K143432 · Cardio-Phoenix, Inc. · DPS · Jun 3, 2015 · Cardiovascular
Device Facts
Record ID
K143432
Device Name
Cardio-TriTest
Applicant
Cardio-Phoenix, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jun 3, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Indications for Use
The intended use of this device is to acquire and record 3 different types of heart bio-signals (ECG, PCG, MCG) and combine the results into a contiguous presentation as an aid to diagnostic interpretation by a Physician in a clinical setting.
Device Story
Device acquires synchronized 12-lead ECG, phonocardiogram (PCG), and mechanocardiogram (MCG) signals; uses 4 electronic stethoscopes and 4 MCG sensors in a harness positioned on thoracic wall; connects via USB to PC-based software; displays signals on a contiguous timeline for physician review; aids diagnostic interpretation of heart symptoms; intended for clinical setting use by physicians or trained personnel; provides visual correlation of electrical and mechanical heart activity; does not provide automated diagnosis.
Clinical Evidence
Bench testing only. Device underwent safety, environmental, and performance testing, including software validation for each functional aspect. Compliance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-1-6, EC53, ISO 13485, and ISO 14971.
Technological Characteristics
3-in-1 system: 12-lead ECG, PCG, and MCG. Uses 4 electronic stethoscopes and 4 MCG sensors with Riester diaphragms/non-chill rings. Connectivity via USB 2.0 to Windows 7/8 PC. Digital filtering: 4 ECG modes, 1 PCG mode, 1 BCG/MCG mode. Standards: IEC 60601-1, IEC 60601-2-25, ISO 13485, ISO 14971.
Indications for Use
Indicated for males and females >= 20 years of age as a standard 12-lead diagnostic ECG, an aid to identify cardiac cycle events, and an aid to detect S1 & S2 heart sounds and murmurs. Not for pediatric use or intensive care unit patients. Requires physician over-read and consideration of other patient data.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
DIGITAL BALLISTOCARDIOGRAPH, MODEL DGB 300 (K081603)
3M Littmann Electronic Stethoscope, Model 4000 (K003723)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2015
Cardio-Phoenix Inc. Marc Bisnaire, President 2102-65 Spring Garden Ave Toronto, Ontario M2N 6H9 Canada
Re: K143432
Trade/Device Name: Cardio-TriTest Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DXR, DQC Dated: April 22, 2015 Received: April 30, 2015
Dear Marc Bisnaire:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K143432
Device Name Cardio-TriTest
Indications for Use (Describe)
The intended use of this device is to acquire and record 3 different types of heart bio-signals (ECG, PCG, MCG) and combine the results into a contiguous presentation as an aid to diagnostic interpretation by a Physician in a clinical setting.
The indication for use is as ...
a. a Standard 12 Lead Diagnostic ECG b. an aid to identify events in the cardiac cvcle c. an aid to detect S1 & S2 hearts sounds and murmurs.
Not intended for use in pediatrics. Not intended for patients in the intensive care units. Intended population: Males and females greater than or equal to 20 years of age.
Any diagnostic interpretation is only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
The Cardio-TriTest is on a prescription basis by a Certified Medical Practitioner.
Type of Use (Select one or both, as applicable)
즈 Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### The assigned 510(k) number is: K143432
| Submitter of 510(k): | Cardio-Phoenix Inc.<br>2102-65 Spring Garden Ave<br>Toronto, M2N 6H9<br>Canada<br>Phone: 416-595-0795<br>Email: marc.bisnaire@cardiophoenix.com |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Marc Bisnaire |
| Date of summary: | November 5th, 2014 |
| Trade/Proprietary name: | Cardio-TriTest™ |
| | |
| Classification name: | EPMCG v5.5 |
| Classification: | DPS – Electrocardiograph<br>DXR – Ballistocardiograph<br>DQC – Phonocardiograph |
### The following device classifications apply to this device:
| Name | Class | CFR |
|---------------------|-------|----------|
| Electrocardiograph | | 870.2340 |
| Ballistocardiograph | | 870.2320 |
| Phonocardiograph | | 870.2390 |
### Classification panel:
Cardiovascular
{4}------------------------------------------------
### Legally marketed predicate devices:
- DIGITAL BALLISTOCARDIOGRAPH, MODEL DGB 300 (K081603)
- Zargis Acoustic Cardioscan-ZAC (K031517) ●
#### Reference device:
- 3M Littmann Electronic Stethoscope, Model 4000 (K003723)
### Device Description
The Cardio-TriTest is three devices in one, combining a 12-Lead Electrocardiograph (ECG), a Phonocardiograph and a Mechanocardiograph. The device will acquire all three types of bio-signals during the same non-invasive test procedure. The three types of signals are synchronized and outputted on the same timeline making it easier and more visual for general practitioners to visually determine a patient's current heart symptoms.
The ECG component is a Standard 12-Lead ECG, conformant with IEA60601-2-25 standards.
