KYPHON (R) HV-R(R) Bone Cement

K150460 · Medtronic Sofamor Danck USA, Inc. · NDN · Apr 28, 2015 · Orthopedic

Device Facts

Record IDK150460
Device NameKYPHON (R) HV-R(R) Bone Cement
ApplicantMedtronic Sofamor Danck USA, Inc.
Product CodeNDN · Orthopedic
Decision DateApr 28, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3027
Device ClassClass 2
AttributesTherapeutic

Indications for Use

KYPHON® HV-R® Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a kyphoplasty or vertebroplasty procedure. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor.

Device Story

Two-component PMMA bone cement system; powder component (PMMA copolymer, barium sulfate radiopacifier, benzoyl peroxide initiator) mixed with liquid component (methylmethacrylate monomer, hydroquinone stabilizer, N,N-dimethyl-p-toluidine promoter) prior to use. Used by physicians in clinical settings during kyphoplasty or vertebroplasty procedures to stabilize vertebral fractures. Provides structural support to the vertebral body; benefits patients by addressing pain and instability associated with pathological fractures.

Clinical Evidence

No clinical data; substantial equivalence based on technological characteristics and performance comparisons to predicate devices.

Technological Characteristics

Two-component PMMA bone cement. Powder: PMMA/methyl-methacrylate-styrene copolymer, barium sulfate, benzoyl peroxide. Liquid: methylmethacrylate monomer, hydroquinone, N,N-dimethyl-p-toluidine. Polymerization-based hardening. Sterilized components.

Indications for Use

Indicated for patients with pathological vertebral body fractures caused by osteoporosis, cancer (multiple myeloma, metastatic lesions from breast/lung cancer, lymphoma), or benign lesions (hemangioma, giant cell tumor) undergoing kyphoplasty or vertebroplasty.

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, represented by three overlapping profiles, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 28, 2015 Medtronic Sofamor Danek USA, Incorporated Ms. Kathy L. Remsen Senior Regulatory Affairs Program Manager 1800 Pyramid Place Memphis, Tennessee 38132 Re: K150460 Trade/Device Name: KYPHON® HV-R® Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: March 10, 2015 Received: March 13, 2015 Dear Ms. Remsen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K150460 Page 1 of 1 #### INDICATIONS FOR USE 510(k) Number (if known): K150460 Device Name: KYPHON® HV-R® Bone Cement Indications for Use: K YPHON® HV-R® Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a kyphoplasty or vertebroplasty procedure. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. | Prescription Use _X_ | AND/OR | Over-The-Counter Use _ _ | |-----------------------------|--------|--------------------------| | (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) | ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ ## KYPHON® HV-R® Bone Cement 510(k) Summary #### February 2015 | Company: | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133<br>Fax: (901) 346-9738 | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Kathy L. Remsen<br>Senior Regulatory Affairs Program Manager | | Proprietary Trade Name: | KYPHON® HV-R® Bone Cement | | Common Name: | Bone Cement | | Classification Name: | Cement, bone, vertebroplasty | | Product Code: | NDN | | Regulation: | 21 CFR 888.3027 | | Classification: | II | Description: KYPHON® HV-R® Bone Cement is provided as a two component system. The powder component consists of a PMMA copolymer (polymethylmethacrylate/ methyl-methacrylate-styrene copolymer) with barium sulfate as a radiopacifier and benzoyl peroxide as an initiator. The liquid component consists of methylmethacrylate monomer, with the addition of hydroquinone as a stabilizer and N,N-dimethyl-p-toluidine as a promoter. The powder and liquid components are mixed prior to use. Indications for Use: KYPHON® HV-R® Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer or benign lesions using a kyphoplasty or vertebroplasty procedure. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Summary of the Technological Characteristics: The subject KYPHON® HV-R® Bone Cement has the same fundamental scientific technology as the predicates KYPHON® HV-R® Bone Cement (K093828 S.E. 8/12/2010) and KYPHON® Xpede™ Bone {4}------------------------------------------------ Cement (K102397 S.E. 2/28/2011). The subject device utilizes equivalent implant materials and sterilization methods. Identification of Legally Marketed Devices: The subject KYPHON® HV-R® Bone Cement is substantially equivalent to the KYPHON® HV-R® Bone Cement (K093828 S.E. 8/12/2010) and KYPHON® Xpede™ Bone Cement (K102397 S.E. 2/28/2011). Conclusion: The design features, device materials, chemical composition, device performance, packaging of the device materials, manufacturing and sterilization methods are substantially equivalent to the previously cleared KYPHON® HV-R® Bone Cement (K093828 S.E. 8/12/2010) and KYPHON® Xpede™ Bone Cement (K102397 S.E. 2/28/2011). The updated indication does not raise new issues of safety or effectiveness. The intended use for KYPHON® HV-R® Bone Cement has not changed and is identical to the predicates KYPHON® HV-R® Bone Cement (K093828 S.E. 8/12/2010) and KYPHON® Xpede™ Bone Cement (K102397 S.E. 2/28/2011).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...