The NavSuite®3 Kit is a computer workstation that, when used with CranialMap Neuro Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing a cranial surgical procedure using stereotactic techniques. The NavSuite®3 Kit is a computer workstation that, when used with SpineMap 3D Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing an orthopaedic spinal surgical procedure using stereotactic techniques.
Device Story
NavSuite3 Kit is a computer workstation for intraoperative surgical guidance; assists in localization of surgical instruments relative to patient-specific images or anatomical landmarks. System uses active optical tracking via infrared light; three linear sensors in a navigation camera detect signals from infrared light-emitting diodes (LEDs) attached to surgical instruments. System displays navigation information on a user-provided monitor. Used in OR by surgeons to perform interventions with high precision. Components (computer, IO tablet, navigation camera) are configured and affixed in the OR suite. Provides real-time tracking of instrument position; helps surgeon navigate target anatomical sites. Benefits include enhanced surgical precision and guidance during cranial and spinal procedures.
Clinical Evidence
Bench testing only. Functional risk analysis performed per ISO 14971. System testing verified shutdown/crash behavior, monitor compatibility (1920x1080 resolution), and mechanical reliability of camera joints and fiber optic cables over 5,687 cycles (exceeding 1,000-cycle lifetime estimate).
Technological Characteristics
Computer workstation; Windows XP, Intel I5 platform with WLAN. Active optical tracking using infrared LEDs and three linear sensors. Connectivity via DVI signal to user-provided monitor. Data/power transmission via fiber optic cables. Fixed installation in OR. Risk management per ISO 14971.
Indications for Use
Indicated for patients undergoing cranial or orthopaedic spinal surgical procedures using stereotactic techniques. Contraindicated for use in an MR environment (MR unsafe).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping faces suggesting community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the department's role and national scope. The logo is simple, using a single color, and is designed to be easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 8, 2015
Stryker Leibinger GmbH & Co. KG Robin Rowe Director, Regulatory/Clinical Affairs Boetzinger Strasse 41 Freiburg, Baden-Wuerttemberg D-79111 Germany
Re: K150301
Trade/Device Name: NavSuite3 Kit Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW, OLO Dated: June 5, 2015 Received: June 8, 2015
Dear Ms. Rowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena =5/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150301
Device Name NavSuite®3 Kit
Indications for Use (Describe)
The NavSuite®3 Kit is a computer workstation that, when used with CranialMap Neuro Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing a cranial surgical procedure using stereotactic techniques.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150301
Device Name NavSuite®3 Kit
#### Indications for Use (Describe)
The NavSuite®3 Kit is a computer workstation that, when used with SpineMap 3D Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing an orthopaedic spinal surgical procedure using stereotactic techniques.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### 510(k) Summary
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the NavSuite®3 Kit is provided below.
| Device Common Name: | Neurological Stereotaxic Instrument | |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device Proprietary Name: | NavSuite®3 Kit | |
| Applicant: | Stryker Leibinger GmbH & Co. KG – Navigation<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Baden-Wuerttemberg, Germany | |
| Contact: | Robin L. Rowe, MS, RAC<br>Director, Regulatory/Clinical Affairs<br>Stryker Corporation<br>4100 E. Milham Ave.<br>Kalamazoo MI 49001<br>Phone: 269.389.3893<br>Cell: 269.806.5779<br>Fax: 269-389-2375 | |
| Date Prepared: | June 5, 2015 | |
| Classification Regulation: | Stereotaxic Instrument, 21 CFR 882.4560, Class II | |
| Panel: | Neurology | |
| Primary Product Code: | HAW - Neurological Stereotaxic Instrument | |
| Secondary Product Code: | OLO - Orthopedic Stereotaxic Instrument | |
| Predicate Device: | Stryker NAV3i Platform (K130874) | |
### Indication for Use:
The NavSuite®3 Kit is a computer workstation that, when used with CranialMap Neuro Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing a cranial surgical procedure using stereotactic techniques.
The NavSuite®3 Kit is a computer workstation that, when used with SpineMap 3D Navigation software, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The clinical setting and target population for the NavSuite®3 Kit is that of a patient undergoing an orthopaedic spinal surgical procedure using stereotactic techniques.
{6}------------------------------------------------
## Device Description:
The Stryker Navigation System is a planning and intraoperative guidance system which assists in various surgical procedures. It allows for the localization of surgical instruments and visualization of their position relative to patient specific images and/or patient specific anatomical landmark information assisting the surgeon in performing the intervention at a high level of precision. For localization, active optical tracking based on infrared light is used. Using three linear sensors, the Navigation Camera detects signals from infrared light emitting diodes which are attached to the instruments to be localized. The Computer Monitor displays the navigation information to the user.
