The Vanguard™ XP Knee System is intended for replacement of a total knee joint and the preservation of the anterior and/or posterior cruciate ligament (ACL/PCL) when used in conjunction with a femoral, tibial and patellar component.
Device Story
Vanguard XP Knee System is a total knee replacement prosthesis; consists of Co-Cr-Mo femoral components, Co-Cr-Mo tibial trays, and E1 poly tibial bearings. Designed for total knee joint replacement with ACL/PCL preservation. Used by orthopedic surgeons in clinical settings. Components feature porous plasma spray (PPS) titanium alloy coating or Interlok coarse blasted finish. PPS components allow for cemented or uncemented biological fixation; non-coated components require cement. System includes accessory femoral pegs and augments. Provides mechanical stability and joint articulation to restore function in patients with degenerative or traumatic knee conditions.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence was determined based on non-clinical bench testing, including tibiofemoral mechanical stability, static and cyclic locking mechanism tests, wear testing, cyclic fatigue of the tibial tray, surface roughness analysis, constraint and contact area analysis, ROM analysis, and surgical instrument cadaver validations.
Indicated for patients with painful and disabled knee joints due to osteoarthritis, rheumatoid arthritis, or traumatic arthritis involving one or more compartments; correction of varus, valgus, or posttraumatic deformity; and correction or revision of failed osteotomy, arthrodesis, or previous total joint replacement. Porous-coated components are for cemented or uncemented biological fixation; non-coated and all-polyethylene patellar components are for cemented use only; Regenerex components are for uncemented biological fixation only.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a stylized font with each letter enclosed in a box. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font. The logo is black and white.
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Biomet Manufacturing Corp. |
| Address | 56 East Bell Drive<br>Warsaw, IN 46581-0857 |
| Phone number | (574) 267-6639 |
| Fax number | (574) 371-1027 |
| Establishment<br>Registration Number | 1825034 |
| Name of contact<br>person | Tracy Bickel Johnson, RAC |
| Date prepared | January 14, 2014 |
| Name of device | |
| Trade or proprietary<br>name | Vanguard™ XP Knee System |
| Common or usual<br>name | Knee Prosthesis |
| Classification name | Knee joint patellofemorotibial metal/polymer porous-<br>coated uncemented prosthesis (§888.3565)<br>Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis (§888.3560)<br>Knee joint patellofemorotibial semi-constrained UHMWPE<br>pegged uncemented polymer/metal/polymer (§888.3560)<br>Knee joint patellofemorotibial<br>polymer+Additive/metal/polymer +Additive semi-<br>constrained cemented prosthesis (§888.3560) |
| Classification panel | Orthopedic |
| Regulation | 21CFR §888.3565<br>21CFR §888.3560 |
| Product Code(s) | JWH, MBH, MBV, OIY |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | K113550 Vanguard™ Knee System (JWH, MBH, OIY)<br>K102580 Vanguard™ Removable Molded Poly Tibia (JWH)<br>K080528 E1™ Tibial Bearings (OIY, JWH, MBH, MBV)<br>K904448 Townley Total Knee (JWH)<br>K833363 Cloutier II Fiber Metal Total Knee (JWH)<br>K031061 NexGen® Total Knee System (MBH)<br>K050222 Vanguard™ Anterior Stabilized Tibial Bearings |
| | (JWH, MBH)<br>K961021 Kirschner Knee Modified (JWH)<br>K921182 Ultra High Molecular Weight Polyethylene (JWH,<br>HRY)<br>K993159 Maxim Knee System (JWH) |
| Reason for 510(k)<br>submission | Additional Vanguard® product line to include an XP femoral,<br>ACL/PCL and PCL tibial tray, tibial bearings (AS, XP) and<br>associated instrumentation. |
| Device description | The Vanguard XP Knee System is a total knee replacement<br>system that consists of a femoral component composed of<br>Co-Cr-Mo, two styles of tibial trays/plates manufactured of<br>Co-Cr-Mo (with locking bar), and dual bearings machined of<br>E1 poly. Biomet® patellae can be used with the Vanguard<br>XP Knee System. Both the XP femoral and the XP-CR tibial<br>components are available with a previously cleared porous<br>plasma spray (PPS) coating of titanium alloy and Biomet's<br>Interlok® coarse blasted finishes. The Vanguard XP-XP<br>tibial components are available in Biomet's Interlok®<br>coarse blasted finish. Porous coated femoral and tibial<br>components are indicated for cemented and uncemented<br>biological fixation application. Non-coated coarse blasted<br>(Interlok®) femoral and tibial components are indicated for<br>cemented application only. Accessory components are<br>available including removable femoral pegs and femoral<br>augments. |