### The Cardio-TriTest device for catching ECG signals uses Standard 12-Lead FDA/CE approved ECG cables by Sino-Hero Company.
The PCG component consists of 4 electronic stethoscopes combined into one Phono recording device.
### PCG/MCG sensors are with FDA/CE approved diaphragms and soft non chill rings manufactured by the Riester Company from Germany. More details:
### http://www.riester.de/cardiophon.431.0.html?&L=79195
The MCG component consists of 4 MCG recording devices.
#### The MCG signal validation is in the Annex B of this application.
The PCG and MCG sensors are housed in common sensor housing for the purpose of being positioned/located on the thoracic wall in the same four (4) primary and standard auscultation points.
The combined PCG/MCG sensors are correctly positioned on the four primary auscultation points on the thoracic wall via a purpose designed harness.
Cardio-Phoenix Inc.
QCL2 v1.1 Page
{5}------------------------------------------------
#### Intended use:
The intended use of this device is to acquire and record 3 different types of heart biosignals (ECG, PCG, MCG) and combine the results into a contiguous presentation as an aid to diagnostic interpretation by a Physician in a clinical setting.
#### Indication for use
The indication for use is as ...
- a Standard 12 Lead Diagnostic ECG a.
- b. an aid to identify events in the cardiac cycle
- an aid to detect S1 & S2 hearts sounds and murmurs. C.
Not intended for use in pediatrics.
Not intended for patients in the intensive care units.
Intended population:
Any diagnostic interpretation is only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
The Cardio-TriTest is on a prescription basis by a Certified Medical Practitioner.
#### Technological characteristics and substantial equivalence:
The Cardio-TriTest (CTT) is a 3-in-1 device. It combines the functionality of an Electrocardiograph (ECG), a Phonocardiograph (PCG) and Mechanocardiograph (MCG) device into 1 device.
Cardio-Phoenix Inc.
{6}------------------------------------------------
The value of the CTT is that it synchronizes the 3 different types of bio-signals recorded by each functional aspect of the device and visually correlates them for combined diagnostic purposes.
The technical characteristics of each functional device are either conformant to FDA approved standards in the case of the ECG functionality or substantially equivalent to their corresponding FDA approved predicate devices for the PCG and MCG functionality.
SAFETY: Being a 3-in-1 device, the safety of the device meets the Standard as per IEA60601-1, in accordance with the standards set in 60601-2-25 for Diagnostic ECG and applies equally to the two other functional aspects of the device.
In accordance with FDA guidelines, the ECG component is conformant with IEA 60601-2-25 (2014) Diagnostic ElectroCardiographs (ECG), (as such no comparison chart is produced with the other predicate devices as they do not include any ECG functionality).
The MCG functionality, acquires and records the mechanical vibrational waveforms produced by the hearts constructions and transmitted to the chest wall is substantially equivalent to the DIGITAL BALLISTOCARDIOGRAPH, MODEL DGB 300.
The PCG functionality, acquires and records the acoustic waveforms produced by the heart and is substantially equivalent to the Zargis Acoustic Cardioscan-ZAC.
### The PCG signal comparison document to the Reference device is in Annex A of this application.
## The MCG signal comparison document to the CTT PCG part is in Annex B of this application.
The Cardio-TriTest (CTT) device and the predicate devices listed above are indicated for use in the clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. They are not intended as a sole means of diagnosis. The interpretations of heart signals offered by CTT are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
Cardio-Phoenix Inc.
QCL2 v1.1 Page
{7}------------------------------------------------
### Technological characteristics comparison table
| Features and<br>functions | DIGITAL<br>BALLISTO<br>CARDIOGRAPH,<br>MODEL DGB 300 | Zargis Acoustic<br>Cardioscan (ZAC) | Cardio-TriTest<br>v5.5 |
|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------|
| Device Type | BCG1 | PCG | ECG/PCG/<br>MCG(BCG) |
| Software | | | |
| Test stored as<br>individual files | YES | YES | YES |
| Save tests | YES | YES | YES |
| File retrieval | From Hard Drive<br>or external drive | From Hard drive | From Hard drive |
| Operating system | Windows XP,<br>Windows Vista | Windows XP SP2<br>or later or Vista<br>32-bit | Windows 7,<br>Windows 8.x<br>(32 or 64bit) |
| PC-based software | YES | YES | YES |
| Printing | YES | YES | YES |
| Networked | NO | N/A | NO |
| Modem data transfer | NO | N/A | NO |
| Display sounds in<br>time and frequency<br>domain | N/A | YES | YES - in time<br>domain |
| Recording test | | | |
| Saving signals just of<br>acceptable quality | N/A | YES | YES |
