Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-Acetylmorphine Urine Calibrator, Immunalysis 6-Acetylmorphine Urine Controls
K150275 · Immunalysis Corporation · DJG · Mar 9, 2015 · Clinical Toxicology
Device Facts
| Record ID | K150275 |
| Device Name | Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-Acetylmorphine Urine Calibrator, Immunalysis 6-Acetylmorphine Urine Controls |
| Applicant | Immunalysis Corporation |
| Product Code | DJG · Clinical Toxicology |
| Decision Date | Mar 9, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3650 |
| Device Class | Class 2 |
Intended Use
The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only. The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are used. Immunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay. Immunalysis 6-Acetylmorphine Urine Calibrator: The Immunalysis 6-Acetylmorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.
Device Story
Device is a homogeneous enzyme immunoassay for qualitative detection of 6-Acetylmorphine in human urine. Input: human urine samples. Principle: competitive binding between drug in sample and enzyme-labeled drug conjugate for recombinant antibodies; G6PDH enzyme activity measured via absorbance change. Output: qualitative positive/negative result based on 10ng/mL cutoff. Used in clinical laboratories by technicians/pathologists on automated clinical chemistry analyzers (e.g., Beckman Coulter AU 400e). Results are preliminary; require confirmation by GC-MS or LC/MS. Clinical decision-making relies on confirmed results and professional judgment. Benefits: rapid screening for 6-Acetylmorphine, a heroin metabolite, to assist in toxicology assessment.
Clinical Evidence
Bench testing only. Precision/cutoff characterization (N=80 per concentration) verified 10ng/mL cutoff performance. Specificity/cross-reactivity evaluated against structurally related and non-similar compounds. Interference testing assessed endogenous compounds, pH, and specific gravity; ascorbic acid, sodium chloride, boric acid, and pH 3.0 identified as potential interferences. Method comparison study (N=80 clinical samples) against LC/MS showed 100% agreement.
Technological Characteristics
Homogeneous enzyme immunoassay. Reagents: recombinant antibodies, G6P, NAD, G6PDH-labeled 6-Acetylmorphine derivative in HEPES buffer with Sodium Azide. Form: liquid, ready-to-use. Cutoff: 10ng/mL. Analyzers: automated clinical chemistry systems. Storage: 2–8°C.
Indications for Use
Indicated for the qualitative analysis of 6-Acetylmorphine in human urine to provide preliminary analytical test results. Intended for laboratory use on automated clinical chemistry analyzers. Prescription use only.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Emit® II Plus 6-Acetylmorphine Assay (K102779)
Related Devices
- K101195 — 6-ACETYLMORPHINE ENZYME IMMUNOASSAY WITH 6-ACETYLMORPHINE CALIBRATORS AND CONTROLS · Lin-Zhi International, Inc. · Jul 6, 2010
- K150925 — Immunalysis Benzoylecgonine Urine Enzyme Immunoassay, Immunalysis Benzoylecgonine Urine Calibrators, Immunalysis Benzoylecgonine Urine Control Set · Immunalysis Corporation · Jun 29, 2015
- K173183 — CEDIA Heroin Metabolite (6-AM) Assay · Microgenics Corporation · Nov 22, 2017
- K192943 — CEDIA Heroin Metabolite (6-AM) assay · Microgenics Corporation · Dec 16, 2019
- K231007 — CEDIA Heroin Metabolite (6-AM) Assay · Microgenics Corporation · Sep 27, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
IMMUNALYSIS CORPORATION JOSEPH GINETE REGULATORY AFFAIRS SPECIALIST 829 TOWNE CENTER DR. POMONA CA 91767
March 9, 2015
Re: K150275
Trade/Device Name: Immunalysis 6-acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-acetylmorphine Urine Calibrator, Immunalysis 6-acetylmorphine Urine Controls
Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DKB, DIF Dated: January 20, 2015 Received: February 4, 2015
Dear Mr. Ginete:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -A
For : Courtney H. Lias, Ph.D.
Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150275
#### Device Name
Immunalysis 6-Acetylmorphine Urine Enzyme Immunalysis 6-Acetylmorphine Urine Controls and Immunalysis 6-Acetylmorphine Urine Calibrator
#### Indications for Use (Describe)
The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is callbrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.
The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are used.
Immunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.