The NavSuite®3 Kit is a minor modification of the previously cleared Stryker NAV3i Platform. The NavSuite®3 Kit consists of the same main components that were cleared as the NAV3i Platform in K130874. The NAV3i Platform (K130874) consisted of a mobile cart, a computer system, a monitor, an IO Tablet and a navigation camera. The subject of this Special 510(k) is that the system is now modified to provide these same components, without the mobile cart and without the monitor. The computer, IO Tablet and navigation camera are now provided as separate components that can be configured and affixed in the operation room suite based on user preferences. The monitor that was provided with the NAV3i Platform is not provided with the NavSuite®3 Kit. The NavSuite®3 Kit is compatible with any monitor that meets the specifications provided in the Instructions for Use.
Like the predicate Nav3i platform, the NavSuite3 Kit platform is compatible with the following previously cleared Stryker Software Application Modules:
- K131214 CranialMap Neuro Navigation (Including CranialMap Express) .
- K141941 SpineMap Navigation
## Performance Data:
A functional risk analysis for the NavSuite®3 Kit was conducted and applied in accordance with ISO 14971, 2nd Edition: Medical Devices – Application of Risk Management to Medical Devices (2007).
The results of the risk analysis for the NavSuite®3 Kit determined that the device modifications which are the subject of this 510(k) do not introduce any new risks compared to the predicate device. All existing risks that were identified for the predicate device are the same for the subject device and are mitigated to a risk level as low as reasonably possible. The effectiveness of all mitigation measures is verified per the testing detailed in Table 1.
| Modification | Description | Tests Performed |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cart with<br>Castors | Cart is not used in the NavSuite®3 Kit. | No testing required. |
| UPS | No UPS is included in NavSuite®3 Kit. | System test:<br>Shutdown and crash behavior was tested with<br>Navigation Software Application Modules<br>integration tests |
| Modification | Description | Tests Performed |
| Operating<br>System | Same as the NAV3i OS, with minor<br>configuration changes. | Component test:<br>To show that the configurations are<br>implemented and effective<br><br>System test:<br>To show that fixed resolution was used on<br>several monitors |
| Monitor | No monitor is provided with NavSuite®3 Kit.<br>Any compatible monitor that meets the<br>specifications provided in the Instructions for<br>Use can be used. | System test:<br>Various compatible monitors were tested with<br>native resolution of 1920 x 1080 pixels, divided<br>on the Navigation Software Application<br>Modules integration tests |
| Navigation<br>Camera<br>Mounting | Camera will be fixed on ceiling-mounted arm<br>using an adapted camera joint. | System test:<br>Automated mechanical stress tests to show<br>reliability by simulated intraoperative use over<br>lifetime, rotational and vertical adjustment and<br>that the camera joint can be moved without<br>restrictions within the range of motion, using<br>statistical methods<br><br>A lifetime of 1000 movement cycles was<br>estimated for a camera joint. This resulted in<br>tests with five test samples of camera joints,<br>each over 5687 test cycles. |
| Navigation<br>Camera<br>Cables | Additional Navigation camera data and power<br>cables are needed.<br>Navigation camera data cables include optical<br>fibers.<br>Camera will be fixed on ceiling-mounted<br>arms | Component Test:<br>To verify voltage and data transfer.<br><br>System Tests:<br>Automated mechanical stress tests to show<br>reliability of the Navigation camera cables by<br>simulated intraoperative use over specified<br>lifetime using statistical methods<br><br>For a camera cable set (power and data), a<br>lifetime of 1000 movement cycles was<br>estimated. This resulted in tests with five test<br>samples of camera cable sets (power and data),<br>each over 5687 test cycles. |
| LiveCAM | Not provided with NavSuite®3 Kit | System test:<br>Verify that applicable Software Application<br>Module is not affected.<br><br>Tested with Navigation Software Application<br>Module integration test. |
Table 1: Summary of Performance Testing
{7}------------------------------------------------
## Substantial Equivalence:
Table 2 provides a comparison of the technological characteristics of the subject device to the predicate device and an assessment of the equivalence of each characteristic.
