K151104 · Spineart · GAD · Jan 6, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151104
Device Name
OTELO LL
Applicant
Spineart
Product Code
GAD · General, Plastic Surgery
Decision Date
Jan 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The OTELO LL Retractor is intended to provide the surgeon access to the spine by dissection and traction of soft and bony tissue
Device Story
OTELO LL is a tubular-based manual surgical retractor system used in spinal surgery. It consists of a retractor base, dilators, blades (right, left, posterior), shim pins, light source, and flex arm attachment. Surgeon operates device to dilate overlying tissues and maintain access to the spine. System allows distraction or articulation of blades to retract tissue; includes self-locking mechanism. Used in conjunction with microscopes, cameras, or light sources. Output is physical retraction of tissue, providing surgeon with visualization of the surgical field and access to the disc/spine. Benefits include improved surgical access and visualization for minimally invasive spinal procedures.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993: cytotoxicity, intracutaneous, acute systemic toxicity, sensitization), cleaning validation (ISO 17664, AAMI TIR 30), steam sterilization validation, and mechanical performance verification (opening ranges, blade integrity, locking mechanism functionality).
Indicated for surgeons requiring access to the spine via dissection and traction of soft and bony tissue during spinal procedures.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
K071771 — INSTRUMED RETRACTORS · Instrumed International, Inc. · Aug 2, 2007
K043602 — QUADRANT RETRACTOR SYSTEM · Medtronic Sofamor Danek · Feb 23, 2005
K161680 — MEDICON Spinal Spreading Systems · Medicon EG · Mar 24, 2017
K112913 — SPOTLIGHT ACCESS SYSTEM · Johnson & Johnson · Dec 16, 2011
K992898 — THE BRIGHT MEDICAL DILATION RETRACTOR SYSTEM · Bright Medical Instruments · Oct 21, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2016
Spineart Mr. Franck Pennesi International Center Cointrin 20 Route de Pre-bois - CP 1813 1215 GENEVA 15 Switzerland
Re: K151104
Trade/Device Name: OTELO LL Regulation Number: 21 CFR 878.4800 Regulation Name: Retractor Regulatory Class: Class I Product Code: GAD Dated: December 2, 2015 Received: December 4, 2015
Dear Mr. Pennesi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# William J. Heetderks -S
for Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151104
Device Name OTELO LL Retractor
#### Indications for Use (Describe)
The OTELO LL Retractor is intended to provide the surgeon access to the spine by dissection and traction of soft and bony tissue
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| 510k | TRADITIONAL |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission | New devices |
| Submitted by | SPINEART<br>International Center Cointrin<br>20 route de pré-bois<br>CP1813<br>1215 GENEVA 15<br>SWITZERLAND |
| Contacts | Franck PENNESI Director of Industry & Quality<br>Phone : +41 22 570 1246 Fax : +41 22 799 4026<br>Mail : fpennesi@spineart.com<br>Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting) idrubaix@nordnet.fr |
| Date Prepared | January 5, 2016 |
| Common Name | Retractor |
| Trade Name | OTELO® LL |
| Classification Name | Retractor |
| Regulation description | Manual surgical instrument for general use. |
| Class | I |
| Product Code | GAD |
| CFR section | 878.4800 |
| Device panel | General & Plastic Surgery |
| Legally marketed predicate devices | Primary predicate: Quadrant Retractor System (K043602) manufactured by<br>Medtronic Sofamor Danek |
| Design features | OTELO LL retractor by SPINEART | MAST QUADRANT Retractor system<br>by Medtronic |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | Present submission | K043602 |
| Product code | GAD | GAD |
| Pictures | Image: OTELO LL retractor | Image: MAST QUADRANT Retractor system |
| 21 CFR | 878.4800 | 878.4800 |
| Common name | Retractor, Manuel Surgical Instrument | Retractor, Manuel Surgical Instrument |
| Design features | OTELO LL retractor by SPINEART | MAST QUADRANT Retractor system<br>by Medtronic |
| Indications for use | The OTELO®LL Retractor is intended to provide<br>the surgeon access to the spine by dissection<br>and traction of soft and bony tissue. | The MAST QUADRANT™ Retractor System is<br>intended to provide surgeons with<br>instruments such as dilators, retractors,<br>light sources and pedicle access needles<br>used to perform a variety of spinal fixation<br>procedures utilizing a minimally invasive<br>approach. The MAST QUADRANT™<br>Retractor System is indicated for<br>visualization of the surgical field in any area<br>of the body cut open during a surgical<br>procedure. When used in the cervical,<br>thoracic, or lumbar spine either from an<br>anterior or posterior direction, for example,<br>the MAST QUADRANT™ Retractors and<br>accessories are intended to aid the<br>surgeon's visualization of the surgical area<br>and allow him/her to perform any type of<br>surgical spinal procedure such as herniated<br>disc repair, visualization of the<br>circumferential decompression of the nerve<br>roots, aiding in the search and removal of<br>nucleus material, spinal fusion, or insertion<br>of spinal implants. |
