SYLPHAR REMESENSE SENSITIVITY RELIEF STRIPS

K132426 · Sylphar N.V. · LBH · Feb 14, 2014 · Dental

Device Facts

Record IDK132426
Device NameSYLPHAR REMESENSE SENSITIVITY RELIEF STRIPS
ApplicantSylphar N.V.
Product CodeLBH · Dental
Decision DateFeb 14, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3260
Device ClassClass 2
AttributesTherapeutic

Intended Use

Sylphar Sensitivity Relief Strips are intended to aid in the relief of dentinal sensitivity resulting from cold, heat, acids, or sweets.

Device Story

Remesense Sensitivity Relief Strips consist of polyethylene strips (30 x 10 mm) coated with a gel containing dipotassium oxalate. Patient applies strips directly to sensitive teeth. Mechanism involves dipotassium oxalate forming crystals that occlude open dentinal tubules; this blocks external stimuli (heat, cold, acids, sweets) from reaching dental nerves, thereby reducing hypersensitivity. Device is for OTC use. Benefits include targeted relief of tooth sensitivity. Safety testing addressed potential choking risk from strips.

Clinical Evidence

Bench testing only. In vitro study using 'Pashley' model compared flow reduction in human dentin between subject device, predicate device, and dipotassium oxalate solution. Results showed no significant difference in flow reduction between subject and predicate. Placebo gel and water controls showed no flow reduction. Elemental analysis confirmed oxalate-based occlusion formation. Usability study confirmed patient comprehension of labeling and instructions. Third-party laboratory study assessed choking risk.

Technological Characteristics

Polyethylene strips (30 x 10 mm) coated with dipotassium oxalate gel. Gel ingredients: glycerin, aqua, potassium oxalate, carbopol, carboxy-methyl-cellulose (CMC), potassium sorbate, sodium benzoate, sodium hydroxide. Mode of action: chemical tubule occlusion. Biocompatibility testing performed per ISO 10993-1, -5, -10 and ISO 7405.