{8}------------------------------------------------
| | Proposed Device | Predicate Device | Equivalence Assessment |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | NavSuite®3 Kit | NAV3i Platform | N/A |
| 510(k) Number | TBD | K130874 | N/A |
| Submitter | Stryker Corporate | Stryker Corporate | Identical |
| Classification<br>Regulation | 882.4560 | 882.4560 | Identical |
| Product Code | HAW, OLO | HAW, OLO | Identical |
| Indication | The NavSuite®3 Kit is a computer<br>workstation that, when used with<br>CranialMap Neuro Navigation<br>software, displays patient specific<br>images and/or patient specific<br>anatomical landmark information<br>and tracks the position and<br>movement of surgical instruments<br>in relation to a target anatomical<br>site on a patient. The clinical<br>setting and target population for<br>the NavSuite®3 Kit is that of a<br>patient undergoing a cranial<br>surgical procedure using<br>stereotactic techniques.<br>The NavSuite®3 Kit is a computer<br>workstation that, when used with<br>SpineMap 3D Navigation software,<br>displays patient specific images<br>and/or patient specific anatomical<br>landmark information and tracks<br>the position and movement of<br>surgical instruments in relation to a<br>target anatomical site on a patient.<br>The clinical setting and target<br>population for the NavSuite®3 Kit<br>is that of a patient undergoing an<br>orthopaedic spinal surgical<br>procedure using stereotactic<br>techniques. | The NAV3i is a computer<br>workstation that, when used with<br>specific Stryker Navigation surgical<br>software, displays patient specific<br>images and/or patient specific<br>anatomical landmark information<br>and tracks the position and<br>movement of surgical instruments<br>in relation to a target anatomical site<br>on a patient.<br>The clinical setting and target<br>population for the NAV3i is that of<br>a patient undergoing a surgical<br>procedure using stereotactic<br>techniques. | Identical |
| Contraindication | The NavSuite®3 Kit is not intended<br>for use in an MR environment. The<br>NavSuite®3 Kit is MR unsafe. | The NAV3i Platform is not intended<br>for use in a MRI environment. The<br>NAV3i Platform is MR unsafe. | Identical |
| | Proposed Device | Predicate Device | Equivalence Assessment |
| Power Supply | External power source:<br>AC Power supply, 100/240V,<br>50/60Hz.<br>No UPS provided. | External power source:<br>AC Power supply, 100/240V,<br>50/60Hz.<br>The system is equipped with an off-<br>the-shelf uninterruptible power<br>supply (UPS) for power<br>interruptions less than or equal to 6<br>minutes. | Equivalent:<br>The subject device and<br>predicate device use the<br>same power supply. The<br>UPS is not needed for the<br>subject device based on it<br>being a fixed installation of<br>the components. |
| Computer | Windows XP, Intel I5 Platform<br>with WLAN capability | Windows XP, Intel I5 Platform with<br>WLAN capability | Identical |
| IO-Tablet and<br>RFID Reader | Touch screen tablet with USB<br>ports and CD/DVD drive and<br>RFID Reader | Touch screen tablet with USB ports<br>and CD/DVD drive and RFID<br>Reader | Identical |
| Monitor Arm | Not provided. | Weight balanced articulated arm,<br>swing range of 300° | Equivalent:<br>The monitor arm is not<br>needed for the subject device<br>because a monitor is not<br>provided. |
| Monitor | The NavSuite®3 Kit is compatible<br>with any monitor that meets the<br>specifications provided in the<br>Instructions for Use. (Resolution of<br>1920x1080 pixels, DVI signal) | 32" flat screen, 16:9 aspect ratio,<br>Resolution of 1920x1080 pixels,<br>DVI signal | Equivalent:<br>Although the subject device<br>does not provide a monitor,<br>the labeling states that it is<br>compatible with any monitor<br>that meets the same<br>specifications stated in the<br>instructions for use. These<br>are the same specifications<br>as that of the monitor<br>provided with the predicate<br>device. |
| | Proposed Device | Predicate Device | Equivalence Assessment |
| Navigation<br>Camera | Navigation System Camera<br>(FP6000)<br>LiveCAM Not provided. | Navigation System Camera<br>(FP6000)<br>With LiveCAM feature. | Equivalent:<br>The subject device and the<br>predicate device use the<br>same navigation camera<br>(FP6000). The LiveCAM<br>was an optional component<br>of the NAV3i used to assist<br>the user in orienting the<br>Navigation Camera to the<br>surgical field. The LiveCAM<br>is not required for optimal<br>performance of the system.<br>The color of the enclosure is<br>white compared to the dark<br>grey enclosure of the<br>NAV3i. This is a cosmetic<br>change only and has no<br>effect on device performance<br>and is not a patient<br>contacting component.<br>The cabling for the<br>Navigation camera has been<br>changed from copper based<br>cables to optical fibers. The<br>fiber optic cable provides<br>better signal integrity over<br>the extended distance. |
| Camera Arm | Camera joint, swing range of 330°<br>when connected to fixed camera<br>arm in the OR | Weight balanced articulated arm,<br>swing range of 330° | Equivalent:<br>The purpose of the<br>articulated arm and camera<br>joint are the same; to allow<br>adjustment of the camera<br>position. When fixed to a<br>ceiling mounted arm in the<br>OR, the camera joint<br>provided with the<br>NavSuite®3 Kit provides the<br>same degrees of freedom as<br>the articulated camera arm of<br>the NAV3i predicate device. |
| Cart with<br>Castors | Not provided | Stainless steel, plastic and fiberglass<br>cart with castors for mobility | Equivalent:<br>A cart is not provided with<br>the NavSuite®3 Kit because<br>the components are in a fixed<br>configuration in the OR and<br>are not intended to be<br>mobile. |
Table 2: Device Comparison Table
{9}------------------------------------------------
{10}------------------------------------------------
# Substantial Equivalence Conclusion:
{11}------------------------------------------------
Based on the identical indications for use and the technological comparison between the subject device and the predicate device, the minor modifications made to the NavSuite®3 Kit since its previous clearance in K130784 do not alter the intended use or the fundamental scientific technology, and do not raise new questions of safety or effectiveness. Therefore the subject device can be found substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.