The Stryker Navigation System - CranialMap Neuro Module is a navigation surgical software module that, when used with a specific Stryker computer workstation, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery. The system is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The system should be operated only by trained personnel such as surgeons and clinic staff.
Device Story
CranialMap Neuro is a navigation software module for the Stryker Navigation System. It processes pre-operative DICOM imaging (CT, MRI, PET, SPECT) and intra-operative tracking data to provide surgical guidance. The system uses infrared optical active sensing: infrared LEDs on tracked instruments are sensed by a camera array to compute spatial coordinates. The software displays 3D models and anatomical overlays on a 32" HD monitor, allowing surgeons to visualize instrument position relative to patient anatomy. It features collision detection, automatic mask registration, and touch-screen control via an IO-Tablet. Used in ORs by surgeons and clinic staff, it assists in procedures like craniotomies, biopsies, and tumor resections. By providing real-time spatial feedback, it aids in precise surgical navigation, potentially improving accuracy and reducing invasiveness.
Clinical Evidence
No clinical testing conducted. Substantial equivalence is supported by non-clinical bench testing, including software validation per IEC 62304 and human factors validation testing.
Technological Characteristics
Infrared optical active sensing; infrared LEDs on wireless, battery-powered instruments. System platform includes Stryker PC-3 computer, 32" 16:9 LCD monitor, and IO-Tablet. Connectivity via DICOM (Query/Retrieve/Storage), CD, DVD, USB. Software developed in C/C++ using object-oriented design. Registration methods include point-to-point and automatic mask registration. Sterilization not applicable (software).
Indications for Use
Indicated for patients requiring computer-assisted planning and intraoperative guidance for cranial, skull base, transnasal, transphenoidal, craniofacial, and orbital procedures, including biopsies, tumor resection, shunt placement, and electrode placement. Suitable for any condition where rigid anatomical reference is identifiable. For use by trained surgeons and clinic staff.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
KB1214
Image /page/0/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are closely spaced together, and the word appears to be a logo or brand name. There is a small superscript symbol after the "r" in "stryker".
Navigation
# 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
# 1. Contact Details
| Applicant Name: | Stryker Leibinger GmbH & Co. KG - Navigation<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany<br>Phone number: +49-761-4512 117<br>Fax number: +49-761-4512 49117 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration No.: | 3007582679 |
| Name of Contact Person: | Lilian Eckert<br>lilian.eckert@stryker.com |
| Date prepared: | 12 September 2013 |
# 2. Device Name
| Trade Name: | CranialMap Neuro | | | | |
|----------------------|-------------------------------------------|-----------------------------------|-------------------|-------|--------------|
| Common Name: | Navigation software for cranial surgeries | | | | |
| Classification Name: | Product Code | Device | Regulation Number | Class | Review Panel |
| | Primary Code:<br>HAW | Neurologic stereotaxic instrument | 21 CFR §882.4560 | II | Neurology |
{2}------------------------------------------------
### 3. Legally Marketed Predicate Device
The legally marketed predicate device is the Cranial Module as cleared in K062640 Stryker Navigation System - Cranial Module.
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|---------------|--------------|------------------------------------|----------------------------------------------------------------------------------------------------------|
| K062640 | HAW | iNtellect Cranial<br>iNtellect ENT | Stryker Leibinger GmbH &<br>Co. KG - Navigation<br>Boetzinger Strasse 41<br>D-79111 Freiburg,<br>Germany |
#### 4. Device Description
The Stryker Navigation System - CranialMap Neuro Module is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery. The system is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative quidance where a reference to a rigid anatomical structure can be identified.
The CranialMap Neuro Module is the Stryker Navigation System for cranial surgeries as described in the indications for use statement.
The subject device is the Stryker Navigation Software - CranialMap Neuro which together with the Stryker Navigation Platform, Smart Instruments and Accessories forms the Stryker Navigation System - CranialMap Module. It is therefore regarded as a component of the Stryker Navigation System - CranialMap Neuro Module. The design modifications introduced with the Stryker Navigation Software CranialMap Neuro are intended to increase the user comfort, to enhance the look and feel of the software and to simplify and extend the provided functions using the latest system platform technology.