Immunalysis 6-Acetylmorphine Urine Calibrator: The Immunalysis 6-Acetylmorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and appears to be in bold font. The red background is a solid color, and the white letters stand out against it.
A. Contact Information
# 510(k) SUMMARY (K150275)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
- 1. Manufacturer: Immunalysis Corporation 2. Contact Name: Joseph Ginete Regulatory Affairs Specialist 3. Contact Title: 4. Address: 829 Towne Center Drive Pomona, CA 91767 (909) 482-0840 5. Phone: (909) 482-0850 6. Fax: 7. Email: iginete@immunalysis.com 8. Summary prepared on: March 6, 2015 B. Device Information 1. Trade Name: Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay Immunalysis 6-AcetyImorphine Urine Controls Immunalysis 6-Acetylmorphine Urine Calibrator 2. Common Name: Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay Immunalysis 6-AcetyImorphine Urine Controls Immunalysis 6-AcetyImorphine Urine Calibrator 3. Device Classification: II 4. Regulation Number: 21 CFR 862.3650 Opiate Test System 21 CFR 862.3200 Clinical Toxicology Calibrator 21 CFR 862.3280 Clinical Toxicology Control Material 5. Panel: Toxicology(91) 6. Product Code: DJG DKB DIF C. Legally Marketed Device to Which We are Claiming Equivalence (807.92(A)(3))
- 1. Predicate Device: Emit® II Plus 6-AcetyImorphine Assay Emit® II Plus 6-AM/ Ecstasy Calibrators Emit® II Plus 6-AM/ Ecstasy Controls 2. Predicate Company: Siemens Healthcare Diagnostics Inc. 3. Predicate K Number: K102779
{4}------------------------------------------------
### D. Device Description
The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes recombinant antibodies to 6-Acetylmorphine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in HEPES buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes 6-AcetyImorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with Sodium Azide as a preservative. Calibrator and controls are sold separately. Reagents are liquid, ready to use
The 6-AcetyImorphine calibrator and controls consist of a cutoff calibrator at 10ng/mL, a LOW control at 7.5ng/mL for the 10ng/mL cutoff and a HIGH control at 12.5ng/mL for the 10ng/mL cutoff.
- E. Intended Use
The Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.
The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liguid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
lmmunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.
Immunalysis 6-AcetyImorphine Urine Calibrator: The Immunalysis 6-AcetyImorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.
| | 6-Acetylmorphine Assay K102779 | Immunalysis 6-Acetylmorphine Urine EIA |
|-------------------|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Intended Use | The qualitative analysis of 6-<br>Acetylmorphine in human urine with<br>automated clinical chemistry<br>analyzers | Same |
| Type of Product | Analytical Reagents | Same |
| Measured Analytes | 6-Acetylmorphine | Same |
| Test Matrix | Urine | Same |
| Cutoff Levels | 10ng/mL of 6-Acetylmorphine | Same |
| Test System | Homogeneous Enzyme<br>Immunoassay | Same |
| Materials | Liquid Ready-to-Use R1 Reagent<br>and Lyophilized R2 Reagent | Liquid Ready-to-Use Antibody/Substrate<br>Reagents and Liquid Ready-to-Use Enzyme<br>Labeled Conjugate |
- F. Comparison of the new device with the predicate device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "IMMUNALYS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects or details in the image. The font is sans-serif and the letters are bold.