| Intended use of the<br>device | The Vanguard™ XP Knee System is intended for replacement<br>of a total knee joint and the preservation of the anterior<br>and/or posterior cruciate ligament (ACL/PCL) when used in<br>conjunction with a femoral, tibial and patellar component. |
| Indications for use | Specific indications are as follows:<br>1. Painful and disabled knee joint resulting from<br>osteoarthritis, rheumatoid arthritis, traumatic arthritis where<br>one or more compartments are involved.<br>2. Correction of varus, valgus, or posttraumatic deformity.<br>3. Correction or revision of unsuccessful osteotomy,<br>arthrodesis, or failure of previous total joint replacement<br>procedure.<br>Femoral components and tibial tray components with porous<br>coatings are indicated for cemented and un-cemented |
56 East Bell Drive Warsaw, IN 46582
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for ROMET MANUFACTURING CORP. The word "ROMET" is in a stylized font with each letter outlined. Below the logo is the text "MANUFACTURING CORP." in a simple, sans-serif font.
・
.
- - - -
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Biomet Manufacturing Corp. The word "BIOMET" is in a stylized font with each letter enclosed in a square. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font.
biological fixation application. Non-coated (Interlok®) devices and all polyethylene patellar components are indicated for cemented application only. Regenerex components are intended only for uncemented biologic fixation application.
## Summary of the Technologies
The Vanguard" XP Knee System is made up of multiple components, including: instrumentation, femoral components, several types of bearings and tibial trays.
The technological characteristics of the Vanguard™ XP Knee System are the same as those of predicate devices (K113550, K102580, K080528, K050222, K904448, K833363, and K031061) in terms of design, material, and principle of operation with the exception of slight modifications as described in this 510(k). The femoral and tibial components of the Vanguard™ XP Knee System utilize the identical manufacturing processes as the predicates (K113550, K080528, and K102580). The previously provided porous plasma spray characterization data in K113550 was used in support of this 510(k). The porous plasma spray coating on the ASTM F75 substrate is for both cemented and uncemented, biologic fixation. Non-clinical testing was conducted to demonstrate that the differences did not adversely affect safety and efficacy, and to demonstrate substantial equivalence to the predicate components. All testing met or exceeded the established acceptance criteria. This information is detailed below in the Performance (Non-clinical) section.
## PERFORMANCE DATA
## SUMMARY OF NON-CLINICAL TESTS
### Performance Test Summary-New Device
The following tests were performed for the new Vanguard® XP Knee System:
- . Tibiofemoral Mechanical Stability
- Static Locking Mechanism
- Fixation and Cyclic Locking Mechanism
- Wear Test .
- Cyclic Fatique of Tibial Tray ●
- Surface Roughness Analysis ●
- Tibiofemoral Constraint .
- . Tibiofemoral Contact Area
- Femoral Section Analysis .
- ROM Analysis
- Tibial/Bearing Assembly
- MRI Analysis and Justification
- Surgical Instrument Cadaver Validations
- Bearing Fatique Tests ●
### SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Biomet Manufacturing Corp. The word "BIOMET" is in a stylized font with a registered trademark symbol. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font.
# SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information: N/A
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
No clinical testing was necessary for a determination of substantial equivalence.
The results of mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2013
Biomet, Incorporated % Ms. Tracy Bickel Johnson, RAC Global Regulatory Project Manger 56 East Bell Drive Warsaw, Indiana 46581
Re: K122160
Trade/Device Name: Vanguard™ XP Knee System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH, MBV, OIY Dated: March 8, 2013 . . Received: March 11, 2013
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Tracy Bickel Johnson, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin J. Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Vanguard™ XP Knee System
Indications For Use:
1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, or traumatic arthritis where one or more compartments are involved.
2. Correction of varus, valgus, or posttraumatic deformity.
3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous total joint replacement procedure.
Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok®) devices and all polyethylene patellar components are indicated for cemented application only. Regenerex components are intended only for uncemented biologic fixation application.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopaedic Devices
Page 1 of 1
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.