{3}------------------------------------------------
#### 5. Indications for use
The Stryker Navigation System - CranialMap Neuro Module is a navigation surgical software module that, when used with a specific Stryker computer workstation, is intended as a planning and intraoperative quidance system to enable open or percutaneous computer assisted surgery. The system is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The system should be operated only by trained personnel such as surgeons and clinic staff. The CranialMap Neuro Navigation system supports, but is not limited to, the following surgical procedures:
- . Cranial biopsies
- Puncture of abscesses .
- Craniotomies .
- . Craniectomies
- Resection of tumors and other lesions ●
- . Removal of foreign objects
- � Skull base procedures
- Transnasal neurosurgical procedures .
- Transphenoidal pituitary surgery .
- Shunt placement, including pediatric shunt placement .
- Placement of electrodes for recording, stimulation and lesion generation .
- . Endoscopic Sinus Surgery (ESS)
- Intranasal procedures .
- Ear implant procedures .
- Craniofacial procedures .
- . Skull reconstruction procedures
- Orbital cavity reconstruction procedures .
{4}------------------------------------------------
#### 6. Substantial Equivalence Comparison
In summary, the indications for use of the modified Stryker CranialMap Neuro Module are identical to the predicate device Cranial Module. Furthermore, the technological characteristics of the modified CranialMap Neuro Application Software are substantially equivalent to the original Cranial Application Software. The differences in the technological characteristics do not raise new questions of safety and effectiveness. Consequently, the CranialMap Neuro Module is substantially equivalent to the predicate device.
The following table gives an overview of the substantial equivalence reflecting all modifications being made between the predicate device and the subject device.
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | Refer to chapter 5<br>"Indications for use" | Refer to chapter 5<br>"Indications for use" | Identical<br>Identical, except for the<br>module name, titles and<br>the order of listed<br>indications for use. |
| Look of the<br>Software<br>User<br>Interface | Grey-style graphical<br>user interface on 4:3<br>screen ratio, one tab<br>per task concept from<br>left to right on top of<br>screen, image box with<br>image tools and task<br>panel on the right | Black-style graphical<br>user interface on 16:9<br>screen ratio, one tab<br>per task concept from<br>left to right on top of<br>screen, image box with<br>image tools and task<br>tablet on the right | Equivalent<br>Refreshed and touch-<br>enabled graphical user<br>interface on HD screen.<br>Changes are minor and do<br>not alter the way the<br>software performs nor<br>affect safety and<br>effectiveness of device. |
| Modes of<br>Operation | Software workflow<br>controlled by tabs from<br>left to right, registration<br>planning tab after<br>planning tab. | Software workflow<br>controlled by tabs from<br>left to right, registration<br>planning tab moved<br>before planning tab. | Equivalent<br>Modification improves the<br>ease of use of the<br>software and does not<br>negatively affect the safety<br>and effectiveness of the<br>proposed device. |
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
| Control<br>Mechanism | Mouse Keyboard Buttons on<br>navigated sterile<br>smart tools. | Mouse Virtual Keyboard Buttons on<br>navigated sterile<br>smart tools. IO-Tablet with touch<br>screen | Equivalent<br>The control mechanism<br>during surgery is identical.<br>The control mechanism<br>before surgery is<br>equivalent: mouse,<br>keyboard and touch<br>screen. |
| Operating<br>Principle | The iNtellect Cranial<br>software is installed on<br>the computer that is<br>part of the Navigation<br>system platform.<br>Images are imported in<br>DICOM format. | The CranialMap<br>software is installed on<br>the computer that is<br>part of the Navigation<br>system platform.<br>Images are imported in<br>DICOM format. | Identical<br>The operating principle on<br>the Stryker Platform is<br>unchanged. |
| Accuracy<br>Statement | Within the camera<br>working space, the<br>system has a mean<br>accuracy of 2 mm for<br>translation and 2° for<br>rotation. | Within the camera<br>working space, the<br>system has a mean<br>accuracy of 2 mm for<br>translation and 2° for<br>rotation. | Identical |