| | 6-Acetylmorphine Assay K102779 | Immunalysis 6-Acetylmorphine Urine ElA |
|-----------------------------------|---------------------------------------------------------|------------------------------------------|
| Mass Spectroscopy<br>Confirmation | Required for preliminary positive<br>analytical results | Same |
| Antibody | Mouse monoclonal antibody to 6-<br>Acetylmorphine | Recombinant antibody to 6-Acetylmorphine |
| Storage | 2 – 8°C until expiration date | Same |
| Calibrator Form | Liquid | Same |
| Calibrator Level | Four (4) Levels (5, 10, 15 and 20<br>ng/mL) | One (1) Level (10ng/mL) |
| Control Levels | Two (2) Levels (7.5 and 12.5ng/mL) | Same |
- G. The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay to the predicate
- 1. Precision/Cutoff Characterization Study was performed for 20 days, 2 runs per day in duplicate (N=80) on concentration of ±25%, ±50%, ±75%, and ±100% of the cutoff. The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. In addition, it also verified product performance relative to the ability of the device to produce the same value during repeated measurements. The instruments used for this was Beckman Coulter AU 400e. y table of the Qualitativ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
| Table 1. Qualitative Analysis (for 10ng/mL cutoff) |
|--------------------------------------------------------------------------------------------------------|
| The following is a summary table of the Qualitative Analysis for the 10ng/mL cutoff test data results. |
| Table 1 - Qualitative Analysis (for 10ng/mL cutoff) | | | |
|-----------------------------------------------------|-------------|---------------------|-----------------|
| Concentration (ng/mL) | % of cutoff | # of determinations | Result |
| 0 | -100% | 80 | 80 Negative |
| 2.5 | -75% | 80 | 80 Negative |
| 5 | -50% | 80 | 80 Negative |
| 7.5 | -25% | 80 | 80 Negative |
| 10 | Cutoff | 80 | 43 Neg / 37 Pos |
| 12.5 | +25% | 80 | 80 Positive |
| 15 | +50% | 80 | 80 Positive |
| 17.5 | +75% | 80 | 80 Positive |
| 20 | +100% | 80 | 80 Positive |
- 2. Specificity and Cross-Reactivity Structurally similar compounds were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The instrument used for this test was a Beckman Coulter AU 400e.
The qualitative result summary table for the 10ng/mL cutoff is outlined below:
| Table 2 - Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative | | | |
|------------------------------------------------------------------------------|------------------------------|----------|----------------------|
| Compound | Concentration Tested (ng/mL) | Result | Cross-Reactivity (%) |
| 6-Acetylmorphine | 10 | Positive | 100 |
| 6-Acetylcodeine | 600 | Positive | 1.7 |
| Buprenorphine | 1,000,000 | Negative | N.D. |
| Codeine | 1,000,000 | Negative | N.D. |
| Dextromethorphan | 1,000,000 | Negative | N.D. |
| Dihydrocodeine | 1,000,000 | Negative | N.D. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters with a drop shadow. The word is set against a red background. The text appears to be bolded and is the central focus of the image.
| Table 2 - Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative | | | |
|------------------------------------------------------------------------------|-----------|----------|----------|
| | | | |
| Ethylmorphine | 1,000,000 | Negative | N.D. |
| Heroin | 1,375 | Positive | 0.7 |
| Hydrocodone | 1,000,000 | Negative | N.D. |
| Hydromorphone | 325,000 | Positive | 0.003 |
| Imipramine | 1,000,000 | Negative | N.D. |
| Levorphanol | 1,000,000 | Negative | N.D. |
| Meperidine | 1,000,000 | Negative | N.D. |
| Morphine | 285,000 | Positive | 0.000035 |
| Morphine 3-D-glucuronide | 1,000,000 | Negative | N.D. |
| Morphine 6-D-glucuronide | 1,000,000 | Negative | N.D. |
| Nalorphine | 80,000 | Positive | 0.0125 |
| Naloxone | 300,000 | Positive | 0.00333 |
| Naltrexone | 390,000 | Positive | 0.00256 |
| Naproxen | 1,000,000 | Negative | N.D. |
| Norbuprenorphine | 100,000 | Negative | N.D. |
| Norcodeine | 1,000,000 | Negative | N.D. |
| Normorphine | 250,000 | Positive | 0.004 |
| Oxycodone | 1,000,000 | Negative | N.D. |
| Oxymorphone | 360,000 | Positive | 0.00277 |
3. Interference - Structurally non-similar compounds, endogenous compounds, the effect of pH and the effect of specific gravity was evaluated by spiking the potential interferent into drug free urine containing the target analyte at ±25% of the cutoff. The instrument used for this test was a Beckman Coulter AU 400e. a. The following is a summary table of the structurally non-similar
| compounds for the 10ng/mL cutoff : | | | | | |
|-------------------------------------------------------------------|---------------------------------|----------------------------------|---------------|-----------------------------------|---------------|
| Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff) | | | | | |
