The CEDIA™ Heroin Metabolite (6-AM) Assay is a homogeneous enzyme immunoassay for the in vitro qualitative and/or semi-quantitative determination of the presence of heroin metabolite (6-AM) in human urine at a cut-off concentration of 10 ng/mL. The assay is intended to be used in laboratories and provides a rapid analytical screening procedure to detect 6-Acetylmorphine in human urine. The assay is designed for use with a number of clinical chemistry analyzers. This product is intended to be used by trained professionals only. The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or Liquid chromatography/ mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only.
Device Story
CEDIA™ Heroin Metabolite (6-AM) Assay is a homogeneous enzyme immunoassay for urine drug screening. Input: human urine sample. Principle: competitive binding between sample 6-AM and 6-AM-conjugated Enzyme Donor (ED) for antibody sites; free ED-6-AM re-associates with Enzyme Acceptor (EA) to form active β-galactosidase; active enzyme cleaves substrate to produce color change measured spectrophotometrically. Output: qualitative or semi-quantitative 6-AM concentration. Used in clinical laboratories; operated by trained professionals on clinical chemistry analyzers (e.g., Horiba Yumizen C1200). Output used by clinicians to identify specimens requiring confirmatory testing via LC-MS/MS or GC/MS. Benefits: rapid preliminary screening for heroin metabolite presence to guide clinical and professional judgment in drug abuse testing.
Clinical Evidence
No clinical studies were performed. Performance was established via bench testing, including precision (n=80 per concentration), linearity, and method comparison against LC-MS/MS using 123 clinical urine samples. Results demonstrated concordance between the candidate device and the reference method in both qualitative and semi-quantitative modes.
Technological Characteristics
Homogeneous enzyme immunoassay using recombinant DNA-derived β-galactosidase fragments. Reagents include mouse monoclonal antibodies, enzyme donor/acceptor fragments, chlorophenol red-β-D-galactopyranoside substrate, buffers, stabilizers, and detergents. Spectrophotometric detection. Designed for use on clinical chemistry analyzers (e.g., Horiba Yumizen C1200).
Indications for Use
Indicated for the qualitative and/or semi-quantitative detection of heroin metabolite (6-AM) in human urine at a 10 ng/mL cutoff. Intended for use by trained professionals in laboratory settings as a preliminary screening procedure. Not for use as a standalone diagnostic; requires confirmation by GC/MS or LC-MS/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K231007
B Applicant
Microgenics Corporation
C Proprietary and Established Names
CEDIA™ Heroin Metabolite (6-AM) Assay
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DJG | Class II | 21 CFR 862.3650 - Opiate Test System | TX - Clinical Toxicology |
## II Submission/Device Overview:
A Purpose for Submission: New device.
B Measurand:
6-Acetylmorphine (6-AM).
C Type of Test:
Qualitative and Semi-quantitative Homogeneous Enzyme Immunoassay
## III Intended Use/Indications for Use:
A Intended Use(s):
The CEDIA™ Heroin Metabolite (6-AM) Assay is a homogeneous enzyme immunoassay for the in vitro qualitative and/or semi-quantitative determination of the presence of heroin metabolite (6-AM) in human urine at a cut-off concentration of 10 ng/mL. The assay is intended to be used in
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
laboratories and provides a rapid analytical screening procedure to detect 6-Acetylmorphine in human urine. The assay is designed for use with a number of clinical chemistry analyzers. This product is intended to be used by trained professionals only.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or Liquid chromatography/ mass spectrometry (LC-MS/MS) is the preferred confirmatory method.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only.
## B Indication(s) for Use:
See Intended use above
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
Performance characteristics studies were conducted on the Horiba Yumizen C1200 clinical chemistry analyzer.
## IV Device/System Characteristics:
### A Device Description:
CEDIA™ Heroin Metabolite (6-AM) assay is supplied as a two liquid (1 and 2) and two lyophilized reagents (1a and 2a) kit that is available in a kit configuration.