| Registration<br>Accuracy | The root mean square<br>value is displayed as<br>mean error of the<br>registration. | The root mean square<br>value is displayed as<br>mean error of the<br>registration. | Identical |
| Check for<br>Sustained<br>Accuracy<br>Intra-<br>operatively | Navigation instrument's<br>tip geometry check on<br>initialization of<br>instrument.<br>Intra-operative<br>interactive landmark<br>checks demanded by<br>the software, e.g.<br>during mandatory<br>registration<br>confirmation step. | Navigation instrument's<br>tip geometry check on<br>initialization of<br>instrument.<br>Intra-operative<br>interactive landmark<br>checks demanded by<br>the software, e.g.<br>during mandatory<br>registration<br>confirmation step. | Identical<br>An additional video<br>guidance how to check the<br>instrument's tip geometry<br>makes the device more<br>users friendly and does<br>not negatively affect the<br>safety and effectiveness of<br>the proposed device. |
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
| System Components | | | |
| Localization<br>Technology<br>and Working<br>Sphere | Infrared optical active<br>sensing technology:<br>Infrared light emitted by<br>diodes placed in a<br>known fashion on<br>instruments (trackers)<br>is sensed by a camera<br>array thus allowing for<br>computation of the<br>spatial information. | Infrared optical active<br>sensing technology:<br>Infrared light emitted by<br>diodes placed in a<br>known fashion on<br>instruments (trackers)<br>is sensed by a camera<br>array thus allowing for<br>computation of the<br>spatial information. | Identical<br>The Navigation Camera is<br>the same camera as used<br>for the predicate device. |
| Navigation<br>System<br>Platform | Stryker Navigation<br>System II - Cart,<br>Stryker eNlite:<br>• Stryker<br>Navigation<br>Camera<br>• Mouse,<br>Keyboard<br>• Stryker PC<br>Computer | Stryker Navigation<br>Platform:<br>• Stryker<br>Navigation<br>Camera<br>• IO-Tablet<br>• Stryker PC-3<br>Computer | Equivalent<br>The same navigation<br>camera and equivalent PC<br>based components are<br>used. The IO-tablet<br>interface to control the<br>application is a state-of-<br>the-art alternative to using<br>a mouse and keyboard for<br>user's convenience. |
| OR Display | Display properties:<br>• 4:3 LCD monitor<br>• 1280 x 1024 pixel<br>resolution<br>• 19" screen | Display properties:<br>• 16:9 LCD monitor<br>• 1980 x 1080 pixel<br>resolution<br>• 32" screen | Equivalent<br>The larger high resolution<br>display does not<br>negatively affect the safety<br>and effectiveness of the<br>subject device. |
| Smart<br>Instrument<br>Technology<br>(Trackers) | Active, battery<br>powered, wireless<br>instrumentation with bi-<br>directional IR-<br>communication and IR-<br>LEDs for tracking | Active, battery<br>powered, wireless<br>instrumentation with bi-<br>directional IR-<br>communication and IR-<br>LEDs for tracking | Identical<br>All smart instruments<br>supported by the predicate<br>device are supported by<br>the subject device as well. |
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
| Calibration of<br>Instrument<br>Dimensions | Workflow to calibrate<br>and validate the<br>navigated instruments<br>during the operation. | Workflow to calibrate<br>and validate the<br>navigated instruments<br>during the operation. | Identical |
| Disposable<br>Parts | Batteries for trackers,<br>registration markers,<br>Frameless Guide,<br>Patient Registration<br>and tracking mask. | Batteries for trackers,<br>registration markers,<br>Frameless Guide,<br>Patient Registration<br>and tracking mask. | Identical |
| Scanner<br>Interface<br>Technology<br>(to imaging<br>devices) | Magneto Optical Disk<br>CD, DVD, USB<br>Network connectivity<br>DICOM storage client | CD, DVD, USB<br>Network connectivity<br>DICOM storage client<br>DICOM query/retrieve | Equivalent<br>Modification to support<br>state-of-the-art scanner<br>interfaces. |
| Supported<br>imaging<br>modalities | CT, CTA<br>MR, MRA, fMRI<br>PET, SPECT | CT, CTA<br>MR, MRA, fMRI<br>PET. SPECT | Equivalent<br>More convenient image<br>modality correlation based<br>on DICOM frame-of-<br>reference ID supported. |
