K143155 · Johnson & Johnson Healthcare Products, Division of Mcneil-Pp · LBH · Jan 29, 2015 · Dental
Device Facts
Record ID
K143155
Device Name
LISTERINE Sensitivity Defense Mouthrinse
Applicant
Johnson & Johnson Healthcare Products, Division of Mcneil-Pp
Product Code
LBH · Dental
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Relieves tooth sensitivity. Provides continual relief from tooth sensitivity due to cold, heat, acids, sweets, or contact. Over-the-counter use.
Device Story
Listerine® Sensitivity Defense™ Mouthrinse (LSDM) is a formulated oral rinse containing Dipotassium oxalate. Used twice daily by patients at home to manage dentinal hypersensitivity. Mechanism involves physical occlusion of dentinal tubules by forming a thin film; this blocks hydrodynamic flow, preventing external stimuli (cold, heat, acids, sweets, contact) from triggering nerve pain within the tooth pulp. Device provides continual relief through this physical sealing action. Healthcare providers recommend the product for patients reporting sensitivity; patients self-administer via rinsing. Benefits include reduced pain response to common triggers and improved quality of life for patients with sensitive teeth.
Clinical Evidence
Multi-center, double-blind, randomized, parallel-group study (n=375; 186 placebo, 189 test). Subjects had qualifying sensitivity to air blast and tactile stimuli. Evaluated at baseline, 2, and 4 weeks. Results showed statistically significant improvement in primary and secondary measures for the test group compared to placebo at 2 and 4 weeks. Adverse events were mild to moderate, self-limiting, and resolved without treatment; no serious adverse events reported.
Technological Characteristics
Formulated oral rinse containing Dipotassium oxalate. Mechanism: physical occlusion of dentinal tubules. Biocompatibility testing per ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 7405:2008. Non-clinical bench testing includes scanning electron micrographs and hydraulic conductance dentin permeability models.
Indications for Use
Indicated for patients experiencing tooth sensitivity due to cold, heat, acids, sweets, or contact. For over-the-counter use.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2015
Johnson & Johnson Healthcare Products Division of McNeil-PPC, Inc. C/O Ms. Angelina M. Hunt, RAC Director, Regulatory Affairs 199 Grandview Road Skillman, New Jersey 08558
Re: K143155
Trade/Device Name: Listerine® Sensitivity Defense™ Mouthwash Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity varnish Regulatory Class: II Product Code: LBH Dated: October 31, 2014 Received: November 3, 2014
Dear Ms. Hunt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### SECTION 4 - INDICATIONS FOR USE STATEMENT
510(k) Number, if known: K143155 Device Name: Listerine® Sensitivity Defense Mouthrinse
#### Indications for Use:
Provides continual relief from tooth sensitivity due to cold, heat, acids, sweets, or contact. Over-the-counter use.
Prescription Use OR Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# SECTION 5 - 510(k) SUMMARY
#### I. Submitter Information
| Name: | Johnson & Johnson Healthcare Products<br>Division of McNeil-PPC, Inc. |
|----------|-----------------------------------------------------------------------|
| Address: | 199 Grandview Rd.<br>Skillman, NJ 08558 |
| Contact: | Angelina M. Hunt, RAC |
|------------|-----------------------|
| Phone: | (908) 874-2943 |
| Facsimile: | (908) 904-3712 |
| Email: | ahunt2@its.jnj.com |
Date Prepared: October 30, 2014
# II. DEVICE
| Name of Device: | Listerine® Sensitivity Defense™ Mouthrinse |
|-----------------------|--------------------------------------------|
| Common or Usual Name: | Mouthwash / Oral Rinse |
| 510(k) Number: | K143155 |
| Regulatory Class: | II |
| Classification Name: | Cavity Varnish (21 CFR 872.3260) |
| Product Code: | LBH |
# III. PREDICATE DEVICE
| Predicate Device: | Colgate® Desensitizing Mouthwash | | |
|-------------------|---------------------------------------------|--|--|
| 510(k) Number: | K140481 | | |
| Applicant: | Colgate-Palmolive Company | | |
| Predicate Device: | Oralief™ OTC Toothpaste for Sensitive Teeth | | |
| 510(k) Number: | K080228 | | |
| Applicant: | NovaMin Technology, Inc. | | |
To our knowledge, the predicate devices have not been subject to a design-related recall.
| Reference Device: | Crest® Sensi-Stop Sensitivity Relief Strips (Sylphar<br>Remesense) |
|-------------------|--------------------------------------------------------------------|
| 510(k) Number: | K132426 |
## COMPANY CONFIDENTIAL
{4}------------------------------------------------
Applicant:
Sylphar N.V.
To our knowledge, the reference device has not been subject to a design-related recall.
# IV. Device Description
Listerine® Sensitivity Defense™ Mouthrinse (LSDM) is a formulated device containing Dipotassium oxalate directed for twice-daily use to provide continual dentinal sensitivity relief. The oxalate formula relieves this painful tooth sensitivity by physically occluding dentinal tubules, preventing hydrodynamic flow, thereby preventing the painful stimulation of nerves within the tooth.
# V. INDICATIONS FOR USE
Intended Use: Relieves tooth sensitivity.
Indications for Use: Provides continual relief from tooth sensitivity due to cold, heat, acids, sweets or contact. For over-the-counter use.
The Indications for Use statement for the LSDM is the same as the listed predicate devices. They are generally all indicated for providing relief from tooth sensitivity due to cold, heat, acids, sweets or contact. Additionally, the subject device and predicate devices are available over-the-counter for home use and have the same intended use for the relief of tooth sensitivity.
