Digital Automatic Blood Pressure Monitor MD2300

K143735 · Grandway Technology (Shenzhen) Limited · DXN · Jul 23, 2015 · Cardiovascular

Device Facts

Record IDK143735
Device NameDigital Automatic Blood Pressure Monitor MD2300
ApplicantGrandway Technology (Shenzhen) Limited
Product CodeDXN · Cardiovascular
Decision DateJul 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The upper arm blood pressure monitor is used for non-invasive measurement and monitoring for adult's arterial blood pressure. You can use it to measure your systolic and diastolic pressure, and pulse rate through an inflatable cuff wrapped around the upper arm. Quickly and easily, storing the results and displaying the progression of readings together with the average.

Device Story

Digital Automatic Blood Pressure Monitor MD2300 is a non-invasive oscillometric device for measuring systolic/diastolic blood pressure and pulse rate. Input: brachial artery pressure oscillations detected via inflatable upper arm cuff. Operation: electric pump inflates/deflates cuff; pressure sensor and electronic control module process signals to derive digital readings. Output: systolic/diastolic pressure and pulse rate displayed on LCD panel; device stores results and calculates averages. Used in clinical or home settings by patients or medical professionals. Benefits: provides quick, easy, and regular blood pressure monitoring to track progression of readings, aiding in hypertension management.

Clinical Evidence

Clinical study conducted per ISO 81060-2:2013 with 100 patients (49 male, 51 female). Standard auscultation method used as reference. Measurements performed alternatively with the MD2300 and auscultation in the same arm. Results demonstrate clinical accuracy equivalent to the predicate device.

Technological Characteristics

Oscillometric measurement; pressure sensor, electric valve, electric pump, CPU, PCB. Materials: ABS plastic, electronic components. Power: 4x1.5V AA batteries or AC adapter. Dimensions/form factor: upper arm cuff. Connectivity: standalone. Software: IEC 62304 compliant. Standards: EN 1060-1/3, IEC 60601-1, IEC 60601-1-2, FCC Part 15, ISO 10993-5/10, IEC 81060-2.