1. EA Reconstitution Buffer: Contains 0.32 mg/L mouse monoclonal antibodies to 6-Acetylmorphine, buffer salts, stabilizer, detergent and preservative
1a. EA Reagent: Contains 0.171 g/L Enzyme Acceptor, buffer salts, detergent and preservative
2. ED Reconstitution Buffer: Contains buffer salts, stabilizer, and preservative
2a. ED Reagent: Contains 16.2 µg/L Enzyme Donor conjugated to 6-Acetylmorphine, 1.67 g/L chlorophenol red-β-D galactopyranoside, stabilizer, detergent, preservative.
### B Principle of Operation:
CEDIA technology uses recombinant DNA technology to produce a unique homogeneous enzyme immunoassay system. The assay is based on the bacterial enzyme β-galactosidase, which has been genetically engineered into two inactive fragments. These
K231007 - Page 2 of 10
{2}
fragments spontaneously re-associate to form fully active enzymes that, in the assay format, cleave a substrate. This generates a color change that can be measured spectrophotometrically.
In the $\mathrm{CEDIA}^{\mathrm{TM}}$ Heroin Metabolite (6-AM), the analyte in the sample competes with 6-AM conjugated to Enzyme Donor (ED) for antibody binding sites. If 6-AM is present in the sample, it binds to antibody, leaving the ED-6-AM conjugate free to re-associate with Enzyme Acceptor (EA) to form active $\beta$ -galactosidase. If no 6-AM is present in the sample, antibody binds to the ED-6-AM conjugate, inhibiting the re-association of inactive - galactosidase fragments, and thus reducing the amount of active enzyme formed. The amount of active enzyme formed, and resultant absorbance change are proportional to the amount of 6-AM present in the sample.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
CEDIA™ Heroin Metabolite (6-AM) assay
B Predicate 510(k) Number(s):
K192943
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K231007 | K192943 |
| --- | --- | --- |
| Device Trade Name | CEDIA™ Heroin Metabolite (6-AM) assay | CEDIA™ Heroin Metabolite (6-AM) assay |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Homogeneous enzyme immunoassay for the in vitro qualitative and/or semi-quantitative determination of the presence of heroin metabolite (6-AM) in human urine at a cut-off concentration of 10 ng/mL. | Same |
| Analyte | 6-Acetylmorphine (6-AM) | 6-Acetylmorphine (6-AM) |
| General Device Characteristic Differences | | |
| Instrument | Horiba Yumizen C1200 | Horiba Pentra C400 |
K231007 - Page 3 of 10
{3}
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3 – Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
All analytical performance studies were conducted on the Horiba Yumizen C1200 clinical chemistry analyzer.
1. Precision/Reproducibility:
The precision study was performed based upon recommendations in CLSI Guideline EP05-A3, at one site with one analyzer and two lots of reagents. Testing was carried out for 20 days with two runs per day, at least two hours apart and two replicates per run in both qualitative and semi-quantitative modes, giving a total of 80 determinants (n = 80). Drug-free negative urine was spiked with 6- Acetylmorphine analyte to final concentrations of -100%, -75%, -50%, -25%, below cutoff, cutoff, and +25%, +50%, +75% and +100%, above cutoff, and the concentrations were confirmed by LC-MS/MS or GC/MS.
For qualitative and semi-quantitative mode results from a representative lot are summarized in the tables below.
