The Kinetis™ Total Knee System components are indicated for use in skeletally mature patients, with severe knee pain and disability, undergoing primary surgery for total knee replacement due to: - Rheumatoid arthritis, osteoarthritic, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle - Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis - Moderate valgus, varus, or flexion deformities. MAKO Kinetis" Total Knee System components are indicated for use only with cement and are single use devices.
Device Story
Kinetis™ Total Knee System; patellofemorotibial polymer/metal/polymer semi-constrained cemented knee prosthesis. Components include femoral component, tibial tray, biomimetic tibial inserts (highly cross-linked vitamin E polyethylene), and patella component. Used in primary total knee replacement surgery where anterior and posterior cruciate ligaments are retained. Operated by orthopedic surgeons in clinical/surgical settings. Provides mechanical joint replacement to alleviate pain and disability; restores joint function. Performance evaluated via non-clinical bench testing including fatigue, fixation, locking mechanism strength, micromotion, range of motion, contact area/stress, and wear.
Clinical Evidence
No clinical data. Device evaluated through non-clinical bench testing including tibial baseplate fatigue/fixation, insert locking mechanism strength, micromotion, tibio-femoral range of motion/constraint/contact area/stress, tibial insert fatigue, patello-femoral subluxation/lateral jump height/contact area/stress, femoral fatigue, wear, and tibial eminence strength.
Technological Characteristics
Patellofemorotibial semi-constrained cemented prosthesis. Materials: metal and highly cross-linked vitamin E polyethylene. Design features include femoral component, tibial tray, and patella component. Intended for use with bone cement. Single-use device.
Indications for Use
Indicated for skeletally mature patients with severe knee pain and disability undergoing primary total knee replacement due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle, or moderate valgus, varus, or flexion deformities. For use only with cement; single use.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized depiction of three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
MAKO Surgical Corporation Mr. Jonathan Reeves Senior Regulatory Affairs Specialist 2555 Davie Road Fort Lauderdale, Florida 33317
Re: K143635
Trade/Device Name: The KINETIS™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: December 19, 2014 Received: December 24, 2014
Dear Mr. Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{2}------------------------------------------------
Page 2 – Mr. Jonathan Reeves
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, black letters. Below "MAKO" is the text "SURGICAL CORP." in smaller, black letters. Above the word "MAKO" is a blue, curved shape that resembles a wave or an abstract mountain range.
555 Davie Road • Ft. Lauderdale, FL 27.2044 • Fax 954.927.0 ww.makosurgical.com
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K143635
Device Name: The KINETIS™ Total Knee System
#### Indications for Use:
The KINETIS™ Total Knee System components are indicated for use in skeletally mature patients, with severe knee pain and disability, undergoing primary surgery for total knee replacement due to:
- Rheumatoid arthritis, osteoarthritic, traumatic arthritis, polyarthritis. ●
- Collagen disorders, and/or avascular necrosis of the femoral condyle ●
- o Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis
- Moderate valgus, varus, or flexion deformities. o
MAKO KINETIS™ Total Knee Replacement System components are indicated for use only with cement and are single use devices
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart
C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, black letters. Below "MAKO" is the text "SURGICAL CORP." in smaller, black letters. Above the word "MAKO" is a blue, curved shape that resembles a wave or an abstract mountain range.
2555 Davie Road • Ft. Lauderdale, FL 33317 Phone 954.927.2044 • Fax 954.927.0446 www.makosurgical.com
#### 510(K) SUMMARY
| Submitter: | MAKO Surgical Corp. |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 2555 Davie Road, Fort Lauderdale, FL 33317 |
| Phone number/ Fax Number: | (Ph) 954-628-0665; (F) 954-927-0446 |
| Contact Person: | Jonathan Reeves |
| Date Prepared: | December 19, 2014 |
| Proprietary Name: | The Kinetis™ Total Knee System |
| Common Name: | Total Knee System |
| Classification: | Class II |
| Product Codes/Classification#: | JWH - Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis (888.3560)<br><br>OIY - Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis (888.3560) |
| Reason for 510(k) submission: | New device submission |
## Device Description:
The Kinetis™ Total Knee System is a patellofemorotibial polymer/metal/polymer semiconstrained cemented knee joint prosthesis. The system includes femoral component, tibial tray, biomimetic tibial inserts, patella component, and associated instruments. The biomimetic tibial inserts are manufactured from highly cross-linked vitamin E polyethylene. The system is for use in cases where the anterior and posterior cruciate ligaments are retained.
## Intended Use:
The Kinetis™ Total Knee System components are indicated for use in skeletally mature patients, with severe knee pain and disability, undergoing primary surgery for total knee replacement due to:
{5}------------------------------------------------
- Rheumatoid arthritis, osteoarthritic, traumatic arthritis, polyarthritis. ●
- Collagen disorders, and/or avascular necrosis of the femoral condyle
- Non-inflammatory degenerative joint disease including osteoarthritis, traumatic ● arthritis, or avascular necrosis
- Moderate valgus, varus, or flexion deformities.
MAKO Kinetis" Total Knee System components are indicated for use only with cement and are single use devices.
## Primary Predicate Device:
MAKO Kinetis "10 Total Knee System is substantially equivalent to the following 510(k) cleared device:
| Device Name | Manufacturer | 510(k) # |
|-------------------------|--------------|----------|
| Vanguard XP Knee System | Biomet | K122160 |
#### Additional Predicate Devices:
| Device Name | Manufacturer | 510(k) # |
|--------------------|--------------|----------|
| Townley Total Knee | BioPro | K904448 |
| Axiom Knee | Orthomet | K926334 |
#### References Devices:
| Device Name | Manufacturer | 510(k) # |
|-------------------|--------------|----------|
| Total Knee System | Pipeline | K123692 |
## Technological Characteristics:
The Kinetis™ Total Knee System is similar to legally marketed devices listed previously in that they share the same indications for use, are manufactured from the same or similar material, have same or similar design/technological characteristics, and have performance characteristics adequate to withstand anticipated physiological loading.
## Performance Data:
The Kinetis™ Total Knee System has been evaluated through non-clinical performance testing for;
- Tibial Baseplate Fatigue ●
- Tibial Baseplate Fixation ●
- Insert Locking Mechanism Strength ●
- Tibial Insert / Baseplate Micromotion
- Tibio-Femoral Range of Motion
- Tibio-Femoral Range of Constraint
- . Tibio-Femoral Contact Area and Stress
- Tibial Insert Fatigue
- Patello Femoral Subluxation
- Patello-Femoral Lateral Jump Height
- Patello-Femoral Contact Area and Stress ●
- Femoral Fatigue ●
- Wear .
{6}------------------------------------------------
- Tibial Eminence Strength .
# Conclusions from Clinical and Non-clinical Data:
The results of performance testing indicated the device performed within the intended use and did not raise any new safety and efficacy issues. The device was found to be substantially equivalent to the predicate devices.
Image /page/5/Picture/3 description: The image shows a light blue abstract shape against a white background. The shape is curved and tapers at both ends, resembling a stylized wing or a swooping arc. The color is a soft, pastel blue, and the overall impression is minimalist and clean.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.