Indications for Use

Indicated for non-invasive measurement and monitoring of arterial blood pressure and pulse rate in adults. Used by medical professionals or home users.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 23, 2015 Grandway Technology (shenzhen) Limited Patrick Chow Block 6 And 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang District, Shenzhen, 518118 CN - Re: K143735 Trade/Device Name: Digital Automatic Blood Pressure Monitor MD2300 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 24, 2015 Received: June 24, 2015 Dear Patrick Chow. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K143735 Device Name Digital Automatic Blood Pressure Monitor MD2300, MD2300 Page 1 of 1 Indications for Use (Describe) The upper arm blood pressure monitor is used for non-invasive measurement and monitoring for adult's arterial blood pressure. You can use it to measure your systolic and diastolic pressure, and pulse rate through an inflatable cuff wrapped around the upper arm. Quickly and easily, storing the results and displaying the progression of readings together with the average. | l ype of Use (Select one or both, as applicable) | | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|--|--|--| | ا Prescription Use (Part 21 CFR 801 Subpart D) | ال Over-The-Counter Use (21 CFR 801 Subpart C) | | | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete | | | | | and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k) Submitter | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED | |--------------------|-----------------------------------------------------------------------------------------------------------------------------| | Address | Block 6 and 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang<br>District, Shenzhen, Guang Dong, People's Republic of China | | Phone Number | (00852)-2851-6789 | | Fax Number | (00852)-2851-6278 | | Contact Person | Mr. Patrick Chow | | Date of Submission | 29-Dec-2014 | # 1. Submitter Identification #### Device Identification 2. | Trade Name | Digital Automatic Blood Pressure Monitor MD2300 | |---------------------|----------------------------------------------------------------------------------------------| | Common Name | Non-invasive Blood Pressure Measurement System | | Classification Name | Non-invasive Blood Pressure Measurement System<br>(CFR 870.1130, Class II, Product Code DXN) | #### Predicate Device 3. | Predicate Device | Digital Automatic Blood Pressure Monitor BPM18 Series | | |------------------|-------------------------------------------------------|--| | Manufacturer | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED | | | 510(k) Number | K133619 | | #### Device Description 4. Digital Automatic Blood Pressure Monitor MD2300 is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel. The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading. The table below illustrate the feature presence in Digital Automatic Blood Pressure Monitor MD2300. | Model | Blood Pressure<br>Measurement | Pulse Rate<br>Measurement | WHO<br>Classification | Irregular<br>Heartbeat | LCD Type | Backlight<br>Illumination | User ×<br>Memory | |--------|-------------------------------|---------------------------|-----------------------|------------------------|--------------------------|---------------------------|------------------| | MD2300 | ✓ | ✓ | ✓ | ✓ | Positive<br>Transmissive | ✓ | $2 \times 60$ | {4}------------------------------------------------ ## 5. Indication for Use The upper arm blood pressure monitor is used for non-invasive measurement and monitoring for adult' s arterial blood pressure. You can use it to measure your systolic and diastolic pressure, and pulse rate through an inflatable cuff wrapped around the upper arm. Quickly and easily, storing the results and displaying the progression of readings together with the average. ## 6. Comparison of Technological Characteristics between New Device and Predicate Devices Digital Automatic Blood Pressure Monitor MD2300 is compared to the predicate device, BPM18 Series (K133619) in the device comparison table below. | Comparison between BPM18 Series and Predicate device | | | | |------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------|------------| | Item | Predicate Device | MD2300 | Comment | | Indication for Use | Digital Automatic Blood Pressure Monitor | The upper arm blood pressure monitor is | Equivalent | | | Series is for use by<br>medical<br>BPM18 | used for non-invasive measurement and | | | | professional or home user. The BPM18 Series | monitoring for adult's arterial blood | | | | is intended to measure the systolic and | pressure. You can use it to measure your | | | | diastolic blood pressure, and pulse rate of an | systolic and diastolic pressure, and pulse | | | | adult individual by using a non-invasive | rate through an inflatable cuff wrapped | | | | technique, in which an inflatable cuff is<br>wrapped around the upper arm. | around the upper arm. Quickly and easily, | | | | | storing the results and displaying the | | | | | progression of readings together with the | | | | | average. | | | Measurement | Non-invasive, Oscillometric | Non-invasive, Oscillometric | Identical | | Method | | | | | IHB Detection | Yes | Yes | Identical | | Patient Population | Adult | Adult | Identical | | BP Measurement | Cuff Pressure: 0 - 300 mmHg | Cuff Pressure: 0 - 300 mmHg | Identical | | Range | Systolic Pressure: 50 - 250 mmHg | Systolic Pressure: 50 - 250 mmHg | | | | Diastolic Pressure: 30 - 200 mmHg | Diastolic Pressure: 30 - 200 mmHg | | | Number of User | 2 independent users | 2 independent users | Identical | | Memory Space | 2 users × 120 memory space | 2 users × 60 memory space | Equivalent | | Resolution of | Blood Pressure: 1 mmHg or 0.1kPa | Blood Pressure: 1 mmHg | Identical | | Measurement | Pulse Rate: 1 beat/ min | Pulse Rate: 1 beat/ min | | | Blood Pressure | ± 3 mmHg or 2% of reading | ± 3 mmHg | Equivalent | | Measurement | | | | | Accuracy | | | | | Pulse Rate | 40 - 180 beats/min | 40 - 180 beats/min | Identical | | Measurement | | | | | Range | | | | | Pulse Rate | ± 5 % of the reading | ± 5 % of the reading | Identical | | Measurement | | | | | Accuracy | | | | | Display Type | LCD | LCD | Identical | | Power Source | 4 × 1.5 V AAA-batteries and/or AC | 4 × 1.5 V AA-batteries and/or AC | Equivalent | | | Adaptor | Adaptor | | | Pressurization | Automatic Inflation | Automatic Inflation | Identical | | Mode | | | | | Deflation Mode | Automatic Exhaust/ Deflation | Automatic Exhaust/ Deflation | Identical | | Operating | Temperature: +5 to +40 °C | Temperature: +5 to +40 °C | Identical | | Condition | Humidity: 15 to 93 % R.H. max | Humidity: 15 to 93 % R.H. max | | | | Atmospheric Pressure: 700 - 1060 hPa | Atmospheric Pressure: 700 - 1060 hPa | | {5}------------------------------------------------ | Comparison between BPM18 Series and Predicate device | | | | |--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Item | Predicate Device | MD2300 | Comment | | Storage and<br>Transportation<br>Condition | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700 - 1060 hPa | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700 - 1060 hPa | Identical | | Material | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU, PCB,<br>cuff ABS button, ABS cabinet, batteries<br>and packaging | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU, PCB,<br>cuff ABS button, ABS cabinet, batteries<br>and packaging | Identical | | Compatibility with<br>Environment and<br>Other Devices | No influence with environment and other<br>device | No influence with environment and other<br>device | Identical | | Applicable<br>Standard | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- EN 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- ISO 10993-5:2009<br>- ISO 10993-10:2010<br>- IEC 62304:2006<br>- IEC 81060-2:2009 | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- EN 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- ISO 10993-5:2009<br>- ISO 10993-10:2010<br>- IEC 62304:2006<br>- IEC 81060-2:2013 | Equivalent | Digital Automatic Blood Pressure Monitor MD2300 is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key components of device are: a pressure sensor, a electric valve and an electronic control module together with an electric pump. The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring blood pressure. ## 7. Clinical and Non-clinical Tests ### Clinical Test Summary Testing to insure clinical accuracy of the device in accordance with ISO 81060-2:2013 as documented in Clinical Test report. One hundred patients (49 males and 51 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2:2013. ### Non-Clinical Test Summary Digital Automatic Blood Pressure Monitor MD2300 has performed several non-clinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows: - EN 1060-1:1995+A2:2009 - EN 1060-3:1997+A2:2009 や - IEC 60601-1:2012 � - ゃ IEC 60601-1-2:2007 - ゃ FCC Part 15 Subpart B - や ISO 10993-5:2009 - ISO 10993-10:2010 や {6}------------------------------------------------ � IEC 62304:2006 All of the clinical and non-clinical testing performed on Digital Automatic Blood Pressure Monitor MD2300 are same as the predicate device. Also, bench testing, IEC 80601-2-30, is conducted to show the performance of Digital Automatic Blood Pressure Monitor MD2300 is equivalent to the predicate device. ### 8. Conclusion Digital Automatic Blood Pressure Monitor MD2300 has the same intended use and same technological characteristics as the predicate device, Digital Automatic Blood Pressure Monitor BPM18 Series (K133619). Moreover both clinical and non- clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Blood Pressure Monitor MD2300 is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%