| Compound | Concentration<br>Tested (ng/mL) | -25% Cutoff (7.5ng/mL)<br>Result | Interference? | +25% Cutoff (12.5ng/mL)<br>Result | Interference? |
| 4-Bromo-2,5,Dimethoxyphenethylamine | 100,000 | Negative | No | Positive | No |
| Acetaminophen | 500,000 | Negative | No | Positive | No |
| Acetylsalicyclic Acid | 500,000 | Negative | No | Positive | No |
| Alprazolam | 100,000 | Negative | No | Positive | No |
| 7-Aminoclonazepam | 100,000 | Negative | No | Positive | No |
| 7-Aminoflurnitrazepam | 100,000 | Negative | No | Positive | No |
| 7-Aminonitrazepam | 100,000 | Negative | No | Positive | No |
| Amitriptyline | 100,000 | Negative | No | Positive | No |
| Amobarbital | 100,000 | Negative | No | Positive | No |
| S-(+) Amphetamine | 100,000 | Negative | No | Positive | No |
| Benzoylecgonine | 500,000 | Negative | No | Positive | No |
| Benzylpiperazine | 100,000 | Negative | No | Positive | No |
| Bromazepam | 100,000 | Negative | No | Positive | No |
| Bupropion | 100,000 | Negative | No | Positive | No |
| Butabarbital | 100,000 | Negative | No | Positive | No |
| Butalbital | 100,000 | Negative | No | Positive | No |
| Caffeine | 500,000 | Negative | No | Positive | No |
| Carbamazepine | 100,000 | Negative | No | Positive | No |
| Chlorpromazine | 100,000 | Negative | No | Positive | No |
| cis Tramadol | 100,000 | Negative | No | Positive | No |
| Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff) | | | | | |
| Compound | Concentration<br>Tested (ng/mL) | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| Clobazam | 100,000 | Negative | No | Positive | No |
| Clomipramine | 100,000 | Negative | No | Positive | No |
| Clonazepam | 100,000 | Negative | No | Positive | No |
| Cannabidiol | 100,000 | Negative | No | Positive | No |
| Cannabinol | 100,000 | Negative | No | Positive | No |
| Carisoprodol | 100,000 | Negative | No | Positive | No |
| Chlordiazepoxide | 100,000 | Negative | No | Positive | No |
| Cocaine | 100,000 | Negative | No | Positive | No |
| Cotinine | 100,000 | Negative | No | Positive | No |
| Cyclobenzaprine | 100,000 | Negative | No | Positive | No |
| Digoxin | 100,000 | Negative | No | Positive | No |
| Desalkyflurazepam | 100,000 | Negative | No | Positive | No |
| Dehydronorketamine | 50,000 | Negative | No | Positive | No |
| Demoxepam | 100,000 | Negative | No | Positive | No |
| Delta-9-THC | 100,000 | Negative | No | Positive | No |
| Desipramine | 100,000 | Negative | No | Positive | No |
| N-desmethyltapentadol | 100,000 | Negative | No | Positive | No |
| Diazepam | 100,000 | Negative | No | Positive | No |
| Diphenhydramine | 500,000 | Negative | No | Positive | No |
| Doxepin | 100,000 | Negative | No | Positive | No |
| Ecgonine | 100,000 | Negative | No | Positive | No |
| Ecgonine methyl ester | 100,000 | Negative | No | Positive | No |
| EDDP | 100,000 | Negative | No | Positive | No |
| 1R,2S(-)-Ephedrine | 100,000 | Negative | No | Positive | No |
| 1S,2R(+)-Ephedrine | 100,000 | Negative | No | Positive | No |
| Ethyl glucuronide | 100,000 | Negative | No | Positive | No |
| Fenfluramine | 100,000 | Negative | No | Positive | No |
| Fentanyl | 100,000 | Negative | No | Positive | No |
| Flunitrazepam | 100,000 | Negative | No | Positive | No |
| Fluoxetine | 100,000 | Negative | No | Positive | No |
| Flurazepam | 100,000 | Negative | No | Positive | No |
| Haloperidol | 100,000 | Negative | No | Positive | No |
| Hexobarbital | 100,000 | Negative | No | Positive | No |
| 11-hydroxy-delta-9-THC | 100,000 | Negative | No | Positive | No |
| Ibuprofen | 500,000 | Negative | No | Positive | No |
| Ketamine | 100,000 | Negative | No | Positive | No |
| Lamotrignine | 100,000 | Negative | No | Positive | No |
| Lidocaine | 100,000 | Negative | No | Positive | No |
| Lorazepam | 100,000 | Negative | No | Positive | No |
| Lorazepam Glucuronide | 50,000 | Negative | No | Positive | No |
| Lormetazepam | 100,000 | Negative | No | Positive | No |
| LSD | 100,000 | Negative | No | Positive | No |
| Maprotiline | 100,000 | Negative | No | Positive | No |
| (+)-MDA | 100,000 | Negative | No | Positive | No |
| MDEA | 100,000 | Negative | No | Positive | No |
| MDMA | 100,000 | Negative | No | Positive | No |
| Meprobamate | 100,000 | Negative | No | Positive | No |
| Methadone | 100,000 | Negative | No | Positive | No |
| S(+)-Methamphetamine | 100,000 | Negative | No | Positive | No |
| Methaquolone | 100,000 | Negative | No | Positive | No |
| Methoxetamine | 100,000 | Negative | No | Positive | No |
| Methylphenidate | 100,000 | Negative | No | Positive | No |
| Midazolam | 100,000 | Negative | No | Positive | No |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly blurred, giving them a three-dimensional appearance. The red background is a solid color, and the overall image has a simple, clean design.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly blurred, giving them a soft appearance. The red background is a solid color, providing a strong contrast to the white text.
| Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff) | | | | | |
|-------------------------------------------------------------------|---------------------------------|------------------------|---------------|-------------------------|---------------|
| Compound | Concentration<br>Tested (ng/mL) | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| Nitrazepam | 100,000 | Negative | No | Positive | No |
| Nordiazepam | 100,000 | Negative | No | Positive | No |
| Norketamine | 100,000 | Negative | No | Positive | No |
| Norpropoxyphene | 100,000 | Negative | No | Positive | No |
| Norpseudoephedrine | 100,000 | Negative | No | Positive | No |
| Nortriptyline | 100,000 | Negative | No | Positive | No |
| Oxazepam | 100,000 | Negative | No | Positive | No |
| Oxazepam glucuronide | 100,000 | Negative | No | Positive | No |
| PCP | 100,000 | Negative | No | Positive | No |
| Pentazocine | 100,000 | Negative | No | Positive | No |
| Phentermine | 100,000 | Negative | No | Positive | No |
| Pentobarbital | 100,000 | Negative | No | Positive | No |
| Phenobarbital | 100,000 | Negative | No | Positive | No |
| Phenylephedrine | 100,000 | Negative | No | Positive | No |
| Phenylpropanolamine | 100,000 | Negative | No | Positive | No |
| Phenytoin | 100,000 | Negative | No | Positive | No |
| PMA | 100,000 | Negative | No | Positive | No |
| Prazepam | 100,000 | Negative | No | Positive | No |
| Proproxyphene | 100,000 | Negative | No | Positive | No |
| Propranolol | 100,000 | Negative | No | Positive | No |
| Protriptyline | 100,000 | Negative | No | Positive | No |
| R,R(-)-Pseudoephedrine | 100,000 | Negative | No | Positive | No |
| S,S(+)-Pseudoephedrine | 100,000 | Negative | No | Positive | No |
| Ranitidine | 100,000 | Negative | No | Positive | No |
| Ritalinic Acid | 100,000 | Negative | No | Positive | No |
| Salicylic Acid | 100,000 | Negative | No | Positive | No |
| Secobarbital | 100,000 | Negative | No | Positive | No |
| Sertraline | 100,000 | Negative | No | Positive | No |
| Sufentanil Citrate | 50,000 | Negative | No | Positive | No |
| 11-nor-9 carboxy THC | 100,000 | Negative | No | Positive | No |
| Temazepam | 100,000 | Negative | No | Positive | No |
| Theophylline | 100,000 | Negative | No | Positive | No |
| Thioridazine | 100,000 | Negative | No | Positive | No |
| Triazolam | 100,000 | Negative | No | Positive | No |
| Trifluoromethylphenyl-piperazine | 100,000 | Negative | No | Positive | No |
| Trimipramine | 100,000 | Negative | No | Positive | No |
| Trazodone | 100,000 | Negative | No | Positive | No |
| Verapamil | 100,000 | Negative | No | Positive | No |
| Venlafaxine | 100,000 | Negative | No | Positive | No |
| Zolpidem Tartrate | 100,000 | Negative | No | Positive | No |
b. The following is a summary table of the endogenous compounds results for the 10ng/mL cutoff:
| Table 4 - Endogenous Compounds (for 10ng/mL cutoff) | | | | | |
|-----------------------------------------------------|----------------|------------------------|---------------|-------------------------|---------------|
| | Concentration | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| Compound | Tested (ng/mL) | Result | Interference? | Result | Interference? |
| Acetone | 1.0 g/dL | Negative | No | Positive | No |
| Ascorbic Acid | 1.5 g/dL | Negative | No | Negative | Yes |
| Bilirubin | 0.002 g/dL | Negative | No | Positive | No |
| Creatinine | 0.5 g/dL | Negative | No | Positive | No |
| Ethanol | 1.0 g/dL | Negative | No | Positive | No |
| Galactose | 0.01 g/dL | Negative | No | Positive | No |
| y-Globulin | 0.5 g/dL | Negative | No | Positive | No |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly shadowed, giving them a three-dimensional appearance. The red background is a solid color, and the word is centered.
