K143662 · Medagent, Inc. · GEI · Mar 12, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143662
Device Name
Bissinger Monopolar Forceps and Cables
Applicant
Medagent, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation. Do not exceed a maximum output of 2000 Vp of your generator.
Device Story
Bissinger Monopolar Forceps and Cables are surgical instruments used to grasp, dissect, and coagulate tissue. The forceps feature straight or bayonet-shaped handles and are connected to high-frequency electrosurgical generators via attached monopolar cables. The cables conduct electrical power from the generator to the instrument. The system requires a neutral electrode for operation. Used by surgeons in clinical settings, the device facilitates tissue coagulation through monopolar current. The output is controlled by the electrosurgical generator, with a maximum voltage limit of 2000 Vp for forceps and 6250 Vp for cables. The device benefits patients by providing precise tissue management during surgery.
Clinical Evidence
Bench testing only. Testing included electrical safety per EN 60601-1:2006 and EN 60601-2-2:2009, validation of steam sterilization, manual cleaning validation, and biocompatibility assessment per ISO 10993-1.
Technological Characteristics
Reusable monopolar forceps (straight/bayonet) and cables. Materials meet ISO 10993-1 biocompatibility requirements. Electrical safety per EN 60601-1 and EN 60601-2-2. Steam sterilizable. Cables feature dual spring connectors compatible with Erbe, Martin, Berchtold, Bovie, Valleylab, and Codmann units.
Indications for Use
Indicated for grasping, dissecting, and coagulating tissue in surgical procedures. Requires connection to a monopolar electrosurgical generator with a neutral electrode.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
EGON FAULHABER Bipolar Forceps, Non Stick Bipolar Forceps and Monopolar Forceps (K101080)
Black and Black Electrosurgical Cutting and Coagulation Forceps and Electrodes (K072124)
Olsen Medical Electrosurgical Cable/Adapters (K111262)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a row. Above the profiles is a curved line that resembles a wing or a stylized wave. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 12, 2015
Medagent Incorporated Mr. Frank Menean Managing Director 112 Corporate Drive, Unit 1 Portsmouth, New Hampshire 03801
Re: K143662
Trade/Device Name: Bissinger Monopolar Forceps and Cables Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 25, 2014 Received: December 23, 2014
Dear Mr. Menean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "MEDAGENT" in a stylized font, with the letters "ME" in black and the letters "AGENT" in gray. The word is enclosed in parentheses, with the left parenthesis in gray and the right parenthesis in gray. Below the word "MEDAGENT" are the words "Individual · Cooperative · Efficient" in a smaller font.
### Indications for Use Statement
510(k) Number: K143622
Device Name: Monopolar Forceps
Indications for Use:
The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation.
Do not exceed a maximum output of 2000 Vp of your generator.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use _____________ (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in a combination of black and gray letters, with the "MED" in black and the "AGENT" in gray. Below the logo, the words "Individual · Cooperative · Efficient" are written in a smaller font.
... Kok ' ** &
## VOLUME 006
# 510(k) Summary
DATE OF APPLICATION: 2014-01-07
### Type of 510(K) submission: Traditional 510(K) submission
APPLICANT: MEDAGENT Inc. 112 Corporate Drive, Unit 1 Portsmouth, NH 03801 USA Tel.: +1.603.570.2039 Fax: +1.603.570.2055 E-Mail: mail@medagent-usa.com Internet: www.medagent-usa.com Establishment Registration Number: 3007219401
- CONTACT PERSON: Franz Menean Tel.: +49.7463.9954.0 E-Mail: franz.menean@medagent.de
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in bold, gray letters, and it is surrounded by parentheses. Below the word, the words "Individual," "Cooperative," and "Efficient" are written in a smaller font, separated by dots.
#### Device Name 1.
| Trade Names: | Bissinger Monopolar Forceps and Cables |
|----------------------|-----------------------------------------------------------------|
| Common Name: | Monopolar Instruments and Accessoires |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories. |
#### Classification Product Code / Subsequent Code 2.
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|------------------------------------------|----------------------|------------------------------|--------------|--------------|-------------------|
| Monopolar Instruments<br>and Accessoires | Part 878 | General & Plastic<br>Surgery | GEI | 2 | 878.4400 |
#### 3. Prior Submissions
There have been no prior submissions of the subject devices so far.