Precision Data in Qualitative Mode
| % of Cutoff | Spiked Conc. (ng/mL) | # of Determinants | # Negative/# Positive |
| --- | --- | --- | --- |
| -100 | 0 | 80 | 80 / 0 |
| -75 | 2.5 | 80 | 80 / 0 |
| -50 | 5 | 80 | 80 / 0 |
| -25 | 7.5 | 80 | 80 / 0 |
| 100 | 10 | 80 | 36 / 44 |
| +25 | 12.5 | 80 | 0 / 80 |
| +50 | 15 | 80 | 0 / 80 |
| +75 | 17.5 | 80 | 0 / 80 |
| +100 | 20 | 80 | 0 / 80 |
Precision Data in Semi-Quantitative Mode
| % of Cutoff | Spiked Conc. (ng/mL) | # of Determinants | # Negative/# Positive | Within-run CV (%) | Total-run CV (%) |
| --- | --- | --- | --- | --- | --- |
| -100 | 0 | 80 | 80 / 0 | N/A | N/A |
K231007 - Page 4 of 10
{4}
K231007 - Page 5 of 10
| -75 | 2.5 | 80 | 80 / 0 | 9.5 | 14.5 |
| --- | --- | --- | --- | --- | --- |
| -50 | 5 | 80 | 80 / 0 | 3.9 | 6.2 |
| -25 | 7.5 | 80 | 80 / 0 | 3.8 | 5.8 |
| 100 | 10 | 80 | 31 / 49 | 3.6 | 5.4 |
| +25 | 12.5 | 80 | 0 / 80 | 3.0 | 4.9 |
| +50 | 15 | 80 | 0 / 80 | 3.4 | 4.9 |
| +75 | 17.5 | 80 | 0 / 80 | 2.7 | 3.7 |
| +100 | 20 | 80 | 0 / 80 | 1.9 | 3.3 |
## 2. Linearity:
To demonstrate linearity of the assay throughout the calibration range of 0 to 20 ng/mL, a drug free-urine pool spiked with 6-AM at 20 ng/mL was serially diluted with drug free urine to generate ten intermediate levels. Each sample was run in replicates of five in semi-quantitative mode and the average was used to determine percent recovery compared to the expected target value.
The observed result (y) and the target expected result (x) were compared using the least squares regression method. The regression equation and correlation obtained are:
$$
y = 0.9008x + 0.3445; R^2 = 0.997
$$
The recovery of the samples from the linear range of the assay (2 ng/mL to 20 ng/mL) prepared by dilution ranged from 90.5% to 111.0%.
## 3. Analytical Specificity/Interference:
A. Cross-reactivity (specificity) of the structurally related and unrelated compounds with the performance of the candidate device was evaluated by adding known amounts of each compound to drug-free negative urine. The samples were tested in duplicate with the candidate device in both the qualitative and semi-quantitative modes. Percent cross-reactivity was calculated as follows:
% Cross-reactivity = (Cutoff concentration / Lowest concentration of cross reactant that gives a positive result) x 100
Cross-Reactivity of 6-Acetylmorphine and Heroin:
| 6-Acetylmorphine and Heroin | Tested Concentration (ng/mL) | Assay Result | Cross-Reactivity (%) |
| --- | --- | --- | --- |
| 6-Acetylmorphine | 10 | Pos | 100% |
| Heroin | 300 | Pos | 3% |
Cross-Reactivity of Opiates and Structurally Related Compounds:
{5}
K231007 - Page 6 of 10
| Structurally related compounds and other opiates | Tested Concentration (ng/mL) | Assay Result | Cross-Reactivity (%) |
| --- | --- | --- | --- |
| 6-Acetylcodeine | 50,000 | Pos | 0.02% |
| Buprenorphine | 100,000 | Neg | No Cross-Reactivity |
| Buprenorphine-3β-D-glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Codeine | 100,000 | Neg | No Cross-Reactivity |
| Dextromethorphan | 90,000 | Pos | 0.01% |
| Dihydrocodeine | 100,000 | Neg | No Cross-Reactivity |
| EDDP | 100,000 | Neg | No Cross-Reactivity |
| EMDP | 100,000 | Neg | No Cross-Reactivity |
| Ethylmorphine | 100,000 | Neg | No Cross-Reactivity |
| Fentanyl | 100,000 | Neg | No Cross-Reactivity |
| Hydrocodone | 100,000 | Neg | No Cross-Reactivity |
| Hydromorphone | 20,000 | Pos | 0.04% |
| Hydromorphone-3β-D-glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Levorphanol | 15,000 | Pos | 0.05% |
| Methadone | 100,000 | Neg | No Cross-Reactivity |