| Description | The OTELO®LL Retractor System is a tubular-<br>based retraction system, designed to provide<br>surgeons with the freedom to retract tissue<br>through any combination of distracting or<br>articulating the blades. The OTELO®LL<br>Retractor System includes instruments used to<br>access the spine by dilating the overlying<br>tissues, as well serving as a retracting device to<br>maintain the access. The system can be used in<br>conjunction with microscopes, light sources,<br>cameras, or other visualization aids. | The MAST QUADRANT™ Retractor System is<br>a tubular-based retraction system, designed<br>to provide surgeons with the freedom to<br>retract tissue through any combination of<br>distracting or articulating the blades. The<br>MAST QUADRANT™ System includes<br>instruments used to access the spine by<br>dilating the overlying tissues, as well serving<br>as a retracting device to maintain the<br>access. The system can be used in<br>conjunction with microscopes, light sources,<br>cameras, or other visualization aids. |
| Material | Stainless Steel, PEEK and PEEK Carbon. | Stainless Steel |
| Design features | Self-locking mechanism and minimal invasive<br>use | Self-locking mechanism and minimal<br>invasive use |
| Operating<br>principle | Gives surgeons an easy access and good<br>visualization of the disc to be treated. | Gives surgeons an easy access and good<br>visualization of the disc to be treated. |
| Technological<br>characteristics | The retractor system is composed of one<br>retractor base associated to different<br>instruments such as dilators tube, blades,<br>wires, retractor, shim pins, light source and<br>flex arm attachment. Range of blades is<br>composed of 10 right, 10 left and 10 posterior<br>blades which varies from 90 to 180 mm length.<br>If needed, a fourth blade distraction system<br>can be adapted in order to increase surgeon's<br>access visibility. This complementary system is<br>composed of 2 wide and 2 narrow<br>supplemental blades which varies from 5 to 10<br>mm width. | The retractor system is composed of one<br>retractor base associated to different<br>instruments such as dilators tube, blades,<br>wires, retractor, light source and flex arm<br>attachment. Range of blades is composed of<br>5 right and 5 left blades which varies from<br>40 to 80 mm length. If needed, a<br>complementary medial /lateral distraction<br>system can be adapted in order to increase<br>surgeon's access visibility. This<br>complementary system is composed of 3<br>wide and 3 narrow supplemental blades<br>which varies from 50 to 90 mm length |
| Design features | OTELO LL retractor by SPINEART | MAST QUADRANT Retractor system<br>by Medtronic |
| Ranges of the<br>retractor opening<br>distance | Antero-posterior opening range: 21 mm to 46<br>mm<br>Cranial-caudal opening range: 21 mm to 100<br>mm | Cranial-caudal opening range: 30 mm to 52<br>mm |
| Ranges of blade<br>angle | Cranial-Caudal inclination angle range:<br>0° to +15° | / |
| Packaging | Instruments are individually packaged in<br>polyethylene sealed pouch with label and IFU<br>included.<br>When packaged as an Instrument set, the set<br>is packaged in polyethylene sealed pouch with<br>label and IFU included. | Same for non sterile instruments. |
| Shelf life | Not applicable. Devices are provided non<br>sterile. | / |
| Sterilization | Steam sterilization | Steam sterilization |
| Biocompatibility | Meets ISO 10993 | Meets ISO 10993 |
{4}------------------------------------------------
{5}------------------------------------------------
# Pre-clinical verification and validation tests results:
| Test | Results | Conclusions |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| ISO MTS Cytotoxicity.<br>Report n° 199740 | Percentage viability of test article: 86.4% | No cytotoxic potential |
| ISO Intracutaneous Study in Rabbits -<br>Two Extracts.<br>Report n° 199741 | The 0.9% NaCl extract of the test article did not<br>induce any erythema or edema reactions after<br>injection by intracutaneous route in<br>the rabbit.<br>The sesame oil extract of the test article did not<br>induce more erythema and edema reactions than<br>the vehicle alone after injection by<br>intracutaneous route in the rabbit. | No erythema or edema<br>reactions |
| ISO Acute Systemic Toxicity Study in<br>Mice - Two Extracts.<br>Report n° 199743 | No body weight loss greater than 10% occurred,<br>no abnormal behaviour was noted and body<br>weight gains were thereafter observed until the<br>end of the study.<br>All animals appeared clinically normal at the<br>beginning and throughout the study. | No evidence of significant<br>systemic toxicity or<br>mortality |
