFR §888.3040 – Smooth or threaded metallic bone<br>fixation fastener<br>21 CFR §878.5000 - Nonabsorbable poly(ethylene<br>terephthalate) surgical suture | # Legally Marketed Predicate Device The Smith & Nephew Absorbable Suture Anchors are substantially equivalent in intended use and Fundamental Scientific Technology to the following legally marketed devices in commercial distribution: | 510(k)<br>Number | Device Name | Clearance<br>Date | |------------------|----------------------------------------------------------------------------------------------------|-------------------| | K011299 | SMITH AND NEPHEW 5.00MM ABSORBABLE POLYMER<br>ANCHOR | 06/27/2001 | | K031685 | TAG ROD BIORAPTOR SUTURE ANCHOR 7209317, 7209318 | 07/01/2003 | | K071586 | Bioraptor 2.3 PK Suture Anchor | 08/17/2007 | | K082215 | Osteoraptor Suture Anchor | 11/03/2008 | | K093844 | SMITH & NEPHEW TWINFIX ULTRA HA SUTURE ANCHOR | 04/01/2010 | | K103309 | OSTEORAPTOR Curved Suture Anchors | 02/03/2011 | | K121861 | SMITH & NEPHEW ULTRA FAST-FIX MENISCAL REPAIR<br>SYSTEM, SMITH & NEPHEW ULTRA FAST-FIX AB MENISCAL | 10/18/2012 | {11}------------------------------------------------ | | REPAIR SYSTEM, SMITH & NEPHEW FAST-FIX 360 MENISCAL<br>REPAIR SYSTEM | | |---------|----------------------------------------------------------------------|------------| | K123393 | HEALICOIL SUTURE ANCHOR WITH REGENESORB | 04/11/2013 | #### Device Description Smith & Nephew Absorbable Suture Anchors are provided in various composite absorbable materials. All of the Smith & Nephew Absorbable Suture Anchors are provided sterile, for single use only. All Smith & Nephew Absorbable Suture Anchors are pre-assembled onto an inserter, sized appropriately to accommodate the indicated procedures. Smith & Nephew Absorbable Suture Anchors come in various configurations, including: with attached non-absorbable suture(s), and with attached stainless steel needle(s). In certain configurations, the Absorbable Suture Anchors are packaged with a guide and threaded dilator. All configurations of the Absorbable Suture Anchors are identical to the identified predicate devices. #### Intended Use #### Smith & Nephew Suture Anchors – 5.0mm Absorbable Polymer Anchor The Smith & Nephew 5.0 Absorbable Polymer Anchor is intended for use only for the reattachment of soft tissue to bone for the following indications: #### Shoulder: - 1. Bankart lesion repairs - SLAP lesion repairs 2. - Acromioclavicular separation repairs ဒေ - Rotator cuff tear repairs 4. - 5. Capsular shift or capsulolabral reconstructions - 6. Biceps tenodesis - 7. Deltoid repairs #### Foot and Ankle: - 1. Hallux valgus repairs - Medial or lateral instability 2. repairs/reconstructions - 3. Achilles tendon repairs/reconstructions - 4. Midfoot reconstructions - 5. Metatarsal ligament/tendon repairs/reconstructions #### Elbow, Wrist, and Hand: - 1. Scapholunate ligament reconstructions - 2. Ulnar or radial collateral ligament reconstructions - ဒေ Lateral epicondylitis repair - 4. Biceps tendon reattachment - Knee: - 1. Extra-capsular repairs: - a. Medial collateral ligament - b. Lateral collateral ligament - Posterior oblique ligament ﻥ - 2. Illiotibial band tenodesis - 3. Patellar realignment and tendon repairs, including vastus medialis obliquous advancement {12}------------------------------------------------ #### Smith & Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317, 7209318 - Bankart Lesion Repair 1. - 2. Rotator Cuff Repair - Capsular Stabilization 3. - Anterior Shoulder Instability Repair 4. - 5. Repair of ligaments and tendons of the elbow, foot, and ankle including the treatment of: - Bunionectomy - - Lateral ankle instability - - Biceps tendon reattachment - # Smith & Nephew Suture Anchors - Bioraptor 2.3 PK Suture Anchor These suture anchors are intended for the fixation of soft tissue to bone in the Hip, Shoulder, Foot, Ankle, Elbow, Wrist, Hand and Knee as follows: # Hip: Hip capsule repair - -Acetabular labrum reattachment # Shoulder: Capsular stabilization - -Bankart repair - Anterior shoulder instability - - -SLAP lesion repairs - Capsular shift or capsulolabral reconstructions - Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis # Foot and Ankle: - Hallux valgus repairs Medial or lateral instability repairs/ reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy # Elbow, Wrist, and Hand: Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair # Knee: Extra-capsular repairs: - Medial collateral ligament - - Lateral collateral ligament - - -Posterior oblique ligament - Patellar realignment and tendon repairs - -Vastus medialis obliquous - advancement - Iliotibial band tenodesis {13}------------------------------------------------ #### Smith & Nephew Suture Anchors - Osteoraptor Suture Anchor The Smith & Nephew OSTEORAPTOR Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: ### Elbow, Wrist, and Hand Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair # Foot and Ankle Hallux valgus repairs Medial or lateral instability repairs/ reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy # Hip Hip capsule repair Acetabular labrum reattachment - ## Knee Extra-capsular repairs: - Medial collateral ligament - -Lateral collateral ligament - -Posterior oblique ligament Patellar realignment and tendon repairs - Vastus medialis obliquous - advancement Iliotibial band tenodesis # Shoulder Capsular stabilization - -Bankart repair - Anterior shoulder instability - - SLAP lesion repairs - - -Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis {14}------------------------------------------------ #### Smith & Nephew Suture Anchors - Twinfix Ultra HA Suture Anchor The Smith & Nephew TWINFIX Ultra HA Suture Anchor is intended for use for the reattachment of soft tissue to bone for the following indications: #### Shoulder: Bankart repairs SLAP lesion repairs Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis # Foot and Ankle: Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions #### Knee: Extra-capsular repairs: - Medial collateral ligament - - Lateral collateral ligament - - Posterior oblique ligament - Patellar realignment and tendon repairs: -Vastus medialis obliquous advancement Iliotibial band tenodesis ### Elbow: Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Biceps tendon reattachment Smith & Nephew Suture Anchors – OSTEORAPTOR Curved Suture Anchors The Smith & Nephew OSTEORAPTOR Curved Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: #### Shoulder: Capsular stabilization - -Bankart repair - Anterior shoulder instability - - SLAP lesion repairs - - Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis {15}------------------------------------------------ #### Smith & Nephew Suture Anchors – Ultra Fast-Fix Meniscal Repair System, Ultra Fast-Fix AB Meniscal Repair System, Fast-Fix 360 Meniscal Repair System The ULTRA FAST-FIX Meniscal Repair System is intended for use as suture retention devices to facilitate percutaneous or endoscopic soft tissue procedures. The ULTRA FAST-FIX System is indicated for use in meniscal repairs and allograft transplant procedures. The ULTRA FAST-FIX System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. The ULTRA FAST-FIX AB Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The ULTRA FAST-FIX AB System is indicated for use in meniscal repairs and allograft transplant procedures. The ULTRA FAST-FIX AB System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. The FAST-FIX 360 Meniscal Repair System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The FAST-FIX 360 System is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures. Smith & Nephew Suture Anchors - Healicoil Absorbable Suture Anchor The Smith & Nephew HEALICOIL Absorbable Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: # Shoulder: Bankart lesion repairs Slap lesion repairs Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis # Foot and Ankle: Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstruction Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions # Knee: Extra-capsular repairs: - Medial collateral ligament - - Lateral collateral ligament - - Posterior oblique ligament । Patellar realignment and tendon repairs: Vastus medialis obliquous advancement -Iliotibial band tenodesis # Elbow: Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Biceps tendon reattachment {16}------------------------------------------------ #### Technological Characteristics Smith & Nephew Absorbable Suture Anchors are substantially equivalent in intended use and fundamental scientific technology to the legally marketed predicate devices in this submission and raises no new issues of safety and efficacy. #### Summary of Performance Data Ship Testing and Post-Shipment mechanical functionaltesting for insertion and pullout (fixation) force demonstrates that the Absorbable Suture Anchors are substantially equivalent to the currently marketed predicate devices. #### Substantial Equivalence Information The substantial equivalence of the Absorbable Suture Anchors is based on identical indications for use, design features, operational principles, material composition, and performance to the predicate devices listed above. Based on the identical features to the predicates, the Absorbable Suture Anchors are substantially equivalent to the predicates.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...