| Microscope<br>interfaces | Zeiss NC4,<br>Zeiss Pentero,<br>Leica M5xx,<br>Leica M720<br>microscope models | Zeiss NC4, Zeiss<br>Pentero, Zeiss<br>Pentero 900, Leica<br>M5xx<br>Leica M720<br>microscope models | Equivalent<br>The additional support of<br>the Zeiss Pentero 900<br>microscope is identical to<br>the Pentero microscope<br>except for a new<br>microscope ID. |
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
| | Operating System and Implementation Technology | | |
| Platform<br>Operating<br>System | Windows 2000<br>Windows XP | Windows XP | Equivalent<br>Higher version of Windows<br>operating system used. |
| Software<br>Implemen-<br>tation | C and C++<br>programming applying<br>object oriented analysis<br>and design methods,<br>Qt GUI design and up-<br>to-date continuous<br>integration and testing<br>tools in a Microsoft<br>development<br>environment. | C and C++<br>programming applying<br>object oriented analysis<br>and design methods,<br>Qt GUI design and up-<br>to-date continuous<br>integration and testing<br>tools in a Microsoft<br>development<br>environment. | Equivalent<br>Additional methods have<br>been applied to the<br>software development of<br>the subject device for<br>state-of-the-art software<br>quality, e.g. code<br>coverage measurements<br>and 100% code reviews. |
| Use of Off-<br>the-Shelf<br>(OTS)<br>Software | OTS software for<br>logging, visualization,<br>algorithms, 3D<br>graphics, multi-<br>threading, GUI<br>components, volume<br>rendering, multi-<br>modality matching,<br>backup on CD and cpp-<br>templates. | OTS software for<br>logging, visualization,<br>algorithms, 3D<br>graphics, multi-<br>threading, GUI<br>components, volume<br>rendering, multi-<br>modality matching,<br>backup on CD and cpp-<br>templates.<br>DICOM Tool Kit | Equivalent<br>For the CranialMap Neuro<br>Software the same Off-<br>the-Shelf software<br>components that have<br>already been used for the<br>predicate device are used<br>with updated versions<br>where appropriate.<br>The new DICOM<br>Query/Retrieve extension<br>is based on the DICOM<br>Tool Kit. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
:
{9}------------------------------------------------
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
|----------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CranialMap Neuro Modifications | | | |
| Touch<br>Screen<br>Functionality | No touch screen GUI operated via<br>Mouse input | Touch on IO-tablet<br>can be used<br>alternatively to<br>mouse control | Equivalent<br>Additional touch screen<br>functionality does not<br>negatively affect the safety<br>and effectiveness |
| LiveCam<br>System<br>Setup | System setup using<br>side and top view of<br>working sphere. | System setup using<br>side and top view of<br>working sphere. Additionally a<br>"localizer's eye" live<br>video view option. | Equivalent<br>The additional option using<br>the LiveCam does not<br>affect the performance or<br>the indications for use<br>of the subject device. |
| Collision<br>Zones | Segments and<br>annotations can be<br>defined | Segments and<br>annotations can be<br>defined Collision check<br>option for<br>segments,<br>annotations and<br>functional overlays<br>during navigation | Equivalent<br>An additional option for<br>audible and visual<br>feedback during<br>navigation is provided.<br>This modification does not<br>negatively affect safety<br>and effectiveness or<br>change the indications for<br>use of the subject device. |
{10}------------------------------------------------
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image series<br>import and<br>handling | Anatomical and functional MRI data is imported as separate image series. The "identity correlation" feature is used to correlate functional data to the appropriate anatomical MRI series. The correctness of the image correlation must be confirmed during image import. | Functional MRI images are detected during image import as overlays of a related anatomical MRI image series automatically. System applies an "identity correlation" based on DICOM reference frame ID. The correctness of the image correlation must be confirmed during image import. Deletion of imported image series. | Equivalent<br>Simplification by reducing number of steps needed to correlate image modalities. Like in the predicate device the user must confirm the result.<br>For convenience an unintentionally imported image series can be removed.<br>The modification does not negatively affect the safety and effectiveness of the subject device. |