| ISO Guinea Pig Maximization<br>Sensitization Test - Two Extracts.<br>Report n° 199742 | 0.9% NaCl extract: sensitization grade 0<br>Sesame oil extract: sensitization grade 0<br>All animals appeared clinically normal throughout<br>the study. | The test article was not<br>considered a sensitizer in<br>the guinea pig<br>maximization model |
| Validation of the Manual Cleaning<br>Process According to the ISO 17664<br>standard and the AAMI TIR 30 Technical<br>Report.<br>Report n° 199900 | No soil was observed.<br>No protein (<6.4 µg/cm²) was detected.<br>No hemoglobin (<2.2 µg/cm²) was detected. | The manual cleaning<br>process is validated. |
| Validation of the Automated Cleaning<br>Process According to the ISO 17664<br>standard and the AAMI TIR 30 Technical<br>Report.<br>Report n° 200220 | No soil was observed.<br>No protein (<6.4 µg/cm²) was detected.<br>No hemoglobin (<2.2 µg/cm²) was detected. | The automated cleaning<br>process is validated. |
| Steam sterilization Study for the<br>SPINEART OTELO Retractor Set.<br>Report n° 154535-002 rev 1 | Sterilization Parameters ( 132° 2') provide a SAL<br>not less than 10⁻⁶.<br>No moisture was observed after dry time of 30'. | Steam sterilization<br>parameters 132°C, 4'and<br>dry time 30' are validated |
{6}------------------------------------------------
| Mechanical performances - Verification<br>Test report n°VRR67-01-00 | Minimal opening of 100 mm is obtained.<br>No breakage of the blades.<br>No disassembly of the blades and the retractor.<br>No disassembly of the arms of the retractor.<br>No breakage or damage of the teeth of the rack.<br>No breakage or damage of the thread feature of<br>the knobs.<br>Minimal opening of 46 mm is obtained.<br>No breakage or damage of the thread feature of<br>the knobs.<br>No rotation or malfunction of the 4th blade<br>assembly.<br>No disassembly of the shim pin from the shim<br>tool.<br>No disassembly of the shim pin from the blade. | Mechanical characteristics<br>are conform to the needs. |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Functionality performances - Validation<br>Test report n°VLR67-08-00 | The light system worked well.<br>Radiolucency is a clear advantage of the system.<br>The arm, the base arm and the attachment<br>worked well. It has been commented that the<br>arm attachment should be connected to the arm<br>before attaching to the retractor body.<br>The multi-function double hex driver was well<br>appreciated for the posterior blade manipulation.<br>The ratcheting mechanism and the toe-ing of the<br>blades worked well.<br>The 4th blade attachment is well designed.<br>Posterior Shim and Shim pin worked correctly | Functionality is conform to<br>the needs. |
## Discussion:
Differences of technical characteristics between subject device and predicate device:
The main differences between the two systems are limited to:
- A the material used for blades (CF/PEEK for OTELO versus stainless steel for predicate device),
- A the length of the range of blades (90 to 150 mm for OTELO versus 40 to 80 mm for predicate device),
- A the complementary distraction system that can be used to increase surgeon 's visibility access,
- > the cranial-caudal opening range (21 mm to 100 mm for OTELO versus 30 to 52 for predicate device).
## Safety and effectiveness:
Material: Based on pre-clinical biocompatibility verification tests (reports n°199740, 199742 and 199743), it has been demonstrated that CF/PEEK material is conforming to ISO 10993-1 for contact inferior to 24 hours.
Length of blade: longer blades (90 to 180 mm) have been designed in order to provide a lateral access to the spine. Based on pre-clinical mechanical verification tests (report n°VRR67-01-00) and pre-clinical validation test (report n° VLR67-08-00), it has been demonstrated that mechanical characteristics for OTELO is conforming to the needs and indications for use.
Complementary distraction system: Based on pre-clinical mechanical verification tests (report n°VRR67-01-00) and pre-clinical validation test (report n° VLR67-08-00), it has been demonstrated
{7}------------------------------------------------
that mechanical characteristics of the complementary system for OTELO is conforming to the needs and indications for use.
Cranial-caudal opening: Based on pre-clinical mechanical verification tests (report n°VRR67-01-00) and pre-clinical validation test (report n° VLR67-08-00), it has been demonstrated that mechanical characteristics for OTELO is conforming to the needs and indications for use.
### Conclusion:
The significant differences between the OTELO retractor system and the other mentioned marketed system which would adversely affect the safety and performances of the product have been tested in order to demonstrate OTELO capabilities.
The Spineart OTELO retractor system is substantially equivalent in design, function, mechanical performances and intended use, to the existing previously cleared spinal devices, MAST QUADRANT Retractor System by Medtronic Sofamor Danek (K043602).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.