K143522 · C.R. Bard, Inc. · MAV · Jan 12, 2015 · Cardiovascular
Device Facts
Record ID
K143522
Device Name
Presto Inflation Device
Applicant
C.R. Bard, Inc.
Product Code
MAV · Cardiovascular
Decision Date
Jan 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Presto™ Inflation Device is indicated for use with angioplasty balloon dilatation catheters to create and monitor the pressure in the angioplasty balloon dilatation catheter and to deflate the angioplasty balloon dilatation catheter.
Device Story
Presto Inflation Device is a 30 mL disposable, one-piece manual syringe system used to inflate, monitor, and deflate angioplasty balloon dilatation catheters. Device features a lever-lock piston control, integrated manometer (0-40 atm range), high-pressure connecting tube with male Luer rotating adapter, and a 3-way stopcock. Operated by clinicians in clinical settings (e.g., cath lab) during angioplasty procedures. Clinician manually manipulates the lever-lock to generate pressure, which is displayed on the manometer to guide balloon inflation. Device facilitates controlled balloon expansion to treat vascular lesions. Benefits include precise pressure management and reliable deflation, ensuring procedural safety and efficacy.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via in vitro testing including syringe volume, rated inflation pressure, fatigue testing, ISO Luer/stopcock hub testing, pressure reading accuracy (±1.6 ATM), tubing tensile strength, bubble visualization, and packaging integrity. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity) performed per ISO 10993-1:2009. Shelf life validated per ASTM F1980-07.
Technological Characteristics
Manual 30 mL disposable inflation syringe; rated for 40 atm. Features lever-lock piston, manometer (1 atm increments), high-pressure tubing, and 3-way stopcock. Materials are latex-free and biocompatible per ISO 10993-1. Mechanical sensing principle. Sterilized via standard methods. No software or electronic components.
Indications for Use
Indicated for use with angioplasty balloon dilatation catheters to create, monitor, and deflate pressure within the catheter during angioplasty procedures.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure in profile, repeated three times to symbolize life and growth. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2015
C.R. Bard, Inc. c/o Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K143522
Trade/Device Name: Presto Inflation Device Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: MAV Dated: December 26, 2014 Received: December 29, 2014
Dear Mr. Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Melissa A. Torres -S
For Bram Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143522
Device Name Presto™ Inflation Device
Indications for Use (Describe)
The Presto™ Inflation Device is indicated for use with angioplasty balloon dilatation catheters to create and monitor the pressure in the angioplasty balloon dilatation catheter and to deflate the angioplasty balloon dilatation catheter.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Presto™ Inflation Device 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(K) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc |
|------------|-------------------------------|
| | 1625 West 3rd Street |
| | Tempe, Arizona 85281 |
Phone: 480-638-2939
Fax: 480-449-2542
Contact: Timothy Wade, Regulatory Affairs Specialist
Date November 21, 2014
#### Subject Device Name:
| Device Trade Name: | Presto™ Inflation Device |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Balloon Inflation Syringe (21 CFR 870.1650;<br>Product Code: MAV) |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
#### Predicate Device:
- SCIMED Encore Inflation Devices (K955869; cleared March 22, 1996) .
Image /page/3/Picture/15 description: The image shows the word "BARD" in a stylized, all-caps font. The letters are bold and outlined in black, giving them a distinct and modern appearance. The letter "A" is designed with a unique triangular shape, setting it apart from the other letters. The overall design is clean and minimalist, making the word easily recognizable.
{4}------------------------------------------------
## Device Description:
The Presto™ Inflation Device is an ultra high pressure, large volume inflation device used to inflate, monitor pressure, and deflate angioplasty balloon dilatation catheters. It is a onepiece, 30 mL disposable inflation device rated for 40 atm with a lever-lock design that controls the piston, a manometer, and a high-pressure connecting tube with a male Luer rotating adapter. Also enclosed is a high-pressure 3-way stopcock to aid in preparation and use of the device. The manometer measures pressures ranging from 0 atm up to 40 atm in 1 atm increments. The accuracy of the manometer has been determined to be within ±1.6 ATM. These products are not made with natural rubber latex.
## Indications for Use of Device:
The Presto™ Inflation Device is indicated for use with angioplasty balloon dilatation catheters to create and monitor the pressure in the angioplasty balloon dilatation catheter and to deflate the angioplasty balloon dilatation catheter.
## Comparison of Indications for Use to Predicate Devices:
The indications for use statement for the Presto™ Inflation Device is similar to the predicate device, with the addition of angioplasty for clarification purposes, and does not raise any new issues of safety and effectiveness as demonstrated through the risk analysis process. Therefore, the subject device, the Presto™ Inflation Device, is substantially equivalent to the predicate device.
#### Technological Comparison to Predicate Devices:
The Presto™ Inflation Device has the following similarities to the predicate device:
- . Same intended use
- Similar indications for use
- . Same target population
- Same operating principle ●
- . Similar materials
- . Same fundamental scientific technology
Image /page/4/Picture/17 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The letter "A" is stylized with a horizontal line through the middle, creating a triangular shape at the top. The overall design is simple and modern.
{5}------------------------------------------------
- . Similar packaging materials and configurations
- . Same method of sterilization
The differences between the predicate and subject device include differences in syringe volume, rated inflation pressure reading accuracy, material composition of components, and changes in packaging configuration.
## Performance Data:
To demonstrate substantial equivalence of the subject device, the Presto™ Inflation Device to the predicate device, the technological characteristics and performance criterion were evaluated. Using FDA Guidance Documents and recognized standards on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- . Syringe Volume
- Rated Inflation Pressure .
- Fatigue Testing .
- ISO Luer and Stop Cock Hub Testing .
- Pressure Reading .
- . Tubing Tensile Strength
- . User Validation (Bubble Visualization)
- Syringe Barrel Graduation Markings .
- Packaging .
- o Visual Inspection
- Dye Penetration o
- o Tray Seal Tensile
Product will have a viable shelf life based upon successful completion of testing performed in accordance with ASTM F1980-07, "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices", FDA Guidance, "Shelf Life of Medical Devices", issued April 1991 and the Bard internal stability program. Additionally, biocompatibility tests were performed in accordance with ISO 10993-1: 2009, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process". The Presto™ Inflation Device, when connected to an angioplasty balloon dilatation catheter, is a closed
Image /page/5/Picture/21 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and outlined in black. The letters are evenly spaced and the word is centered in the image. The overall impression is one of strength and simplicity.
{6}------------------------------------------------
system and does not deliver contrast/saline to the circulatory system. Successful testing for Cytotoxicity, Sensitization and Intracutaneous Reactivity show the devices are biocompatible for their intended use. Additionally, pyrogenicity testing is conducted to support labeling claims.
The results from these tests demonstrate that the technological characteristics and performance criteria of the Presto™ Inflation Device is comparable to the predicate devices and that it can perform in a manner equivalent to balloon inflation devices currently on the market with the same intended use.
# Conclusions:
The subject device, the Presto™ Inflation Device, met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Presto™ Inflation Device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.