| Meperidine | 100,000 | Neg | No Cross-Reactivity |
| Mitragynine | 100,000 | Neg | No Cross-Reactivity |
| 7-Hydroxymitragynine | 100,000 | Neg | No Cross-Reactivity |
| Morphine | 13,500 | Pos | 0.07% |
| Morphine-3β-D-Glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Morphine-6β-D-Glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Nalorphine | 10,500 | Pos | 0.10% |
| Naloxone | 100,000 | Neg | No Cross-Reactivity |
| Naltrexone | 100,000 | Neg | No Cross-Reactivity |
| Norbuprenorphine | 100,000 | Neg | No Cross-Reactivity |
| Norbuprenorphine Glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Norcodeine | 100,000 | Neg | No Cross-Reactivity |
| Norhydrocodone | 100,000 | Neg | No Cross-Reactivity |
| Normorphine | 50,000 | Pos | 0.02% |
| Norpropoxyphene | 100,000 | Neg | No Cross-Reactivity |
| Noroxycodone | 100,000 | Neg | No Cross-Reactivity |
| Noroxymorphone | 100,000 | Neg | No Cross-Reactivity |
| Oxymorphone-3β-D-glucuronide | 100,000 | Neg | No Cross-Reactivity |
| Oxycodone | 100,000 | Neg | No Cross-Reactivity |
{6}
B. Interference: The interference study was performed using one lot of reagents, calibrators and controls. Structurally unrelated compounds were evaluated by adding each substance to 6-Acetylmorphine spiked at low (7.5 ng/mL) and high (12.5 ng/mL) controls at the concentrations indicated. As shown in the tables below, all the pharmacologic compounds evaluated exhibited no cross-reactivity at the concentrations tested.
- Samples were tested in replications of five (n=5) in both qualitative and semiquantitative modes.
| Interferents | Spiked Concentration (ng/mL) | Spiked 6-Acetylmorphine Level | |
| --- | --- | --- | --- |
| | | Low Control | High Control |
| 10,11 Dihydrocarbamazepine | 85,000 | Neg | Pos |
| 11-nor-delta9-THC-COOH | 10,000 | Neg | Pos |
| Acetaminophen | 500,000 | Neg | Pos |
| Acetylsalicylic Acid | 500,000 | Neg | Pos |
| Amitriptyline | 50,000 | Neg | Pos |
| Amoxicillin | 500,000 | Neg | Pos |
| Amphetamine | 100,000 | Neg | Pos |
| Amisulpride | 100,000 | Neg | Pos |
| Benztropine Mesylate | 50,000 | Neg | Pos |
| Benzoylecgonine | 100,000 | Neg | Pos |
| Brompheniramine | 75,000 | Neg | Pos |
| Caffeine | 500,000 | Neg | Pos |
| Captopril | 500,000 | Neg | Pos |
| Chlordiazepoxide | 100,000 | Neg | Pos |
| Chlorpromazine | 10,000 | Neg | Pos |
| Clomipramine | 150,000 | Neg | Pos |
| Chloroquine | 500,000 | Neg | Pos |
| Cimetidine | 500,000 | Neg | Pos |
| Desipramine | 50,000 | Neg | Pos |
| Diazepam | 100,000 | Neg | Pos |
| Digoxin | 100,000 | Neg | Pos |
| Diphenhydramine | 20,000 | Neg | Pos |
| Doxepin HCl | 10,000 | Neg | Pos |
| Enalapril | 500,000 | Neg | Pos |
| Fluoxetine | 500,000 | Neg | Pos |
| Fluophenazine | 500,000 | Neg | Pos |
| Haloperidol | 50,000 | Neg | Pos |
K231007 - Page 7 of 10
{7}
Interference with Endogenous Substances
| Compound | Spiked Concentration (mg/dL) | Spiked 6-Acetylmorphine Level | |
| --- | --- | --- | --- |
| | | Low Control | High Control |
| Acetone | 1000 | Neg | Pos |
| Ascorbic acid | 1500 | Neg | Pos |
| Creatinine | 500 | Neg | Pos |
| Ethanol | 1000 | Neg | Pos |
| Galactose | 10 | Neg | Pos |
| Y-globulin | 500 | Neg | Pos |
| Glucose | 1000 | Neg | Pos |
| Hemoglobin | 300 | Neg | Pos |
| Human serum albumin | 500 | Neg | Pos |
| Oxalic acid | 100 | Neg | Pos |
| Riboflavin | 7.5 | Neg | Pos |
| Sodium Chloride | 6000 | Neg | Pos |
| Urea | 2000 | Neg | Pos |
| Hydroxychlroquine | 100,000 | Neg | Pos |
| --- | --- | --- | --- |
| Hydroxyzine | 125,000 | Neg | Pos |
| Ibuprofen | 500,000 | Neg | Pos |