| Reporting errors<br>during recording | N/A | YES – voice<br>prompt | YES – written<br>prompt |
| Indication of the<br>failure cause and<br>suggested corrective<br>action | N/A | YES | YES |
| Re-recording ability | N/A | YES | YES |
| ECG -BCG<br>Synchronized Stacked<br>Display | Yes (ECG &<br>3 Axis BCG) | N/A | Yes (ECG & PCG &<br>MCG(BCG)) |
| Rhythm Strip | YES | N/A | YES |
| ECG | No, Single Lead<br>Rhythm Strip | N/A | YES, 12-lead ECG |
| X axis BCG | YES | N/A | YES |
| Y axis BCG | YES | N/A | YES |
| Z axis BCG | YES | N/A | YES |
| XYZ combined | NO | N/A | YES |
| Preset recording time | User configurable<br>10, 30, 60 sec | User configurable<br>1 – 20 sec | User configurable<br>30, 60, 120, 180,<br>300 seconds |
| Pre exercise test | YES | N/A | YES |
| Post exercise test | YES | N/A | N/A |
| Data recording | N/A | standard .wav and<br>Native .zac file | 24bit binary and<br>zipped |
| Acoustic sensor | N/A | Electronic<br>stethoscope | Electronic<br>stethoscope |
| Number of sensors | N/A | 1 | 4 |
| Waveform review | | | |
| Waveform markers | Yes, User Applies<br>Markers for ACC<br>recognized waves<br>(Mitral Valve Close,<br>J, Mitral Valve<br>Open, Aortic<br>Value Open, Aortic<br>Valve Close, Early<br>Diastole) | N/A | NO<br>Not required for<br>Intended Use |
| Scroll BCG | YES | N/A | YES |
| Zoom BCG | YES | N/A | YES |
| Measurement | YES | N/A | YES |
| Signal averaging | YES | N/A | NO |
| Zoom option | YES | YES | YES |
| Scroll option | YES | YES | YES |
| Digital filtering | N/A | YES – 3 modes:<br>Bell mode,<br>Diaphragm mode<br>and Extended<br>mode | YES<br>4 kind of ECG<br>filter, 1 PCG, 1<br>BCG/MCG |
| Heart sound analysis | YES | YES | NO |
| Lung sound analysis | NO | NO | NO |
| Number of display<br>panels | N/A | 4 | 3 |
<sup>1</sup> References to BCG and MCG refer to the same accelerometer generated signal but a different naming convention is used, Other names include Vibrocardiograph (VCG), Seismocardiograh (SCG) and all are substantially equivalent to MCG. The literature is confusing as they are often used interchangeably.
Cardio-Phoenix Inc.
{8}------------------------------------------------
Cardio-Phoenix Inc.
QCL2 v1.1 Page
{9}------------------------------------------------
| Identifying presented<br>specific heart sounds | N/A | S1, S2 | S1, S2 |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------|--------------------------------------------|
| Graphically displays<br>heart wave intensity,<br>timing and location | N/A | YES | NO |
| Spectrogram view | N/A | YES | NO |
| Wavelet view | N/A | YES | NO |
| Hardware | | | |
| Portable unit | Portable hand held<br>unit | N/A | Transportable<br>unit |
| Battery power | YES | N/A | NO |
| Tri-axial<br>accelerometer | YES | N/A | YES |
| ECG Lead Wires | No (ECG<br>adhesive patches<br>applied to sensor<br>and patient) | N/A | YES |
| Data transmitting | Wireless Bluetooth<br>transmission from<br>dBG 300 device to<br>dBG Software on a<br>PC | N/A | USB 2.0 |
| PC connection – input<br>type | N/A | USB wireless<br>dongle | USB port |
| Display requirement | N/A | 1024 x 600<br>graphic display or<br>better | 1024 x 600<br>graphic display or<br>better |
| Connectivity | N/A | USB 1.1 port or<br>better | USB 2.0 + |
| Bluetooth | N/A | YES | N/A |
Any differences in technical characteristics, such as differences in Windows OS systems, type and number of sensors, preset recording time, data transmission type, power supply type, recording format, number of display panels, PC connection type do not raise any significant concern with respect to the safety or on the effectiveness of the operating results.
### Safety and performance
The Cardio-TriTest is a 3-in-1 device. The criteria for safety is governed best by the ECG functionality which is subject to FDA approved safety standard set out in
{10}------------------------------------------------
IEA60601-1. The Cardio-TriTest device has passed all the required safety standards and has also been subjected to extensive safety, environmental and performance testing.
Final testing for the device included various performance tests, including software validation corresponding to each functional aspect of the device, to ensure that all safety and performance specifications were met.
The Cardio-TriTest device has also been tested to assure compliance to the requirement of various published standards including the following:
- IEC 60601-1
- IEC 60601-1-2 ●
- IEC 60601-2-25 ●
- IEC 60601-1-6 ●
- EC53 ●
- ISO 13485
- ISO 14971
### Conclusion:
Based upon the each predicates corresponding indications for use, their technological characteristics and their safety and performance testing, the Cardio-TriTest has been shown to be substantially equivalent to the corresponding above listed legally marketed predicate devices under the Federal Food, Drug, and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.