#### Predicate Device 4.
Bissinger's Monopolar Forceps and Cables are substantially equivalent to the following predicate devices, already cleared by the FDA:
| Predicate Device | 510(k) Number | 510(k) Holder |
|-----------------------------------------------------------------------------------------|---------------|-----------------------------------------------|
| EGON FAULHABER Bipolar Forceps, Non<br>Stick Bipolar Forceps and Monopolar For-<br>ceps | K101080 | EGON FAULHABER Surgical Instruments Pinzetten |
| Black and Black Electrosurgical Cutting and<br>Coagulation Forceps and Electrodes | K072124 | Black & Black surgical Inc. |
| Olsen Medical Electrosurqical Ca-<br>ble/Adapters | K111262 | Olsen Medical |
| Suffer Electrosurqical Cables | K073450 | Suffer Medizintechnik GmbH |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "MEDAGENT" in a stylized font, enclosed in parentheses. The first three letters, "MED", are in black, while the last four letters, "AGENT", are in gray. Below the word, there are three words separated by dots: "Individual", "Cooperative", and "Efficient". The words are in a smaller font size than "MEDAGENT". The overall design appears to be a logo or branding element.
#### Description of the Device 5.
Bissinger's Monopolar Forceps are used in various surgical applications. Therefore they are varying in size and build-up, though the general principle remains unchanged.
By using the firmly attached connecting cable, the monopolar forceps can be connected to high-frequency electrosurgical generators.
#### Monopolar Forceps 5.1.
Monopolar Forceps are used for grasping, dressing and coagulating of tissue. They must be connected to a monopolar output of a high frequency generator by using the attached monopolar cable. The application of monopolar current and connection with a neutral electrode is mandatory.
Bissinger's monopolar forceps either have a straight or bayonet-shaped handle. They are equipped with a firmly attached connecting cable with dual spring connector.
#### Monopolar Cables 5.2.
Bissinger Cables are a line of non-sterile, reusable Monopolar Cables fitting Erbe, Martin, Berchtold, Bovie, Valleylab, Codmann Electrosurgical units. They are designed to conduct electrical power from the output of a high frequency generator to the instrument.
#### Indications for Use 6.
#### 6.1. Monopolar Forceps
The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation.
Do not exceed a maximum output of 2000 Vp of your generator
#### 6.2. .Monopolar Cables
Cables for electrosurgery are designed to conduct electrical power from the output of a high-frequency electrosurgical generator to the electrosurgical instrument.
Do not exceed a maximum output of 6250 Vp of your generator.
#### Technological Characteristics 7.
The technological characteristics of Bissinger's Forceps and Cables are identical or substantial equivalent as the previously cleared predicate devices stated in section 4 in terms of design, dimensions, intended use and materials.
#### 8. Testing
Testing in order to proof safety and effectiveness of Bissinger's Monopolar Forceps and Cables has been performed according to recognized consensus Standards and results are conforming to the respective requirements.
#### 8.1. Electrical Safety
The devices subject to this submission have been tested according to the requirements of EN 6061-1:2006 and EN 60601-2-2:2009.
<sup>001 510</sup>K Summary v.4
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in a sans-serif font, with the "MED" portion in black and the "AGENT" portion in gray. The word is enclosed in parentheses. Below the word, the words "Individual · Cooperative · Efficient" are written in a smaller font.
#### 8.2. Sterilization
The sterilization process has been validated under consideration of recognized standards. Testing shows that the Products can be steam sterilized with a sufficient sterlity assurance level by use of standard sterilization parameters.
#### 8.3. Reprocessing
Reprocessability was tested and validated under consideration of recognized standards.
#### Manual cleaning 8.4.
Manual cleaning validation was performed under consideration of recognized standards.
#### Biocompatibility 9.
All requirements of biocompatibility are met through the composition of the used materials which demonstrate the appropriate levels of biocompatibility for its clinical use. The used materials are also used in many other medical devices and have an established history of safe use and biocompatibility outlined in ISO 10993-1.
### 10. Substantial Equivalence Summary / Conclusion
Based on available 510(k) information provided herein, Bissinger's Monopolar Forceps and Cables are considered substantial equivalent to the predicate devices in terms of indications for use, material, technoloqy, design and performance specifications.
There are no differences between the devices which would raise new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.