| Table 4 - Endogenous Compounds (for 10ng/mL cutoff) | | | | | |
|-----------------------------------------------------|---------------------------------|------------------------|---------------|-------------------------|---------------|
| Compound | Concentration<br>Tested (ng/mL) | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| Glucose | 2.0 g/dL | Negative | No | Positive | No |
| Hemoglobin | 0.115 g/dL | Negative | No | Positive | No |
| Human Serum Albumin | 0.5 g/dL | Negative | No | Positive | No |
| Oxalic Acid | 0.1 g/dL | Negative | No | Positive | No |
| Riboflavin | 0.0075 g/dL | Negative | No | Positive | No |
| Sodium Azide | 1% w/v | Negative | No | Positive | No |
| Sodium Chloride | 6.0 g/dL | Negative | No | Negative | Yes |
| Sodium Fluoride | 1% w/v | Negative | No | Positive | No |
| Urea | 6.0 g/dL | Negative | No | Positive | No |
## c. The following is a summary table of the additional ascorbic acid and sodium chloride testing at ±50% of the cutoff:
| Table 5 - Endogenous Compounds (for 10ng/mL cutoff) | | | | | |
|-----------------------------------------------------|---------------------------------|----------------------|---------------|-----------------------|---------------|
| Compound | Concentration<br>Tested (ng/mL) | -50% Cutoff (5ng/mL) | | +50% Cutoff (15ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| Ascorbic Acid | 1.5 g/dL | Negative | No | Negative | Yes |
| Sodium Chloride | 6.0 g/dL | Negative | No | Positive | No |
## d. The following is a summary table of Boric Acid for the 10ng/mL cutoff results:
| Table 6 - Boric Acid (for 10ng/mL cutoff) | | | | | | |
|-------------------------------------------|---------------------------------|------------------------|---------------|-------------------------|---------------|--|
| Compound | Concentration<br>Tested (ng/mL) | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | | |
| Boric Acid | 1% w/v | Result | Interference? | Result | Interference? | |
| | | Negative | No | Negative | Yes | |
## e. The following is a summary table of the additional boric acid testing at +50% of the cutoff:
| Table 7 - Boric Acid (for 10ng/mL cutoff) | | | | | | |
|-------------------------------------------|---------------------------------|----------------------|---------------|-----------------------|---------------|--|
| Compound | Concentration<br>Tested (ng/mL) | -50% Cutoff (5ng/mL) | | +50% Cutoff (15ng/mL) | | |
| | | Result | Interference? | Result | Interference? | |
| Boric Acid | 1% w/v | Negative | No | Negative | Yes | |
#### f. The following is a summary table of the effect of pH results for the 10ng/mL cutoff:
| Table 8 - Effect of pH (for 10ng/mL cutoff) | | | | | |
|---------------------------------------------|------------|------------------------|---------------|-------------------------|---------------|
| Test Parameter | Value | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| pH | <b>3.0</b> | Negative | No | Negative | Yes |
| pH | 4.0 | Negative | No | Positive | No |
| pH | 5.0 | Negative | No | Positive | No |
| pH | 6.0 | Negative | No | Positive | No |
| pH | 7.0 | Negative | No | Positive | No |
| pH | 8.0 | Negative | No | Positive | No |
| pH | 9.0 | Negative | No | Positive | No |
| pH | 10.0 | Negative | No | Positive | No |
| pH | 11.0 | Negative | No | Positive | No |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color, and the white letters stand out clearly against it. The font is sans-serif and appears to be bolded.