| Imipramine | 20,000 | Neg | Pos |
| Levothyroxine | 50,000 | Neg | Pos |
| Methamphetamine | 100,000 | Neg | Pos |
| Maprotiline | 500,000 | Neg | Pos |
| Nalbuphine | 100,000 | Neg | Pos |
| Naproxen | 500,000 | Neg | Pos |
| Nortriptyline | 175,000 | Neg | Pos |
| Nifedipine | 500,000 | Neg | Pos |
| Nordiazepam | 60,000 | Neg | Pos |
| Oxazepam | 100,000 | Neg | Pos |
| Perphenazine | 150,000 | Neg | Pos |
| Phencyclidine | 7,500 | Neg | Pos |
| Phenobarbital | 100,000 | Neg | Pos |
| Procyclidine | 150,000 | Neg | Pos |
| Propoxyphene | 25,000 | Neg | Pos |
| Protriptyline | 10,000 | Neg | Pos |
| Ranitidine | 500,000 | Neg | Pos |
| Salicyluric Acid | 500,000 | Neg | Pos |
| Secobarbital | 100,000 | Neg | Pos |
| Sulpiride | 500,000 | Neg | Pos |
| Thioridazine | 250,000 | Neg | Pos |
| Triprolidine | 125,000 | Neg | Pos |
| Verapamil | 500,000 | Neg | Pos |
K231007 - Page 8 of 10
{8}
Effect of pH: Drug free urine spiked with 7.5 or 12.5 ng/mL of 6-Acetylmorphine was adjusted to the indicated pH (3, 4, 5, 6, 7, 8, 9, 10, and 11). Samples were tested in replicates of five (n=5) in both qualitative and semiquantitative modes. The results demonstrated that the tested range of pH did not affect the performance of the candidate test.
## Effect of Specific Gravity:
Urine that spanned a specific gravity range of 1.000–1.030 were spiked with 7.5 or 12.5 ng/mL of 6-Acetylmorphine (6-AM). Samples were tested in replicates of five (n=5) in both qualitative and semiquantitative modes. The results demonstrated that the specific gravity range evaluated did not affect the performance of the candidate test.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The device is traceable to a commercially available standard that was verified by LC/MS-MS or GC/MS.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
See section VII A. 1
## B Comparison Studies:
Method Comparison with Predicate Device: A Method comparison study was performed using one hundred and twenty-three unaltered clinical samples analyzed using the candidate test in one replicate, in both qualitative and semi-quantitative modes. The results were compared to LC-MS/MS (Liquid chromatography-tandem mass spectroscopy). The results obtained in the qualitative and semi-quantitative modes are summarized below.
Qualitative Mode:
| CEDIA Heroin Metabolite (6-AM) Assay | < 50% of Cutoff concentration by LC-MS/MS (< 5ng/mL) | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration as determined by LC-MS/MS) (5 – 9.9 ng/mL) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration as determined by LC-MS/MS) (10 – 15.0 ng/mL) | High Positives (Greater than 50% above cutoff concentration (>15.0 ng/mL) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 6 | 53 |
| Negative | 57 | 7 | 0 | 0 |
K231007 - Page 9 of 10
{9}
Semi-Quantitative Mode:
| CEDIA Heroin Metabolite (6-AM) Assay | < 50% of Cutoff concentration by LC-MS/MS (< 5ng/mL) | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration as determined by LC-MS/MS) (5 – 9.9 ng/mL) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration as determined by LC-MS/MS) (10 – 15.0 ng/mL) | High Positives (Greater than 50% above cutoff concentration (>15.0 ng/mL) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 5 | 53 |
| Negative | 57 | 6 | 1 | 0 |
1. Matrix Comparison: Not applicable.
C Clinical Studies:
1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): None.
D Clinical Cut-Off: Not applicable
E Expected Values/Reference Range: Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K231007 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.