# VI. COMPARISON OF TECHNICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The intended use, oral application and fundamental technology of the Listerine Sensitivity Defense Mouthrinse device is substantially equivalent to the predicate devices, Colgate Desensitizing Mouthwash and Oralief toothpaste for Sensitive Teeth. These devices contain a formulated occlusion technology that blocks the hydrodynamic flow within the open tubules. Open dentin tubules allow the fluid to transmit external stimuli to the nerves within the dentin pulp and to trigger a pain response, resulting in dentinal hypersensitivity. The components in the LSDM device physically form a thin film that acts like a seal, thereby physically restricting fluid movement through the dentin tubules and preventing external stimuli from triggering the pain response. The physical mechanism of action is responsible for the product's ability to achieve its intended purpose.
Technological characteristics differ in each formula composition between the subject device and the predicates; however, the fundamental scientific technology responsible for how the device achieves its intended use is the same.
{5}------------------------------------------------
| Device | Classification<br>Name | Intended<br>Use | Technical<br>Composition/<br>Occluding<br>Component | Fundamental<br>Technology /<br>Mechanism<br>of Action | Usage<br>Form |
|---------------------------------------------------------------------------------------------------------------------|------------------------|-------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------|-------------------------------|
| Listerine®<br>Sensitivity<br>Defense™<br>Mouthrinse<br>(LSDM)<br>(K143155) | Cavity<br>Varnish | Relief<br>from tooth<br>sensitivity | Dipotassium<br>Oxalate | Occlusion of<br>dentinal<br>tubules | Mouthwash/<br>Oral rinse |
| Predicate<br>Colgate®<br>Desensitizing<br>Mouthwash<br>(K140481) | Cavity<br>Varnish | Relief<br>from tooth<br>sensitivity | Arginine | Occlusion of<br>dentinal<br>tubules | Mouthwash/<br>Oral rinse |
| Predicate<br>Oralief™ OTC<br>Toothpaste for<br>Sensitive Teeth<br>(K080228) | Cavity<br>Varnish | Relief<br>from tooth<br>sensitivity | Calcium<br>sodium<br>phosphosilicate<br>(Novamin<br>Technology) | Occlusion of<br>dentinal<br>tubules | Non-<br>aqueous<br>Toothpaste |
| Reference<br>Device<br>Crest® Sensi-<br>Stop<br>Sensitivity<br>Relief Strips<br>(Sylphar<br>Remesense)<br>(K132426) | Cavity<br>Varnish | Relief<br>from tooth<br>sensitivity | Oxalic Acid,<br>Potassium Salt | Occlusion of<br>dentinal<br>tubules | Gel strips |
Table1: Intended Use and Technical Characteristic Comparison
{6}------------------------------------------------
# PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The biocompatibility evaluation for the LSDM device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standards (ISO) ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," and ISO-7405:2008 - Dentistry - Preclinical evaluation and biocompatibility of medical devices used in dentistry, as recognized by the FDA.
Biocompatibility results indicate that all components in the LSDM device comply with the ISO standards:
- ISO-10993-10:2010 Maximization Test for Delayed-Type Hypersensitivity -
- ISO-10993-10:2010 Oral Mucosa Irritation Test -
- ISO-10993-5:2009 Cytotoxicity -
## Bench Performance Testing
The LSDM device was tested in in-vitro, using recognized methods for dentin occlusion including scanning electron micrographs. Hydraulic conductance testing of the subject device using Dentin Permeability model demonstrated physical occlusion of dentin tubules compared to similar device controls based on various occlusion technologies.
## Clinical Performance Testing
A multi-center, double-blind, randomized, parallel-group study, controlled clinical study was conducted to determine the safety and effectiveness for the LSDM device in reducing and controlling dentinal hypersensitivity.
375 subjects were evaluated with 186 in the placebo group and 189 in the test product group. The subjects had at least two teeth that satisfied the qualifying response to stimuli for sensitivity measures (air blast and tactile). Evaluations occurred at baseline, 2 and 4 weeks. Results indicated that the test product group provided a statistically significant improvement relative to the placebo group for all primary and secondary measures at 2 and 4 weeks. In summary, in this study, the LSDM device was demonstrated to be superior to the placebo group in both the primary and the key secondary efficacy endpoints.
During the study, 7 of the subjects in the test group and 8 in the placebo control group experienced self-limiting adverse events which all resolved. All of the adverse events considered related to the test product group were mild to moderate and resolved without treatment or a change in study product usage. These were not unexpected adverse events for an oral rinse product. There were no serious adverse events during the study.
## COMPANY CONFIDENTIAL
{7}------------------------------------------------
The results of the study conclude that the LSDM device showed statistically significant reductions in hypersensitivity at 2 and 4 weeks with twice-daily use supporting device effectiveness. All mild to moderate adverse events in the study groups resolved without treatment or change in study product usage and no serious adverse events were observed.
#### Summary
This study demonstrated that the LSDM device used twice daily as an adjunct to toothbrushing provided significantly greater efficacy in reducing and controlling dentinal hypersensitivity compared to a placebo mouth rinse. There were no serious adverse events during the study and the LSDM device was well tolerated. The LSDM device has been tested in non-clinical and clinical testing and has been shown to be safe and effective for its intended use.
#### VII. Conclusion
Based on the results of non-clinical and clinical testing on the LSDM device, the mouthrinse device is at least as safe and effective as the predicate devices through toxicological safety assessments, stability and in-vitro and clinical performance testing, therefore supporting that the LSDM device meets its intended use and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.