| Table 9 - Effect of pH (for 10ng/mL cutoff) | | | | | |
|---------------------------------------------|-------|----------------------|---------------|-----------------------|---------------|
| Test Parameter | Value | -50% Cutoff (5ng/mL) | | +50% Cutoff (15ng/mL) | |
| | | Result | Interference? | Result | Interference? |
| pH | 3.0 | Negative | No | Positive | No |
- g. The following is a summary table of the additional pH 3.0 testing at ±50%
- h. The following is a summary table of the effect of specific gravity results for 10ng/mL cutoff:
| Table 10 - Effect of Specific Gravity (for 10ng/mL cutoff) | | | | | |
|------------------------------------------------------------|-------|------------------------|---------------|-------------------------|---------------|
| Test Parameter | Value | -25% Cutoff (7.5ng/mL) | | +25% Cutoff (12.5ng/mL) | |
| Specific Gravity | 1.000 | Result | Interference? | Result | Interference? |
| Specific Gravity | 1.000 | Negative | No | Positive | No |
| Specific Gravity | 1.002 | Negative | No | Positive | No |
| Specific Gravity | 1.005 | Negative | No | Positive | No |
| Specific Gravity | 1.010 | Negative | No | Positive | No |
| Specific Gravity | 1.015 | Negative | No | Positive | No |
| Specific Gravity | 1.020 | Negative | No | Positive | No |
| Specific Gravity | 1.025 | Negative | No | Positive | No |
| Specific Gravity | 1.030 | Negative | No | Positive | No |
- i. Ascorbic acid, sodium chloride, boric acid and pH of 3 are considered interferences
- 4. Method Comparison Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device. The study verified that the product performance can be verified by Mass Spectrometry. The instrument used for this test was a Beckman Coulter AU 400e and an Agilent 6430 Liquid Chromatography Tandem Mass Spectrometry.
- a. The following is a comparison table of qualitative assay performance for the 10ng/mL cutoff:
Table 11 - Method Comparison (for 10ng/mL cutoff) - Qualitative
| | LC/MS Confirmation | | |
|-------------|--------------------|-----|----|
| | (+) | (-) | |
| Test Device | (+) | 40 | 0 |
| | (-) | 0 | 40 |
- b. The following is a summary table of qualitative assay performance for the 10ng/mL cutoff:
| Type | 6-Acetylmorphine Concentration | | | | Agreement (%) |
|-----------------------|--------------------------------|-----------------|---------------|------------|---------------|
| | < 5ng/mL | 5.0 ~ 9.9 ng/mL | 10 ~ 15 ng/mL | > 15 ng/mL | |
| Qualitative/ Positive | 0 | 0 | 4 | 36 | 100 |
| Qualitative/ Negative | 36 | 4 | 0 | 0 | 100 |
- 5. Stability -
A closed accelerated stability study was performed on reagents, calibrator and controls at 25°C to establish the initial expiration dating. The stability study supported an initial expiration date of 1 year for reagents. This stability study supported an initial expiration date of 12 months for calibrator and controls. The instrument used for this test was a Beckman Coulter AU 400e. Real stability studies are ongoing.
- 6. Calibrator and Control Traceability all components of the calibrator and controls have been traced to a commercially available standard solution.
- 7. Calibrator and Control Stability An open accelerated stability study was performed at 25°C to establish the initial open vial expiration dating. The stability
{11}------------------------------------------------
study supported an initial open vial expiration date of 12 months. The instrument used for this test was a Beckman Coulter AU 400e. All calibrator level(s) (10ng/mL) and control level(s) (7.5ng/mL and 12.5ng/mL) were within specifications for Day 0, 2, 8, 16, 24, 30, 32, and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
- Calibrator and Control Value Assignment calibrator and controls are 8. manufactured and are tested by mass spectrometry. If any of the analytes are out of the acceptable range, then the calibrator and control is adjusted and retested. Values are assigned to the calibrator and controls once the mass spectrometry results are within the acceptable ranges.
- H. Conclusion
The information provided in this pre-market notification demonstrates that the Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use.