| 3D model<br>visualization | Advanced virtual surgical planning based on 3D DICOM data, e.g. a CT scanned implant possible. | Planning based on 3D model data, e.g. STL models is possible. The 3D models can be loaded and aligned with anatomical image data. | Equivalent<br>Advanced virtual planning is enhanced but no change of indications for use. Change does not negatively affect safety and effectiveness. |
{11}------------------------------------------------
| Topic | Predicate Device<br>iNtellect Cranial<br>Software | Subject Device<br>CranialMap Neuro<br>Software | Equivalence assessment |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Automatic<br>intra-<br>operative<br>mask (AIM)<br>registration | • Automatic mask<br>registration and<br>tracking based on<br>skin surface<br>extraction. | • Automatic mask<br>registration and<br>tracking based on<br>skin surface<br>extraction.<br><br>• The mask is<br>attached on the<br>patient during the<br>CT scan and AIM<br>registration is able<br>to detect the LED<br>positions of the<br>mask in the scan. | Equivalent<br>Additional mask<br>registration method using<br>a CT scan while the mask<br>is attached to the patient.<br>The established point-to-<br>point registration<br>algorithms are used. No<br>change of the indications<br>for use. Change does not<br>negatively affect safety<br>and effectiveness. |
#### 7. Non-clinical Testing
Validation activities, including human factors validation testing, have been conducted to provide assurance that the device meets the performance requirements under its indications for use conditions. In addition, system software validation acc, to IEC 62304 "Medical device software -Software life cycle processes", have been performed to show that the device is safe and effective.
# 8. Clinical Testing
No clinical testing has been conducted.
# 9. Conclusions
The conclusions drawn from the nonclinical tests demonstrate that the Stryker Navigation Application Software CranialMap Neuro performs as safely and effectively as the legally marketed device identified in chapter 3. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the Stryker Navigation System - CranialMap Neuro Module is substantially equivalent to the predicate device with respect to its indications for use, technological characteristics and performance characteristics.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 18, 2013
Stryker Corporate Ms. Lilian Eckert Boetzinger Str. 41 Freiburg, Baden-Wuerttemberg, GM 79111
Re: K131214
Trade/Device Name: Cranial Map Neuro Module Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: HAW Dated: September 12, 2013 Received: September 16, 2013
Dear Ms. Eckert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{13}------------------------------------------------
Page 2 - Ms. Lilian Eckert
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K131214
Device Name:
Stryker CranialMap Neuro Module
Indications For Use:
The Stryker Navigation System - CranialMap Neuro Module is a navigation surgical software module that, when used with a specific Styker computer workstation, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery.
The system is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
The system should be operated only by trained personnel such as surgeons and clinic staff.
The CranialMap Neuro Navigation system supports, but is not limited to, the following surgical procedures:
- Cranial biopsies ●
- Puncture of abscesses .
- Craniotomies .
- Craniectomies .
- Resection of tumors and other lesions .
- Removal of foreign objects .
- Skull base procedures �
- Transnasal neurosurgical procedures �
- . Transphenoidal pituitary surgery
- Shunt placement, including pediatric shunt placement .
- Placement of electrodes for recording, stimulation and lesion generation �
- Endoscopic Sinus Surgery (ESS) �
- Intranasal procedures .
- Ear implant procedures .
- Craniofacial procedures .
- Skull reconstruction procedures .
- Orbital cavity reconstruction procedures .
{15}------------------------------------------------
Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ・
Concurrence of Center for Devices and Radiological Health (CDRH)
Joyce M. Whang -S
Page 2 of _ _ 2__
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.