MUSE System

K151001 · Medigus , Ltd. · ODE · Jun 5, 2015 · Gastroenterology, Urology

Device Facts

Record IDK151001
Device NameMUSE System
ApplicantMedigus , Ltd.
Product CodeODE · Gastroenterology, Urology
Decision DateJun 5, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medigus Ultrasonic Surgical Endostapler (MUSE™) System is intended for endoscopic placement of surgical staples in the soft tissue of the esophagus and stomach in order to create anterior partial fundoplication for treatment of symptomatic chronic Gastro Esophageal Reflux Disease in patients who require and respond to pharmacological therapy.

Device Story

MUSE system performs transoral endoscopic fundoplication to treat chronic GERD. Device consists of endoscopic stapler, control console, and accessories. Stapler features distal tip with video camera, ultrasonic range finder, illumination, irrigation/insufflation ports, and staple anvil. Operator retroflexes distal tip to align with staple cartridge in shaft; alignment pin ensures positioning. System fires five titanium staples in three staggered rows per application. Procedure performed via mouth without incisions. Control console manages insufflation, light, and camera electronics. Healthcare provider uses visual feedback from camera and ultrasonic range finding to position device for stapling fundus to esophagus in two or more locations. Creates anterior partial fundoplication to reduce reflux symptoms. Benefits include minimally invasive surgical alternative to traditional fundoplication.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including dimensional comparisons, mechanical testing, and biocompatibility testing.

Technological Characteristics

Endoscopic stapling system. Distal tip includes video camera, ultrasonic range finder, illumination, irrigation/insufflation ports, and staple anvil. Uses 4.8 mm titanium surgical staples. Single-use, EtO sterilized stapler. Control console provides insufflation, light, and camera electronics. Includes overtube, irrigation bottle, silicon tubing, and cartridge replacement tools.

Indications for Use

Indicated for patients with symptomatic chronic Gastro Esophageal Reflux Disease (GERD) who require and respond to pharmacological therapy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing right, layered on top of each other, with a flowing ribbon-like element extending from the bottom profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 5, 2015 Medigus, Ltd. % Sheila S. Stevens, Ph.D. US Clinical and Regulatory Affairs Consultant 2121 North California Blvd. Suite 290 Walnut Creek, CA 94596 Re: K151001 Trade/Device Name: Medigus Ultrasonic Surgical Endostapler (MUSE™) System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODE Dated: May 15, 2015 Received: May 18, 2015 Dear Sheila S. Stevens, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ Page 2 - Sheila S. Stevens, Ph.D. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, **Herbert P. Lerner -S** for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # I. Statement of Indications for Use 510(k) Number (if known): K151001 Device Name: Medigus Ultrasonic Surgical Endostapler (MUSE™) System Indications for Use: The Medigus Ultrasonic Surgical Endostapler (MUSE™) System is intended for endoscopic placement of surgical staples in the soft tissue of the esophagus and stomach in order to create anterior partial fundoplication for treatment of symptomatic chronic Gastro Esophageal Reflux Disease in patients who require and respond to pharmacological therapy. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 {3}------------------------------------------------ ## II. 510(k) Summary ### Medigus Ultrasonic Surgical Endostapler (MUSE™) System #### I. SUBMITTER Medigus Ltd. Building 7A Omer Industrial Park Omer, 84965 Israel Phone: +972-8-6466880 Fax: +972-8-6466770 Sheila S. Stevens, PhD Contact Person: US Regulatory and Clinical Affairs Consultant Date Prepared: May 26, 2015 #### II. DEVICE Name of Device: Medigus Ultrasonic Surgical Endostapler (MUSE) System Common or Usual Name: Endoscopic Stapling System Classification Name: Endoscopic Suture/Plication System, Gastroesophageal Reflux Disease (GERD) 21 CFR 876.1500- Endoscope and Accessories Regulatory Class: II Product Code: ODE Special Controls: N/A #### III. PREDICATE DEVICE K143634 Medigus SRS Endoscopic Stapling System (Trade Name MUSE) This predicate has not been subject to a design-related recall. No reference devices were used in this submission. #### IV. DEVICE DESCRIPTION The Medigus Ultrasonic Surgical Endostapler (MUSE™) System enables the operator to staple the fundus of the stomach to the esophagus, in 2 or more locations around the esophageal circumference, thereby creating a permanent surgical fundoplication. The procedure is completed entirely through the mouth, without any incisions. The system consists of three main parts: the endoscopic stapler, the control console and several accessories. The endoscopic stapler is a single use, sterile device (EtO) which resembles an endoscope in appearance and material construction. The distal tip of the device contains a video camera, ultrasonic range finding sight, illumination, irrigation port, {4}------------------------------------------------ insufflation port, and the staple anvil. The distal tip is retroflexed to align with the staple cartridge located in the shaft of the stapler. An alignment pin in the distal tip is used for initial positioning of the anvil against the cartridge. The cartridge is provided sterile (EtO) and contains standard, 4.8 mm titanium surgical staples. Each application of the device fires five staples in 3 staggered rows. A new cartridge is loaded for each application. The control console includes the insufflation, light and camera electronics. The associated accessories include: - Irrigation bottle with liquids for irrigation of the camera lens - Silicon tubes for connecting the console and other accessories to the ● stapler - Standard overtube for protecting patient's pharynx ● - Tweezers to remove and replace staple cartridge ● - Staple cartridges ● #### V. INDICATIONS FOR USE The Medigus Ultrasonic Surgical Endostapler (MUSE™) System is intended for endoscopic placement of surgical staples in the soft tissue of the esophagus and stomach in order to create anterior partial fundoplication for treatment of symptomatic chronic Gastro Esophageal Reflux Disease in patients who require and respond to pharmacological therapy. The Indications for Use statement for the modified device is identical to that of the predicate. ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE The modified device shares all technological characteristics with the predicate with the exception of the following differences: - The subject device contains an overtube manufactured by Medigus, ● whereas the predicate device contained a commercially available overtube not manufactured by Medigus. #### VII. PERFORMANCE DATA The following performance data summaries were provided in support of the substantial equivalence determination: - Dimensional comparisons - Mechanical Testing - Biocompatibility Testing ● {5}------------------------------------------------ ### VIII. CONCLUSIONS The modified Medigus Ultrasonic Surgical Endostapler (MUSE™) System has the same intended use, indications, principles of operation and fundamental technology as the cleared, predicate version of the device. The minor differences in the subject device's overtube do not raise any new or different questions of safety or effectiveness. Performance data demonstrates that the subject device performs comparably to the predicate device that is currently marketed for the same intended use. Thus, the subject device is substantially equivalent to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...