Indications for Use

Indicated for patients suffering from dentinal sensitivity caused by cold, heat, acids, or sweets. For over-the-counter use.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132426 FEB 1 4 2014 ## Attachment 9 # 510(k) Summary in accordance with 21 CFR 807.92(c) | Device Name: | Sylphar Remesense Sensitivity Relief Strips | | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Type of 510(k) submission: | Traditional | | | Date of Submission: | 29 August 2013 | | | Manufacturer: | Sylphar N.V.<br>Xavier de Cocklaan 42<br>B-9831 Deurle<br>Belgium | | | FDA Registration Number: | 3004847139 | | | 510(k) Owner: | Sylphar N.V.<br>Xavier de Cocklaan 42<br>B-9831 Deurle<br>Belgium | | | Phone:<br>Fax: | +32 (0)9 321 70 80<br>+32 (0)9 321 70 90 | | | FDA Registration Number: | 3006673777 | | | Owner/Operator Number: | 10034642 | | | 510(k) Submitter and Contact: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy<br>Phone: +39 06 578 2665<br>Fax: +30 06 574 3786<br>Email: rgray@donawa.com | | | FDA Product Code: | LBH | | | FDA Regulation Number: | 872.2360 | | | FDA Classification Name: | Varnish, Cavity | | | Classification Panel: | Dental | | | Common Name: | Tooth Desensitizer | | | FDA Panel: | Dental | | | FDA Classification: | Class II | | | FDA Identification: | Cavity varnish is a device that consists of a compound intended<br>to coat a prepared cavity of a tooth before insertion of restorative<br>materials. The device is intended to prevent penetration of<br>restorative materials, such as amalgam, into the dentinal tissue. | | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for "Sulphar". The logo consists of a circular symbol with the letters "S" and "P" intertwined, followed by the word "Sulphar" in a stylized font. The logo is black and white. Indications for Use: Remesense Sensitivity Relief Strips are intended to aid in the relief of dentinal sensitivity resulting from cold, heat, acids, or sweets. #### Device Description: Remesense Sensitivity Relief Strips are polyethylene strips with dimensions of approximately 30 x 10 mm. On the strips is a gel which contains a tooth desensitizing agent, dipotassium oxalate. The directions for use provided with the device instruct the patient to apply one or more strips on the of any teeth that are causing sensitivity problems. The dipotassium oxalate in the Remesense Sensitivity Relief Strips device forms oxalate crystals which block the open dentinal tubules, thus preventing the stimuli that may cause the sensitivity. By blocking the tubules, the crystals at the tooth surface stop the signals from hot or cold food reaching the dental nerve, providing a remedy for hypersensitivity. The component parts of the subject device are: - · Outer box with inner plastic sleeve - · Foil pouches (2 or 18 per pack), each containing a single desensitizing strip, mounted on a plastic release backing - The gel on each strip contains glycerin, aqua, potassium oxalate, carbopol, carboxy-methyl-. cellulose (CMC), potassium sorbate, sodium benzoate, sodium hydroxide (NaOH) (50% solution) - Instruction sheet . #### Performance Data: An in vitro test utilizing the 'Pashley' model has been conducted to demonstrate that the specific gel formulation used in the Remesense Sensitivity Relief Strips causes oxalate crystals to form in the tubules of teeth. Test results demonstrate no significant difference in flow reduction between the subject device, Remesense Sensitivity Relief Strips, and the predicate device, Remesense for Sensitive Teeth, when applied for 10 minutes, as per the instructions for use of both devices. In summary, Remesense Sensitivity Relief Strips (subject device), Remesense For Sensitive Teeth (predicate device), and a dipotassium oxalate solution at the same concentration and pH as the Remesense products, all produced highly significant reductions in flow through human dentin after treatment. The occlusions formed from all three oxalate treatments were similar in shape, texture and location. Elemental analysis showed the occlusions formed from the treatment with Remesense Sensitivity Relief Strips are oxalate based. The same study also assessed the occlusion potential of a Remesense Sensitivity Relief Strip placebo gel control (no dipotassium oxalate) and a water control. Neither the placebo gel nor the water provided any quantitative reduction in flow, and cross sectional analysis of the tested dentin did not show any evidence of occlusions, demonstrating that any occlusion and desensitizing effect is caused only by the oxalate component and not by the gel. In order to confirm comprehension of the device labelling and the minimization of the potential for use error, a usability study was carried out. The conclusion from the usability study was that patients suffering from tooth hypersensitivity are able to understand the labelling and instructions, and use in accordance with the instructions for use provided by the manufacturer. One new risk of use identified was the risk of choking on the polyethylene strips, especially if a number of strips were used at the same time. In order to assess this risk, a study was carried out by a third party laboratory. #### Biocompatibility: Biocompatibility testing has been carried out in accordance with the recommendations of ISO 10993-1, specifically tests have been conducted in accordance with ISO 10993-1:2009, ISO 10993-5:2009, ISO {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "Sulphar" in a stylized font, with a line above it. To the left of the word is a circular logo with the letters "SP" inside. The logo is mostly black, with the letters "SP" in white. The word "Sulphar" is in black. 10993-10:2010. The requirements of ISO 7405:2008 have also been considered in establishing the biocompatibility test regime for the device. #### Comparison with predicate devices: The predicate device selected for comparison with the Sylphar Remesense Sensitivity Relief Strips is: | Predicate Device: | Sylphar Remesense for Sensitive Teeth | |----------------------|---------------------------------------| | 510(k) Sponsor: | Sylphar N.V. | | 510(k) Number: | K122708 | | Clearance Date: | 14 January 2013 | | FDA Product Code: | LBH | | Classification Name: | Varnish, Cavity | | Regulation No: | 872.3260 | This 510(k) submission describes a delivery system for the desensitizing ingredients of the subject device formulation that is different from that used in the predicate. Sylphar Remesense for Sensitive Teeth. Sylphar Remesense for Sensitive Teeth used foam pads impregnated with the desensitizing gel, and the pads were mounted in a mouth-shaped tray, so that the desensitizing agent was applied to a number of teeth at the same time. Remesense Sensitivity Relief Strips uses a gel with a very similar formulation coated on a polyethylene strip, which may be applied directly to only the sensitive areas. Multiple strips are included in the pack, so that a number of sensitive teeth can be treated at the if required. Remesense Sensitivity Relief Strips is intended for over-the-counter (OTC) use, which is the same as for the predicate device. The subiect device and the predicate device use the same mode of action (tubule occlusion) and desensitizing substance (potassium oxalate), and have identical indications for use statements. The subject device and the predicate device are very similar with the only significant difference being the desensitizing gel delivery system (gel-coated polyethylene strips instead of foam impregnated pads). This difference has no significant effect on safety or effectiveness, as demonstrated by the results of in vitro tests on both the subject device and predicate device. The incidental excipients included in the gel formulation for the subject device are different from those in the predicate device, because the viscosity of the gels used in the subject device and predicate device needs to be different, due to the different methods of delivery. This difference has no significant effect on safety or effectiveness. #### Conclusion: Based on the information contained within this submission, it is concluded that 'Sylphar Remesense Sensitivity Relief Strips' is substantially equivalent to the identified predicate device already in interstate commerce within the USA. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 14, 2014 Sylphar N.V. C/0 Mr. Roger Gray Vice President, Quality and Regulatory Donawa Lifescience Consulting Piazza Albania 10 Rome, Italy 00153 Re: K132426 Trade/Device Name: Sylphar Remesense Sensitivity Relief Strips Regulation Number: 21 CFR 872.2360 Regulation Name: Tooth Desensitizer Regulatory Class: Class II Product Code: LBH Dated: January 14, 2014 Received: January 16, 2014 Dear Mr. Gray : We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Mr. Roger Gray comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Image /page/4/Picture/7 description: The image contains a logo with the letters "FDA" in a stylized font. Above the logo, there is text that reads "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director." Below the logo, the word "DAGRID" is written. The word "FOR" is in the bottom right corner of the image. Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): K132426 Device Name: Sylphar Remesense Sensitivity Relief Strips #### Indications for Use: Sylphar Sensitivity Relief Strips are intended to aid in the relief of dentinal sensitivity resulting from cold, heat, acids, or sweets. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Image /page/5/Picture/8 description: The image contains a single, bolded 'X' symbol. The 'X' is enclosed within a square box, suggesting it might be a checkbox or a similar interactive element. The symbol is simple and lacks any additional context or surrounding elements. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of Device Evaluation (ODE) Mary S. Runner -S Susan R DDS, MA 2014.02.14 12:59:21-05'00' Page 1 of 1 Rev.2 Apr 4, 2008
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%